TLP150 ADVANCEDENERGY | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 7

Technical content

©2022 Advanced Energy Industries, Inc. ARTESYN TLP150 MEDICAL SERIES Single Output Total Power: 150 W Input Voltage: 85 to 264 VAC # of Outputs: Single DATA SHEETAdvanced Energy's Artesyn TLP150 series of open-frame AC-DC power supplies comprises two single output models, offering a main voltage of 12 V or 24 V. Each model also provides 12 V fan and 5 V standby outputs. Both models feature medical safety approvals and accept a universal input of 85 to 264 VAC. Comprehensive overcurrent, overvoltage and overtemperature protection is provided as standard. Additional features include main output remote sense, active current sharing and integrated control and monitoring facilities. TLP150-M series power supplies provide 100 W of output power with free air convection cooling and 150 W with 200 LFM of forced air. Less than 1U high, they are ideal for systems that implement distributed power architectures (DPA) and point-of-load (POL) schemes, as well as for powering electromechanical devices and non-patient contact and non-patient critical medical applications. SPECIAL FEATURES  150 W on main channel with only

200 LFM

 Low profile fits 1U applications  Active PFC and EN61000-3-2 compliant  Integrated control and monitoring

features

 Overcurrent, overvoltage and overtemperature protection  Compliance to EN55022-B conducted noise standard  12 V fan output  5 V standby output (optional)  RoHS compliant  Two years warranty SAFETY  TUV EN60601-1, IEC60601-1  VDE CB Certificate and Report  UL 62368-1  UKCA Mark

2 advancedenergy.com TLP150M ELECTRICAL SPECIFICATIONS Conducted emissions EN55022, FCC part 15 Level B Harmonic current correction EN61000-3-2 Compliant ESD air EN61000-4-2 Level 3 ESD contact EN61000-4-2 Level 3 Radiated immunity EN61000-4-3 Level 3 Fast transients EN61000-4-4 Level 4 Surge EN61000-4-5 Level 3 Conducted immunity EN61000-4-6 Level 3 EMC CHARACTERISTICS (5) Input Input voltage range Universal Input 85 to 264 VAC Input frequency range 47 to 63 Hz Input surge current 264 VAC (cold start) 40 A max. Safety ground leakage current 264 VAC, 50 Hz 150 μA Input current 120 VAC @ 150 W

230 VAC @ 150 W

1.8 A rms

0.8 A rms

Input fuse: UL/IEC127 T 3.15 A, 250 VAC Output Adjustment range ± 10% Total regulation (line and load) Main output Auxiliary outputs Fan output ± 3% ± 5% ± 10% Turn-on delay @120 VAC Input 2.0 s max.. Transient response Main output 25% to 75% step at 0.5 A/μs max. dev., 1 ms max. recovery to 1% Temperature coefficient ±0.02%/°C Overvoltage protection Main outputs 125%, ± 5% Short circuit protection Current limited Continuous Minimum output current Singles 0 A Fan voltage output See Note 9 12 V @ 0.5 A Standby output See Note 9 5 V @ 1.0 A

advancedenergy.com 3 TLP150M Hold-up time 850 VAC @ 60 Hz 20 ms @ 150 W Efficiency 115 VAC @ 150 W 81% typ. 84% typ. Isolation voltage Input/output Input/chassis

4000 VAC

1500 VAC

Weight 260 g (9.2 oz) MTBF (@ 25 ºC) Telcordia SR-332 MIL-HDBK-217F 900,000 hours min. 350,000 hours min. GENERAL SPECIFICATIONS Thermal performance Operating ambient, (See derating curve) 0 °C to +70 °C Non-operating -40 °C to +85 °C 0 °C to 50 °C ambient, 200 LFM forced air 150 W 0 °C to 50 °C ambient, convection cooled 100 W 50 °C to 70 °C ambient, convection cooled Derate linearly to 50% load Relative humidity Non-condensing 5 to 95% RH Altitude Operating 10,000 feet max. Non-operating 30,000 feet max. Vibration (See Note 7) 5 - 500 Hz 2.4 G rms peak Shock per MIL-STD-810E 516.4 Part IV ENVIRONMENTAL SPECIFICATIONS

4 advancedenergy.com TLP150M

ORDERING INFORMATION

Ripple Total Regulation Model Numbers (8, 10) Min Max (free air) (1, 4) Max (forced air) (2, 4) 12 V 0 A 8.4 A 12.5 A 120 mV ± 3.0% TLP150N-99S12J 24 V 0 A 4.2 A 6.3 A 240 mV ± 3.0% TLP150N-99S24J Notes: 1. Free air convection. Maximum continuous output power not to exceed 100 W. Refer to Figure 1 for the derating curve. 2. 200 LFM forced air cooling from the ac input side. Maximum continuous output power not to exceed 150 W. 3. Figure is peak-to-peak for room temperature rating. Output noise measurements are made across a 20 MHz bandwidth using a 6 inch twisted pair, terminated with a 10 μF tantalum capacitor and a 0.1 μF ceramic capacitor. 4. CAUTION: Allow a minimum of 1 second after disconnecting line power when making thermal measurements. For optimum reliability no part of the heatsink should exceed 115 °C and no semi-conductor case temperature should exceed 120 °C. 5. No external filtering required during conducted emissions testing but some applications may require additional filtering to achieve system compliance. Compliance with radiated EMI specifications may require mounting in a suitable enclosure. 6. This product is only for inclusion by professional installers within other equipment and must not be operated as a stand alone product. 7. Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G. 8. Replace the ‘J’ at the end of the model number with ‘FJ’ when the optional standby output and/or remote ON/OFF control is required e.g. TLP150N-99S12FJ. 9. 12 V (fan) present when main output is present. An optional 5 Vsb (standby) output is available whenever ac input is present, regardless of remote ON/OFF signal status. 10. The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant. 11. NOTICE: Some models do not support all options. Please contact your local Artesyn representative or use the on-line model number search tool at http://www.artesyn.com to find a suitable alternative. 12. Power good signal required 100 mA load on the main output. 13. This product is a Component Power Supply and is only for inclusion by professional installers within other equipment and must not be operated as a standalone product. EMC compliance to appropriate standards must be verified at the system level. This product is for sale to OEMs and System Integrators, including through Distribution Channels. It is not intended for sale to End Users. Figure 1: Derating Curve DERATING CURVE Output Power (Watts)

0 C 10 C 20 C 30 C 40 C 50 C 60 C 70 C

200 LFM (1.0 m/s) FORCED AIR COOLING NATURAL CONVECTION COOLING 50 W 75 W 0 W Dimensions in Inches (mm) ø0.156 (4 Places) 0.225 (5.72) 0.225 (5.72) FAN 4.550 (115.57) 5.000 (127.00) 3.000 (76.20) 2.550 (64.77) 1.250 (31.75) Outline Dimensions

1 J301

Figure 2: Mechanical Drawing

advancedenergy.com 5 TLP150M CONNECTOR AND MATING CONNECTOR TYPES Connector Type Mating Connector Type J1 Molex 09-65-2038 (5273 series) void pin 2 or equivalent Molex 09-52-4034 (5239 series) or equivalent Molex 08-52-0072 (2478 series) or equivalent crimp terminals J3 Molex 09-65-2068 (5273 series) or equivalent Molex 09-52-4064 (5239 series) or equivalent Molex 08-52-0072 (2478 series) or equivalent crimp terminals J301 (Optional) Leoco 2421P04H000 (2421 series) or equivalent Leoco 2420S04000 (2420 series) or equivalent Leoco 2453TPB00V1 (2453T series) or equivalent crimp terminals or JST EHR-4 (EH series) or equivalent JST SEH-001T-P0.6 (EH series) or equivalent crimp terminals Fan Molex 22-23-2021(6373 series) or equivalent Molex 22-01-3027 (2695 series) or equivalent Molex 08-50-0113 (2759 series) or equivalent crimp terminals PIN CONNECTIONS J1 Pin Connections Pin 1 Neutral Pin 3 Live J2 Tab Connections Tan Ground/Earth

6 advancedenergy.com TLP150M PIN CONNECTIONS (CONTINUED) J3 Pin Connections Pin 1 RTN Main Return Pin 2 RTN Main Return Pin 3 RTN Main Return Pin 4 Vo +Main Output Pin 5 Vo +Main Output Pin 6 Vo +Main Output J4 Pin Connections Pin 1 -S -Vo Remote Sense Pin 2 DC OK DC Power Good Signal Pin 3 PW OK Power Good Pin 4 LS Load Share Signal Pin 5 +S +Vo Remote Sense Pin 6 SGND Signal Common J301 Pin Connections (Optional) Pin 1 5 Vsb Standby Voltage Pin 2 SGND Signal Common Pin 3 Reserved Do Not Connect Pin 4 PS OFF Remote ON/OFF Signal Fan Pin Connections Pin 1 + 12 V Fan Voltage Pin 2 +SGND Return

ENG-TLP150M-235-01 4.20 For international contact information, visit advancedenergy .com. powersales@aei.com (Sales Support) productsupport.ep@aei.com (Technical Support) +1 888 412 7832 ABOUT ADVANCED ENERGY Advanced Energy (AE) has devoted more than three decades to perfecting power for its global customers. AE designs and manufactures highly engineered, precision power conversion, measurement and control solutions for mission-critical applications and processes. Our products enable customer innovation in complex applications for a wide range of industries including semiconductor equipment, industrial, manufacturing, telecommunications, data center computing, and medical. With deep applications know-how and responsive service and support across the globe, we build collaborative partnerships to meet rapid technological developments, propel growth for our customers, and innovate the future of power. Specifications are subject to change without notice. Not responsible for errors or omissions. ©2022 Advanced Energy Industries, Inc. All rights reserved. Advanced Energy®, AE® and Artesyn™ are U.S. trademarks of Advanced Energy Industries, Inc.