2680 Series
Technical data subject to change © B&K Precision Corp. 2019 www.bkprecision.com The 2680 Series of spectrum analyzers delivers performance and functionality in a lightweight, compact design, suitable for lab and field use. The large 10.1” wide-screen color display allows the user to visualize the waveform and make precision measurements such as third order intercept, occupied bandwidth, 2D and 3D spectrum monitor. The 2680 Series provides a standard pre-amplifier and tracking generator in both the 2.1 and 3.1 GHz models. The series also includes
1 Hz minimum RBW and advanced measurements,
which make these analyzers perfect for applications in 2 way radio, site surveying, EMI pre-compliance, characterizing the frequency response of RF devices and more. Features & benefits n Frequency range: 9 kHz to 2.1 or 3.2 GHz n High Sensitivity -161 dBm/Hz displayed average noise level (DANL) n Low phase noise of -98 dBc/Hz @ 10 kHz offset n Low level uncertainty of ±0.7 dB n 1 Hz minimum resolution bandwidth (RBW) n Preamplifier and tracking generator standard on all models n 10.1” wide-screen 1024 x 600 color display n LAN and USBTMC connectivity n USB host port to store and recall waveform data, setups, and screen captures Options n Reflection measurement n EMI pre-compliance Models 2682 2683 Frequency Range 9 kHz to 2.1 GHz 9 kHz to 3.2 GHz Tracking Generator √ √ Preamplifier √ √ Advanced Measurements √ √ USB (USBTMC) LAN
1 Hz minimum resolution bandwidth
(RBW) Low resolution bandwidth helps differentiate between adjacent signals RBW 1 Hz Advanced measurements n Channel Power n Adjacent Channel Power n Occupied Bandwidth n Total Power n Third-Order-Intercept n 2D and 3D Spectrum Monitor Standard
www.bkprecision.com Front panel Side & rear panel LCD display 10.1 inch TFT high resolution color display to view critical detail in the spectrum Function keys Setup measurement modes and activate context based help for each key which also includes SCPI information where applicable RF inputTracking generator output For gain/loss, antenna, and cable measurements. Standard on all models Earphone jack Listen to demodulated AM or FM signals Trigger input External TTL signal to initiate sweep Control keys Use the rotary knob, arrow keys or alpha numeric keys to adjust values USB host port Store or recall analyzer settings and save screen captures Kensington security slot Secure your spectrum analyzer and prevent theft Power switch Remote interface USB and LAN connections Menu control keys Provide quick access to the most common functions in the current mode
10 MHz REF input/
www.bkprecision.com Operation highlights Low displayed average noise level (DANL) Phase noise -98 dBc/Hz@ 1 GHz, offset 10 kHz. Four independent traces and markers Capture snapshots, continuously update the maximum or minimum value, and perform math on all 4 individually colored traces. Delta markers Powerful delta markers can be used to select amplitude, span, stop, start or center frequency, measure noise level, amplitude or frequency. Take advantage of the preamp and -161 dBm DANL to measure low level signals accurately. Low phase noise for accurate measurements The 2D and 3D spectrum monitor features are standard on the 2.1 GHz and 3.2 GHz models. The 3D spectrum monitor can be displayed using the provided PC software, while the 2D is viewable on the spectrum analyzer screen and in the PC software. This feature shows how the frequency content of a signal changes over time by representing the power intensity with a color gradient. In today’s crowded spectrum, ACPR measurements are critical to ensure compliance with regulations. The 2680 series displays the main channel power, left and right channel power as well as bandwidth for each channel on screen for ease of determining the total power being transmitted and the spectrum being used. Spectrum monitor Adjacent channel power ratio (ACPR)
www.bkprecision.com Options PC software Use 3D spectrum monitoring with the PC software.Generate test reports Expand control of the spectrum analyzer with front panel emulation. Create, load or save user defined limit and correction files, save screen captures and store readings from the included software. Reflection bridge Quasi-peak detection with dwell time helps identify non-compliant emissions. Use the provided EMI software (available for download at www.bkprecision.com) to configure the spectrum analyzer, perform prescan, peak search, final scan and generate reports of your pre-compliance tests. Visualize return loss, reflection coefficient, and VSWR of your DUT. This option enables VSWR, reflection coefficient, and return loss measurements for tuning and determining the efficiency of antennas, filters, or RF transmission modules. Reflection measurement option This option enables the instrument’s EMI measurement function which includes pre-defined bandwidth set points of 200 Hz, 9 kHz and 120 kHz, a -6dB EMI filter, and the quasi-peak detector as specified by CISPR 16-1. EMI pre-compliance option The RF energy radiating from a device can be detected and measured with near field probes and the spectrum analyzer. The wide band amplifier can be connected between the probe and the
2680 Series to increase the dynamic range of the
measurement system. The probes can also be used to test RF immunity by inducing signal into the circuit. Near field probe kit Use near field probes to help track down emissions. Magnetic (H) and electric (E) near field probes with 40 dB pre-amplifier Order information for instrument options Order number Description EMI2680 License key, activates EMI measurements with Quasi-peak RFL2680 License key, activates reflection measurements RB2680B Reflection bridge with adapters PR262 1 electric and 3 magnetic field probes with amplifier and SMA cable Buy now, upgrade later Install the licenses at any time or try before you buy with the 30 day trial license on each instrument. Installation is quick and easily done within the spectrum analyzer menu. To purchase a license key, please fill out the license request form which can be found on the 2680 Series accessory page on our website www.bkprecision.com.
www.bkprecision.com Specifications Series 2682 2683 Frequency characteristics Frequency Range 9 kHz to 2.1 GHz 9 kHz to 3.2 GHz Frequency Resolution 1 Hz Frequency Span 0 Hz, 100 Hz to 2.1 or 3.2 GHz Frequency Span Accuracy ±Span / (number of sweep points -1) Internal reference source Reference Frequency 10 MHz Initial Calibration Accuracy <1 ppm Temperature Stability <1 ppm/year, 0 °C to +50 °C Frequency Aging Rate <0.5 ppm/first year, 3.0 ppm/20 year Frequency Reference Accuracy ±[(time since last adjustment × frequency aging rate) + temperature stability + calibration accuracy] Marker Marker Resolution Span / (number of sweep points -1) Marker Uncertainty ± [frequency indication x frequency reference uncertainty + 1% x span + 10% x resolution bandwidth + marker resolution] Frequency Counter Resolution 1 Hz Frequency Counter Uncertainty ± [frequency indication x frequency reference accuracy + counter resolution] Bandwidths Resolution Bandwidth (-3 dB) 1 Hz to 1 MHz, in 1-3-10 sequence Resolution Filter Shape Factor <4.8:1 (60 dB: 3 dB), Gaussian-like RBW Uncertainty <5% Video Bandwidth (-3 dB) 1 Hz to 3 MHz, in 1-3-10 sequence VBW Uncertainty <5% Amplitude and level Measurement Range (preamplifier off) DANL to +10 dBm, 100 kHz to 1 MHz DANL to +20 dBm,1 MHz to 3.2 GHz Reference Level -100 dBm to +30 dBm, 1 dB steps Preamplifier 20 dB (nom.), 9 kHz to 3.2 GHz Input Attenuation 0 to 51 dB, 1 dB steps Maximum Input DC Voltage ±50 Vdc Maximum Average RF Power 30 dBm, 3 minutes, fc >equal to 10 MHz, attenuation >20 dBm, preamplifier off Maximum Damage Level 33 dBm, fc >equal 10 MHz, attenuation >20 dBm, preamplifier off Specifications are valid under the following conditions: The instrument is within the calibration period, has been stored between 0 and 50°C for at least 2 hours prior to use, and has been powered on and warmed up for at least 40 minutes. The specifications include the measurement uncertainty, unless otherwise noted. Specifications: All products are guaranteed to meet published specifications when operating temperatures from 5 to 45°C, unless otherwise noted. Typical: Performance deemed typical implies that 80 percent of the measurement results will meet the typical published performance with a 95th percentile confidence level at room temperature (approximately 25°C). Typical performance is not warranted and does not include measurement uncertainty. Nominal: The expected performance or design attribute. Displayed average noise level (DANL) 20 °C to 30 °C ,attenuation = 0 dB, sample detector, trace average >50 RBW=10 Hz Normalization to 1 Hz Preamp Off 9 kHz to 100 kHz -100 dBm (nom.) -100 dBm (nom.) 100 kHz to 1 MHz -97 dBm, -101 dBm (typ.) -107 dBm, -111 dBm (typ.)
1 MHz to 10 MHz -122 dBm,
-126 dBm (typ.) -132 dBm, -136 dBm (typ.)
10 MHz to 200 MHz -127 dBm,
-131 dBm (typ.) -137 dBm, -141 dBm (typ.) 200 MHz to 2.1 GHz -125 dBm, -129 dBm (typ.) -135 dBm, -139 dBm (typ.) 2.1 GHz to 3.2 GHz -116 dBm, -122 dBm (typ.) -126 dBm, -132 dBm (typ.) Preamp On 9 kHz to 100 kHz -107 dBm (nom.) -117 dBm (nom.) 100 kHz to 1 MHz -122 dBm, -127 dBm (typ.) -132 dBm, -137 dBm (typ.)
1 MHz to 10 MHz -138 dBm,
-144 dBm (typ. -148 dBm, -154 dBm (typ.)
10 MHz to 200 MHz -146 dBm,
-151 dBm (typ.) -156 dBm, -161 dBm (typ.) 200 MHz to 2.1 GHz -145 dBm, -148 dBm (typ.) -155 dBm, -158 dBm (typ.) 2.1 GHz to 3.2 GHz -135 dBm, -139 dBm (typ.) -145 dBm, -149 dBm (typ.) Phase noise Carrier Offset fc=1 GHz, 20 ºC ~30 ºC 10 kHz <-95 dBc/Hz, <-98 dBc/Hz (typ.) 100 kHz <-96 dBc/Hz, <-97 dBc/Hz (typ.) 1 MHz <-115 dBc/Hz , <-117 dBc/Hz (typ.) Level display Logarithmic Level Axis 10 dB to 100 dB Linear Level Axis 0 to reference level Units of Level Axis dBm, dBmV, dBµV, dBµA , V, W Number of Display Points 751 Number of Traces 4 Trace Detectors Positive-Peak, Negative-Peak, Sample, Normal, Average (Voltage/RMS/Video), Quasi-Peak (with EMI option) Trace Functions Clear Write, Max Hold, Min Hold, View, Blank, Average
www.bkprecision.comv081220 Frequency response Preamplifier Off ±0.8 dB, ±0.4 dB typ. On ±0.9 dB, ±0.5 dB typ. Error and accuracy Resolution Bandwidth Switching Uncertainty
1 Hz RBW
Logarithmic resolution ±0.2 dB, Linear resolution ±0.01, nom. Input Attenuation Switching Uncertainty 20 °C to 30 °C, fc = 50 MHz, preamp off, Relative to 20 dB, 1 to 51 dB attenuation ±0.5 dB Absolute Amplitude Accuracy Preamplifier off: ±0.4 dB, input signal -20 dBm Preamplifier off: ±0.5 dB, input signal -40 dBm Total Amplitude Accuracy ±0.7 dB 20 °C to 30 °C , Fc>100 kHz, input signal -50 dBm to 0 dBm, RBW = 1 kHz, VBW = 1 kHz, peak detector, attenuation = 20 dB,preamp off, 95th percentile reliability RF Input VSWR <1.5 nom. Input attenuation 10 dB, 1 MHz to 3.2 GHz Distortion and spurious responses Second Harmonic Distortion -65 dBc fc ≥50 MHz, Mixer Level -30 dBm, attenuation = 0 dB, preamp off, 20 °C to 30 °C Third-Order Intercept +10 dBm fc ≥50 MHz, two -20 dBm tones at input mixer spaced by 100 kHz, attenuation = 0 dB, preamp off, 20 °C to 30 °C 1 dB Gain Compression >-5 dBm, nom. fc ≥50 MHz, attenuation = 0 dB, preamp off, 20 °C to 30 °C Residual Response <-90 dBm, typ. input terminated = 50 Ω,attenuation = 0 dB, 20 °C to 30 °C Input Related Spurious <-65 dBc Mixer level = -30 dBm, 20 °C to 30 °C Sweep and trigger Sweep Time 1 ms to 3000 s Sweep Accuracy Accuracy, Speed Sweep Mode Sweep, FFT Sweep Rule Single, Continuous Trigger Source Free, Video, External External Trigger 5 V TTL level, 1 kΩ, BNC-female, rising edge/falling edge Specifications (continued) Tracking generator Frequency Range 100 kHz to 2.1 GHz 100 kHz to 3.2 GHz Output Level -20 dBm to 0 dBm Output Level Resolution 1 dB Output Flatness ±3 dB Output Maximum Reserve Level Mean power: 30 dBm, DC: ±50 Vdc EMI Pre-compliance option (EMI2680) Resolution Bandwidth (6 dB) 200 Hz, 9 kHz, 120 kHz Detector Quasi-peak (following CISPR 16-1-1) Dwell Time 0 µs to 10 s Reflection measurement option (RFL2680) Measurements VSWR, Return loss, Reflect coefficient RF and 10 MHz input/output Front panel RF input 50 Ω, N-female Front panel TG output 50 Ω, N-female
10 MHz reference output 10 MHz, >0 dBm, 50 Ω, BNC-female
10 MHz reference input 10 MHz, -5 dBm to +10 dBm, 50 Ω, BNC-female
AC Input 100 V - 240 V, 50 Hz/60 Hz/400 Hz AC Display TFT LCD, 1024 × 600 (waveform area 751 × 501), 10.1” I/O Interface USB host (type A) USB 2.0 USB device (type B) USB 2.0 LAN 10/100 Base T, RJ45 Temperature Operating: 0 °C to 50 °C Storage: -20 °C to 70 °C Humidity 0 °C to 30 °C , ≤95% RH 30 °C to 50 °C , ≤75% RH Safety EN 61010-1:2010, Low Voltage Directive (LVD) 2014/35/EU Electromagnetic Compatibility EN 61326-1:2013, EMC Directive 2014/30/EU Dimensions (W x H x D) 15.47” x 8.15” x 4.59” (393 mm x 207 mm x 116.5 mm) Weight 10.1 lb (4.60 kg) Warranty 3 years Included Accessories Power cord, certificate of calibration Optional Accessories EMC Near-field probes (PR262), reflection bridge (RB2680B)