TRV2M TRACOPOWER
Document overview
- PDF pages: 5
Technical content
Page 1
DC/DC Medical Converter
- I/O isolation 5000 VAC (reinforced)
- Wide 1.5:1 input voltage range
- Operating temperature range -40 to +80 °C without derating
- Certification according to IEC/EN/ES
ES 60601-1 IEC 60601-1 UL 62368-1 IEC 62368-1
60601-1 3rd edition for 2xMOPP and
IEC/EN/UL 62368-1
The TRV 2M is a series of 2 Watt DC/DC converters in a compact SIP-9 package with reinforced isolation of 5000 VAC for medical and industrial applica-
- Short circuit protection
tions. The series offers a 1.5:1 input voltage range with a nominal input ranging
- Regulated outputs
between 5 and 24 VDC. With a continuous short circuit protection and a low leakage current of less than 2 µA, this converter is especially suited to protect
- Low leakage current < 2 μA
any connected interfaces or applied parts to patients. Featuring almost fully regulated outputs this series provides great level of regulation without affecting
- Efficiency up to 84 %
the cost efficiency. It is an ideal solution for applications where an unregulated
- Operation up to 5000 m altitude
DC/DC converter would not meet your regulation requirements but cost still is a critical factor. With an operating temperature range from -40 to +80°C without
- 5-year product warranty
derating and certifications acc. IEC/EN/ES 60601-1 3rd ed. for 2xMOPP and IEC/EN/UL 62368-1 this series is suitable for many different applications where a medical isolation system and short circuit protection is needed.
Models
| Order Code | Input Voltage Range | Output 1 Vnom Imax | Output 2 Vnom Imax | Efficiency typ. | ||
|---|---|---|---|---|---|---|
| TRV 2-0510M TRV 2-0511M TRV 2-0512M TRV 2-0513M TRV 2-0521M TRV 2-0522M TRV 2-0523M | 4.5 - 7 VDC (5 VDC nom.) | 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC | 600 mA 400 mA 167 mA 134 mA 200 mA 83 mA 67 mA | –5 VDC –12 VDC –15 VDC | 200 mA 83 mA 67 mA | 78 % 81 % 83 % 83 % 82 % 83 % 81 % |
| TRV 2-1210M TRV 2-1211M TRV 2-1212M TRV 2-1213M TRV 2-1221M TRV 2-1222M TRV 2-1223M | 9.6 - 14.4 VDC (12 VDC nom.) | 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC | 600 mA 400 mA 167 mA 134 mA 200 mA 83 mA 67 mA | –5 VDC –12 VDC –15 VDC | 200 mA 83 mA 67 mA | 79 % 81 % 84 % 83 % 81 % 83 % 82 % |
| TRV 2-1510M TRV 2-1511M TRV 2-1512M TRV 2-1513M TRV 2-1521M TRV 2-1522M TRV 2-1523M | 12 - 18 VDC (15 VDC nom.) | 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC | 600 mA 400 mA 167 mA 134 mA 200 mA 83 mA 67 mA | –5 VDC –12 VDC –15 VDC | 200 mA 83 mA 67 mA | 79 % 81 % 84 % 83 % 81 % 83 % 80 % |
| TRV 2-2410M TRV 2-2411M TRV 2-2412M TRV 2-2413M TRV 2-2421M TRV 2-2422M TRV 2-2423M | 19.2 - 28.8 VDC (24 VDC nom.) | 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC | 600 mA 400 mA 167 mA 134 mA 200 mA 83 mA 67 mA | –5 VDC –12 VDC –15 VDC | 200 mA 83 mA 67 mA | 78 % 80 % 82 % 82 % 81 % 81 % 80 % |
Note - If the input will be switched electromechanically, use an external 47 μF/63 V E/C. to avoid voltage transient.
Page 1 / 5
Page 2
Input Specifications
Input Current
- At no load
- At full load
Surge Voltage
Recommended Input Fuse
Input Filter
Output Specifications
Voltage Set Accuracy
Regulation
- Input Variation (Vmin - Vmax)
- Load Variation - Voltage Balance (symmetrical load) - Cross Regulation (25% / 100% asym. load)
5 Vin models: 5 mA typ. (3.3 Vout model)
5 mA typ. (5 Vout model) 20 mA typ. (12 Vout model) 20 mA typ. (15 Vout model) 5 mA typ. (5 / -5 Vout model) 20 mA typ. (12 / -12 Vout model) 25 mA typ. (15 / -15 Vout model)
12 Vin models: 4 mA typ. (3.3 Vout model)
4 mA typ. (5 Vout model) 10 mA typ. (12 Vout model) 10 mA typ. (15 Vout model) 4 mA typ. (5 / -5 Vout model) 10 mA typ. (12 / -12 Vout model) 10 mA typ. (15 / -15 Vout model)
15 Vin models: 4 mA typ. (3.3 Vout model)
4 mA typ. (5 Vout model) 8 mA typ. (12 Vout model) 8 mA typ. (15 Vout model) 4 mA typ. (5 / -5 Vout model) 8 mA typ. (12 / -12 Vout model) 8 mA typ. (15 / -15 Vout model)
24 Vin models: 3 mA typ. (3.3 Vout model)
3 mA typ. (5 Vout model) 6 mA typ. (12 Vout model) 6 mA typ. (15 Vout model) 3 mA typ. (5 / -5 Vout model) 6 mA typ. (12 / -12 Vout model) 6 mA typ. (15 / -15 Vout model)
5 Vin models: 510 mA typ. 12 Vin models: 210 mA typ. 15 Vin models: 170 mA typ. 24 Vin models: 105 mA typ. 5 Vin models: 15 VDC max. (1 s max.) 12 Vin models: 25 VDC max. (1 s max.) 15 Vin models: 25 VDC max. (1 s max.) 24 Vin models: 35 VDC max. (1 s max.) 5 Vin models: 1'000 mA (slow blow) 12 Vin models: 500 mA (slow blow) 15 Vin models: 500 mA (slow blow) 24 Vin models: 315 mA (slow blow)
(The need of an external fuse has to be assessed in the final application.) Internal Capacitor
±3% max. (60% load: 3.3, 5, ±5 Vout models) ±3% max. (90% load: other models)
single output models: 2% max. dual output models: 2% max. See application note: www.tracopower.com/overview/trv2m dual output models: 4% max.
dual output models: 4% max.
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
Page 2 / 5
Page 3
Ripple and Noise
- single output
(20 MHz Bandwidth)
- dual output
Capacitive Load
- single output
- dual output
Minimum Load Temperature Coefficient Start-up Time Short Circuit Protection
Safety Specifications
Safety Standards
- IT / Multimedia Equipment
- Medical Equipment
- Certification Documents
Pollution Degree Over Voltage Category
EMC Specifications
EMI Emissions
- Conducted Emissions
- Radiated Emissions
EMS Immunity
- Electrostatic Discharge
- RF Electromagnetic Field - EFT (Burst) / Surge
- Conducted RF Disturbances - PF Magnetic Field
General Specifications
Relative Humidity
3.3 Vout models: 100 mVp-p typ. 5 Vout models: 100 mVp-p typ. 12 Vout models: 125 mVp-p typ. 15 Vout models: 125 mVp-p typ. 5 / -5 Vout models: 100 / 100 mVp-p typ. 12 / -12 Vout models: 125 / 125 mVp-p typ. 15 / -15 Vout models: 125 / 125 mVp-p typ. 3.3 Vout models: 2'000 µF max. 5 Vout models: 820 µF max. 12 Vout models: 470 µF max. 15 Vout models: 470 µF max. 5 / -5 Vout models: 470 / 470 µF max. 12 / -12 Vout models: 220 / 220 µF max. 15 / -15 Vout models: 220 / 220 µF max.
Not required ±0.03 %/K max. 20 ms max. Continuous, Automatic recovery
EN 62368-1 IEC 62368-1 UL 62368-1 EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) www.tracopower.com/overview/trv2m PD 2 Not mains connected
EN 60601-1-2 edition 4 (Medical Devices) EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter) EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter)
External filter proposal: www.tracopower.com/overview/trv2m
EN 55024 (IT Equipment) EN 60601-1-2 edition 4 (Medical Devices) Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A EN 61000-4-3, 10 V/m, perf. criteria A EN 61000-4-4, ±2 kV, perf. criteria A EN 61000-4-5, ±2 kV, perf. criteria A
External filter proposal: www.tracopower.com/overview/trv2m
EN 61000-4-6, 10 Vrms, perf. criteria A Continuous: EN 61000-4-8, 100 A/m, perf. criteria A 1 s: EN 61000-4-8, 1000 A/m, perf. criteria A
95% max. (non condensing)
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
Page 3 / 5
Page 4
Temperature Ranges
- Operating Temperature - Case Temperature - Storage Temperature
Power Derating
- High Temperature
–40°C to +95°C +105°C max. –55°C to +125°C Depending on model
See application note: www.tracopower.com/overview/trv2m
Cooling System Altitude During Operation Switching Frequency Insulation System Isolation Test Voltage
- Input to Output, 60 s - Input to Output, 1 s
Creepage
- Input to Output
Clearance
- Input to Output
Isolation Resistance
- Input to Output, 500 VDC
Isolation Capacitance
- Input to Output, 100 kHz, 1 V
Leakage Current
- Touch Current
Reliability
- Calculated MTBF
Washing Process
Natural convection (20 LFM) 5'000 m max. 180 - 360 kHz (PFM) Reinforced Insulation 5'000 VAC 7'000 VDC 8 mm min. 8 mm min. 10'000 MΩ min. 10 pF typ. 20 pF max. 2 µA max. (240 VAC, 60 Hz) 10'410'000 h (MIL-HDBK-217F, ground benign) Allowed (hermetical product)
See Cleaning Guideline:
Environment
- Vibration
- Mechanical Shock
- Thermal Shock
Housing Material Potting Material Pin Material Pin Foundation Plating Pin Surface Plating Housing Type Mounting Type Connection Type Footprint Type Soldering Profile
Weight Environmental Compliance - REACH Declaration
- RoHS Declaration
Supporting Documents
Overview Link (for additional Documents)
www.tracopower.com/info/cleaning.pdf MIL-STD-810F 7.6 g, 3 axis, 60 min, 20-2000 Hz MIL-STD-810F 50 g, 3 axis, 11 ms MIL-STD-810F -55°C to +125°C, 72 cycles, 30 min each Non-conductive Plastic (UL 94 V-0 rated) Silicone (UL 94 V-0 rated) Tinned Copper Nickel (2 - 3 µm) Tin (3 - 5 µm), matte Plastic Case PCB Mount THD (Through-Hole Device) SIP9 Wave Soldering 260°C / 6 s max. 4.8 g www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.)
www.tracopower.com/overview/trv2m
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
Page 4 / 5
Page 5
Outline Dimensions
| Pinout | ||
| Pin | Single | Dual |
| 1 | +Vin | +Vin |
| 2 | in | in |
| 7 | out | out |
| 8 | No Pin | Common |
| 9 | +Vout | +Vout |


Tolerances: x.x ±0.5 (±0.02) x.xx ±0.25 (±0.01) Pin diameter ±0.1 (±0.004)
© Copyright 2022 Traco Electronic AG
Specifications can be changed without notice.
Rev. November 30, 2022