R7446 HAMAMATSU

Document overview

  • PDF pages: 4

Technical content

Page 1

PHOTOMULTlPLlER TUBES

R7446

R7446P (For Photon Counting)

High Cathode Sensitivity with Low Noise Photocathode

PDF page 1 diagram

FEATURES

Spectral Response ..................................... 160 to 680 nm Cathode Sensitivity Luminous ........................................................ 60 µA/lm Radiant at 400nm ............................................ 60 mA/W Anode Sensitivity (at 1000V) Luminous ......................................................... 400 A/lm Radiant at 400nm ..................................... 4.0 × 105 A/W Low Dark Current .................................................... 0.1 nA Low Dark Counts (R7446P) .................................... 10 cps

APPLICATIONS

Environmental Monitoring Atomic Emission Spectrometer Atomic Absorption Spectrometer

GENERAL

PDF page 1 diagram

Figure 1: Typical Spectral Response

TPMSB0193EA

PDF page 1 diagram

WAVELENGTH (nm)

Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©1999 Hamamatsu Photonics K.K

Page 2

PHOTOMULTlPLlER TUBES R7446, R7446P (For Photon Counting)

MAXIMUM RATINGS (Absolute Maximum Values)

PDF page 2 diagram
PDF page 2 diagram

NOTES

A: Averaged over any interval of 30 seconds maximum. B: The light source is a tungsten filament lamp operated at a distribution temperature of 2856K. Supply voltage is 150 volts between the cathode and all other electrodes connected together as anode. C:The value is cathode output current when a blue filter(Corning CS-5-58 polished to 1/2 stock thickness) is interposed between the light source and the tube under the same condition as Note B. D:Measured with the same light source as Note B and with the anode-to- cathode supply voltage and voltage distribution ratio shown in Table 1 be- low. E: Measured with the same supply voltage and voltage distribution ratio as

Note D after removal of light. F: Measured at the voltage producing the gain of 1 × 106. G:ENI is an indication of the photon-limited signal-to-noise ratio. It refers to the amount of light in watts to produce a signal-to-noise ratio of unity in the output of a photomultiplier tube.

2q.ldb.G. f ENI = S

where q = Electronic charge (1.60 × 10-19 coulomb). ldb = Anode dark current(after 30 minute storage) in amperes. G = Gain. f = Bandwidth of the system in hertz. 1 hertz is used. S = Anode radiant sensitivity in amperes per watt at the wavelength of peak response.

H:The rise time is the time for the output pulse to rise from 10% to 90% of the peak amplitude when the entire photocathode is illuminated by a delta function light pulse. J: The electron transit time is the interval between the arrival of delta function light pulse at the entrance window of the tube and the time when the anode output reaches the peak amplitube. In measurement, the whole photocathode is illuminated.

K: Also called transit time jitter. This is the fluctuation in electron transit time between individual pulses in the signal photoelectron mode, and may be defined as the FWHM of the frequency distribution of electron transit times. L: Hysteresis is temporary instability in anode current after light and voltage are applied.

lmax. lmin.

Hysteresis =

× 100(%) li

l max.

il l min. TIME

5 6 7 (minutes)

0

TPMSB0002EA

(1)Current Hysteresis The tube is operated at 750 volts with an anode current of 1 micro-ampere for 5 minutes. The light is then removed from the tube for a minute. The tube is then re-illuminated by the previous light level for a minute to measure the variation. (2)Voltage Hysteresis The tube is operated at 300 volts with an anode current of 0.1 micro-ampere for 5 minutes. The light is then removed from the tube and the supply voltage is quickly increased to 800 volts. After a minute, the supply voltage is then reduced to the previous value and the tube is re-illuminated for a minute to measure the variation. Table 1:Voltage Distribution Ratio

Electrode K Dy1 Dy2 Dy3 Dy4 Dy5 Dy6 Dy7 Dy8 Dy9 P
Distribution Ratio 1 1 1 1 1 1 1 1 1 1

SuppIy Voltage : 1000Vdc K : Cathode, Dy : Dynode, P : Anode

Page 3

Figure 2: Typical Gain and Anode Dark Current

TPMSB0026EA

PDF page 3 diagram

SUPPLY VOLTAGE (V)

Figure 4: Typical ENI vs. Wavelength

TPMSB0194EA

PDF page 3 diagram

WAVELENGTH (nm)

Figure 6: Typical Single Photon Height Distribution

for R7446P

TPMSB0195EA

PDF page 3 diagram

CHANNEL NUMBER (CH)

Figure 3: Typical Time Response

TPMSB0004EB

PDF page 3 diagram

SUPPLY VOLTAGE (V)

Figure 5: Typical EADCI (Equivalent Anode Dark Current

Input) vs. Supply Voltage

TPMSB0028EA

PDF page 3 diagram

SUPPLY VOLTAGE (V)

Data shown here, which is given from a relation among supply voltage,

anode sensitivity and dark current, serves as a good reference in order

to determine the most suitable supply voltage or its range.

Figure 7: Typical Temperature Characteristics

of Dark Count for R7446P

TPMSB0030EA

PDF page 3 diagram

TEMPERATURE (°C)

Page 4

PHOTOMULTlPLlER TUBES R7446, R7446P (For Photon Counting)

Figure 8: Dimensional Outline and Basing Diagram(Unit: mm)

PDF page 4 diagram

DY6 DY7

DY5 6 5 7

DY4

DY8

4 8

DY3 3

9 DY9

2 10 P

DY2 1 11 K DY1

DIRECTION OF LIGHT

BOTTOM VIEW

(BASING DIAGRAM)

TPMSA0001EA

Figure 9: D Type Socket Assembly E717-63 (Option)

PDF page 4 diagram

SOCKET

PDF page 4 diagram

\* Hamamatsu also provides C4900 series compact high voltage power

supplies and C6270 series DP type socket assemblies which incorporate a DC to DC converter type high voltage power supply.

Figure 10: Socket E678-11A (Option)

PDF page 4 diagram

TACCA0064EA

SIGNAL OUTPUT RG-174/U (BLACK) POWER SUPPLY GND AWG22 (BLACK)

AWG22 (VIOLET)

TACCA0002EG

PDF page 4 diagram

HOMEPAGE URL http://www.hamamatsu.com

HAMAMATSU PHOTONICS K.K., Electron Tube Center

314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 United Kingdom: Hamamatsu Photonics UK Limited: Lough Point, 2 Gladbeck Way, Windmill Hill, Enfield, Middlesex EN2 7JA, United Kingdom, Telephone: 44(20)8-367-3560, Fax: 44(20)8-367-6384

North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01

TPMS1068E01

Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741

OCT. 1999