TRV1M TRACOPOWER

Document overview

  • PDF pages: 4

Technical content

Page 1

DC/DC Medical Converter

  • I/O isolation 5000 VAC (reinforced)
  • Short circuit protection
  • Semi-regulated outputs
  • Input voltage ranges (±10%):

ES 60601-1 IEC 60601-1 UL 62368-1 IEC 62368-1

5, 12, 15, 24 VDC

The TRV 1M is a series of 1 Watt DC/DC converters in a compact SIP-9 pack-

  • Operating temperature range

age with reinforced isolation of 5000 VAC for medical and industrial applica-

-40 to +95 °C without derating

tions. The series offers models with different input voltages (±10%) between 5

  • Certification according to IEC/EN/ES

and 24 VDC. With a continuous short circuit protection and a low leakage current of less than 2 µA, this converter series is especially suited to protect any

60601-1 3rd edition for 2xMOPP and

connected interfaces or applied parts to patients. Featuring semi-regulated out-

IEC/EN/UL 62368-1

puts this series provides a good level of regulation without affecting the cost ef-

  • Low leakage current < 2 μA

ficiency. It is an ideal solution for applications where a completely unregulated DC/DC converter would not meet you regulation requirements and therefore

  • Efficiency up to 85%

opens up the overall application range of this series. Together with an operat-

  • Operation up to 5000 m altitude

ing temperature range from -40 to +95°C without derating and certifications according to IEC/EN/ES 60601-1 3rd ed. for 2xMOPP and IEC/EN/UL

  • 5-year product warranty

62368-1 this series is suitable for many different applications where a medical isolation system and short circuit protection is needed.

Models

Order Code Input Voltage Range Output 1 Vnom Imax Output 2 Vnom Imax Efficiency typ.
TRV 1-0510M TRV 1-0511M TRV 1-0512M TRV 1-0513M TRV 1-0521M TRV 1-0522M TRV 1-0523M 4.5 - 5.5 VDC (5 VDC nom.) 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC 303 mA 200 mA 83 mA 67 mA 100 mA 42 mA 34 mA –5 VDC –12 VDC –15 VDC 100 mA 42 mA 34 mA 80 % 82 % 85 % 84 % 85 % 85 % 84 %
TRV 1-1210M TRV 1-1211M TRV 1-1212M TRV 1-1213M TRV 1-1221M TRV 1-1222M TRV 1-1223M 9.6 - 14.4 VDC (12 VDC nom.) 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC 303 mA 200 mA 83 mA 67 mA 100 mA 42 mA 34 mA –5 VDC –12 VDC –15 VDC 100 mA 42 mA 34 mA 80 % 82 % 84 % 83 % 82 % 83 % 83 %
TRV 1-1510M TRV 1-1511M TRV 1-1512M TRV 1-1513M TRV 1-1521M TRV 1-1522M TRV 1-1523M 12 - 18 VDC (15 VDC nom.) 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC 303 mA 200 mA 83 mA 67 mA 100 mA 42 mA 34 mA –5 VDC –12 VDC –15 VDC 100 mA 42 mA 34 mA 79 % 83 % 84 % 84 % 82 % 83 % 83 %
TRV 1-2410M TRV 1-2411M TRV 1-2412M TRV 1-2413M TRV 1-2421M TRV 1-2422M TRV 1-2423M 19.2 - 28.8 VDC (24 VDC nom.) 3.3 VDC 5 VDC 12 VDC 15 VDC +5 VDC +12 VDC +15 VDC 303 mA 200 mA 83 mA 67 mA 100 mA 42 mA 34 mA –5 VDC –12 VDC –15 VDC 100 mA 42 mA 34 mA 78 % 82 % 83 % 83 % 80 % 81 % 81 %

Note - 5 Vin models: If the input will be switched electromechanically, use an external 100 μF/10 V E/C. to avoid voltage transient.

Page 1 / 4

Page 2

Input Specifications

Input Current

  • At no load

Surge Voltage

Recommended Input Fuse

Output Specifications

Voltage Set Accuracy

Regulation

  • Input Variation (1% Vin step)
  • Load Variation - Cross Regulation (25% / 100% asym. load)

Ripple and Noise

  • 20 MHz Bandwidth

Capacitive Load

  • single output
  • dual output

Minimum Load Temperature Coefficient Short Circuit Protection

Safety Specifications

Safety Standards

  • IT / Multimedia Equipment
  • Medical Equipment
  • Certification Documents

Pollution Degree

5 Vin models: 30 mA typ. 12 Vin models: 30 mA typ. 15 Vin models: 15 mA typ. 24 Vin models: 10 mA typ. 5 Vin models: 6 VDC max. (1 s max.) 12 Vin models: 25 VDC max. (1 s max.) 15 Vin models: 25 VDC max. (1 s max.) 24 Vin models: 40 VDC max. (1 s max.) 5 Vin models: 500 mA (slow blow) 12 Vin models: 315 mA (slow blow) 15 Vin models: 315 mA (slow blow) 24 Vin models: 160 mA (slow blow)

(The need of an external fuse has to be assessed in the final application.)

±3.5% max. (60% load: 3.3, 5, ±5 Vout models) ±3.5% max. (90% load: other models)

single output models: 0.2% max. dual output models: 0.2% max. See application note: www.tracopower.com/overview/trv1m dual output models: 6% max.

75 mVp-p typ. 100 mVp-p max.

3.3 Vout models: 2'000 µF max. 5 Vout models: 820 µF max. 12 Vout models: 470 µF max. 15 Vout models: 470 µF max. 5 / -5 Vout models: 470 / 470 µF max. 12 / -12 Vout models: 220 / 220 µF max. 15 / -15 Vout models: 220 / 220 µF max.

Not required ±0.03 %/K max. Continuous, Automatic recovery

EN 62368-1 IEC 62368-1 UL 62368-1 EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) www.tracopower.com/overview/trv1m PD 2

All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.

Page 2 / 4

Page 3

EMC Specifications

EMI Emissions

  • Conducted Emissions
  • Radiated Emissions

EN 60601-1-2 edition 4 (Medical Devices) EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter) EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter)

External filter proposal: www.tracopower.com/overview/trv1m

EMS Immunity

  • Electrostatic Discharge
  • RF Electromagnetic Field - EFT (Burst) / Surge

EN 55024 (IT Equipment) EN 60601-1-2 edition 4 (Medical Devices) Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A EN 61000-4-3, 10 V/m, perf. criteria A EN 61000-4-4, ±2 kV, perf. criteria A EN 61000-4-5, ±2 kV, perf. criteria A

External filter proposal: www.tracopower.com/overview/trv1m

  • Conducted RF Disturbances - PF Magnetic Field

General Specifications

Relative Humidity Temperature Ranges

  • Operating Temperature - Case Temperature - Storage Temperature

Cooling System Altitude During Operation Switching Frequency Insulation System Working Voltage (rated) Isolation Test Voltage

  • Input to Output, 60 s

Creepage

  • Input to Output

Clearance

  • Input to Output

Isolation Resistance

  • Input to Output, 500 VDC

Isolation Capacitance

  • Input to Output, 100 kHz, 1 V

Leakage Current

  • Touch Current

Reliability

  • Calculated MTBF

Washing Process

EN 61000-4-6, 10 Vrms, perf. criteria A Continuous: EN 61000-4-8, 100 A/m, perf. criteria A 1 s: EN 61000-4-8, 1000 A/m, perf. criteria A

95% max. (non condensing) –40°C to +95°C (without derating) +105°C max. –55°C to +125°C Natural convection (20 LFM) 5'000 m max. 220 - 380 kHz (PWM) Reinforced Insulation 250 VAC 5'000 VAC 8 mm min. 8 mm min. 10'000 MΩ min. 15 pF typ. 20 pF max. 2 µA max. 19'360'000 h (MIL-HDBK-217F, ground benign) Allowed (hermetical product)

See Cleaning Guideline:

Environment

  • Vibration - Mechanical Shock - Thermal Shock

Housing Material Potting Material Pin Material Pin Foundation Plating Pin Surface Plating Housing Type Mounting Type Connection Type Footprint Type

www.tracopower.com/info/cleaning.pdf MIL-STD-810F MIL-STD-810F MIL-STD-810F Non-conductive Plastic (UL 94 V-0 rated) Silicone (UL 94 V-0 rated) Brass Nickel (1 - 2 µm) Tin (3 - 5 µm), matte Plastic Case PCB Mount THD (Through-Hole Device) SIP9

All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.

Page 3 / 4

Page 4

Weight Environmental Compliance - REACH Declaration

  • RoHS Declaration

Supporting Documents

Overview Link (for additional Documents)

Outline Dimensions

PDF page 4 diagram
PDF page 4 diagram

4.8 g www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.)

www.tracopower.com/overview/trv1m

Pinout
Pin Single Dual
1 +Vin (Vcc) +Vin (Vcc)
2 –Vin (GND) –Vin (GND)
7 –Vout –Vout
8 No Pin Common
9 +Vout +Vout

Tolerances: x.x ±0.5 (±0.02) x.xx ±0.25 (±0.01) Pin diameter ±0.1 (±0.004)

© Copyright 2022 Traco Electronic AG

Specifications can be changed without notice.

Rev. November 30, 2022

Page 4 / 4