TRV2M TRACOPOWER

Document overview

  • PDF pages: 5

Technical content

Page 1

DC/DC Medical Converter

  • I/O isolation 5000 VAC (reinforced)
  • Wide 1.5:1 input voltage range
  • Operating temperature range -40 to +80 °C without derating
  • Certification according to IEC/EN/ES

ES 60601-1 IEC 60601-1 UL 62368-1 IEC 62368-1

60601-1 3rd edition for 2xMOPP and

IEC/EN/UL 62368-1

The TRV 2M is a series of 2 Watt DC/DC converters in a compact SIP-9 package with reinforced isolation of 5000 VAC for medical and industrial applica-

  • Short circuit protection

tions. The series offers a 1.5:1 input voltage range with a nominal input ranging

  • Regulated outputs

between 5 and 24 VDC. With a continuous short circuit protection and a low leakage current of less than 2 µA, this converter is especially suited to protect

  • Low leakage current < 2 μA

any connected interfaces or applied parts to patients. Featuring almost fully regulated outputs this series provides great level of regulation without affecting

  • Efficiency up to 84 %

the cost efficiency. It is an ideal solution for applications where an unregulated

  • Operation up to 5000 m altitude

DC/DC converter would not meet your regulation requirements but cost still is a critical factor. With an operating temperature range from -40 to +80°C without

  • 5-year product warranty

derating and certifications acc. IEC/EN/ES 60601-1 3rd ed. for 2xMOPP and IEC/EN/UL 62368-1 this series is suitable for many different applications where a medical isolation system and short circuit protection is needed.

Models

Order Code Input Voltage Range Output 1 Vnom Imax Output 2 Vnom Imax Efficiency typ.
TRV 2-0510M TRV 2-0511M TRV 2-0512M TRV 2-0513M TRV 2-0521M TRV 2-0522M TRV 2-0523M 4.5 - 7 VDC (5 VDC nom.) 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC 600 mA 400 mA 167 mA 134 mA 200 mA 83 mA 67 mA –5 VDC –12 VDC –15 VDC 200 mA 83 mA 67 mA 78 % 81 % 83 % 83 % 82 % 83 % 81 %
TRV 2-1210M TRV 2-1211M TRV 2-1212M TRV 2-1213M TRV 2-1221M TRV 2-1222M TRV 2-1223M 9.6 - 14.4 VDC (12 VDC nom.) 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC 600 mA 400 mA 167 mA 134 mA 200 mA 83 mA 67 mA –5 VDC –12 VDC –15 VDC 200 mA 83 mA 67 mA 79 % 81 % 84 % 83 % 81 % 83 % 82 %
TRV 2-1510M TRV 2-1511M TRV 2-1512M TRV 2-1513M TRV 2-1521M TRV 2-1522M TRV 2-1523M 12 - 18 VDC (15 VDC nom.) 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC 600 mA 400 mA 167 mA 134 mA 200 mA 83 mA 67 mA –5 VDC –12 VDC –15 VDC 200 mA 83 mA 67 mA 79 % 81 % 84 % 83 % 81 % 83 % 80 %
TRV 2-2410M TRV 2-2411M TRV 2-2412M TRV 2-2413M TRV 2-2421M TRV 2-2422M TRV 2-2423M 19.2 - 28.8 VDC (24 VDC nom.) 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC 600 mA 400 mA 167 mA 134 mA 200 mA 83 mA 67 mA –5 VDC –12 VDC –15 VDC 200 mA 83 mA 67 mA 78 % 80 % 82 % 82 % 81 % 81 % 80 %

Note - If the input will be switched electromechanically, use an external 47 μF/63 V E/C. to avoid voltage transient.

Page 1 / 5

Page 2

Input Specifications

Input Current

  • At no load
  • At full load

Surge Voltage

Recommended Input Fuse

Input Filter

Output Specifications

Voltage Set Accuracy

Regulation

  • Input Variation (Vmin - Vmax)
  • Load Variation - Voltage Balance (symmetrical load) - Cross Regulation (25% / 100% asym. load)

5 Vin models: 5 mA typ. (3.3 Vout model)

5 mA typ. (5 Vout model) 20 mA typ. (12 Vout model) 20 mA typ. (15 Vout model) 5 mA typ. (5 / -5 Vout model) 20 mA typ. (12 / -12 Vout model) 25 mA typ. (15 / -15 Vout model)

12 Vin models: 4 mA typ. (3.3 Vout model)

4 mA typ. (5 Vout model) 10 mA typ. (12 Vout model) 10 mA typ. (15 Vout model) 4 mA typ. (5 / -5 Vout model) 10 mA typ. (12 / -12 Vout model) 10 mA typ. (15 / -15 Vout model)

15 Vin models: 4 mA typ. (3.3 Vout model)

4 mA typ. (5 Vout model) 8 mA typ. (12 Vout model) 8 mA typ. (15 Vout model) 4 mA typ. (5 / -5 Vout model) 8 mA typ. (12 / -12 Vout model) 8 mA typ. (15 / -15 Vout model)

24 Vin models: 3 mA typ. (3.3 Vout model)

3 mA typ. (5 Vout model) 6 mA typ. (12 Vout model) 6 mA typ. (15 Vout model) 3 mA typ. (5 / -5 Vout model) 6 mA typ. (12 / -12 Vout model) 6 mA typ. (15 / -15 Vout model)

5 Vin models: 510 mA typ. 12 Vin models: 210 mA typ. 15 Vin models: 170 mA typ. 24 Vin models: 105 mA typ. 5 Vin models: 15 VDC max. (1 s max.) 12 Vin models: 25 VDC max. (1 s max.) 15 Vin models: 25 VDC max. (1 s max.) 24 Vin models: 35 VDC max. (1 s max.) 5 Vin models: 1'000 mA (slow blow) 12 Vin models: 500 mA (slow blow) 15 Vin models: 500 mA (slow blow) 24 Vin models: 315 mA (slow blow)

(The need of an external fuse has to be assessed in the final application.) Internal Capacitor

±3% max. (60% load: 3.3, 5, ±5 Vout models) ±3% max. (90% load: other models)

single output models: 2% max. dual output models: 2% max. See application note: www.tracopower.com/overview/trv2m dual output models: 4% max.

dual output models: 4% max.

All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.

Page 2 / 5

Page 3

Ripple and Noise

  • single output

(20 MHz Bandwidth)

  • dual output

Capacitive Load

  • single output
  • dual output

Minimum Load Temperature Coefficient Start-up Time Short Circuit Protection

Safety Specifications

Safety Standards

  • IT / Multimedia Equipment
  • Medical Equipment
  • Certification Documents

Pollution Degree Over Voltage Category

EMC Specifications

EMI Emissions

  • Conducted Emissions
  • Radiated Emissions

EMS Immunity

  • Electrostatic Discharge
  • RF Electromagnetic Field - EFT (Burst) / Surge
  • Conducted RF Disturbances - PF Magnetic Field

General Specifications

Relative Humidity

3.3 Vout models: 100 mVp-p typ. 5 Vout models: 100 mVp-p typ. 12 Vout models: 125 mVp-p typ. 15 Vout models: 125 mVp-p typ. 5 / -5 Vout models: 100 / 100 mVp-p typ. 12 / -12 Vout models: 125 / 125 mVp-p typ. 15 / -15 Vout models: 125 / 125 mVp-p typ. 3.3 Vout models: 2'000 µF max. 5 Vout models: 820 µF max. 12 Vout models: 470 µF max. 15 Vout models: 470 µF max. 5 / -5 Vout models: 470 / 470 µF max. 12 / -12 Vout models: 220 / 220 µF max. 15 / -15 Vout models: 220 / 220 µF max.

Not required ±0.03 %/K max. 20 ms max. Continuous, Automatic recovery

EN 62368-1 IEC 62368-1 UL 62368-1 EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) www.tracopower.com/overview/trv2m PD 2 Not mains connected

EN 60601-1-2 edition 4 (Medical Devices) EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter) EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter)

External filter proposal: www.tracopower.com/overview/trv2m

EN 55024 (IT Equipment) EN 60601-1-2 edition 4 (Medical Devices) Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A EN 61000-4-3, 10 V/m, perf. criteria A EN 61000-4-4, ±2 kV, perf. criteria A EN 61000-4-5, ±2 kV, perf. criteria A

External filter proposal: www.tracopower.com/overview/trv2m

EN 61000-4-6, 10 Vrms, perf. criteria A Continuous: EN 61000-4-8, 100 A/m, perf. criteria A 1 s: EN 61000-4-8, 1000 A/m, perf. criteria A

95% max. (non condensing)

All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.

Page 3 / 5

Page 4

Temperature Ranges

  • Operating Temperature - Case Temperature - Storage Temperature

Power Derating

  • High Temperature

–40°C to +95°C +105°C max. –55°C to +125°C Depending on model

See application note: www.tracopower.com/overview/trv2m

Cooling System Altitude During Operation Switching Frequency Insulation System Isolation Test Voltage

  • Input to Output, 60 s - Input to Output, 1 s

Creepage

  • Input to Output

Clearance

  • Input to Output

Isolation Resistance

  • Input to Output, 500 VDC

Isolation Capacitance

  • Input to Output, 100 kHz, 1 V

Leakage Current

  • Touch Current

Reliability

  • Calculated MTBF

Washing Process

Natural convection (20 LFM) 5'000 m max. 180 - 360 kHz (PFM) Reinforced Insulation 5'000 VAC 7'000 VDC 8 mm min. 8 mm min. 10'000 MΩ min. 10 pF typ. 20 pF max. 2 µA max. (240 VAC, 60 Hz) 10'410'000 h (MIL-HDBK-217F, ground benign) Allowed (hermetical product)

See Cleaning Guideline:

Environment

  • Vibration
  • Mechanical Shock
  • Thermal Shock

Housing Material Potting Material Pin Material Pin Foundation Plating Pin Surface Plating Housing Type Mounting Type Connection Type Footprint Type Soldering Profile

Weight Environmental Compliance - REACH Declaration

  • RoHS Declaration

Supporting Documents

Overview Link (for additional Documents)

www.tracopower.com/info/cleaning.pdf MIL-STD-810F 7.6 g, 3 axis, 60 min, 20-2000 Hz MIL-STD-810F 50 g, 3 axis, 11 ms MIL-STD-810F -55°C to +125°C, 72 cycles, 30 min each Non-conductive Plastic (UL 94 V-0 rated) Silicone (UL 94 V-0 rated) Tinned Copper Nickel (2 - 3 µm) Tin (3 - 5 µm), matte Plastic Case PCB Mount THD (Through-Hole Device) SIP9 Wave Soldering 260°C / 6 s max. 4.8 g www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.)

www.tracopower.com/overview/trv2m

All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.

Page 4 / 5

Page 5

Outline Dimensions

Pinout
Pin Single Dual
1 +Vin +Vin
2 in in
7 out out
8 No Pin Common
9 +Vout +Vout
PDF page 5 diagram
PDF page 5 diagram

Tolerances: x.x ±0.5 (±0.02) x.xx ±0.25 (±0.01) Pin diameter ±0.1 (±0.004)

© Copyright 2022 Traco Electronic AG

Specifications can be changed without notice.

Rev. November 30, 2022

Page 5 / 5