DC/DC Medical Converter
- I/O isolation 5000 VAC (reinforced)
- Short circuit protection
- Semi-regulated outputs
- Input voltage ranges (±10%):
ES 60601-1 IEC 60601-1 UL 62368-1 IEC 62368-1
5, 12, 15, 24 VDC
The TRV 1M is a series of 1 Watt DC/DC converters in a compact SIP-9 pack-
- Operating temperature range
age with reinforced isolation of 5000 VAC for medical and industrial applica-
-40 to +95 °C without derating
tions. The series offers models with different input voltages (±10%) between 5
- Certification according to IEC/EN/ES
and 24 VDC. With a continuous short circuit protection and a low leakage current of less than 2 µA, this converter series is especially suited to protect any
60601-1 3rd edition for 2xMOPP and
connected interfaces or applied parts to patients. Featuring semi-regulated out-
IEC/EN/UL 62368-1
puts this series provides a good level of regulation without affecting the cost ef-
- Low leakage current < 2 μA
ficiency. It is an ideal solution for applications where a completely unregulated DC/DC converter would not meet you regulation requirements and therefore
- Efficiency up to 85%
opens up the overall application range of this series. Together with an operat-
- Operation up to 5000 m altitude
ing temperature range from -40 to +95°C without derating and certifications according to IEC/EN/ES 60601-1 3rd ed. for 2xMOPP and IEC/EN/UL
- 5-year product warranty
62368-1 this series is suitable for many different applications where a medical isolation system and short circuit protection is needed.
Models
| Order Code | Input Voltage Range | Output 1 Vnom Imax | Output 2 Vnom Imax | Efficiency typ. | ||
|---|---|---|---|---|---|---|
| TRV 1-0510M TRV 1-0511M TRV 1-0512M TRV 1-0513M TRV 1-0521M TRV 1-0522M TRV 1-0523M | 4.5 - 5.5 VDC (5 VDC nom.) | 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC | 303 mA 200 mA 83 mA 67 mA 100 mA 42 mA 34 mA | –5 VDC –12 VDC –15 VDC | 100 mA 42 mA 34 mA | 80 % 82 % 85 % 84 % 85 % 85 % 84 % |
| TRV 1-1210M TRV 1-1211M TRV 1-1212M TRV 1-1213M TRV 1-1221M TRV 1-1222M TRV 1-1223M | 9.6 - 14.4 VDC (12 VDC nom.) | 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC | 303 mA 200 mA 83 mA 67 mA 100 mA 42 mA 34 mA | –5 VDC –12 VDC –15 VDC | 100 mA 42 mA 34 mA | 80 % 82 % 84 % 83 % 82 % 83 % 83 % |
| TRV 1-1510M TRV 1-1511M TRV 1-1512M TRV 1-1513M TRV 1-1521M TRV 1-1522M TRV 1-1523M | 12 - 18 VDC (15 VDC nom.) | 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC | 303 mA 200 mA 83 mA 67 mA 100 mA 42 mA 34 mA | –5 VDC –12 VDC –15 VDC | 100 mA 42 mA 34 mA | 79 % 83 % 84 % 84 % 82 % 83 % 83 % |
| TRV 1-2410M TRV 1-2411M TRV 1-2412M TRV 1-2413M TRV 1-2421M TRV 1-2422M TRV 1-2423M | 19.2 - 28.8 VDC (24 VDC nom.) | 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC | 303 mA 200 mA 83 mA 67 mA 100 mA 42 mA 34 mA | –5 VDC –12 VDC –15 VDC | 100 mA 42 mA 34 mA | 78 % 82 % 83 % 83 % 80 % 81 % 81 % |
Note - 5 Vin models: If the input will be switched electromechanically, use an external 100 μF/10 V E/C. to avoid voltage transient.
Page 1 / 4
Page 2
Input Specifications
Input Current
- At no load
Surge Voltage
Recommended Input Fuse
Output Specifications
Voltage Set Accuracy
Regulation
- Input Variation (1% Vin step)
- Load Variation - Cross Regulation (25% / 100% asym. load)
Ripple and Noise
- 20 MHz Bandwidth
Capacitive Load
- single output
- dual output
Minimum Load Temperature Coefficient Short Circuit Protection
Safety Specifications
Safety Standards
- IT / Multimedia Equipment
- Medical Equipment
- Certification Documents
Pollution Degree
5 Vin models: 30 mA typ. 12 Vin models: 30 mA typ. 15 Vin models: 15 mA typ. 24 Vin models: 10 mA typ. 5 Vin models: 6 VDC max. (1 s max.) 12 Vin models: 25 VDC max. (1 s max.) 15 Vin models: 25 VDC max. (1 s max.) 24 Vin models: 40 VDC max. (1 s max.) 5 Vin models: 500 mA (slow blow) 12 Vin models: 315 mA (slow blow) 15 Vin models: 315 mA (slow blow) 24 Vin models: 160 mA (slow blow)
(The need of an external fuse has to be assessed in the final application.)
±3.5% max. (60% load: 3.3, 5, ±5 Vout models) ±3.5% max. (90% load: other models)
single output models: 0.2% max. dual output models: 0.2% max. See application note: www.tracopower.com/overview/trv1m dual output models: 6% max.
75 mVp-p typ. 100 mVp-p max.
3.3 Vout models: 2'000 µF max. 5 Vout models: 820 µF max. 12 Vout models: 470 µF max. 15 Vout models: 470 µF max. 5 / -5 Vout models: 470 / 470 µF max. 12 / -12 Vout models: 220 / 220 µF max. 15 / -15 Vout models: 220 / 220 µF max.
Not required ±0.03 %/K max. Continuous, Automatic recovery
EN 62368-1 IEC 62368-1 UL 62368-1 EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) www.tracopower.com/overview/trv1m PD 2
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
Page 2 / 4
Page 3
EMC Specifications
EMI Emissions
- Conducted Emissions
- Radiated Emissions
EN 60601-1-2 edition 4 (Medical Devices) EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter) EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter)
External filter proposal: www.tracopower.com/overview/trv1m
EMS Immunity
- Electrostatic Discharge
- RF Electromagnetic Field - EFT (Burst) / Surge
EN 55024 (IT Equipment) EN 60601-1-2 edition 4 (Medical Devices) Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A EN 61000-4-3, 10 V/m, perf. criteria A EN 61000-4-4, ±2 kV, perf. criteria A EN 61000-4-5, ±2 kV, perf. criteria A
External filter proposal: www.tracopower.com/overview/trv1m
- Conducted RF Disturbances - PF Magnetic Field
General Specifications
Relative Humidity Temperature Ranges
- Operating Temperature - Case Temperature - Storage Temperature
Cooling System Altitude During Operation Switching Frequency Insulation System Working Voltage (rated) Isolation Test Voltage
- Input to Output, 60 s
Creepage
- Input to Output
Clearance
- Input to Output
Isolation Resistance
- Input to Output, 500 VDC
Isolation Capacitance
- Input to Output, 100 kHz, 1 V
Leakage Current
- Touch Current
Reliability
- Calculated MTBF
Washing Process
EN 61000-4-6, 10 Vrms, perf. criteria A Continuous: EN 61000-4-8, 100 A/m, perf. criteria A 1 s: EN 61000-4-8, 1000 A/m, perf. criteria A
95% max. (non condensing) –40°C to +95°C (without derating) +105°C max. –55°C to +125°C Natural convection (20 LFM) 5'000 m max. 220 - 380 kHz (PWM) Reinforced Insulation 250 VAC 5'000 VAC 8 mm min. 8 mm min. 10'000 MΩ min. 15 pF typ. 20 pF max. 2 µA max. 19'360'000 h (MIL-HDBK-217F, ground benign) Allowed (hermetical product)
See Cleaning Guideline:
Environment
- Vibration - Mechanical Shock - Thermal Shock
Housing Material Potting Material Pin Material Pin Foundation Plating Pin Surface Plating Housing Type Mounting Type Connection Type Footprint Type
www.tracopower.com/info/cleaning.pdf MIL-STD-810F MIL-STD-810F MIL-STD-810F Non-conductive Plastic (UL 94 V-0 rated) Silicone (UL 94 V-0 rated) Brass Nickel (1 - 2 µm) Tin (3 - 5 µm), matte Plastic Case PCB Mount THD (Through-Hole Device) SIP9
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
Page 3 / 4
Page 4
Weight Environmental Compliance - REACH Declaration
- RoHS Declaration
Supporting Documents
Overview Link (for additional Documents)
Outline Dimensions


4.8 g www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.)
www.tracopower.com/overview/trv1m
| Pinout | ||
| Pin | Single | Dual |
| 1 | +Vin (Vcc) | +Vin (Vcc) |
| 2 | –Vin (GND) | –Vin (GND) |
| 7 | –Vout | –Vout |
| 8 | No Pin | Common |
| 9 | +Vout | +Vout |
Tolerances: x.x ±0.5 (±0.02) x.xx ±0.25 (±0.01) Pin diameter ±0.1 (±0.004)