R3809U-50 HAMAMATSU
Document overview
- PDF pages: 8
Technical content
Page 1

Compact MCP-PMT Series Featuring
Variety of Spectral Response with Fast Time Response
FEATURES
GHigh Speed Rise Time: 150 ps TTS (Transit Time Spread) A: 25 ps (FWHM) GLow Noise GCompact Profile Useful Photocathode: 11 mm diameter (Overall length: 70.2 mm Outer diameter: 45.0 mm)
APPLICATIONS
GMolecular Science Analysis of Molecular Structure GMedical Science Optical Computer Tomography GBiochemistry Fast Gene Sequencing GMaterial Engineering Semiconductor Analysis Crystal Research
Figure 1: Spectral Response Characteristics

WAVELENGTH (nm)

Figure 2: Transit Time Spread

TIME (ps)
Figure 3: Block Diagram of TTS Mesuring System

TPMHC0078ED
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2007 Hamamatsu Photonics K.K.
Page 2
SPECIFICATIONS
PHOTOCATHODE SELECTION GUIDE
| Suffix Number | Spectral Response (nm) | Photocathode Material | Window Material | |
| Range | Peek Wavelength | |||
| 50 | 160 to 850 | 430 | Multialkali | Synthetic Silica |
| 51 | 160 to 910 | 600 | Extended Red Multialkali | Synthetic Silica |
| 52 | 160 to 650 | 400 | Bialkali | Synthetic Silica |
| 53 | 160 to 320 | 230 | Cs-Te | Synthetic Silica |
| 57 | 115 to 320 | 230 | Cs-Te | MgF2 |
| 58 | 115 to 850 | 430 | Multialkali | MgF2 |
GENERAL
| Parameter | Description / Value | Unit |
|---|---|---|
| Photocathode Useful Area in Diameter | 11 | mm |
| MCP Channel Diameter | 6 | µm |
| Dynode Structure B | 2-Stage Filmed MCP | — |
| Capacitance Between Anode and MCP out | 3 | pF |
| Weight | 98 | g |
| Operating Ambient Temperature C | -50 to +50 | °C |
| Storage Temperature | -50 to +50 | °C |
MAXIMUM RATINGS (Absolute Maximum Values)
| Parameter | Value | Unit |
|---|---|---|
| Supply Voltage | -3400 | V |
| Average Anode Current | 100 | nA |
| Pulsed Peak Current D | 350 | mA |
ELECTRICAL CHARACTERISTICS (R3809U-50) at 25 °C E
| Parameter | Min. | Typ. | Max. | Unit | |
|---|---|---|---|---|---|
| Cathode Sensitivity | Luminouse F | 100 | 150 | — | µA/lm |
| Radiant at 430 nm | — | 50 | — | mA/W | |
| Gain at -3000 V | 1 × 105 | 2 × 105 | — | — | |
| Anode Dark Counts at -3000 V | — | — | 2000 | s-1 | |
| Voltage Divider Current at -3000 V | — | — | 75 | µA | |
| Time Response | Rise Time G | — | 150 | — | ps |
| Fall Time H | — | 360 | — | ps | |
| IRF (FWHM) I | — | 45 J | — | ps | |
| TTS (FWHM) | — | — | 25 K | ps | |
NOTES
A Transit-time spread (TTS) is the fluctuation in transit time between individual pulse and specified as an FWHM (full width at half maximum) with the incident light having a single photoelectron state. B Two microchannel plates (MCP) are incorporated as a standard but we can provide it with either one or three MCPs as an option depending upon your request. C We recommend use R3809U-51 with thermoelectric cooling unit to reduce dark counts (Refer to Figure 5) D This is specified under the operating conditions that the repetition rate of light input is 100 Hz or below and its pulse width is 70 ps. E This data is based on R3809U-50. All other types (suffix number 51 through 58) have different characteristics on cathode sensitivity and anode dark counts.
F The light source used to measure the luminous sensitivity is a tungsten filament lamp operated at a distribution temperature of 2856 K. The incident light intensity is 10–4 lumen and 100 V is applied between the photocathode and all other electrodes connected as an anode. G This is the mean time difference between the 10 % and 90 % amplitude points on the output waveform for full cathode illumination. H This is the mean time difference between the 90 % and 10 % amplitude points on the tailing edge of the output waveform for full cathode illumination. I IRF stands for Instrument Response Function which is a convolution of the δ pulse function (H(t)) of the measuring system and the excitation function (E(t)) of a
laser. The IRF is given by the following formula: IRF = H(t) × E(t) J We specify the IRF as an FWHM of the time distribution taken by using the measuring system in Figure 13 that is Hamamatsu standard IRF measurement. It can be temporary estimated by the following equation: (IRF (FWHM))2 = (TTS)2 + (Tw)2 + (Tj)2 where Tw is the pulse width of the laser used and Tj is the time jitter of all equipments used. An IRF data is provided with the tube purchased as a standard. K TTS stands for Transit Time Spread (seeA above). Assuming that a laser pulse width (Tw) and time jitter of all equipments (Tj) used in Figure 3 are negligible, IRF can be estimated as equal to TTS (seeJ) above. Therefore, TTS can be estimated to be 25 picoseconds or less.
Page 3
TECHNICAL REFERENCE DATA
Figure 4: Typical Gain

SUPPLY VOLTAGE (kV)
Figure 6: Typical Output Deviation as a Function of Anode DC Current

ANODE CURRENT (nA)
Figure 5: Variation of Dark Counts Depending on Ambient Temperature

AMBIENT TEMPERATURE (°C)
Figure 7: Typical Output Deviation as a Function of Anode Count Rate

COUNT RATE (s-1)
Page 4
Figure 8: Typical Output Waveform

TIME (0.2 ns/div)
Figure 10: Typical Pulse Height Distribution (PHD)

PULSE HEIGHT (CHANNEL NUMBER)
Figure 9: Block Diagram of Output Waveform Measuring System

TRIGGER
TPMHC0079EE
Figure 11: Block Diagram of PHD Measuring System


COMPUTER
MCA NAIG NAIG E-511A
CANBERRA 2005
TPMHC0080ED
Page 5
Figure 12: Typical Instrument Response Function (IRF)

TIME (0.2 ns/div)
Figure 14: Dimensional Outline (Unit: mm)
EFFECTIVE
PHOTOCATHODE
WINDOW
DIAMETER
Figure 13: Block Diagram of IRF Measuring System

HAMAMATSU MODEL#PLP-01 WAVELENGTH: 410 nm PULSE WIDTH (FWHM) : 35 ps
TPMHC0081ED
70.2±0.3
52.5±0.1
-HV INPUT

TPMHA0352EB
Page 6
PRECAUTIONS FOR PROPER OPERATION
Handling on set-up 1) The photomultiplier tube (PMT) is a glass product under high vacuum. EXCESSIVE PRESSURE, VIBRATIONS OR SHOCKS TO THE TUBE FROM THE SURROUNDING COULD CAUSE A PERMANENT DAMAGE. Please pay special attention on insuring proper handling. 2) DO NOT PLACE ANY OBJECTS OF GROUND POTENTIAL CLOSER THAN 5mm TO THE PHOTOCATHODE WINDOW when negative high voltage is applied to the photocathode. It could generate extra noise and damage the photocathode permanently. 3) DO NOT EXPOSE THE PHOTOCATHODE TO SUNLIGHT DIRECTLY and any light stronger than the room light even during of no operation. 4) NEVER TOUCH THE INPUT WINDOW WITH YOUR BARE HANDS. In case the window contaminated by dust or grease,
wipe it off using alcohol and a soft cloth or dust free tissue. 5) DO NOT OPERATE OR STORE IN A PLACE OF UNSPECIFIED TEMPERATURE AND HUMIDITY.
Supplying high voltage 1) DO NOT SUPPLY ANY VOLTAGE HIGHER THAN SPECIFIED. Also make sure the output current does NOT EXCEED THE MAXIMUM CURRENT specified. 2) This device is very sensitive even with weak light input. When applying high voltage to the tube, GRADUALLY (IDEALLY 100 Vdc STEP BUT 500 Vdc STEP IS OK) AND CAREFULLY INCREASE THE VOLTAGE while monitoring the output using an ammeter or oscilloscope. Also make sure before use that the polarity of the applied voltage is correct. 3) DO NOT REMOVE OR CONNECT ANY INPUT OR OUTPUT CABLES WHILE HIGH VOLTAGE IS APPLIED. If a high voltage is applied when its output is opened, DO NOT CONNECT ANY READOUT CIRCUIT TO THE TUBE IMMEDIATELY
after turning the high voltage off. Ground the anode of the tube before connecting in order to avoid possible damage to the readout circuit due to an excessive electron charge flowing from its anode. 4) IT IS RECOMMENDED TO TURN HIGH VOLTAGE OFF WHILE NOT BEING USED FOR MEASUREMENTS. This is to avoid shortening its period of life time as well as a risk of damage due to an exposure of excessive incident light.
Incident light amount 1) KEEP THE INCIDENT LIGHT AMOUNT AS LOWS AS POSSIBLE to extend its period of life time. 2) In a case of photon counting application, it is recommended to KEEP THE SIGNAL COUNT RATE LESS THAN 20kcps. 3) ILLUMINATE PHOTOCATHODE EFFECTIVE AREA AS LARGE AS POSSIBLE to keep better linearity characteristics and avoid an excessive stress in partial area, which may result in a reduction of sensitivity partially.
Usage in vacuum 1) DO NOT USE A PMT AS AN INTERFACE BETWEEN VACUUM AND ENVIRONMENTAL PRESSURE. Standard MCP-PMT is not designed for vacuum-tight construction. 2) KEEP THE TUBE CLEAN. Unless otherwise, it would cause outgassing in a vacuum. 3) DO NOT SUPPLY HIGH VOLTAGE UNLESS THE VACUUM LEVEL REACHES 1 × 10–3 Pa OR HIGHER. 4) DO NOT PROCEED BAKING VACUUM INSTRUMENTS WHILE THE TUBE IS PLACED INSIDE.
OTHERS
1) If the tube won't be used with a cooler, it is recommended to LEAVE THE TUBE IN DARKNESS (YOUR INSTRUMENT WITHOUT ANY INPUT LIGHT) FOR 30 MINUTES OR SO before start any measurements because it occasionally takes a little while until its dark noise settles down.
WARRANTY
The detectors indicated in this data sheet are warranted to the original purchaser for a period of 12 MONTHS following the date of shipment. The warranty is limited to repair or replacement of any defective material due to defects in workmanship or materials used in manufacture.
1) Any claim for damage of shipment must be made directly to the delivering carrier within five days. 2) Customer must inspect and test all detectors within 30 days after shipment. Failure to accomplish said incoming inspection shall limit all claims to 75% of invoice value. 3) No credit will be issued for broken detector unless in the opinion of Hamamatsu the damage is due to a manufacturing defect. 4) No credit will be issued for any detector which in the judgement of Hamamatsu has been damaged, abused, modified or whose serial number or type number have been obliterated or defaced. 5) No detector will be accepted for return unless permission has been obtaind from Hamamatsu in writing, the shipment has
been returned repaired and insured, the detector is packed in their original box and accompanied by the original data sheet furnished to the customer with the tube, and a full written explanation of the reason for rejection of detector.
Page 7
ACCESSORIES
THERMOELECTRIC COOLING UNIT C10373
HOLDER E3059-500
| Left: Power Supply Right: Cooled PMT Housing | |
| Parameter | Description / Value |
| Cooling Methode | Thermoelectric cooling using peltier module |
| Heat Exchange Medium | Water |
| Amount of Cooling Water | 1 L/min to 3 L/min (water pressure: below 0.3 MPa) |
| Cooling Temperature (with cooling water at +20 °C) | Approx. -30 °C |
| Temperature Controllable Range (with cooling water at +20 °C) | -30 °C to 0 °C (continuously adjustable) |
| Cooling Time | Approx. 120 min |
| Optical Window Material | Evacuated double-pane synthetic silica window with heater (185 nm to 2200 nm) |
| Operating Ambient Temperature A | +5 °C to +40 °C / Below 75 % |
| Storage Temperature A | -15 °C to +50 °C / Below 80 % |
| Weight | 5.5 kg |
| NOTE: ANo condensation HIGH SPEED AMPLIFIER C5594 Series BENCH-TOP HIGH VOLTAGE POWER SUPPLY C4840 Series | |

Suffix numbers and input / output connectors
| Input Connectors | Output Connectors | |
| SMA Jack | BNC Jack | |
| SMA Plug (male) | C5594-12 | C5594-14 |
| SMA Receptacle (female) | C5594-22 | C5594-24 |
| BNC Plug (male) | C5594-32 | C5594-34 |
| BNC Receptacle (female) | C5594-42 | C5594-44 |
Specifications
| Parameters | Description / Value | ||
|---|---|---|---|
| Frequency Response Range | 50 kHz to 1.5 GHz | ||
| Gain | Typ. | 36 dB | |
| Input / Output Impedance | 50 Ω | ||
| Noise Figure (NF) | Typ. | 5 dB | |
| Recommend Input Voltage | +12 V to 16 V | ||
| Supply Current | Typ. | 95 mA | |
| Absolute Maximum Ratings | Supply Voltage | +17 V | |
| Input Power | +10 mW | ||

Specifications
| Parameter | Description / Value | ||
|---|---|---|---|
| Output Voltage | 0 V to ±3000 V | ||
| Maximum Output Current | 10 mA | ||
| Line Regulation Against ±10 % Line Voltage Change AB | Max. | ±(0.005 % + 10 mV) | |
| Load Regulation Against 0 % to 100 % Load Change A | Max. | ±(0.01 % + 50 mV) | |
| Ripple / Noise (p-p) AB | Max. | 0.0007 % | |
| Drift (after 1 h Warm-up) AB | Max. | ±(0.02 % + 10 mV)/8 h | |
| Temperature Coefficient AB | Max. | ±0.01 % / °C | |
| AC Input Voltage | C4840-01 | 120 V (±10 %) (50 / 60 Hz) | |
| C4840-02 | 230 V (±10 %) (50 / 60 Hz) | ||
| Power Consumption AB | Approx. 100 V·A | ||
| Operating Ambient Temperature / Humidity C | 0 °C to +40 °C / below 80 % | ||
| Storage Temperature / Humidity C | -20 °C to +50 °C / below 85 % | ||
NOTE: AAt maximum output voltage BAt maximum output current CNo condensation
Page 8
WEB SITE www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Electron Tube Division 314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: info@hamamatsu.co.uk North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it
TPMH1067E08 AUG. 2007. IP