Handheld Digital Storage Oscilloscopes

2510 Series

www.bkprecision.comTechnical data subject to change © B&K Precision Corp. 2016 The 2510 Series handheld digital storage oscilloscopes provide floating measurement and recorder capabilities with a built-in digital multimeter (DMM), all in one portable and lightweight package. These versatile 60 MHz and 100 MHz bandwidth scopes offer 1 GSa/s sample rates, 2 Mpts waveform memory, 32 automatic measurements, and multiple recording functions to capture transient or long-term signal behavior. The built-in 6000-count multimeter allows users to quickly switch over from an oscilloscope to a DMM to measure DC/AC voltage and current, resistance, and capacitance, including diode and continuity tests. These handheld scopes feature many useful recording functions such as trend plot, which allows data logging from the scope or multimeter. Additionally, the scope recorder function offers users 7 Mpts record length on a single channel or 3.5 Mpts on dual channel. The 2510 Series handheld oscilloscopes are ideal for industrial applications, power systems, electronics design, and field test and service. Features and Benefits I 60 MHz (2511/2515) and 100 MHz (2512/2516) bandwidth I 1 GSa/s sample rate I Deep waveform memory up to 2 Mpts I 2 fully isolated and floating 1,000 V CAT II,

600 V CAT III rated inputs (isolated models

2515 and 2516) I 300 V CAT II rated inputs (non-isolated models 2511 and 2512) I Built-in 6000-count DMM with dedicated terminals for current measurement I Scope and meter trend plot functions for data logging I Bright 5.7” color display I Compact and lightweight – 3.4 lbs (1.54 kg) I FFT including four additional math functions - Add, Subtract, Multiply, and Divide I 32 automatic measurements I USB host port for saving and recalling waveform setups, data, and screenshots on a USB flash drive I Software provided for remote PC control I Advanced tools include digital filters with adjustable limits, scope and waveform recorder mode I Multi-language user interface I Up to four hours of continuous battery operation Model 2511 2512 2515 2516 Bandwidth 60 MHz 100 MHz 60 MHz 100 MHz Channels 2 non-isolated 2 fully isolated Typical Applications General electronics Power electronics and industrial

www.bkprecision.com Handheld Digital Storage Oscilloscopes 5.7” color display Scope and meter mode buttons Recorder function button Use the scope recorder or trend plot functions to log oscilloscope or multimeter measurements. Trigger menu button Multimeter input terminals Dedicated terminals provided for current measurements. Scope input terminals Durable rubber bezel Provides grip and protection against accidental bumps. Auto setup Vertical, horizontal, and trigger controls are automatically adjusted for fast signal display. CH2 selection button CH1 selection button CH2 vertical controls Horizontal controls CH1 vertical controls Á Side View Front Panel DC power adapter input USB host port Connect your USB flash drive to conveniently store and recall waveform data, setups, and screenshots. Mini USB device port Compensate your probe or enable remote control from a PC. ÁUSB ports T op View

www.bkprecision.com Handheld Digital Storage Oscilloscopes Filter out unwanted signal components such as various types of noise with built-in digital filters. Choose from Low-Pass, High-Pass, Band-Pass, and Band-Stop filters. PC software provided (free download from www.bkprecision.com) for seamless integration between the oscilloscope and PC. Capture and transfer waveforms, screen images, setups and measurement results to a Windows PC via the USB device port on the side of the instrument. A USB host port is also available for quick and easy screen saving. Monitor and analyze long-term signal behavior by recording data continuously over a period of time. These modes allow recorded data to be played back for post acquisition analysis. Display and measure the input signal’s frequency spectrum. Select one of the 4 FFT windows: Rectangular, Hanning, Hamming, and Blackman. Use cursors to measure the spectral component’s magnitude and frequency. Capture intermittent errors in your system. The trend plot function can be used with the oscilloscope or built-in DMM to plot measurement values over time. Up to two voltage or time parameters can be selected by the scope, and any one of the multimeter’s measurement functions can be graphed. These data points can then be exported to a CSV file for further analysis. Quickly troubleshoot in the field using battery powered operation. Built for portability, the 2510 Series handheld digital oscilloscopes are rugged, compact, and lightweight. All models come standard with travel case for safe transport on the road. See more details in your waveform with deep memory. When enabled, waveforms can be captured in high resolution while maintaining a high sample rate over a wider period of time than other comparable scopes. Speed up troubleshooting with the built-in 6000-count multimeter. Measurement functions include AC/DC voltage and current, resistance, capacitance, diode, and continuity test. The tools you need Digital Filtering PC Connectivity Scope and Waveform Recorder ModesPowerful Measurement Functions Scope and Meter Trend Plot Functions Portable Operation

2 Mpts Deep Memory

Built-in Digital Multimeter

www.bkprecision.com Handheld Digital Storage Oscilloscopes Fully Isolated Channel Design for Safe Measurements (models 2515/2516 only) Models 2515 and 2516 offer two CAT III 600 V input channels for floating measurements and feature an electrically isolated circuit design between inputs and the digital acquisition circuit. Isolating the ground references eliminate ground loops and help reduce channel noise and crosstalk. Input Signal and Float Voltage Safety Ratings Model 2511 / 2512 2515 / 2516 Maximum signal input safety rating with included probe 300 Vrms CAT II 1000 Vrms CAT II, 600 Vrms CAT III Maximum signal input safety rating without probe 300 Vrms CAT II 300 Vrms CAT II Maximum reference floating safety rating 30 Vrms 1000 Vrms CAT II, 600 Vrms CAT III All 2510 Series models come standard with high bandwidth, safety certified passive probes (one per channel) to help you get the most out of your scope. Safety Rated High Bandwidth Oscilloscope Probes Floating and Differential Measurements Many industrial applications such as power electronics require measurements of high voltages and currents that are not referenced to ground. This poses an issue with traditional line-powered oscilloscopes, which typically have signal common connected to the chassis of the oscilloscope. This means all measurements must be made relative to earth ground, preventing users from making differential measurements where none of the test points are referenced to ground. As a workaround, some people choose to float an oscilloscope by removing the connection between the instrument's chassis and power line ground. Floating a scope is not recommended as it can put the user at a safety risk. Parasitic capacitance is also induced in the measurement which can cause ringing and invalidate the measurement. The 2510 Series allows engineers and technicians to make accurate and safe measurements when the signal reference is floating. Model Included Probes

2511 Two 150 MHz bandwidth, x1/x10

probes rated for 300 V CATII measurements2512

2515 Two touch-protected 250 MHz bandwidth,

x10 probes rated for 1000 V CATII,

600 V CATIII measurements2516

www.bkprecision.com Handheld Digital Storage Oscilloscopes Models 2511 2512 2515 2516 Performance Characteristics Bandwidth 60 MHz 100 MHz 60 MHz 100 MHz Real Time Sampling Rate 1 GSa/s (half-channel interleaved)(1), 500 MSa/s (per channel) Channels 2 non-isolated 2 isolated Rise Time < 5.8 ns < 3.5 ns < 5.8 ns < 3.5 ns Ch-to-Ch Isolation (both channels at same V/div setting) > 100:1 at 50 MHz Memory Depth 40 kpts (half-channel interleaved)(1)(2), 20 kpts (per channel) Deep Memory (3) 2 Mpts (half-channel interleaved)(1), 1 Mpts (per channel) Vertical Resolution 8 bits Vertical Sensitivity 2 mV/div – 100 V/div (1-2-5 order) 5 mV/div – 100 V/div (1-2-5 order) DC Gain Accuracy 5 mV/div-100 V/div: ≤ ± 4 % Max. BNC Input Voltage CATII 300 Vrms from BNC signal to BNC shell Max. Input Voltage for Probe PR150SA (4) : 1x/10x CAT II 300 V PR250SA (5)(6): 10x CAT III 600 V , CATII 1000 V Channel Voltage Offset Range 2 mV - 200 mV : ±1.6 V 206 mV - 10 V : ±40 V

10.2 V - 100 V : ±400 V

5 mV - 200 mV : ±1.6 V 206 mV - 10 V : ±40 V Bandwidth Limit 20 MHz (-3 dB) Horizontal Scan Range 5.0 nS/div - 50 S/div 2.5 nS/div - 50 S/div 5.0 nS/div - 50 S/div 2.5 nS/div - 50 S/div Timebase Accuracy ± 50 ppm measured over 1 ms interval Input Coupling AC, DC, GND Input Impedance 1 MΩ +/- 2 % || 18 pF ± 3 pF 1 MΩ +/- 2 % || 16 pF ± 3 pF Probe Attenuation Selectable Factors 1X, 5X, 10X, 50X, 100X, 500X, 1000X Vertical and Horizontal Zoom Vertically or horizontally expand or compress a live or stopped waveform I/O Interface USB USB host port support USB flash drives (FAT format) mini-USB device port for PC connectivity and probe compensation Acquisition Modes Sampling Display sample data only Peak Detect Capture the maximum and minimum values of a signal Average Waveform averaged, selectable from 4, 16, 32, 64, 128, 256 Trigger System Trigger Types Edge, Pulse Width, Video*, Slope, Alternative *Support signal formats: PAL/SECAM, NTSC Trigger condition: odd field, even field, all lines, or line number Trigger Modes Auto, Normal, Single Trigger Coupling AC, DC, LF reject, HF reject Trigger Source CH1, CH2 Trigger Level Range ± 6 divisions from center of display Trigger Displacement Pre-trigger: Memory depth / 2* sampling Delay Trigger: 268.04 div Pulse Width Trigger Positive slope (>, <, =), Negative slope (>, <, =), Time: 20 ns - 10 s Slope Trigger Positive slope (>, <, =), Negative slope (>, <, =), Time: 20 ns - 10 s Alternate Trigger CH1 trigger type: Edge, Pulse, Video, Slope CH2 trigger type: Edge, Pulse, Video, Slope Digital Storage Oscilloscope Specifications

www.bkprecision.com Handheld Digital Storage Oscilloscopes Digital Storage Oscilloscope Specifications (cont.) Hardware Frequency Counter Reading Resolution 1 Hz Range DC couple, 10 Hz to MAX bandwidth Signal Types Satisfying all trigger signals (except pulse width trigger and video trigger) Waveform Math and Measure Math Operation Add, Subtract, Multiply, Divide, FFT FFT Window mode: Hanning, Hamming, Blackman, Rectangular Sampling points: 1024 Measure Vpp, Vmax, Vmin, Vamp, Vtop, Vbase, Vavg, Mean, Crms, Vrms, ROVShoot, FOVShoot, RPREShoot, FPREShoot, Rise, Fall, Freq, Prd, +Wid, -Wid, +Dut, -Dut, BWid, Phas, FRR, FRF , FFR, FFF , LRR, LRF , LFR, LFF Cursors Types Voltage, Time Measurements ∆V , ∆T , 1/∆T (frequency) Display System Display 5.7” Color TFT , 320 x 234 resolution, 64K color Display Contrast (Typical) 150:1 Backlight Intensity (Typical) 300 nits Wave Display Range 8 x 12 div Wave Display Mode Dots, Vector Persistence Off, 1 sec, 2 sec, 5 sec, Infinite Menu Display 2 sec, 5 sec, 10 sec, 20 sec, Infinite Waveform Interpolation Sin(x)/x, Linear Measure Display Modes Main, Window zoom, Scan, X-Y X-Y Sampling Frequency Support 25 kSa/s - 250 MSa/s sampling rate (1-2.5-5 order) Color Mode Normal, Invert Model 2511 2512 2515 2516 Environmental and Safety T emperature Operating: 32 °F to 104 °F (0 °C to +40 °C) Not operating: -4 °F to 158 °F (-20 °C to +70 °C) Humidity Operating: 85% RH, 104 °F (40 °C), 24 hours Altitude Operating: 9,842.5 ft (3,000 m) Electromagnetic Compatibility EMC Directive 2004/108/EC, EN61326:2006 Safety Low voltage directive 2006/95/EC, EN61010-1:2001 General Storage Memory 2 reference waveforms, 20 setups, 10 waveforms AC Adapter Power Requirements Input: 100-240 VAC, 50/60 Hz Output: 9V DC, 4 A Battery Rating 5000 mAh, 7.4 VDC Battery Charge Time Approx. 4 hrs Battery Operating Time Approx. 4 hrs Weight Approx. 3.4 lbs (1.54 kg) including battery Three-Year Warranty (1) Half channel operation means that only Ch1 or Ch2 is active. (2) When sampling rate is 1 GSa/s. For sampling rate ≤500 MSa/s, the maximum memory depth is 20 kpts. (3) When sampling rate is < 500 MSa/s and maximum data depth mode is enabled. (4) Probe included with models 2511 and 2512 only. (5) Probe included with models 2515 and 2516 only. (6) Refer to respective probe’s manual for more information on the specification.

www.bkprecision.com Handheld Digital Storage Oscilloscopes Multimeter and Recorder Specifications I All specifications are based on operating at temperatures 23 ± 5°C and relative humidity < 75%. I Accuracy is based on ± (% of reading + offset). Multimeter Display Resolution 6000 counts Measurement Function DC voltage, AC voltage, resistance, diode, continuity, capacitance, DC current, AC current Max. Input Voltage AC: 750 V (30 Hz - 1 kHz) DC: 1000 V Max. Input Current(1) AC: 10 A (30 Hz - 1 kHz) DC: 10 A Input Impedance 10 MΩ Max. Input Voltage Between Multimeter Input Reference and Ground CAT II 600 V CAT III 300 V DC Voltage Range Resolution Accuracy 60.00 mV 10 µV ± (1 % + 15 digits) 600.0 mV 100 µV ± (1 % + 5 digits)

6.000 V 1 mV

60.00 V 10 mV

600.0 V 100 mV

1000 V 1 V

AC Voltage(2) Range Resolution Accuracy 60.00 mV 10 µV ± (1 % + 15 digits) 600.0 mV 100 µV ± (1 % + 5 digits)

750 V 1 V

Multimeter (cont.) Resistance Range Resolution Accuracy 600.0 Ω 0.1 Ω ± (1 % + 5 digits) 6.000 kΩ 1 Ω 60.00 kΩ 10 Ω 600.0 kΩ 100 Ω

6.000 MΩ 1 kΩ

60.00 MΩ 10 kΩ ± (3 % + 5 digits)

Diode and Continuity Measure Diode 0 – 2 V Continuity < 50 Ω alarm Capacitance Range Resolution Accuracy 40.00 nF 10 pF ± (3 % + 10 digits) 400.0 nF 100 pF ± (4 % + 5 digits) 4.000 µF 1 nF 40.00 µF 10 nF 400.0 µF 100 nF DC and AC Current(3)(4) Range Resolution Accuracy 60.00 mA 10 µA ± (1.5 % + 5 digits) 600.0 mA 100 µA

6.000 A 1 mA

± (2.5 % + 5 digits)

10.00 A 10 mA

Display Mode Full view, Normal Record Length 800k points, > 24 hours Number of Channels 2 Multimeter Trend Plot Display Mode Full view, Normal Record Length 1.2M dots, > 24 hours Number of Channels 1 Scope Recorder Display Mode Full view, Normal Max. Record Length Single Channel: 7 M pts Dual Channel: 3.5 M pts Number of Channels 2 Maximum Record Size to External Storage 4 GB, 3000 hours (1) Current input terminals protected with internal 250 V rated fuse. (2) For frequency range 30 Hz to 1 kHz. (3) For 10 A terminal, > 6 A DC or AC rms for 10 seconds ON and 15 minutes OFF . (4) For AC current ranges, frequency is verified for 30 Hz to 1 kHz. v032416 Included Accessories User manual, passive probes (one per channel), pair of DMM test leads, 7.4 V Li-ion battery BP2510, USB cable, probe compensation connector, AC power adapter, travel case LC2510B and certificate of calibration