Features

  • An aff ordable gas sensing solution for OEMs.
  • A reliable sensor design based on 15 years of engineering and manufacturing expertise.
  • Flexible CO2 sensor platform designed to interact with other microprocessor devices.
  • Eliminates the need for calibration in most applications with Telaire’s patented ABC Logic™ software. Lifetime calibration warranty. g GE Sensing & Inspection Technologies

Specifi cations Method Non Dispersive Infrared (NDIR), gold plated optics, diff usion or fl ow through sampling (with Telaire’s Patented ABC Logic Self Calibrated Algorithm) Measurement Range 400 to 2000 ppm Dimensions 2.25 in X 1.365 in X 0.60 in (57.15 mm X 34.67 mm X 15.24 mm) Accuracy* @72°F (22°C) when compared against a certifi ed factory reference ±40 ppm + 2% of reading up to 1250 ppm Temperature Dependence 0.2% FS per °C Stability < 2% of FS over life of sensor (15 year typical) Pressure Dependence 0.135 of reading per mm Hg Calibration Interval Not required Response Time < 2 minutes for 90% step change typical Signal Update Every 4 seconds Warm Up Time

  • < 2 minutes (operational)
  • 10 minutes (maximum accuracy) Operating Conditions
  • 32°F to 122°F (0°C to 50°C)
  • 0 to 95% RH, non condensing Storage Conditions -40°F to 158°F (-40°C to 70°C) Output Digital UART @ 19200 Baud (Please call for detailed product specifi cations). Analog 0.8 to 4 VDC Power Supply Requirements

5 VDC regulated (±5%)

  • 0.90 watts peak
  • 0.165 watts average Interface Connections Designed for 12 pin male header with 0.1 in (2.54 mm) spacing. Header not included. Flow Rates (Via Flow Ports) Flow through version 40 to 50 cc/min Warranty Terms 12 months * Handling and OEM assembly may a ff ect factory calibration. Specifi ed accuracy is after 21 days of continuous operation with ABC Logic. 0.38 in (9.65 mm) 1.24 in (31.5 mm) 1.365 in (34.67 mm) 2.25 in (57.15 mm) 1.125 in (28.575 mm) 0.15 in (3.81 mm) 1.75 in (44.45 mm) 2 Mounting Holes Ø.105 (2.667) 0.062 in (0.1575 mm) 0.54 in (13.7 mm) 0.60 in (15.24 mm)

Automatic Background Logic, or ABC Logic, is a patented self-calibration technique that is designed to be used in applications where concentrations will drop to outside ambient conditions (400 ppm) at least three times in a 14 day period, typically during unoccupied periods. *Full accuracy to be achieved utilizing ABC Logic. With ABC Logic enabled, the sensor will typically reach its operational accuracy after 25 hours of continuous operation at a condition that it was exposed to ambient reference levels of air at 400 ppm ±10 ppm CO 2. Sensor will maintain accuracy specifi cations with ABC Logic enabled, given that it is at least four times in 21 days exposed to the reference value and this reference value is the lowest concentration to which the sensor is exposed. ABC Logic requires continuous operation of the sensor for periods of at least 24 hours. Note: Applies when used in typical residential ambient air. Consult GE if other gases or corrosive agents are part of the application environment. Pin Designations Connector Pinout Function A TX (UART) B RX (UART) C V+ (5 VDC)

1 V+ (5 VDC)

2 GND

3 GND

4 AV OUT (0 to 4 VDC)

5 No Connect

6 No Connect

7 No Connect

8 No Connect

9 No Connect

10 TX (UART)

11 RX (UART)

12 GND

Module 6613-F (Flow Through) The diff usion membrane is replaced with a non- permeable seal. The two fl ow ports can be used for fl ow through sampling. Flow rate of 40 to 50 cc/min required. Handling and Installation The T6613 CO2 OEM modules are infrared gas sensors on a printed circuit board and should be treated carefully. Precautions should be taken to observe specifi ed limits and prevent damage from electrostatic discharge or rough handling. Please refer to ANSI/ESD S20. 20-1999 for more information on preventing ESD damage and IPC 610 Rev D for more information on proper electronic assembly practices. Calibration GE does not recommend the customer recalibrate the T6613 module after installation. The sensor ABC Logic algorithm will begin after 24 hours, adjusting the sensor measurement resulting in sensor-to-sensor consistency. If immediate consistency is desired, the sensors may be single point calibrated using pre-mixed reference gas or ambient air measured by a reference sensor. Available Models Part Number Sampling Method T6613 Di ff usion T6613-F Flow Through

© 2008 General Electric Company. All Rights Reserved. Specifi cations are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affi liated with GE. 920-448B www.gesensinginspection.com g