DATASHEET SEARCH SITE | WWW.ALLDATASHEET.COM

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 6

Technical content

The Fluke 1587 and 1577 Insulation Multimeters combine a digital insulation tester with a full-featured, true-rms digital multimeter in a single compact, handheld unit, which provides maximum versatility for both troubleshooting and preventative maintenance. Like other tools that you have come to expect from Fluke, the 1587 and 1577 are rugged, reliable, and easy to use. Whether you work on motors, generators, cables, or switch-gear, the Fluke 1587/1577 Insulation Multimeters are ideally suited to help you with your tasks.

  • Large display with backlight
  • Insulation test (1587: 0.01 MΩ to 2 GΩ ) 1577: 0.1 MΩ to 600 MΩ )
  • Insulation test voltages (1587: 50 V, 100 V, 250 V,

500 V, 1000 V),(1577: 500 V, 1000 V) for many

applications

  • Live circuit detection prevents insulation test if volt- age > 30 V is detected for added user protection
  • Auto-discharge of capacitive voltage for added user protection
  • AC/DC voltage, DC milliVolts, AC/DC milliamps, Resistance (Ω ), Continuity
  • Filter for motor drive measurements (1587 only)
  • Capacitance, diode test, temperature, Min/Max, frequency (Hz) (1587 only)
  • Auto power off to save battery power
  • CAT III 1000 V, CAT IV 600 V measurement category
  • Included accessories: Remote probe, test leads and probes, alligator clips, (K-type thermocouple, 1587 only)
  • Accepts optional Fluke TPAK™ magnetic hanging system to free your hands for other work
  • Rugged, utility hard case allows you to bring everything you need for the job
  • 3-year warranty Insulation Multimeters Two powerful tools in one.

2 Fluke Corporation 1587/1577 Insulation Multimeters

Range Resolution 50 Hz to 60 Hz 60 Hz to 5000 Hz 11 kHz bandwidth 1587 lowpass filter Voltage Range Resolution 50 Hz to 60 Hz 60 Hz to 400 Hz 1577 accuracy Range Resolution 50 Hz to 60 Hz

1000 V 1 V + (2 % + 3)

AC conversion: Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are true-rms responding and spec ified from 5 % to 100 % of range. Input signal crest factor can be up to 3 at full scale up to 500 V, decreasing linearly to crest factor < = 1.5 at 1000 V. For non-sinusoidal waveforms add ± (2 % reading + 2 % FS) typical, for a crest factor up to 3. Input impedance: 10 MΩ (nominal), < 100 pF, ac-c oupled Common mode rejection ratio ( 1 kΩ unbalanced): > 60 dB at dc, 50 or 6 0 Hz Overload protection: 1000 V rms or dc, 107 V Hz Max DC voltage measurement Range Resolution Accuracy 15871 Accuracy 15771 6.000 V dc 0.001 V 0.09 % + 2 0.2 % + 2 60.00 V dc 0.01 V 0.09 % + 2 0.2 % + 2 600.0 V dc 0.1 V 0.09 % + 2 0.2 % + 2 1000 V dc 1 V 0.09 % + 2 0.2 % + 2 1Accuracies apply to ± 100 % of range Input impedance: 10 MΩ (nominal), < 100 pF Normal mode rejection ratio: > 60 dB @ 50 Hz or 60 Hz Common rode rejection ratio: > 120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance) Overload protection: 1000 V rms or dc DC millivolts measurement Range Resolution Accuracy 1587 Accuracy 1577 600.0 mV dc 0.1 mV 0.1 % + 1 0.2 % + 1 1587/1577 Specifications

1587/1577 Insulation Multimeters Fluke Corporation 3 DC and ac current measurement Accuracy 1587 Accuracy 1577 Burden Voltage Range Resolution ± (% of Rdg+Digits) ± (% of Rdg+Digits) (Typical) AC 400 mA 0.1 mA ± (1.5 % + 2)1 ± (2 % + 2)1 2 mV/mA 45 to 1000 Hz 60 mA 0.01 mA ± (1.5 % + 2)1 ± (2 % + 2)1 11 kHz bandwidth Overload: 600 mA for 2 minutes maximum Overload protection: 440 mA, 1000 V, FAST fuse AC conversion: Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are true-rms responding and spec ified from 5 % to 100 % of range. Input signal crest factor can be up to 3 at full scale up to 300 mA, decreasing linearly to crest factor < = 1.5 at 600 mA. For non-sinusoidal waveforms add + (2 % reading + 2 % FS) typical, for a crest factor up to 3. Ohms measurement Accuracy 15871 Accuracy 15771 Range Resolution ± (% of Rdg+Digits) ± (% of Rdg+Digits) 600.0 Ω 0.1 Ω 6.000 kΩ 0.001 kΩ 60.00 kΩ 0.01 kΩ 0.9 % + 2 1.2 % + 2 600.0 kΩ 0.1 kΩ 6.000 MΩ 0.001 MΩ 50.0 MΩ 0.01 MΩ 1.5 % + 3 2.0 % + 3 1Accuracies apply from 0 to 100 % of range Overload protection: 1000 V rms or dc Open circuit test voltage: < 8.0 V dc Short circuit current: < 1.1 mA Diode test (1587 Only) Diode test indication: Display voltage drop: 0.6 V at 1.0 mA nominal test current Accuracy: + (2 % + 1) Continuity test Continuity indication: Continuous audible tone for test resistance below 25 Ω and off above 100 Ω . Maximum reading; 1 000 Ω Open circuit voltage: < 8.0 V Short circuit current: 1.0 mA typical Overload protection: 1000 V rms Response time: > 1 m sec Frequency measurement (1587 only) Accuracy Range Resolution ± (% of Rdg+Digits) 9.999 kHz 0.001 kHz ± (0.1 % + 1) 99.99 kHz 0.01 kHz ± (0.1 % + 1)

4 Fluke Corporation 1587/1577 Insulation Multimeters

1587/1577 Specifications cont. Frequency counter sensitivity V ac Sensitivity (RMS Sinewave)1 DC Trigger Levels Input Range 5 Hz to 20 kHz 20 kHz to 100 kHz to 20 kHz2 600.0 mV ac 150.0 mV 150.0 mV N/A 1Maximum input for specified accuracy = 10x range (1000 V max). Noise at low frequencies and amplitudes may affect accuracy. 2Usable to 100 kHz with full scale input. Capacitance (1587 only) Range Resolution ± (% of Rdg+Digits) 1000 nF 1 nF 100.0 µF 0.1 µF 9999 µF 1 µF ± (1.2 % +/- 90 counts) Temperature measurement (1587 only) Range Resolution Accuracy1 -40 ° C to 537 °C 0.1 °C 1 % + 10 counts -40 °F to 998 °F 0.1 °F 1 % + 18 counts 1Accuracies apply following 90 minutes settling time after a change in the ambient temperature of the instrument Insulation specifications Measurement range: 1587: 0.01 MΩ to 2 GΩ , 1577: 0.1 MΩ to 600 MΩ Test voltages: 50, 100, 250, 500, 1000 V model 1 587, 500 and 1000 V model 1577 Test voltage accuracy: + 20 %, - 0 % Short-circuit test current: 1 mA nominal Auto discharge: Discharge time < 0.5 second for C = 1 µ F or less Live circuit detection: Inhibit test if terminal voltage > 30 V prior to initialization of test Maximum capacitive load: Operable with up to 1 µF load

1587/1577 Insulation Multimeters Fluke Corporation 5 Model 1587 Output Voltage Display Range Resolution Test Current Resistance Accuracy 100 V 0.01 to 6.00 MΩ 0.01 MΩ (0 % to + 20 %) 6.0 to 60.0 MΩ 0.1 MΩ 1 mA @ 100 kΩ 3 % + 5 counts 60 to 100 MΩ 1 MΩ 250 V 0.1 to 60.0 MΩ 0.1 MΩ 1 mA @ 250 kΩ 1.5 % + 5 counts(0 % to + 20 %) 60 to 250 MΩ 1 MΩ 500 V 0.1 to 60.0 MΩ 0.1 MΩ 1 mA @ 500 kΩ 1.5 % + 5 counts(0 % to + 20 %) 60 to 500 MΩ 1 MΩ 1000 V 0.1 to 60.0 MΩ 0.1 MΩ 1.5 % + 5 counts (0 % to + 20 %) 60 to 600 MΩ 1 MΩ 1 mA @ 1 MΩ 0.6 to 2.0 GΩ 100 MΩ 10 % + 3 counts Model 1577 Output Voltage Display Range Resolution Test Current Resistance Accuracy 500 V 0.1 to 60.0 MΩ 0.1 MΩ 1 mA @ 500 kΩ 2.0 % + 5 counts (0 % to + 20 %) 60 to 500 MΩ 1 MΩ 1000 V 0.1 to 60.0 MΩ 0.1 MΩ 1 mA @ 1 MΩ 2.0 % + 5 counts (0 % to + 20 %) 60 to 600 MΩ 1 MΩ Maximum voltage applied to any terminal: 1000 V ac rms or dc Storage temperature: -40 °C to 60 °C (-40 °F to 140 °F) Operating temperature: -20 °C to 55 °C (-4 °F to 131 °F) Temperature coefficient: 0.05 x (specified accuracy) per °C for temperatures < 18 ° C or > 28 ° Relative humidity, non-condensing:< °C 0 % to 95 % @ 1 0 °C to 30 °C (50 °F to 86 °F) 0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F) 0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F) Vibration: Random, 2 g, 5-500 Hz per MIL -PRF-28800F, Class 2 instrument Shock: 1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor) Electromagnetic compatibility: In an RF field of 3 V/M, accurac y = specified accuracy except in temperature: specified accuracy ± 5 °C (9 °F). (EN 61326-1:1997) Safety: Complies with ANSI/ISA 82.02.01 (61010-1) 2004, C AN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement CAT III 1000 V and CAT IV 600 V Certifications: CSA per standard CSA/CAN C22.2 N o. 61010.1-04; TUV per standard EN 61010 Part 1-1002 Batteries: Four AA batteries (NEDA 15A or IEC LR6) Battery life: Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of

000 V into 1 MΩ with a duty cycle of 5 seconds on and 25

W x 8.00 in L) Weight: 550 g (1.2 lb) IP rating: IP40 Altitude (operating): 2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V, CAT I II 600 V Storage: 12,000 m Over-range capability : 11 0 % of range except for capacitance which is 1 % Compliance to EN 61557:IEC61557-1, IEC61557-2 1587/1577 General Specifications

For additional accessories for these products, see the Top 10 Accessories for Insulation Multimeters at www.fluke.com/top10 or ask for Fluke literature c ode 2427793.

6 Fluke Corporation 1587/1577 Insulation Multimeters

PO Box 9090, Everett, WA USA 98206 Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The N etherlands For more information call: In the U.S.A. (800) 443-5853 or F ax (425) 446-5116 In Europe/M-East/Africa (31 40) 2 675 200 or Fax (31 40) 2 675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com/ ©2002 Fluke Corporation. All rights reserved. Printed in U.S.A. 4/2005 2434792 D-US-N Rev B Fluke.Keeping your world up and running Comparison Chart 1587 1577 Insulation test voltages 50 V, 100 V, 250 V, 500 V, 1000 V • Insulation test voltages 500 V, 1000 V • Insulation test: 0.01 MΩ to 2.0 GΩ • Insulation test: 0.1 MΩ to 600 MΩ • Auto-discharge of capacitive voltage •• Insulation test smoothing reading • Frequency • Capacitance • Diode test • Temperature • Min/Max • Low pass filter (for work on VSDs) • AC/DC Voltage •• DC Millivolts •• AC/DC milliAmps •• Resistance (0.1 Ω to 50 MΩ) •• Continuity •• Three-year warranty •• Remote probe, test leads, alligator clips •• K-type thermocouple • Rugged, utility hard case •• Auto power off • •

Ordering Information

Remote probe, test leads, alligator clips, K -type thermo- couple (1587 only), hard case, user documentation Optional accessories TPAK Magnetic Tool Hanger i400s Clamp with Adapter C25 Soft Case 1577 1587 TPAK i400s C25