Technical content
- Two stage electrical setup provides significant attenuation from 10kHz to 30MHz
- Low profile design allows for a space-saving design solution
- Compliant with internal spacing requirements for UL60601-1, 3rd edition for medical applications
- Designed for easy connection and installation
APPLICATIONS
- Dialysis machines
- Diagnostic equipment
- Medical laser
- CAT scanners
- MRT scanners
- Dentist equipment
- Operating room equipment
- Physical therapy equipment
ORDERING INFORMATION
The HT Series filters are designed to provide significant attenuation of RFI noise in the frequency ranges from 10kHz to 30MHz for low leakage current applications. The new 20A versions come in a chassis mount style and offer spade or threaded bolt termination options. STANDARDS AND SPECIFICATIONS
- UL Recognized
- VDE Approved
- CSA Pending ELECTRICAL SPECIFICATIONS Hipot rating (one minute): Line to Ground: 1500 VAC Line to Line: 1450 VDC Rated Voltage: 250 VAC Operating Frequency: 50/60 Hz Rated Current: 20A max. ELECTRICAL SCHEMATIC PRODUCT OFFERING 4-1609037-0: 20EHT1 20A HT Series filter with spade terminals 3-1609037-5: 20EHT6 20A HT Series filter with threaded bolt terminals Samples Available LINE LOAD
20 E HT 1
1 – .250 [6.3 ] spade t erminals 6 – Threaded bolts E – Lo w Leakage (<0.5 mA) 20A HT Series Input / Output Style Leakage Current Designation Current Rating 1. Does not apply to 20EHT7 version Corcom HT Series corcom.com 1-1773704-3 JG2 PDF 09/2013 © 2013 Tyco Electronics Corporation, a TE Connectivity Ltd. Company. All Rights Reserved. Corcom, TE Connectivity and TE connectivity (logo) are trademarks. Other logos, product and/or company names might be trademarks of their respective owners. USA: 1-800-522-6752 Latin/S. America: +54 (0) 11-4733-2200 France: +33 (0) 1-3420-8686 Canada: +1-905-475-6222 Germany: +49 (0) 6251-133-1999 Netherlands: +31 (0) 73-6246-999 Mexico: +52 (0) 55-1106-0800 UK: +44 (0) 800-267666 China: +86 (0) 400-820-6015
A (max) B (max) C (max) D ± .015 ± .38 20EHT1 4.55 3.12 2.18 2.0 115.6 79.2 55.4 50.8 20EHT6 4.55 3.12 2.18 2.0 115.6 79.2 55.4 50.8 20EHT1 C A D B 20EHT6 C A D B Minimum Insertion Loss - Measured in closed 50 Ohm system Current Rating Frequency — MHz 20A 2 10 15 24 29 26 20 14 11 9 5 Common Mode / Asymmetrical (Line to Ground) Differential Mode / Symmetrical (Line to Line) Dimensions in inches and millimeters unless otherwise shown. Values in italics are metric equivalents. Dimenions are shown for reference purposes only. Specifications subject to change. Customer drawing should be used for all design activity. Current Rating Frequency — MHz 20A 10 17 13 51 70 70 70 70 65 65 50 USA: 1-800-522-6752 Latin/S. America: +54 (0) 11-4733-2200 France: +33 (0) 1-3420-8686 Canada: +1-905-475-6222 Germany: +49 (0) 6251-133-1999 Netherlands: +31 (0) 73-6246-999 Mexico: +52 (0) 55-1106-0800 UK: +44 (0) 800-267666 China: +86 (0) 400-820-6015 Corcom HT Series 20A Power Line FiltersTE Connectivity corcom.com 1-1773704-3 JG2 PDF 09/2013 © 2013 Tyco Electronics Corporation, a TE Connectivity Ltd. Company. All Rights Reserved. Corcom, TE Connectivity and TE connectivity (logo) are trademarks. Other logos, product and/or company names might be trademarks of their respective owners.