TB-4310 NI | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 24
Technical content
GETTING STARTED GUIDE AND TERMINAL BLOCK SPECIFICATIONS PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) 8-channel, 16-bit, 400 kS/s/ch, Ch-Ch Isolated Analog Input Module This document explains how to install, configure, and set up the PXIe-4310 module with the TB-4310 (10V)/TB-4310 (600V) terminal block. Driver support for the PXIe-4310 was first available in NI-DAQmx 17.1. For the list of devices supported by a specific release, refer to the NI-DAQmx Readme, available on the version-specific download page or installation media. Note The keying of this terminal block prevents it from connecting to other modules that could be damaged by the voltage present on the terminal block. However, you should only use these terminal blocks with their supported modules. Caution Do not supply hazardous voltages (>30 Vrms/42.4 Vpk/60 VDC) to the terminal block without the terminal block being connected to the PXIe-4310. Caution This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash.
Contents
2 | ni.com | PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) Electromagnetic Compatibility Guidelines This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic environment. This product is intended for use in industrial locations. However, harmful interference may occur in some installations, when the product is connected to a peripheral device or test object, or if the product is used in residential or commercial areas. To minimize interference with radio and television reception and prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation. Furthermore, any modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules.
PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) | © National Instruments | 3 Caution To ensure the specified EMC performance, operate this product only with shielded cables and shielded accessories. Use only twisted, shielded pair cables (Belden 83319 or equivalent) for channel connections. Refer to the PXIe-4310 User Manual for specific guidance on terminating cable shields. Safety Guidelines for Hazardous Voltages If hazardous voltages are connected to the module, take the following precautions. A hazardous voltage is a voltage greater than 42.4 Vpk or 60 VDC to earth ground. Caution Ensure that hazardous voltage wiring is performed only by qualified personnel adhering to local electrical standards. Caution You must install mating connectors according to local safety codes and standards and according to the specifications provided by the manufacturer. You are responsible for verifying the safety compliance of third-party connectors and their usage according to the relevant standard(s), including UL and CSA in North America and IEC and VDE in Europe. Caution Make sure that devices and circuits connected to the module are properly insulated from human contact. Caution When module terminals are hazardous voltage LIVE (>42.4 Vpk /60 VDC), you must ensure that devices and circuits connected to the module are properly insulated from human contact. Caution Do not mix hazardous voltage circuits and human-accessible circuits on the same module. What Y ou Need to Get Started To set up and use a PXIe-4310 module with a TB-4310 (10V)/TB-4310 (600V) terminal block, you need the following items: Hardware – PXIe-4310 module – TB-4310 (10V)/TB-4310 (600V) terminal block – PXI Express chassis – Cabling and sensors as re quired for your application Tools – Number 1 and 2 Philli ps-head screwdrivers – 1/8 in. flathead screwdriver – Long-nose pliers
4 | ni.com | PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) – Wire cutter – Wire insulation stripper Documentation – PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) Getting Started Guide and Terminal Block Specifications – PXIe-4310 User Manual – PXIe-4310 Specifications – PXI Express chas sis user manual You can download needed documents from ni.com/manuals. Installation Install the Software Note You must be an administrator to install NI software and devices on your computer. Ensure that the following software is installed before installing the PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) hardware: 1. Your application software, such as LabVIEW, LabWindows™/CVI™, or .NET. 2. NI-DAQmx—PXIe-4310 modules were first supported in NI-DAQmx 17.1. Note For detailed NI software version support, refer to the NI-DAQmx Readme. Back up any applications before upgrading software or modifying the application. Unpack and Install the Module Remove the packaging and inspect the module. Contact NI if the module is damaged. Do not install a damaged module. Caution The module is static sensitive. Always properly ground yourself and the equipment when handling or connecting to the module. Complete the following steps to install the PXIe-4310 module while referring to Figures 1 and 2: Note To maintain forced air cooling in the PXI Express system, refer to the Maintain Forced-Air Cooling Note to Users. 1. Plug in your chassis before ins talling the PXIe-4310. The power cord grounds the chassis and protects it from electrical damage while you install the module. 2. Make sure the cha ssis is powered off.
- Touch a metal part on the chassis to discharge any accumulated static electricity.
- Identify a supported PXIe slot in the chassis. PXIe-4310 modu les can be placed only in the
various slot types. Refer to your chassis documentation for more details. Figure 1. PXIe/PXI Chassis Slot Type Symbol Location
- Remove the filler pa nel, and touch any metal part of the chassis to discharge static
- Place the module edges into the module guides at the top and bottom of the slot.
- Slide the module along the gui des until it reaches the rear connector then seat the module
by pushing the front panel until it is flush with the front panel of the chassis.
1 Slot Type Symbols Locations
Table 1. PXIe-4310 Supported Slot Types
- Secure the module to the chassis using the front-panel captive screws, shown in Figure 2.
Tighten the screws to 0.31 N · m (2.7 lb · in.).
- Install a filler pan el in any empty slots.
Figure 2. Installing PXIe-4310 Modules terminal block without the terminal block being connected to the PXIe-4310. Interfaces and selecting Device Pinouts.
- Loosen the captive top cover screws and remove the top cover.
- Loosen the strain-relief screw s and remove the strain-relief bar.
- Prepare the shielded signal wi re by stripping the insulation no more than 8 mm (0.24 in.).
- Run the shielded signal wires through the strain-relief opening.
Figure 3. TB-4310 (10V)/TB-4310 (600V) Parts Locator Diagram
- Insert the stripped end of the shielded signal wires fully into the appropriate terminal. Refer
1 Strain-Relief Screws
2 Strain-Relief Bar
3 Tie Wrap Holes
4 Ground Lugs
5 Terminal Block to Module Connector
6 Captive Top Cover Screws
8 Tie Wraps
9 Shielded Signal Wires
Figure 4. TB-4310 (10V)/TB-4310 (600V) Circuit Board Parts Locator Diagram connect sensors to the TB-4310 (10V)/TB-4310 (600V). Note Refer to the PXIe-4310 User Manual for details about shielding the signals.
- Reinstall the strain-relief b ar and tighten the strain-relief screws.
- Use tie wraps to c onnect the shielded signal wires to the tie-wrap holes for additional strain
- Reinstall the top c over and tighten the captive top cover screws.
- Move the terminal bl ock into position in front of the module and engage the alignment
feature with the guide on the associated module. Figure 5. Installing the Terminal Block on the Module
- Attach the termina l block to the module by pushing the terminal block straight into the
module. A spring mechanism will lock in the bottom of the terminal block.
- Tighten the mounting screw at the top of the terminal block to attach it to the module.
1 PXI Express Controller
2 PXI Express Chassis
3 Mounting Screw
4 Terminal Block
5 Module
6 Alignment Feature
COOLING CLEARANCE AND FAN FILTER MAINTENANCE REQUIRED. SEE MANUAL. COOLING CLEARANCE AND FAN FILTER MAINTENANCE REQUIRED. SEE MANUAL.
10 | ni.com | PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) Confirm PXIe-4310 Module Recognition To confirm module recognition, complete the following steps: 1. Power on the system. 2. Launch MAX. 3. Expand Devices and Interfaces, and expand the chassis in which the module is located to confirm that MAX detects the module and terminal block. The terminal block should appear beneath its associated module. If your module or terminal block is not listed, press <F5> to refresh MAX. If the module is still not recognized, go to ni.com/support/ daqmx.
PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) | © National Instruments | 11 Tip Using an NI-DAQmx simulated device, you can test NI-DAQmx applications without installing hardware. Refer to the Create a Simulated Device section for instructions for creating NI-DAQmx simulated devices. 4. Right-click the modul e name and select Self-Test. When the self-test completes, a verification message appears. If an error occurs, refer to ni.com/support/daqmx. Device Self-Calibration NI recommends that you self-calibrate your PXIe-4310 module after installation and whenever the ambient temperature changes. Self-calibration should be performed after the device has warmed up for the recommended time period. Refer to the device specifications to find your device warm-up time. This function measures the onboard reference voltage of the device and adjusts the self-calibration constants to account for any errors caused by short-term fluctuations in the environment. You can initiate self-calibration using Measurement & Automation Explorer (MAX), by completing the following steps: 1. Launch MAX. 2. Select My System»Devices and Interfaces»your device. 3. Initiate self-c alibration using one of the following methods:
- Select the PXIe-4310 in the MAX configuration tree and click the Self-Calibrate button located on the MAX toolbar.
- Right-click the name of the dev ice in the MAX configuration tree and select Self-Calibrate from the drop-down menu. Note You can also programmatically self-calibrate your device with NI-DAQmx, as described in Device Calibration in the NI-DAQmx Help or the LabVIEW Help.
12 | ni.com | PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) Run Test Panels 1. In MAX, expand Devices and Interfaces select the chassis in which the module is located. 2. Right-click the d evice, and select Test Panels. 3. Click Start to test device functions, or Help for operating instructions. To troubleshoot errors, refer to the NI-DAQmx Help, or go to ni.com/support. Take an NI-DAQmx Measurement NI-DAQmx Channels and Tasks Refer to the NI-DAQmx Help for complete information about channels and tasks. Use the DAQ Assistant to configure virtual channels and tasks in MAX or in your application. Configure a Task Using the DAQ Assistant in MAX Complete the following steps to create a task using the DAQ Assistant in MAX: 1. In MAX, right-click Data Neighborhood and select Create New to open the DAQ Assistant. 2. Select NI-DAQmx Task and click Next. 3. Select Acquire Signals. 4. Select the measurement type, s uch as Analog Input - V oltage. 5. Select the phys ical channel(s) to use and click Next. 6. Name the task and click Finish.
- Configure the individual cha nnel settings. Each physical channel you assign to a task
triggering for your task. Click Run. All Programs»National Instruments»NI-DAQmx»NI-DAQmx Documentation. NI-DAQmx includes example programs to help you get started developing an application. LabVIEW and CVI examples are located at Help»Find Examples in your application software. develop a new application or add example code to an existing application. For other examples, go to ni.com/info and enter the Info Code daqmxexp. Table 2. DAQ Assistant Tutorial Locations Programs»National Instruments»NI SignalExpress.
Note The chassis can be powered on or off when removing the terminal block.
- Ensure that the devices and ci rcuits connected to the terminal block are powered down and
are no longer hazardous live.
- Loosen the terminal block mounting screw located at the top of the terminal block.
- To remove the terminal block:
a. Raise the latch release using a flathead screwdriver to disen gage the latch. Figure 6. Removing the Terminal Block from the Module
- Store the terminal block in an antistatic protective bag.
1 Mounting Screw
2 Latch Release
3 PXI Express Controller
4 PXI Express Chassis
5 Terminal Block
COOLING CLEARANCE AND FAN FILTER MAINTENANCE REQUIRED. SEE MANUAL.
Maintain Forced-Air Cooling Note to Users.
- Make sure that the c hassis is powered off before removing the module.
- Loosen the captive screws on the top and bottom of the module.
- Pull the captive screws to uns eat the module and slowly slide the module along the guides.
- Install a filler panel in the empty slot.
- Store the module in an an tistatic protective bag.
Figure 7. Removing PXIe-4310 Modules
- Right click Devices and Interfaces»Create New.
- From the dialog, select Simulated NI-DAQmx Device or Modular Instrument and click
- Type 4310 in the text box at the top of the window.
- Select the device from the list provided.
- Right-click the sim ulated PXIe module and select Configure.
- Under Connector 0 on the Access ory tab, select the desired terminal block to simulate and
Figure 8. TB-4310 (10V)
Figure 9. TB-4310 (600V) resources are online at ni.com/gettingstarted. documents for supported modules are included on the NI-DAQmx media.
- Go to ni.com/kb.
- If you need to return your Natio nal Instruments hardware for repair or device calibration, go to ni.com/info and enter rdsenn to start the Return Merchandise Authorization (RMA) process. Specifications All specifications are for both the TB-4310 (10V) and the TB-4310 (600V) unless otherwise noted. Note PXIe-4310 module specifications are located in the PXIe-4310 Specifications document. All performance specifications are typical unless otherwise noted. These specifications are valid within the full operating temperature range.
18 | ni.com | PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) Calibration Electrical Analog input ranges Input impedance (device on) TB-4310 (600V) Typical Accuracy (TB-4310 600V) Stability Nominal Range Residual Gain Error (ppm of Reading) Residual Offset (ppm of Range) Offset Tempco (ppm of Range/°C) Random Noise σ (mVrms) Typical Accuracy at Full Scale (mV) Positive Full Scale Negative Full Scale 600 -600 335 16 9 10.3 298 300 -300 336 21 9 5.5 149 120 -120 339 42 10 2.9 61 60 -60 345 75 13 2.3 33
PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) | © National Instruments | 19 AI Typical Accuracy Equation Accuracy = Reading * (GainError) + Range * (Offset Error) + Self Heating + Noise Uncertainty Gain Error = Residual Gain Error + Gain Tempco * (TempChangeFromLastInternalCal) + Reference Tempco * (TempChangeFromLastExternalCal) Offset Error = Residual Offset Error + Offset Tempco * (TempChangeFromLastInternalCal) + INL_Error Self Heating = If Reading > 120 V , Self-Heating * (Reading - 120 V), otherwise 0. Noise Uncertainty = for a covera ge factor of 3 σ and averaging 10000 points. Typical Accuracy Example Accuracy at full scale is determined using the following assumptions:
- TempChangeFromLastExternalCal = 10 °C
- TempChangeFromLastInternalCal = 1°C
- Number_of_readings = 10000
- CoverageFactor = 3 σ For example, on the 600 V range, the accuracy at full scale is as follows: Gain Error = 335 ppm + 15 ppm * 1 °C + 5 ppm * 10 °C = 400 ppm Offset Error = 16 ppm + 9 ppm * 1 °C + 51 ppm = 76 ppm Self Heating = 26 µV/V * (600 V - 120 V) = 12.48 mV Noise Uncertainty = = 309 µV Typical Accuracy = 600 V * (Gain Error) + 600 V * (Offset Error) + Self Heating + Noise Uncertainty = 298 mV Random Noise * 3 10000 10.3 mV * 3 10000
Figure 10. Terminal Block Dimensions hardware is completely dry and free from contaminants before returning it to service.
PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) | © National Instruments | 21 Storage Environment (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.) (Tested in accordance with IEC 60068-2-56.) Shock and Vibration (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.) Random vibration (Tested in accordance with IEC 60068-2-64. Non-operating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) Safety Voltages Maximum Input Voltage Connect only voltages that are within the following limits. TB-4310 (10V) TB-4310 (600V)
300 Vrms, Measurement Category II
22 | ni.com | PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) Isolation Voltages Caution Do not connect the TB-4310 (10V)/TB-4310 (600V) to signals or use for measurements within Measurement Categories III, or IV . Caution The protection provided by the TB-4310 (10V)/TB-4310 (600V) can be impaired if it is used in a manner not described in this document. Safety Standards This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1 Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Gr oup 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Gr oup 1, Class A emissions 1 Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV . MAINS is a hazardous live electrical supply system that powers equipment. Measurement Category O is for measurements of voltages from specially protected secondary circuits, such as signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics. Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 120 V for U.S. or 240 V for Europe.
PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) | © National Instruments | 23
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B : Class A emissions
- ICES-001: Class A emissions Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations. Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section. CE Compliance This product meets the essential requirements of applicable European Directives as follows:
- 2014/35/EU; Low-V oltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC) Online Product Certification Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
© 2017 National Instruments. All rights reserved. 377029B-01 May17 Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste and Electronic Equipment, visit ni.com/environment/ weee. Worldwide Support and Services The NI website is your complete resource for technical support. At ni.com/support you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers. Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and other services. Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI. A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European Communities using the manufacturer’s declaration of conformity. This system affords the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC for your product by visiting ni.com/certification. If your product supports calibration, you can obtain the calibration certificate for your product at ni.com/calibration. NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. ⊩ ˄Ё RoHS˅ Ёᅶ᠋ National InstrumentsՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ (RoHS)Ѣ National Instruments ЁRoHSᙃˈ䇋ⱏᔩ ni.com/ environment/rohs_chinaDŽ(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)