780055-01 NI | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 32
Technical content
Datasheet sections
PCI/PXI/USB-6289 Specifications NI 6289 Specifications Analog Input Number of channels 16 differential or 32 single ended ADC resolution 18 bits DNL No missing codes guaranteed INL Refer to the Al Absolute Accuracy section Sample rate Single channel maximum 625 kS/s Multichannel maximum (aggregate) 500 kS/s Minimum No minimum Timing accuracy 50 ppm of sample rate Timing resolution 50 ns Input coupling DC Input range +0.1V, 0.2 V, 0.5 V, +1 V, +2 V, +5 V, +10 V Maximum working voltage for analog inputs (signal+ +11 V of AlGND common mode) ©National instruments 3
PCI/PXI/USB-6289 Specifications CMRR (DC to 60 Hz) 110 dB Input impedance Device on Al+ to Al GND >10 GQ in parallel with 100 pF Al- to AIGND >10 GQ in parallel with 100 pF Device off Al+ to AlGND 8200 Al- to AlGND 8200 Input bias current +100 pA Crosstalk (at 100 kHz) Adjacent channels -75 dB Non-adjacent channels -95 dB Small signal bandwidth (-3 dB) 750 kHz filter off, 40 kHz filter on Input FIFO size 2,047 samples Scan list memory 4,095 entries Data transfers PCI/PXI DMA (scatter-gather), interrupts, programmed I/O 4 hicom
Table 5. Settling Time for Multichannel Measurements Figure 1. Al Settling Error versus Time for Different Source Impedances
0 SSS Sa
Figure 2. Al Small Signal Bandwidth Figure 3. AlCMRR
120 TTS SL
100 Tra
Table 5. Al Absolute Accuracy (Filter On)
Table 5. Al Absolute Accuracy (Filter Off) for a coverage factor of 3 o and averaging 100 points.
PCI/PXI/USB-6289 Specifications Al Absolute Accuracy Example Absolute accuracy at full scale on the analog input channels is determined using the following assumptions: = TempChangeFromLastExternalCal = 10 °C = TempChangeFromLastinternalCal = 1 °C = number_of_readings = 100 = CoverageFactor =30 For example, on the 10 V range of the Filter On accuracy table, the absolute accuracy at full scale is as follows: = GainError=40 ppm +17 ppm-1+1ppm-10=67 ppm = OffsetError =8 ppm+11ppm-1+10 ppm=29 ppm = NoiseUncertainty = 60 yV_ 3 100 =18uV = AbsoluteAccuracy = 10 V- (GainError) + 10 V- (OffsetError) + NoiseUncertainty = 980 pV Analog Triggers Number of triggers 1 Source Al <0..31>, APFI <0, 1> Functions Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Convert Clock, Sample Clock Timebase Source level Al <0..31> +Full scale © National instruments 9
PCI/PXI/USB-6289 Specifications APFI <0, 1> +10V Resolution 10 bits, 1 in 1,024 Modes Analog edge triggering, analog edge triggering with hysteresis, and analog window triggering Bandwidth (-3 dB) Al <0..31> 700 kHz filter off, 40 kHz filter on APFI <0, 1> 5 MHz Accuracy +1% APFI <0, 1> characteristics Input impedance 10kQ Coupling DC Protection, power on +30V Protection, power off +15V Analog Output Number of channels 4 DAC resolution 16 bits DNL +1LSB 10 ni.com
PCI/PXI/USB-6289 Specifications Monotonicity 16 bit guaranteed Accuracy Refer to the AO Absolute Accuracy section Maximum update rate 1 channel 2.86 MS/s 2 channels 2.00 MS/s per channel 3 channels 1.54 MS/s per channel 4 channels 1.25 MS/s per channel Timing accuracy 50 ppm of sample rate Timing resolution 50 ns Output range (offset + reference) Calibrated ranges +1V,+2V,+5V,+10V Offset sources OV, 5 V, APFI<0, 1>, AO <0,.3>[] Reference sources 1V,2V,5V, 10 V, APFI <0, 1>, AO <0..3>/2] Maximum output level +11V Output coupling Dc Output impedance 0.20 Output current drive +5 mA ©National instruments 11
PCI/PXI/USB-6289 Specifications Overdrive protection 425V Overdrive current 20mA Power-on statel3) +5 mV Power-on glitch 2.3 V peak for 1.2 s Output FIFO size 8,191 samples shared among channels used Data transfers PCI/PXI DMA (scatter-gather), interrupts, programmed I/O USB USB Signal Stream, programmed I/O AO waveform modes Non-periodic waveform, periodic waveform regeneration mode from onboard FIFO, periodic waveform regeneration from host buffer including dynamic update Settling time, full-scale step, 15 ppm (1 LSB) 3 us Slew rate 20 V/s Glitch energy at midscale transition, +10 V range Magnitude 15mvV Duration 0.5 us External Reference APFI<0,1> characteristics 12 nicom
Figure 4. AO External Reference Bandwidth
00 Tk 10k 100k 1M
Table 5. AO Absolute Accuracy
PCI/PXI/USB-6289 Specifications Direction control Each terminal individually programmable as input or output Pull-down resistor 50 kQ typical, 20 kQ minimum Input voltage protection +20V on up to two pinsl4l Waveform Characteristics (Port 0 Only) Terminals used Port 0 (P0.<0..31>) Port/sample size Up to 32 bits Waveform generation (DO) FIFO 2,047 samples Waveform acquisition (DI) FIFO 2,047 samples DI Sample Clock frequency PCI/PXI 0 MHz to 10 MHz, system and bus activity dependent USB 0 MHz to 1 MHz, system and bus activity dependent DO Sample Clock frequency PCI/PXI Regenerate from FIFO 0 MHz to 10 MHz Streaming from memory 0 MHz to 10 MHz, system and bus activity dependent USB Regenerate from FIFO 0 MHz to 10 MHz ©National instruments 15
PCI/PXI/USB-6289 Specifications Streaming from memory 0 MHz to 1 MHz, system and bus activity dependent Data transfers PCI/PXI DMA (scatter-gather), interrupts, programmed I/O USB USB Signal Stream, programmed I/O DI or DO Sample Clock sourcel®) Any PFI, RTSI, Al Sample or Convert Clock, AO Sample Clock, Ctrn Internal Output, and many other signals PFI/Port 1/Port 2 Functionality Functionality Static digital input, static digital output, timing input, timing output Timing output sources Many Al, AO, counter, DI, DO timing signals Debounce filter settings 125 ns, 6.425 us, 2.56 ms, disable; high and low transitions; selectable per input Recommended Operating Conditions Level Minimum Maximum Input high voltage (Vi) 2.2V 5.25V Input low voltage (Vj) ov 0.8V Output high current (Io) P0.<0..31> - -24mA Output high current (Io4) PFI <0..15>/P1/P2 - -16 mA Output low current (Io,) P0.<0..31> - 24mA Output low current (Io,) PFI <0..15>/P1/P2 _ 16mA 16 ni.com
PCI/PXI/USB-6289 Specifications Output applications Pulse, pulse train with dynamic updates, frequency division, equivalent time sampling Internal base clocks 80 MHz, 20 MHz, 0.1 MHz External base clock frequency 0 MHz to 20 MHz Base clock accuracy 50 ppm Inputs Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down Routing options for inputs Any PFI, RTSI, PXI_TRIG, PXI_STAR, analog trigger, many internal signals FIFO 2 samples Data transfers PCI/PXI Dedicated scatter-gather DMA controller for each counter/timer; interrupts, programmed I/O USB USB Signal Stream, programmed I/O Frequency Generator Number of channels 1 Base clocks 10 MHz, 100 kHz Divisors 1to16 Base clock accuracy 50 ppm ©National instruments 19
PCI/PXI/USB-6289 Specifications Output can be available on any output PFI or RTSI terminal. Phase-Locked Loop (PLL) | Note Pci/PXxI devices only. Number of PLLs 1 Reference signal PXI_STAR, PXI_CLK10, RTSI <0..7> Output of PLL 80 MHz Timebase; other signals derived from 80 MHz Timebase including
20 MHz and 100 kHz Timebases
Source Any PFI, RTSI, PXI_TRIG, PXI_LSTAR Polarity Software-selectable for most signals Analog input function Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Convert Clock, Sample Clock Timebase Analog output function Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase Counter/timer function Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down Digital waveform generation (DO) Sample Clock function Digital waveform acquisition (DI) | Sample Clock function 20 nicom
PCI/PXI/USB-6289 Specifications Device-to-Device Trigger Bus PCI RTS! <0..7>16) PX! PXI_TRIG <0..7>, PXILSTAR USB source None Output selections 10 MHz Clock, frequency generator output, many internal signals Debounce filter settings 125 ns, 6.425 ps, 2.56 ms, disable; high and low transitions; selectable per input Bus Interface PCI/PXI 3.3 Vor 5 Vsignal environment USB USB 2.0 Hi-Speed or full-speed!”1. [8] DMA channels (PCI/PXI) 6, can be used for analog input, analog output, digital input, digital output, counter/timer 0, counter/timer 1 USB Signal Stream 4, can be used for analog input, analog output, digital input, digital output, counter/timer 0, counter/timer 1 The PXI device supports one of the following features: » May be installed in PXI Express hybrid slots = Or, may be used to control SCXI in PXI/SCXI combo chassis M Series Part Number | SCX! Control in PXI/SCXI Combo PX! Express Hybrid Slot Chassis Compatible 191501C-01 No Yes 191501C-11 Yes No © National instruments 21.
Table 5. PXI/SCXI Combo and PXI Express Chassis Compatibility
PCI/PXI/USB-6289 Specifications USB Power supply requirements 11 to 30 VDC, 20 W, locking or non-locking power jack with 0.080 in. diameter center pin, 5/16-32 thread for locking collars Power supply fuse 2A,250V Current Limits Caution Exceeding the current limits may cause unpredictable behavior by the device and/or PC/chassis. Pcl +5 V terminal (connector 0) 1A maximum!22) +5 V terminal (connector 1) 1Amaximum!22) PXI +5 V terminal (connector 0) 1Amaximum!20) +5 V terminal (connector 1) 1Amaximum!22] PO/PFI/P1/P2 and +5 V terminals combined 2Amaximum USB +5 Vterminal 1Amaximum!2) PO/PFI/P1/P2 and +5 V terminals combined 2Amaximum ©National instruments 23
PCI/PXI/USB-6289 Specifications Physical Characteristics Dimensions PXI printed circuit board Standard 3U PXI Mass Termination 18.8 cm x 17.09 cm x Screw Terminal (includes connectors) 26.67 cm x 17.09 cm x USB OEM Refer to the NI USB-622x/625x/628x OEM User Guide Weight PCI 167 g (5.9 oz) PXI 237 g (8.4 02) USB Mass Termination 1.06 kg (2 [b5.5 0) USB Screw Terminal 1.52 kg (3 [b5.5 02) USB OEM 274 g (9.6 02) I/O connectors Mass Termination 2 68-pin SCSI USB OEM 2 34-pin IDC, 2 50-pin IDC 24 nicom
PCI/PXI/USB-6289 Specifications Screw terminal wiring 16 AWG to 28 AWG Calibration Recommended warm-up time PCI/PXI/PCI Express/PXI Express 15 minutes USB 30 minutes Calibration interval 2 years Maximum Working Voltage Connect only voltages that are below these limits. Channel-to-earth 11V, Measurement Category | Measurement Category | is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics. | Caution Do not use for measurements within Categories Il, Ill, or IV. Note Measurement Categories CAT | and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV. © National Instruments 25
PCI/PXI/USB-6289 Specifications Environmental Operating temperature PCI/PXI 0°C to 55°C USB 0°C to 45°C Storage temperature -20 °C to 70 °C Humidity 10% RH to 90% RH, noncondensing Maximum altitude 2,000 m Pollution Degree (indoor use only) 2 Indoor use only. Shock and Vibration (PXI Only) Operational shock 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) Random vibration Operating 5 Hz to 500 Hz, 0.3 gems Nonoperating 5 Hz to 500 Hz, 2.4 gms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) 26 nicom
PCI/PXI/USB-6289 Specifications Safety Compliance Standards This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use: » IEC 61010-1, EN 61010-1 = UL61010-1, CSA C22.2 No. 61010-1 Note For safety certifications, refer to the product label or the Product Certifications and Declarations section. Electromagnetic Compatibility CE Compliance C € = 2011/65/EU; Restriction of Hazardous Substances (RoHS) Product Certifications and Declarations Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link. Environmental Management Nl is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. © National instruments 27
PCI/PXI/USB-6289 Specifications EU and UK Customers . ps4 Waste Electrical and Electronic Equipment (WEEE)—At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee. BTS as Ah BED (FE Rois) + 666 FH RoHS— NI HAPHRHESH ope RX AS [AHES(RoHS)o AF NIFH RoHS AMES, IER ni.com/environment/ rohs_chinao (For information about China RoHS compliance, go to ni.com/ environment/rohs_china.) 28 nicom
Figure 9. NI PCI/PXI-6289 Pinout
Figure 10. NI USB-6289 Mass Termination Pinout
Figure 11. NI USB-6289 Screw Terminal Pinout ‘AIGND sla] S22 AlSnD ‘AIGND 38| (Ql) Si] 54 ALGND. 1 An AO channel cannot be a reference or offset to itself.
2 When the USB Screw Terminal device is powered on, the analog output signal is
not defined until after the USB configuration is complete.
3 For all USB Screw Terminal devices, when powered on, the analog output signal is
not defined until after USB configuration is complete.
4 Stresses beyond those listed under Input voltage protection may cause
permanent damage to the device.
PCI/PXI/USB-6289 Specifications 5 The digital subsystem does not have its own dedicated internal timing engine. Therefore, a sample clock must be provided from another subsystem on the device or an external source. 6 In other sections of this document, RTSI refers to RTSI <0..7> for the PCI devices or PXI_TRIG <0..7> for PXI devices.
7 If you are using an USB M Series device in full-speed mode, device performance
will be lower and you will not be able to achieve maximum sample/update rates. 8 Operating on a full-speed bus may result in lower performance. 9 Does not include P0/PFI/P1/P2 and +5 V terminals.
10 Older revisions have a self-resetting fuse that opens when current exceeds this
specification. Newer revisions have a traditional fuse that opens when current exceeds this specification. This fuse is not customer-replaceable; if the fuse permanently opens, return the device to NI for repair.
1 Has a user-replaceable socketed fuse that opens when current exceeds this