PMC71 EH
Document overview
- Manufacturer or author: Provided By www.digicamel.com(free datasheet download site)
- PDF pages: 124
Technical content
Pressure transmitter with ceramic and metal sensors
Applications
The device is used for the following measuring tasks:
- Absolute pressure and gauge pressure measurement in gases, steams or liquids in all areas of process engineering and process measurement technology
- Level, volume or mass measurements in liquids
- High process temperatures – up to 150 °C (302 °F) without diaphragm seal – up to 400 °C (752 °F) with typical diaphragm seals
- High pressures up to 700 bar (10 500 psi)
- MID part certificate according to OIML R117-1 Edition 2007 (E) and EN 12405-1/A1 Edition 2006 Your benefits
- Very good reproducibility and long-term stability
- High reference accuracy up to ±0.025 %
- Turn down up to 100:1, higher on request
- Used for process pressure monitoring up to SIL 3, certified to IEC 61508 by TÜV SÜD
- High level of safety during operation thanks to function monitoring from the measuring cell to the electronics
- The patented TempC membrane for the diaphragm seal reduces measured errors caused by environmental and process temperature influences to a minimum
- Easy electronic replacement guaranteed with HistoROM®/M-DAT
- Uniform platform for differential pressure, hydrostatics and pressure (Deltabar S – Deltapilot S – Cerabar S)
- Practical user navigation for quick and easy commissioning
- Extensive diagnostic functions Products Solutions Services Technical Information Cerabar S PMC71, PMP71, PMP75 Process pressure measurement TI00383P/00/EN/26.16 71331990
Cerabar S PMC71, PMP71, PMP75
2 Endress+Hauser
Applications suitable for custody transfer measurement ... 10 Performance characteristics of PMP71/PMP75 metallic pressure transmitter Performance characteristics of PMC71 ceramic pressure transmitter Performance characteristics - calculation Measuring arrangement for devices without diaphragm Measuring arrangement for devices with diaphragm seals Process temperature limits of capillary armoring: PMP75 .. 51
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 3 Process connections for PMC71 with internal process Process connections for PMC71 with internal process Process connections for PMC71 with internal process Process connections for PMC71 with flush-mounted Process connections for PMC71 with flush-mounted Process connections for PMC71 with flush-mounted Process connections for PMC71 with flush-mounted Hygienic process connections for PMC71 with flush- Process connections for PMP71 with internal process Process connections for PMP71 with internal process Process connections for PMP71 with internal process Process connections for PMP71 with flush-mounted Process connections for PMP71 with flush-mounted Process connections for PMP71 with flush-mounted Process connections for PMP71 with flush-mounted Process connections for PMP71 with flush-mounted Process connections for PMP75 with flush-mounted Process connections for PMP75 with flush-mounted Process connections for PMP75 with flush-mounted Process connections for PMP75 with flush-mounted Hygienic process connections for PMP75 with flush- Hygienic process connections for PMP75 with flush- Hygienic process connections for PMP75 with flush- Process connections for PMP75 with flush-mounted Process connections for PMP75 with flush-mounted Separate housing: Wall and pipe mounting with mounting Planning instructions, diaphragm seal systems .. 107 Functional safety SIL/ IEC 61508 Declaration of Classification of process sealing between electrical systems and (flammable or combustible) process fluids in
Cerabar S PMC71, PMP71, PMP75
4 Endress+Hauser
Document function The document contains all the technical data on the device and provides an overview of the accessories and other products that can be ordered for the device. Symbols used Safety symbols Symbol Meaning DANGER DANGER! This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. CAUTION CAUTION! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury. NOTICE NOTE! This symbol contains information on procedures and other facts which do not result in personal injury. Electrical symbols Symbol Meaning Symbol Meaning Protective ground connection A terminal which must be connected to ground prior to establishing any other connections. Ground connection A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system. Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred. Forbidden Procedures, processes or actions that are forbidden. Tip Indicates additional information. Reference to documentation A Reference to page Reference to graphic Visual inspection
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 5 Symbols in graphics Symbol Meaning 1, 2, 3 ... Item numbers , …,1. 2. 3. Series of steps A, B, C, ... Views A-A, B-B, C-C, ... Sections Documentation See the "Supplementary documentation" section → 122 The document types listed are available: In the Downloads area of the Endress+Hauser website: www.endress.com → Downloads Safety Instructions (XA) See the "Safety instructions" section → 122
Cerabar S PMC71, PMP71, PMP75
6 Endress+Hauser
p LRV URV A0029505 Item Term/abbreviation Explanation 1 OPL The OPL (over pressure limit = sensor overload limit) for the measuring device depends on the lowest-rated element, with regard to pressure, of the selected components, i.e. the process connection has to be taken into consideration in addition to the measuring cell. Also observe pressure-temperature dependency. For the relevant standards and additional information, see the "Pressure specifications"→ 52 section. The OPL may only be applied for a limited period of time. 2 MWP The MWP (maximum working pressure) for the sensors depends on the lowest-rated element, with regard to pressure, of the selected components, i.e. the process connection has to be taken into consideration in addition to the measuring cell. Also observe pressure-temperature dependency. For the relevant standards and additional information, see the "Pressure specifications"→ 52 section. The MWP may be applied at the device for an unlimited period. The MWP can also be found on the nameplate.
3 Maximum sensor
This sensor measuring range is equivalent to the maximum calibratable/adjustable span.
4 Calibrated/adjusted
Factory setting: 0 to URL Other calibrated spans can be ordered as customized spans. p - Pressure - LRL Lower range limit - URL Upper range limit - LRV Lower range value - URV Upper range value - TD (turn down) Turn down Example - see the following section.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 7 Turn down calculation LRV URLURVLRL 1 = 2 3 A0029545
1 Calibrated/adjusted span
2 Zero point-based span
3 URL sensor
- Sensor:10 bar (150 psi)
- Upper range value (URL) = 10 bar (150 psi) Turn down (TD):
- Calibrated/adjusted span: 0 to 5 bar (0 to 75 psi)
- Lower range value (LRV) = 0 bar (0 psi)
- Upper range value (URV) = 5 bar (75 psi) TD = URL |URV - LRV| TD = 10 bar (150 psi) = 2 |5 bar (75 psi) - 0 bar (0 psi)| In this example, the TD is 2:1. This span is based on the zero point. Registered trademarks HARTâ Registered trademark of the FieldComm Group, Austin, USA PROFIBUSâ Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany FOUNDATIONTMFieldbus Registered trademark of the FieldComm Group, Austin, Texas, USA
Cerabar S PMC71, PMP71, PMP75
8 Endress+Hauser
Function and system design Device features PMC71 PMP71 PMP75 A0020461 A0020463 A0020464 With capacitance measuring cell and ceramic process isolating diaphragm (Ceraphire®) With piezoresistive measuring cell and metallic welded process isolating diaphragm With diaphragm seal Field of application
- Gauge pressure and absolute pressure
- Level Process connections
- Diverse thread
- DN 25 – DN 80
- ASME 1 ½" – 4"
- JIS 50 A – 100 A
- Diverse thread
- DN 25 – DN 80
- ASME 1 ½" – 4"
- JIS 25 A – 100 A
- Oval flange adapter
- Prepared for diaphragm seal mount Wide range of diaphragm seals Measuring ranges from –100/0 to 100 mbar (–1.5/0 to 1.5 psi) to –1/0 to 40 bar (–15/0 to 600 psi) from –400/0 to 400 mbar (–6/0 to 6 psi) to –1/0 to 700 bar (–15/0 to 10500 psi) from –400/0 to 400 mbar (–6/0 to 6 psi) to –1/0 to 400 bar (–15/0 to 6000 psi) OPL max. 60 bar (900 psi) max. 1 050 bar (15 750 psi) max. 600 bar (9 000 psi) Process temperature range 1) (depends on filling oil) Ambient temperature range
- Without LCD display: –40 to +85 °C (–40 to +185 °F) 3)
- With LCD display: –20 to +70 °C (–4 to +158 °F)
- Separate housing: –20 to +60 °C (–4 to +140 °F)
- Diaphragm seal systems depending on the version Reference accuracy
- Up to ±0.05 % of the set span
- PLATINUM version: up to ±0.025 % of the set span Up to ±0.075 % of the set span Supply voltage non-Ex
- 4 to 20 mA HART: 10.5 to 45 V DC
- PROFIBUS PA and FOUNDATION Fieldbus: 9 to 32 V DC Supply voltage Ex ia 10.5 to 30 V DC Output 4 to 20 mA with superimposed HART protocol, PROFIBUS PA or FOUNDATION Fieldbus Options • Gold-rhodium coated process isolating diaphragm
- NACE-compliant materials
- Inspection certificate 3.1
- HistoROM®/M-DAT memory chip
- Separate housing Specialties • Metal-free measurement with PVDF connection
- Special cleaning of the transmitter to remove paint-wetting substances, for use in paint shops
- Process connections with minimum oil volume
- Gas-tight, elastomer-free
- Wide range of diaphragm seals
- For high media temperatures
- Process connections with minimum oil volume
- Completely welded versions 1) (temperature at process connection) 2) High-temperature version, see Product Configurator section "Additional options 1" or 110 "Additional options 2", option "T" 3) PMP71 and PMP75: lower temperatures on request
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 9 Measuring principle Devices with ceramic process isolating diaphragm (Ceraphire®) 1 2 3 4 p A0020465
1 Air pressure (relative pressure sensors)
2 Ceramic substrate
3 Electrodes
4 Ceramic process isolating diaphragm
The ceramic sensor is an oil-free sensor, i.e. the process pressure acts directly on the robust ceramic process isolating diaphragm and causes it to deflect. A pressure-dependent change in capacitance is measured at the electrodes of the ceramic substrate and the process isolating diaphragm. The measuring range is determined by the thickness of the ceramic process isolating diaphragm. Advantages:
- Guaranteed overload resistance up to 40 times the nominal pressure (see "OPL" column in table) → 12)
- The ultrapure 99.9% ceramic (Ceraphire®, see also "www.endress.com/ceraphire") ensures: – Extremely high chemical durability – Less relaxation – High mechanical durability
- Suitable for vacuums
- Secondary containment for enhanced integrity
- Process temperatures up to 150 °C (302 °F) Devices with metallic process isolating diaphragm p A0016448
1 Silicon measuring element, substrate
2 Wheatstone bridge
3 Channel with fill fluid
4 Metal process isolating diaphragm
The process pressure deflects the metal process isolating diaphragm of the sensor and a fill fluid transfers the pressure to a Wheatstone bridge (semiconductor technology). The pressure-dependent change in the bridge output voltage is measured and evaluated.
Cerabar S PMC71, PMP71, PMP75
10 Endress+Hauser
Advantages:
- Can be used for process pressures up to 700 bar (10 500 psi)
- High long-term stability
- Guaranteed overload resistance up to 4 times the nominal pressure
- Secondary containment for enhanced integrity
- Significantly reduced thermal effect e.g. compared to diaphragm seal systems with capillaries PMP75 The operating pressure acts on the process isolating diaphragm of the diaphragm seal and is transferred to the process isolating diaphragm of the sensor by a diaphragm seal fill fluid. The process isolating diaphragm is deflected and a fill fluid transfers the pressure to a resistance measuring bridge. The pressure-dependent change in the bridge output voltage is measured and evaluated. Advantages:
- Depending on the version, can be used for process pressures up to 400 bar (6 000 psi) and for extreme process temperatures
- High long-term stability
- Guaranteed overload resistance up to 4 times the nominal pressure
- Secondary containment for enhanced integrity Product design Level measurement (level, volume and mass): h = p ρ g h A0020466 h Height (level) p Pressure ρ Density of the medium g Gravitational constant Your benefits
- Selection of the level operating mode which is optimum for your application in the device software.
- Volume and mass measurements in any container shapes by means of a freely programmable characteristic curve.
- Choice of diverse level units with automatic unit conversion.
- A customized unit can be specified.
- Has a wide range of uses, e.g. – in the event of foam formation – in containers with agitators or screen fittings – in the event of liquid gases Applications suitable for custody transfer measurement The Parts Certificate is issued on the basis of the following standards:
- WELMEC guide 8.8 "General and Administrative Aspects of the Voluntary System of Modular Evaluation of Measuring instruments under the MID".
- OIML R117-1 Edition 2007 (E) "Dynamic measuring systems for liquids other than water".
- EN 12405-1/A1 Edition 2006 "Gas meters – Conversion devices – Part 1: Volume conversion".
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 11 Communication protocol • 4 to 20 mA with HART communication protocol
- PROFIBUS PA – The Endress+Hauser devices meet the requirements of the FISCO model. – Due to a low current consumption of 13 mA ± 1 mA, the following number of devices can be operated on one bus segment if installing as per FISCO: up to 7 devices for Ex ia, CSA IS and FM IS applications or up to 27 devices for all other applications e.g. in non-hazardous areas, Ex nA etc. Further information on PROFIBUS PA can be found in Operating Instructions BA00034S "PROFIBUS DP/PA: Guidelines for planning and commissioning" and in the PNO Guideline.
- FOUNDATION Fieldbus – The Endress+Hauser devices meet the requirements of the FISCO model. – Due to a low current consumption of 15.5 mA ± 1 mA, the following number of devices can be operated on one bus segment if installing as per FISCO: up to 6 devices for Ex ia, CSA IS and FM IS applications or up to 24 devices for all other applications e.g. in non-hazardous areas, Ex nA etc. Further information on FOUNDATION Fieldbus, such as requirements for bus system components, can be found in Operating Instructions BA00013S "FOUNDATION Fieldbus Overview".
Cerabar S PMC71, PMP71, PMP75
12 Endress+Hauser
Measured variable Measured process variables
- Absolute pressure
- Gauge pressure Calculated process variables Level (level, volume or mass) Measuring range PMC71 – with ceramic process isolating diaphragm (Ceraphire®) for gauge pressure Nominal value Range limit Lowest calibratable Span 1) MWP OPL Vacuum resistance Option 2) lower (LRL) upper (URL) [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [barabs (psiabs)] 1) Turndown > 100:1 on request or can be set on device 2) Product Configurator, order code for "Sensor range; sensor over pressure limit" PMC71 – with ceramic process isolating diaphragm (Ceraphire®) for absolute pressure Nominal value Range limit Lowest calibratable Span 1) MWP OPL Vacuum resistance Option 2) lower (LRL) upper (URL) [barabs (psiabs)] [barabs (psiabs)] [bar (psi)] [barabs (psiabs)] [barabs (psiabs)] [barabs (psiabs)] 2 bar (30 psi) 0 +2 (+30) 0.02 (0.3) 12 (180) 18 (270) 0 2K 10 bar (150 psi) 0 +10 (+150) 0.1 (1.5) 26.7 (400.5) 40 (600) 0 2P 40 bar (600 psi) 0 +40 (+600) 0.4 (6) 40 (600) 60 (900) 0 2S 1) Turndown > 100:1 on request or can be set on device 2) Product Configurator, order code for "Sensor range; sensor over pressure limit"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 13 PMP71 and PMP75 – metallic process isolating diaphragm for gauge pressure Nominal value Range limit Lowest calibratable Span 1) MWP OPL Vacuum resistance 2) Option 3) lower (LRL) upper (URL) Silicone oil/ Inert oil [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [barabs (psiabs)] 40 bar (600 psi) –1 (–15) +40 (+600) 0.4 (6) 100 (1500) 160 (2400) 1S 100 bar (1 500 psi) –1 (–15) +100 (+1500) 1.0 (15) 100 (1500) 400 (6000) 1U 400 bar (6 000 psi) –1 (–15) +400 (+6000) 4.0 (60) 400 (6000) 600 (9000) 1W 700 bar (10 500 psi) 4) –1 (–15) +700 (+10500) 7.0 (105) 700 (10500) 1050 (15750) 1X 1) Turndown > 100:1 on request or can be set on device 2) The vacuum resistance applies to the measuring cell under reference operating conditions. In the case of the PMP75, the pressure and temperature application limits of the selected filling oil must also be observed → 109. 3) Product Configurator, order code for "Sensor range; sensor over pressure limit" 4) PMP71 only, PMP75 on request PMP71 and PMP75 – metallic process isolating diaphragm for absolute pressure Nominal value Range limit Lowest calibratable Span 1) MWP OPL Vacuum resistance 2) Option 3) lower (LRL) upper (URL) Silicone oil/ Inert oil [barabs (psiabs)] [barabs (psiabs)] [bar (psi)] [barabs (psiabs)] [barabs (psiabs)] [barabs (psiabs)] 10 bar (150 psi) 0 +10 (+150) 0.1 (1.5) 26.7 (400.5) 40 (600) 2P 40 bar (600 psi) 0 +40 (+600) 0.4 (6) 100 (1500) 160 (2400) 2S 100 bar (1 500 psi) 0 +100 (+1500) 1.0 (15) 100 (1500) 400 (6000) 2U 400 bar (6 000 psi) 0 +400 (+6000) 4.0 (60) 400 (6000) 600 (9000) 2W 700 bar (10 500 psi) 4) 0 +700 (+10500) 7.0 (105) 700 (10500) 1050 (15750) 2X 1) Turndown > 100:1 on request or can be set on device 2) The vacuum resistance applies to the measuring cell under reference operating conditions. In the case of the PMP75, the pressure and temperature application limits of the selected filling oil must also be observed → 109. 3) Product Configurator, order code for "Sensor range; sensor over pressure limit" 4) PMP71 only, PMP75 on request
Cerabar S PMC71, PMP71, PMP75
14 Endress+Hauser
PMP71 - metallic process isolating diaphragms for absolute pressure with MID parts certificate Nominal value Range limit Min. WP for gas applications suitable for custody transfer measurement Min. WP for liquid applications suitable for custody transfer measurement MWP OPL Vacuum resistance 1) Option 2) lower (LRL) 3) upper (URL) 4) Silicone oil/ Inert oil [bar (psi)] [barabs (psiabs)] [barabs (psiabs)] [bar (psi)] [bar (psi)] [barabs (psiabs)] [barabs (psiabs)] [bar (psi)] (400.5) 40 (600) 0.01/0.04 (0.15/1) MP 50 (750) 0 +50 (750) 10 (150) 2.5 (37.5) 100 (1500) 400 (6000) 0.01/0.04 (0.15/1) MT 100 (1500) 0 +100 (1500) 5 (75) 5 (75) 100 (1500) 400 (6000) 0.01/0.04 (0.15/1) MU 1) The vacuum resistance applies to the measuring cell under reference operating conditions 2) Product Configurator, order code for "Sensor range; sensor over pressure limit" 3) By default, the device is set to a lower range limit of 0 bar. Please specify in the order if the lower range limit is to be set to a different default value. 4) Max. WP for gas and liquid applications suitable for custody transfer measurement
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 15 Output Output Signal • 4 to 20 mA with superimposed digital communication protocol HART, 2-wire
- Digital communication signal PROFIBUS PA (Profile 3.0), 2-wire – Signal coding: Manchester Bus Powered (MBP): Manchester II – Transmission rate: 31.25 KBit/s voltage mode
- Digital communication signal FOUNDATION Fieldbus, 2-wire – Signal coding: Manchester Bus Powered (MBP): Manchester II – Transmission rate: 31.25 KBit/s voltage mode Output Internal + LCD External + LCD Internal on on off off damp. damp. Display Display Sensor Sensor HART HART R R FIELD COMMUNICATION PROTOCOL FIELD COMMUNICATION PROTOCOL SW /min SW /min E+–E+– A0021280 Option 1) 4 to 20mA HART B A C 4 to 20mA HART, Li=0 E D F PROFIBUS PA N M O FOUNDATION Fieldbus Q P R 1) Product Configurator, order code for "Display, operation: " Signal range 4 to 20 mA 3.8 mA to 20.5 mA Signal on alarm As per NAMUR NE43 4 to 20 mA HART
- Max. alarm: Can be set between 21 to 23 mA (factory setting: 22 mA)
- Hold measured value: last measured value is held
- Min. alarm: 3.6 mA PROFIBUS PA Can be set in the Analog Input Block. Options:
- Last Valid Out Value (factory setting)
- Fail Safe Value
- Status bad FOUNDATION Fieldbus Can be set in the Analog Input Block. Options:
- Last Good Value
- Fail Safe Value (factory setting)
- Wrong Value
Cerabar S PMC71, PMP71, PMP75
16 Endress+Hauser
In order to guarantee sufficient terminal voltage in two-wire devices, a maximum load resistance R (including line resistance) must not be exceeded depending on the supply voltage U0 of the supply unit. In the following load diagrams, observe the position of the jumper and the explosion protection: U – 11.5 V U – 12 V 23 mA 23 mA [ ]/c87 [ ]/c87 11.5 1239 1217 1456 1434 804 782 369 347 T est /c163 /c163 B B U – 10.5 V U – 11 V R R L L max max 23 mA 23 mA /c163 /c163 10.5 U U U U [V] [V] [V] [V] 1282 1260 1500 1478 847 826 413 391 [ ]/c87 [ ]/c87 R R L L max max T est A A R R L L max max R R L L max max A0020467 A Jumper for 4 to 20 mA test signal set to "Non-test" position B Jumper for 4 to 20 mA test signal set to "Test" position 1 Power supply 10.5 (11.5) to 30 V DC for 1/2 G, 1 GD, 1/2 GD, FM IS, CSA IS, IECEx ia, NEPSI Ex ia 2 Power supply 10.5 (11.5) to 45 V DC for devices for non-hazardous areas, 1/2 D, 1/3 D, 2 G Ex d, 3 G Ex nA, FM XP, FM DIP, FM NI, CSA XP, CSA dust ignition-proof, NEPSI Ex d
3 Power supply 11 (12) to 45 V DC for PMC71, Ex d[ia], NEPSI Ex d[ia]
4 RLmax maximum load resistance
When operating via a handheld terminal or via a PC with an operating program, a minimum communication resistance of 250 Ω must be taken into account. Dead time, time constant Presentation of the dead time and the time constant:
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 17 I 63 % 100 % tt1 t2 90 % A0019786 Dynamic behavior, current output Type Measuring cell Dead time (t1) [ms] Time constant T63 (t2) [ms] Time constant T90 (t3) [ms] PMC71 max. All 90 120 276 PMP71 max. • 400 mbar (6 psi)
- ≥1 bar (15 psi) 45 • 70
- 35
- 161
- 81 PMP75 max. PMP71 + influence of the diaphragm seal Dynamic behavior digital output ( HART electronics) A typical burst rate of 300 ms results in the following behavior: Type Measuring cell Dead time (t1) [ms] Dead time (t1) [ms] + Time constant T63 (t2) [ms] Dead time (t1) [ms] + Time constant T90 (t3) [ms] PMC71 Min. All 250 370 436 max. 1050 1170 1236 PMP71 Min. • 400 mbar (6 psi)
- ≥1 bar (15 psi) 205 • 275
- 240
- 321
- 241 max. • 400 mbar (6 psi)
- ≥1 bar (15 psi) 1005 • 1075
- 1040
- 1121
- 1041 PMP75 max. PMP71 + influence of the diaphragm seal Reading cycle
- Acyclic: max. 3/s, typically 1/s (depending on command # and number of preambles)
- Cyclic (burst): max. 3/s, typically 2/s The device commands the BURST MODE functionality for cyclical value transmission via the HART communication protocol. Cycle time (update time) Cyclic (burst): min. 300 ms Response time
- Acyclic: min. 330 ms, typically 590 ms (depending on command # and number of preambles)
- Cyclic (burst): min. 160 ms, typically 350 ms (depending on command # and number of preambles) Dynamic behavior PROFIBUS PA A typical PLC cycle time of 1 s results in the following behavior: Type Measuring cell Dead time (t1) [ms] Dead time (t1) [ms] + Time constant T63 (t2) [ms] Dead time (t1) [ms] + Time constant T90 (t3) [ms] PMC71 Min. All 125 245 311 max. 1325 1445 1511
Cerabar S PMC71, PMP71, PMP75
18 Endress+Hauser
Type Measuring cell Dead time (t1) [ms] Dead time (t1) [ms] + Time constant T63 (t2) [ms] Dead time (t1) [ms] + Time constant T90 (t3) [ms] PMP71 Min. • 400 mbar (6 psi)
- ≥1 bar (15 psi) 80 • 150
- 115
- 196
- 116 max. • 400 mbar (6 psi)
- ≥1 bar (15 psi) 1280 • 1350
- 1315
- 1396
- 1316 PMP75 max. PMP71 + influence of the diaphragm seal Reading cycle (PLC)
- Acyclic: Typically 25/s
- Cyclic: Typically 30/s (depending on the number and type of the function blocks used in the closed-control loop) Cycle time (update time) Min. 200 ms The cycle time in a bus segment in cyclic data communication depends on the number of devices, on the segment coupler used and on the internal PLC cycle time. A new measured value can be determined up to five times a second. Response time
- Acyclic: Approx. 60 ms to 70 ms (depending on Min. Slave Interval)
- Cyclic: Approx. 10 ms to 13 ms (depending on Min. Slave Interval) Dynamic behavior FOUNDATION Fieldbus A typical configuration for the macro cycle time (host system) of 1 s results in the following behavior: Type Measuring cell Dead time (t1) [ms] Dead time (t1) [ms] + Time constant T63 (t2) [ms] Dead time (t1) [ms] + Time constant T90 (t3) [ms] PMC71 Min. All 135 255 321 max. 1135 1255 1321 PMP71 Min. • 400 mbar (6 psi)
- ≥1 bar (15 psi) 90 • 160
- 125
- 206
- 126 max. • 400 mbar (6 psi)
- ≥1 bar (15 psi) 1090 • 1160
- 1125
- 1206
- 1126 PMP75 max. PMP71 + influence of the diaphragm seal Reading cycle
- Acyclic: Typically 10/s
- Cyclic: max. 10/s (dependent on the number and type of function blocks used in a closed-control loop) Cycle time (update time) Cyclic: Min. 100 ms Response time
- Acyclic: Typically 100 ms (for standard bus parameter settings)
- Cyclic: max. 20 ms (for standard bus parameter settings) Damping A damping affects all outputs (output signal, display):
- Via onsite display, handheld terminal or PC with operating program, continuous from 0 to 999 s
- Also for HART and PROFIBUS PA: Via DIP switch on the electronic insert, switch position "on" = set value and "off"
- Factory setting: 2 s
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 19 Alarm current Description Option 1) Min alarm current J HART burst mode PV Min alarm current + HART burst mode PV 1) Product Configurator, order code for "Additional options 1" and "Additional options 2" Firmware version Description Option 1) 02.20.zz, HART 7, DevRev22 72 02.11.zz, HART 5, DevRev21 73 04.00.zz, FF, DevRev07 74 04.01.zz, PROFIBUS PA, DevRev03 75 02.10.zz, HART 5, DevRev21 76 03.00.zz, FF, DevRev06 77 04.00.zz, PROFIBUS PA 78 1) Product Configurator, order code for "Firmware version" HART protocol-specific data Manufacturer ID 17 (11 hex) Device type code 24 (18 hex) Device revision • 21 (15 hex) - SW version 02.1y.zz - HART specification 5
- 22 (16 hex) - SW version 02.2y.zz - HART specification 7 HART specification • 5
- 7 DD revision • 4 (Russian in language selection) for device revision 21
- 3 (Dutch in language selection) for device revision 21
- 1 for device revision 22 Device description files (DTM, DD) Information and files at:
- www.endress.com
- www.hartcomm.org HART load Min. 250 Ω HART device variables The measured values are assigned to the device variables as follows: Measured values for PV (primary variable)
- Pressure
- Level
- Tank content Measured values for SV, TV (second and third variable) Pressure Measured values for QV (fourth variable) Temperature Supported functions • Burst mode
- Additional transmitter status
- Device locking
- Alternative operating modes PROFIBUS PA protocol- specific data Manufacturer ID 17 (11 hex) Identification number 1541 hex
Cerabar S PMC71, PMP71, PMP75
20 Endress+Hauser
Profile version 3.0
- SW version 03.00.zz
- SW version 04.00.zz 3.02 SW version 04.01.zz ( device revision 3) Compatibility with SW version 03.00.zz and higher. GSD revision • 4 (SW version 3.00.zz and 4.00.zz)
- 5 (device revision 3) DD revision • 1 (SW version 3.00.zz and 4.00.zz)
- 1 (device revision 3) GSD file Information and files at:
- www.endress.com
- www.profibus.org DD files Output values Measured value for PV (via Analog Input Function Block)
- Pressure
- Level
- Tank content Measured value for SV
- Pressure
- Temperature Input values Input value sent from PLC, can be shown on display Supported functions • Identification & maintenance Simplest device identifier on the control system and nameplate
- Condensed status (only with Profile Version 3.02)
- Automatic ID number adjustment and switchable to the following ID numbers (only with Profile Version 3.02): – 9700: Profile-specific transmitter identification number with the "Classic" or "Condensed" status". – 1501: Compatibility mode for the old Cerabar S generation (PMC731, PMP731, PMC631, PMP635). – 1541: Identification number for the new Cerabar S generation (PMC71, PMP71, PMP75).
- Device locking: The device can be locked by hardware or software. FOUNDATION Fieldbus protocol-specific data Manufacturer ID 452B48 hex Device type 1007 hex Device revision • 6 - SW version 03.00.zz
- 7 - SW version 04.00.zz (FF-912) DD revision • 3 (device revision 6)
- 2 (device revision 7) CFF revision • 4 (device revision 6)
- 1 (device revision 7) DD files Information and files at:
- www.endress.com
- www.fieldbus.org CFF files Device tester version (ITK version) • 5.0 (device revision 6)
- 6.01 (device revision 7) Number of ITK test campaign • IT054600 (Device Revision 6)
- IT085500 (Device Revision 7) Link Master (LAS) capable Yes Choice of "Link Master" and "Basic Device" Yes, factory setting is Basic Device Node address Factory setting: 247 (F7 hex)
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 21 Supported functions Field diagnostics profile (only with FF912) The following methods are supported:
- Restart
- Configure error as warning or alarm
- HistoROM
- Peakhold
- Alarm info
- Sensor trim Number of VCRs • 44 (device revision 6)
- 24 (device revision 7) Number of link objects in VFD 50 Virtual communication references (VCRs) Device revision 6 Device revision 7 Permanent entries 44 1 Client VCRs 0 0 Server VCRs 5 10 Source VCRs 8 43 Sink VCRs 0 0 Subscriber VCRs 12 43 Publisher VCRs 19 43 Link settings Device revision 6 Device revision 7 Slot time 4 4 Min. Inter PDU delay 12 10 Max. response delay 10 10 Transducer Blocks Block Content Output values TRD1 Block Contains all parameters related to the measurement
- Pressure or level (channel 1)
- Process temperature (channel 2) Service Block Contains service information • Pressure after damping (channel 3)
- Pressure peakhold indicator (channel 4)
- Counter for max. pressure transgressions (channel Diagnostic Block Contains diagnostic information Error code via DI channels (channel 0 to 16) Display Block Contains parameters to configure the onsite display No output values
Cerabar S PMC71, PMP71, PMP75
22 Endress+Hauser
Execution time Functionality Device Revision 6 Device Revision 7 Device Revision 6 Device Revision 7 Resource Block This block contains all the data that uniquely identifies the device; it is an electronic version of a nameplate for the device. 1 enhanced enhanced Analog Input Block 1 Analog Input Block 2 The AI Block receives the measuring data from the Sensor Block, (selectable via a channel number) and makes the data available to other function blocks at its output. Enhancement: Digital outputs for process alarms, fail safe mode 2 45 ms 45 ms (without trend and alarm reports) enhanced enhanced Digital Input Block This block contains the discrete data of the Diagnose Block (selectable via a channel number 0 to 16) and provides them for other blocks at the output. 1 40 ms 30 ms standard enhanced Digital Output Block This block converts the discrete input and thus initiates an action (selectable via a channel number) in the DP Flow Block or in the Service Block. Channel 1 resets the counter for max. pressure transgressions. 1 60 ms 40 ms standard enhanced PID Block This block is used as a proportional-integral-derivative controller and can be used universally for closed-loop-control in the field. It enables cascade mode and feedforward control. Input IN can be indicated on the display. The selection is performed in the Display Block (DISPLAY_MAIN_LINE_CONTENT). 1 120 ms 70 ms standard enhanced Arithmetic Block This block is designed to permit simple use of popular measurement math functions. The user does not have to know how to write equations. The math algorithm is selected by name, chosen by the user for the function to be performed. 1 50 ms 40 ms standard enhanced Input Selector Block The Input Selector Block facilitates the selection of up to four inputs and generates an output based on the configured action. This block normally receives its inputs from AI Blocks. The block performs maximum, minimum, average and ‘first good’ signal selection. Inputs IN1 to IN4 can be indicated on the display. The selection is performed in the Display Block (DISPLAY_MAIN_LINE_CONTENT). 1 35 ms 35 ms standard enhanced Signal Characterizer Block The Signal Characterizer Block has two sections, each with an output that is a non-linear function of the respective input. The non-linear function is generated by a single look-up table with 21 arbitrary x-y pairs. 1 30 ms 40 ms standard enhanced Integrator Block The Integrator Block integrates a variable as a function of the time or accumulates the counts from a Pulse Input Block. The block can be used as a totalizer that counts up until a reset, or as a batch totalizer whereby the integrated value is compared against a target value generated before or during the control routine and generates a binary signal when the target value is reached. 1 35 ms 40 ms standard enhanced Analog Alarm Block This block contains all process alarm conditions (working like a comparator) and represents them at the output. 1 35 ms 35 ms standard enhanced Additional function block information: Instantiatable function blocks YES YES Number of additional instantiatable function blocks 11 5
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 23 Power supply LWARNING Electrical safety is compromised by an incorrect connection. ‣ When using the measuring device in hazardous areas, the relevant national standards and regulations as well as the Safety Instructions or Installation or Control Drawings must be observed → 122. ‣ All explosion protection data are given in separate documentation which is available upon request. The Ex documentation is supplied as standard with all Ex-systems → 122. ‣ Devices with integrated overvoltage protection must be grounded → 27. ‣ Protective circuits against reverse polarity, HF influences and overvoltage peaks are integrated. Terminal assignment 4 to 20 mA HART 4... 20mA Test Test Test 4... 20mA Test A0019989
1 Housing
2 Supply voltage
4 Devices with integrated overvoltage protection are labeled "OVP" (overvoltage protection) here.
5 External ground terminal
6 4 to 20 mA test signal between positive and test terminal
7 Internal ground terminal
8 Jumper for 4 to 20 mA test signal → 24
PROFIBUS PA and FOUNDATION Fieldbus A0020158 3 Devices with integrated overvoltage protection are labeled "OVP" (overvoltage protection) here.
4 External ground terminal
5 Internal ground terminal
Cerabar S PMC71, PMP71, PMP75
24 Endress+Hauser
Supply voltage 4 to 20 mA HART Electronic version Jumper for 4 to 20 mA test signal in "Test" position (delivery status) Jumper for 4 to 20 mA test signal in "Non-test" position Version for non- hazardous area 11.5 to 45 V DC 10.5 to 45 V DC Intrinsically safe 11.5 to 30 V DC 10.5 to 30 V DC
- Other types of protection
- Devices without certificate 11.5 to 45 V DC (versions with 35 V DC plug-in connection) 10.5 to 45 V DC (versions with 35 V DC plug-in connection) Measuring a 4 to 20 mA test signal Jumper position for test signal
Description
- Measurement of 4 to 20 mA test signal via the positive and test terminal: Possible. (Thus, the output current can be measured without interruption via the diode.)
- Delivery status
- Minimum supply voltage: 11.5 V DC T est A0019993
- Measurement of 4 to 20 mA test signal via positive and test terminal: Not possible.
- Minimum supply voltage: 10.5 V DC PROFIBUS PA
- Version for non-hazardous areas: 9 to 32 V DC
- Ex ia: 10.5 to 30 V DC FOUNDATION Fieldbus
- Version for non-hazardous areas: 9 to 32 V DC
- Ex ia: 10.5 to 30 V DC Current consumption • PROFIBUS PA: 13 mA ±1 mA, switch-on current corresponds to IEC 61158-2, Clause 21
- FOUNDATION Fieldbus: 15.5 mA ±1 mA, switch-on current corresponds to IEC 61158-2, Clause
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 25 Electrical connection PROFIBUS PA The digital communication signal is transmitted to the bus via a 2-wire connection. The bus also provides the power supply. For further information on the network structure and grounding, and for further bus system components such as bus cables, see the relevant documentation, e.g. Operating Instructions BA00034S "PROFIBUS DP/PA: Guidelines for planning and commissioning" and the PNO Guideline. FOUNDATION Fieldbus The digital communication signal is transmitted to the bus via a 2-wire connection. The bus also provides the power supply. For further information on the network structure and grounding and for further bus system components such as bus cables, see the relevant documentation, e.g. Operating Instructions BA00013S "FOUNDATION Fieldbus Overview" and the FOUNDATION Fieldbus Guideline. Terminals • Supply voltage and internal ground terminal: 0.5 to 2.5 mm2 (20 to 14 AWG)
- External ground terminal: 0.5 to 4 mm2 (20 to 12 AWG) Cable entries Approval Cable gland Clamping area Standard, II 1/2 G Ex ia, IS Plastic M20x1.5 5 to 10 mm (0.2 to 0.39 in) ATEX II 1/2 D, II 1/3 D, II 1/2 GD Ex ia, II 1 GD Ex ia, II 3 G Ex nA Metal M20x1.5 (Ex e) 7 to 10.5 mm (0.28 to 0.41 in) For additional technical data, see section on housing → 54 Device plug Devices with Harting plug Han7D Han7D + – – + A B A0019990 A Electrical connection for devices with Harting plug Han7D B View of the plug-in connection on the device Material: CuZn, gold-plated plug-in jack and plug Cable version connection PE 1 2 3 4 A0019991 1 rd = red 2 bk = black 3 gnye = green 4 4 to 20 mA
Cerabar S PMC71, PMP71, PMP75
26 Endress+Hauser
1 Signal +
2 Not assigned
3 Signal –
4 Ground
Endress+Hauser offers the following accessories for devices with an M12 plug: Plug-in jack M 12x1, straight
- Material: Body PA; coupling nut CuZn, nickel-plated
- Degree of protection (fully locked): IP67
- Order number: 52006263 Plug-in jack M 12x1, elbowed
- Material: Body PBT/PA; coupling nut GD-Zn, nickel-plated
- Degree of protection (fully locked): IP67
- Order number: 71114212 Cable 4x0.34 mm2 (20 AWG) with M12 socket, elbowed, screw plug, length 5 m (16 ft)
- Material: Body PUR; coupling nut CuSn/Ni; cable PVC
- Degree of protection (fully locked): IP67
- Order number: 52010285 Devices with 7/8" plug A0011176
1 Signal –
2 Signal +
3 Not assigned
4 Shielding
External thread: 7/8 - 16 UNC
- Material: 316L (1.4401)
- Degree of protection: IP68 Cable specification HART
- Endress+Hauser recommends using shielded, twisted-pair two-wire cables.
- Cable outer diameter: 5 to 9 mm (0.2 to 0.35 in) depending on the cable entry used → 25 PROFIBUS PA Use a twisted, shielded twin-core cable, preferably cable type A. For further information regarding cable specifications, see the Operating Instructions BA00034S "PROFIBUS DP/PA: Guidelines for planning and commissioning", the PNO guideline 2.092 "PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP).
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 27 FOUNDATION Fieldbus Use a twisted, shielded twin-core cable, preferably cable type A. For further information on the cable specifications, see Operating Instructions BA00013S "FOUNDATION Fieldbus Overview", FOUNDATION Fieldbus Guideline and IEC 61158-2 (MBP). Start-up current 12 mA Residual ripple Without influence on 4 to 20 mA signal up to ±5% residual ripple within the permitted voltage range [according to HART hardware specification HCF_SPEC-54 (DIN IEC 60381-1)]. Overvoltage protection (optional)
- Overvoltage protection: – Nominal functioning DC voltage: 600 V – Nominal discharge current: 10 kA
- Surge current check î = 20 kA satisfied as per DIN EN 60079-14: 8/20 μs
- Arrester AC current check I = 10 A satisfied Ordering information: Product Configurator, order code for "Additional options 1" or Additional options 2", version "M" NOTICE Device could be destroyed! ‣ Devices with integrated overvoltage protection must be grounded. Influence of power supply ≤0.0006 % of URL/1 V
Cerabar S PMC71, PMP71, PMP75
28 Endress+Hauser
Performance characteristics of PMP71/PMP75 metallic pressure transmitter (Sensor module + Electronic module) Reference operating conditions
- As per IEC 60770
- Ambient temperature TA = constant, in the range +21 to +33 °C (+70 to +91 °F)
- Humidity j = constant, in the range: 5 to 80 % RH ± 5 %
- Ambient pressure pA = constant, in range: 860 to 1 060 mbar (12.47 to 15.37 psi)
- Position of measuring cell: horizontal ±1° (see also "Influence of the installation position" section → 42)
- Input of LOW SENSOR TRIM and HIGH SENSOR TRIM for lower range value and upper range value
- Zero based span
- PMP71 material of the process isolating diaphragm: AISI 316L (1.4435)
- PMC71 material of the process isolating diaphragm: Al2O3 aluminum oxide ceramic FDA, ultrapure 99.9 %
- PMP71 filling oil: silicone oil
- Supply voltage: 24 V DC ±3 V DC
- Load with HART: 250 Ω
- Turn down, TD = URL /|URV |; zero based span Preamble The performance characteristic of the transmitter refers to "Accuracy of the transmitter". The factors influencing accuracy can be divided into 2 groups
- Total performance of transmitter → 30
- Installation dependent influencing factors → 42 Total performance of transmitter Transmitter total performance → 30 comprises the reference accuracy, ambient temperature effect and it is calculated by the formula Total performance = ± √ ( (E1)² + (E2)² ) E1 = Reference accuracy → 29 E2 = Ambient temperature effect per 28°C (per 50°F) → 31
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 29 Reference accuracy [E1] Reference accuracy 1) Measuring cell PMP71 PMP75 2) Standard Reference accuracy Platinum 3) Reference accuracy Standard Reference accuracy 400 mbar (6 psi) • TD 1:1
- TD > 1:1 ±0.05 ±0.05 · TD
- TD 1:1
- TD > 1:1 = ±0.025 ±0.05 · TD
- TD 1:1
- TD > 1:1 ±0.15 ±0.15 · TD 1 bar (15 psi) • TD ≤ 2.5:1
- TD > 2.5:1 ±0.05 ±0.02 · TD
- TD 1:1
- TD > 1:1 = ±0.025 See standard
- TD ≤ 2.5:1
- TD > 2.5:1 ±0.075 ±0.03 · TD 2 bar (30 psi) • TD ≤ 5:1
- TD > 5:1 ±0.05 ±0.01 · TD
- TD 1:1
- TD > 1:1 = ±0.025 See standard
- TD ≤ 5:1
- TD > 5:1 ±0.075 ±0.015 · TD 4 bar (60 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.025 See standard
- TD ≤ 10:1
- TD > 10:1 ±0.075 ±0.0075 · TD 10 bar (150 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.025 See standard
- TD ≤ 10:1
- TD > 10:1 ±0.075 ±0.0075 · TD 40 bar (600 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.025 See standard
- TD ≤ 10:1
- TD > 10:1 ±0.075 ±0.0075 · TD 100 bar (1 500 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.035 See standard
- TD ≤ 10:1
- TD > 10:1 ±0.075 ±0.0075 · TD 400 bar (6 000 psi) • TD ≤ 5:1
- TD > 5:1 ±0.1 ±0.02 · TD
- TD 1:1
- TD > 1:1 = ±0.065 See standard
- TD ≤ 5:1
- TD > 5:1 ±0.15 ±0.03 · TD 700 bar (10 500 psi) • TD ≤ 5:1
- TD > 5:1 ±0.1 ±0.02 · TD
- TD 1:1
- TD > 1:1 = ±0.065 See standard
- TD ≤ 5:1
- TD > 5:1 ±0.15 ±0.03 · TD 1) The reference accuracy contains the non-linearity [DIN EN 61298-2 3.11] including the hysteresis [DIN EN 61298-23.13] and non-repeatability [DIN EN 61298-2 3.11] in accordance with the limit point method as per [DIN EN 60770]. The specifications refer to the calibrated span/Upper Range Value (URV). 2) PMP75: Total performance excludes diaphragm seal errors. Diaphragm seal errors can be calculated separatly in applicator diaphragm seal calculation module. Link to online tool Applicator: www.endress.com/applicator → Sizing Diaphragm Seal 3) Platinum version not for flush-mounted process connections G ½ and M20 Please refer to the next chapter “Detailed performance explanation and calculation” for further explanations on "Ambient temperature effect".
Cerabar S PMC71, PMP71, PMP75
30 Endress+Hauser
Total performance – Specification values Total performance - PMP71 Measuring cell Standard 1) 2) 3) Platinum 1) 2) 3) TD 1:1 TD 2:1 TD 3:1 TD 4:1 TD 5:1 TD 1:1 TD 2:1 TD 3:1 TD 4:1 TD 5:1 1) PMP75: Total performance excludes diaphragm seal errors. Those diaphragm seal errors will be calculated separately in the applicator diaphragm seal calculation module. Link to online tool Applicator: www.endress.com/applicator → Sizing Diaphragm Seal 2) The specification values apply for the temperature range per 28 °C (per 50 °F) with reference to + 25 °C (77 °F) for all measuring cells. The specification values are applicable to analog output (i.e. inclusion of electronics error). 3) The specifications refer to the calibrated span/Upper Range Value (URV). Long-term stability Long term stability - PMP71 Measuring cell Standard & platinum % of the upper range limit / 1 year 5 years 10 years 400 mbar (6 psi) ± 0.05 ± 0.07 ± 0.10 1 bar (15 psi) ± 0.05 ± 0.07 ± 0.10 2 bar (30 psi) ± 0.07 ± 0.12 ± 0.15 4 bar (60 psi) ± 0.05 ± 0.07 ± 0.10 10 bar (150 psi) ± 0.05 ± 0.07 ± 0.10 40 bar (600 psi) ± 0.05 ± 0.07 ± 0.10 100 bar (1 500 psi) ± 0.05 ± 0.07 ± 0.10 400 bar (6 000 psi) ± 0.05 ± 0.07 ± 0.10 700 bar (10 500 psi) ± 0.05 ± 0.07 ± 0.10
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 31 Detailed Performance Explanation and Calculation To calculate the total performance outside the –3 to +53 °C (+27 to +127 °F) temperature range or for a process isolating diaphragm material other than 1.4435/316L or Alloy C 276, please refer to the following sections: "Ambient temperature effect" and "Calculation of total performance". Ambient temperature effect [E2]
- E2 = (E2M · CF1 ) + E2E
- E2M = Main temperature error
- CF1 = Temperature range correction factor
- E2E = Electronics error for analog current output E2M - Main temperature error The output changes due to the effect of ambient temperature [IEC 61298-3] with respect to reference temperature [DIN 16086]. The values specify the maximum error due to min./max. ambient or process temperature conditions. The specified error is applicable to the membrane material AISI 316L. Ambient temperature effect per 28 °C 1) (Per 50 °F) - PMP71 2) Measuring cell Standard Platinum 700 bar (10 500 psi) ± (0.015 · TD + 0.06) ± (0.015 · TD + 0.06) 1) Valid for the temperature range 25 °C ± per 28 °C (77 °F ± per 50 °F). In contrast, the electronics error applies for the entire transmitter temperature range –40 to +85 °C (–40 to +185 °F). 2) PMP75: Total performance excludes diaphragm seal errors. Those diaphragm seal errors will be calculated separately in the applicator diaphragm seal calculation module. Link to online tool Applicator: www.endress.com/applicator → Sizing Diaphragm Seal CF1 - Temperature range correction factor Measuring cell Temperature range Factor, CF1 400 mbar (6 psi), 1 bar (15 psi), 2 bar (30 psi), 4 bar (60 psi), 10 bar (150 psi), 40 bar (600 psi), 100 bar (1 500 psi), 400 bar (6 000 psi), 700 bar (10 500 psi) 25°C ± per 28°C (77°F ± per 50°F) 1 –32 to –4 °C (–26 to +25 °F) and E2E - Electronics error Measuring cell Electronics Error (%) For all measuring cells Digital output (HART/PA/FF) 0 % Analog output (4 to 20 mA/1 to 5V) 0.05 % Total error Total error = total performance + long-term stability
Cerabar S PMC71, PMP71, PMP75
32 Endress+Hauser
Total performance in % of the set span with TD1:1 → 30 Long-term stability in % of the upper range limit (URL) → 30 Total error - PMP71 Measuring cell Standard 1) 2) Platinum 1) 2) AISI 316L or Alloy C % of the upper range limit / AISI 316L or Alloy C % of the upper range limit / 1 year 5 years 10 years 1 year 5 years 10 years 1) Valid for membrane material AISI 316L (1.4435), Alloy C 276. 2) The specification values are applicable to the temperature range per 28 °C (per 50 °F) with reference to + 25°C (77°F) for all measuring cells. The specification values are applicable to analog output (i.e. inclusion of electronics error).
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 33 Performance characteristics of PMC71 ceramic pressure transmitter (Sensor module + Electronic module) Reference operating conditions
- As per IEC 60770
- Ambient temperature TA = constant, in the range +21 to +33 °C (+70 to +91 °F)
- Humidity j = constant, in the range: 5 to 80 % RH ± 5 %
- Ambient pressure pA = constant, in range: 860 to 1 060 mbar (12.47 to 15.37 psi)
- Position of measuring cell: horizontal ±1° (see also "Influence of the installation position" section → 42)
- Input of LOW SENSOR TRIM and HIGH SENSOR TRIM for lower range value and upper range value
- Zero based span
- PMC71 material of the process isolating diaphragm: Al2O3 aluminum oxide ceramic FDA, ultrapure 99.9 %
- Supply voltage: 24 V DC ±3 V DC
- Load with HART: 250 Ω
- Turn down, TD = URL /|URV |; zero based span Preamble The performance characteristic of the transmitter refers to "Accuracy of the transmitter". The factors influencing accuracy can be divided into 2 groups
- Total performance of transmitter → 33
- Installation dependent influencing factors → 42 Total performance of transmitter Transmitter total performance → 35 comprises the reference accuracy, ambient temperature effect and it is calculated by the formula Total performance = ± √ ( (E1)² + (E2)² ) E1 = Reference accuracy → 34 E2 = Ambient temperature effect per 28°C (per 50°F) → 36
Cerabar S PMC71, PMP71, PMP75
34 Endress+Hauser
Reference accuracy [E1] Measuring cell PMC71 standard 1) Gauge pressure and absolute pressure Reference accuracy PMC71 platinum 1) Gauge pressure and absolute pressure Reference accuracy 100 mbar (1.5 psi) • TD ≤ 10:1
- TD > 10:1 ±0.075 ±0.0075 · TD
- TD 1:1
- TD > 1:1 = ±0.05 See standard 250 mbar (3.75 psi) • TD ≤ 10:1
- TD > 10:1 ±0.075 ±0.0075 · TD
- TD 1:1
- TD > 1:1 = ±0.05 See standard 400 mbar (6 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.035 See standard 1 bar (15 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.035 See standard 2 bar (30 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.025 See standard 4 bar (60 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.025 See standard 10 bar (150 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.035 See standard 40 bar (600 psi) • TD ≤ 10:1
- TD > 10:1 ±0.05 ±0.005 · TD
- TD 1:1
- TD > 1:1 = ±0.035 See standard 1) The reference accuracy contains the non-linearity [DIN EN 61298-2 3.11] including the hysteresis [DIN EN 61298-23.13] and non-repeatability [DIN EN 61298-2 3.11] in accordance with the limit point method as per [DIN EN 60770]. The specifications refer to the calibrated span/Upper Range Value (URV).
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 35 Total performance – Specification values Measuring cell PMC71 standard 1) 2) Gauge pressure and absolute pressure PMC71 platinum 2) Gauge pressure and absolute pressure TD 1:1 TD 2:1 TD 3:1 TD 4:1 TD 5:1 TD 1:1 TD 2:1 TD 3:1 TD 4:1 TD 5:1 1) The specification values are applicable to the temperature range per 28 °C (per 50 °F) with reference to + 25°C (77°F) for all measuring cells. The specification values are applicable to analog output (i.e. inclusion of electronics error). 2) The specifications refer to the calibrated span/Upper Range Value (URV). Long-term stability Measuring cell PMC71 - % of the upper range limit / Gauge pressure PMC71 - % of the upper range limit / Absolute pressure 1 year 5 years 10 years 1 year 5 years 10 years
Cerabar S PMC71, PMP71, PMP75
36 Endress+Hauser
Explanation and Calculation To calculate the total performance outside the –3 to +53 °C (+27 to +127 °F) temperature range or for a process isolating diaphragm material other than 1.4435/316L or Alloy C 276, please refer to the following sections: "Ambient temperature effect" and "Calculation of total performance". PMC71 standard version - Ambient temperature effect [E2]
- E2 = (E2M · CF1 ) + E2E
- E2M = Main temperature error
- CF1 = Temperature range correction factor
- E2E = Electronics error for analog current output E2M - Main temperature error The output changes due to the effect of ambient temperature [IEC 61298-3] with respect to reference temperature [DIN 16086]. The values specify the maximum error due to min./max. ambient or process temperature conditions. Ambient temperature effect per 28 °C 1) (per 50 °F) Measuring cell Standard Gauge pressure and absolute pressure Platinum Gauge pressure and absolute pressure 1) Valid for the temperature range 25 °C ± per 28 °C (77 °F ± per 50 °F) CF1 - Temperature range correction factor Measuring cell Temperature range Factor, CF1 100 mbar (1.5 psi) 250 mbar (3.75 psi) 400 mbar (6 psi) 1 bar (15 psi) 2 bar (30 psi) 4 bar (60 psi) 10 bar (150 psi) 40 bar (600 psi) 25°C ± per 28°C (77°F ± per 50°F) 1 –20 to –4 °C (–4 to +25 °F) and 2.4 +83 to +125 °C (+181 to +257 °F) 2.4 E2E - Electronics error Measuring cell Electronics Error (%) For all measuring cells Digital output (HART/PA/FF) 0.05 % Analog output (4 to 20 mA/1 to 5V) 0.05 %
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 37 PMC71 high-temperature version - Ambient temperature effect [E2]
- E2 = E2M
- E2M = Main temperature error E2M - Main temperature error The output changes due to the effect of ambient temperature [IEC 61298-3] with respect to reference temperature [DIN 16086]. The values specify the maximum error due to min./max. ambient or process temperature conditions. Influence of ambient temperature–10 to +60 °C (+14 to +140 °F) Measuring cell Standard Gauge pressure and absolute pressure % of the set span Platinum Gauge pressure and absolute pressure % of the set span Influence of ambient temperature+60 to +150 °C (+140 to +302 °F) Measuring cell Standard % of the set span Platinum % of the set span Gauge pressure Absolute pressure Gauge pressure Absolute pressure
Cerabar S PMC71, PMP71, PMP75
38 Endress+Hauser
Total error Total error = total performance + long-term stability Total performance in % of the set span with TD1:1 → 35 Long-term stability in % of the upper range limit (URL) → 35 Measuring cell Gauge pressure - Standard 1) Gauge pressure - platinum 1) PMC71 - % of the upper range limit / PMC71 - % of the upper range limit / 1 year 5 years 10 years 1 year 5 years 10 years 1) The specification values are applicable to the temperature range per 28 °C (per 50 °F) with reference to + 25°C (77°F) for all measuring cells. The specification values are applicable to analog output (i.e. inclusion of electronics error). Measuring cell Absolute pressure - Standard 1) Absolute pressure - Platinum PMC71 - % of the upper range limit / PMC71 - % of the upper range limit / 1 year 5 years 10 years 1 year 5 years 10 years 1) The specification values are applicable to the temperature range per 28 °C (per 50 °F) with reference to + 25°C (77°F) for all measuring cells. The specification values are applicable to analog output (i.e. inclusion of electronics error).
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 39 Performance characteristics - calculation example and additional information Calculating the total performance in 4 steps Data (example with PMP71) Measuring Conditions / Device configuration Measuring range (URV) 5 bar (75 psi) Min./max. temp. of pressure transmitter (ambient/process) Ambient temp.:0 to 45 °C (32 to 113 °F) Max. process temp.: 50 °C (122 °F) Membrane material AISI 316L Reference accuracy (± 0.05%) Standard PMP71 - Suitable measuring cell (upper range limit, URL) 10 bar (150 psi) with TD 2:1 Output Signal 4 to 20 mA Formula Total performance = ± √ ( (E1)² + (E2)² ) E1 = Reference accuracy → 29 E2 = Ambient temperature effect per 28°C (per 50°F) → 31 Calculation Step 1: Calculation of the turn down → 7 Turn down, TD = URL / [URV] = 10 bar (150 psi)/5 bar (75 psi) = TD = 2:1 Step 2: Calculation of the reference accuracy (E1) → 29 For the measuring conditions, Reference accuracy, E1 = ± 0.05 (% of the set span) = ±0.0025 bar (0.03625 psi) E1 = ± 0.05 (% of the set span) (or) ±0.0025 bar (0.03625 psi)
Cerabar S PMC71, PMP71, PMP75
40 Endress+Hauser
Step 3: Calculation of the ambient temperature effect [E2 = (E2M · CF1) + E2E] → 31 For the measuring conditions, Main temperature error, E2M ± (0.03 · TD + 0.03) % of the set span Temperature correction factor, CF1 = 1 Electronics error , E2E = 0.05 % Ambient temperature effect, E2 = ± [(0.03 · TD + 0.03) · 1] +0.05 = ± 0.14 (% of the set span) = ±0.007 bar (0.1015 psi) E2 = ± 0.14 (% of the set span) (or) ±0.007 bar (0.1015 psi) A B C 0.0 0.05 0.1 0.15 0.2 0.25 0.3 [%] -40 -40 -30 -22 -10 +14 +32 +10 +50 +20 +68 +30 +86 +40 +104 +50 +122 +60 +140 +70 +158 +80 +176 +90 +194 -20 [°C] [°F] D A0031299 A Error (% of the set span) B Temperature C 316L process isolating diaphragm D Ambient temperature effect: 0.14 (% of the set span)
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 41 Step 4: Calculation of the total performance Total performance = ± √ ((E1)² + (E2)² ) E1 = Reference accuracy E2 = Ambient Temperature effect per 28°C Total performance, if calculated in mbar = ± √ ((0.0025)² + (0.007)² ) = ± 0.0074 bar (0.1073 psi) or 7.4 mbar (0.1073 psi) ~ 0.149 % of 5 bar (75 psi) (or) Total performance, if calculated in % of the upper range limit = ± √ ((0.05)2 + (0.142)) = ± 0.149 (% of the set span) Total performance = ± 0.149 (% of the set span) (or) ±0.0074 bar (0.1073 psi) A B 0.00 0.20 0.40 0.60 0.80 [%] 4530 105 120 135 150 [bar] [psi] TD 2:1 TD 5:1 TD 4:1 TD 3:1 TD 1:1 C A0031339 A Error (% of the set span) B Set span [bar] C Standard measuring cell
Cerabar S PMC71, PMP71, PMP75
42 Endress+Hauser
Installation factors Some of the influencing factors are:
- Influence of installation position → 42
- Uncertainty of measurement for small absolute pressure ranges → 42 Influence of installation position
- PMP71: Device rotated 180°, process connection pointing upwards. The value is doubled for devices with inert oil. – Process connection thread G 1 A, G 1 ½, G 2, 1 ½ MNPT, 2 MNPT, M 44x1.25, EN/DIN, ASME and JIS flanges: ≤ 10 mbar (0.15 psi). – Process connection thread: G ½, ½ MNPT, JIS G ½, JIS R ½, M20x1.5: ≤ 4 mbar (0.06 psi).
- PMC71: ≤ 0.18 mbar (0.003 psi). Device rotated 180°, process connection pointing upwards. A position-dependent zero point shift can be corrected. Please refer to operating instructions and chapter Commissioning → Position adjustment. Different tightening torques (e.g. for Clamp or Varivent connections) can merely cause a shift in the zero point. This effect is corrected by position adjustment during commissioning. Uncertainty of measurment for small absolute pressure ranges The smallest expanded uncertainty of measurement that can be returned by our calibration standards is:
- in the range 1 to 30 mbar (0.0145 to 0.435 psi): 0.4 % of (current) reading
- in the range < 1 mbar (0.0145 psi): 1 % of (current) reading. Performance Characteristics Conformance All performance characteristics are in conformance to ≥ ± 3 sigma. Resolution Current output: 1 μA Warm-up period
- 4 to 20 mA HART: < 10 s
- PROFIBUS PA: 6 s
- FOUNDATION Fieldbus: 50 s
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 43 Installation General installation instructions
- For PMP75: → 107 "Installation instructions" section.
- A position-dependent zero point shift can be corrected directly at the device via operating keys, and also in hazardous areas in the case of devices with external operation. Diaphragm seals also shift the zero point, depending on the installation position→ 110.
- The device housing can be rotated up to 380° .
- Endress+Hauser offers a mounting bracket for installing the device on pipes or walls → 43.
- Use flushing rings for flange and cell diaphragm seals if buildup or clogging can be expected at the diaphragm seal connection. The flushing ring can be fitted between the process connection and diaphragm seal. Material buildup in front of the process isolating diaphragm can be flushed away, and the pressure chamber vented, via the two lateral flushing holes.
- When measuring in media containing solids, such as dirty liquids, installing separators and drain valves is useful for capturing and removing sediment.
- Point the cable and connector downwards where possible to prevent moisture from entering (e.g. rain or condensation water). Measuring arrangement for devices without diaphragm seals – PMC71, PMP71 Cerabar S transmitters without diaphragm seals are mounted as per the norms for a manometer (DIN EN 837‑2). We recommend the use of shutoff devices and siphons. The orientation depends on the measuring application. Pressure measurement in gases Mount Cerabar S with shutoff device above the tapping point so that any condensate can flow into the process. Pressure measurement in vapors Use siphons for pressure measurement in steam. The siphon reduces the temperature to almost ambient temperature. Fill the siphon with liquid before commissioning. Preferably mount the Cerabar S with the siphon below the tapping point. Advantages:
- defined water column only causes minimal/negligible measured errors
- only minimal/negligible thermal effects on the device Mounting above the tapping point is also possible. Note the max. permitted ambient temperature of the transmitter. Pressure measurement in liquids Mount Cerabar S with shutoff device below or at the same level as the tapping point. Level measurement
- Mount Cerabar S below the lowest measuring point.
- Do not install the device in the following positions: In the filling curtain, in the tank outlet or at a point in the container which could be affected by pressure pulses from an agitator or a pump.
- The calibration and functional test can be carried out more easily if you mount the device downstream from a shutoff device. Measuring arrangement for devices with diaphragm seals – PMP75 → 107 Orientation The orientation may cause a zero point shift, see → 42. This position-dependent zero point shift can be corrected directly at the device via the operating key, and also in hazardous areas in the case of devices with external operation (position adjustment). Wall and pipe mounting Endress+Hauser offers a mounting bracket for installing the device on pipes or walls. The mounting bracket can be installed on pipes with a diameter of 1¼" to 2" or on walls. Ordering information:
- Product Configurator, order code for "Additional options 2", version "U" or Product Configurator, order code for "Accessory enclosed", version "PA" or
- as a separate accessory (part no.: 71102216).
Cerabar S PMC71, PMP71, PMP75
44 Endress+Hauser
Dimensions → 94. Heat insulation – PMC71 high-temperature version The PMC71 high-temperature version must only be insulated up to a certain height. The maximum permitted insulation height is indicated on the devices and applies to an insulation material with a heat conductivity ≤ 0.04 W/(m x K) and to the maximum permitted ambient and process temperature (see table below). The data were determined under the most critical application "quiescent air". The data were determined under the most critical application "quiescent air". A B 1 2 A0021075 A Ambient temperature B Process temperature
1 Insulation height
2 Insulation material
Ambient temperature ≤ 70 °C (158 °F) Process temperature ≤ 150 °C (302 °F) Mounting of PVDF screw-in fittings LWARNING Risk of damage to process connection! Risk of injury! ‣ PVDF process connections with threaded connections must be installed with the enclosed mounting bracket. The mounting bracket can be installed on pipes with a diameter of 1¼" to 2" or on walls. Dimensions → 53.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 45 "Separate housing" version With the "separate housing" version, you are able to mount the housing with the electronics insert at a distance from the measuring point. This version facilitates trouble-free measurement:
- Under particularly difficult measuring conditions (at installation locations that are cramped or difficult to access)
- If rapid cleaning of the measuring point is required and
- If the measuring point is exposed to vibrations You can choose between different cable versions:
- PE: 2 m (6.6 ft), 5 m (16 ft) and 10 m (33 ft)
- FEP: 5 m (16 ft). Ordering information: Product Configurator, order code for "Additional options 2", version "G". Dimensions → 94 In the case of the "separate housing" version, the sensor is delivered with the process connection and cable ready mounted. The housing and a mounting bracket are enclosed as separate units. The cable is provided with a socket at both ends. These sockets are simply connected to the housing and the sensor. 2 31 r /c179120 (4.72) A0019994
1 Process connection with sensor - For degrees of protection, see the following section
2 Process connection adapter (weight: 0.93 kg (2.05 lb)) 3 Cable (weight: 0.05 kg/meter (0.11 lb)), both ends are fitted with a socket
4 Mounting bracket provided, suitable for pipe and wall mounting
5 Housing with electronic insert - degrees of protection → 54
Engineering unit mm (in) Degree of protection for the process connection and sensor with the use of
- FEP cable: – IP 69 1) – IP 66 NEMA 4/6P – IP 68 (1.83 mH2O for 24 h) NEMA 4/6P
- PE cable: – IP 66 NEMA 4/6P – IP 68 (1.83 mH2O for 24 h) NEMA 4/6P Technical data of the PE and FEP cable:
- Minimum bending radius: 120 mm (4.72 in)
- Cable extraction force: max. 450 N (101.16 lbf)
- Resistance to UV light Use in hazardous area:
- Intrinsically safe installations (Ex ia/IS)
- FM/CSA IS: for Div. 1 installation only 1) Designation of the IP protection class according to DIN EN 60529. Previous designation "IP69K" according to DIN 40050 Part 9 is no longer valid (standard withdrawn on November 1, 2012). The tests required by both standards are identical.
Cerabar S PMC71, PMP71, PMP75
46 Endress+Hauser
Reduction of the installation height If the separate housing is used, the installation height of the process connection is reduced compared to the dimensions of the standard version. 81 (3.19) 62 (2.44)T14 T17 ø54.1 (2.13) 45 (1.77) r /c179120 (4.72) r /c179120 (4.72) A0019995
1 Process connection adapter
Engineering unit mm (in) Turning the housing The housing can be rotated up to 380° by loosening the Allen screw. Your benefits
- Easy mounting due to optimum alignment of housing
- Good, accessible device operation
- Optimum readability of the onsite display (optional). T14 T15 T17 2 3 A0019996
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 47 Oxygen applications Oxygen and other gases can react explosively to oils, grease and plastics, such that, among other things, the following precautions must be taken:
- All components of the system, such as measuring devices, must be cleaned in accordance with the BAM (DIN 19247) requirements.
- Dependent on the materials used, a certain maximum temperature and a maximum pressure for oxygen applications must not be exceeded. The devices suitable for gaseous oxygen applications are listed in the following table with the specification pmax. HB = Cleaned for oxygen service Order code for devices 1), cleaned for oxygen applications pmax for oxygen applications Tmax for oxygen
Devices with sensors, nominal value < 10 bar (150 psi) Over pressure limit (OPL) 2) 3) of sensor 60 °C (140 °F) Devices with sensors, nominal value ≥ 10 bar (150 psi) 30 bar (450 psi) 60 °C (140 °F) depends on the lowest-rated element, with regard to pressure, of the selected components: over pressure limit (OPL) of the sensor, process connection (1.5 x PN) or fill fluid (160 bar (2 400 psi)) 85 °C (185 °F) depends on the lowest-rated element, with regard to pressure, of the selected components: over pressure limit (OPL) of the sensor, process connection (1.5 x PN) or fill fluid (160 bar (2 400 psi)) 85 °C (185 °F) 1) Devices only, not accessories or enclosed accessories. 2) Product Configurator, order code for "Sensor range; sensor over pressure limit (= OPL)" 3) PMC71 with PVDF threaded connection: Only mount with the enclosed mounting bracket. MWP 10 bar (150 psi), OPL max. 15 bar (225 psi). Process temperature range –10 to +60 °C (+14 to +140 °F) Silicone-free applications Special cleaning of the transmitter to remove paint-wetting substances, e.g. for use in paint shops. Ordering information: Product Configurator, order code for "Seal", option "L" or "M". Ultrapure gas applications Endress+Hauser also offers devices for special applications, such as ultrapure gas, cleaned from oil and grease. No special restrictions regarding the process conditions apply to these devices. Ordering information:
- Product Configurator, order code for "Seal" or
- Product Configurator, order code for "Fill fluid". Applications with hydrogen A ceramic process isolating diaphragm or a gold-plated metal process isolating diaphragm offers universal protection against hydrogen diffusion, both in gas applications and in applications with aqueous solutions. Applications with hydrogen in aqueous solutions A gold/rhodium-plated metal process isolating diaphragm (AU/Rh) offers effective protection against hydrogen diffusion.
Cerabar S PMC71, PMP71, PMP75
48 Endress+Hauser
Ambient temperature range Version PMC71 High-temperature version PMC71 PMP71 1) PMP75 1) Without LCD display –20 to +70 °C (–4 to +158 °F) With LCD display 2) With M12 connector, elbowed With separate housing Diaphragm seal systems 3) MID parts certificate — — –25 to +55 °C (–13 to +131 °F) 1) Lower temperatures on request 2) Extended temperature application range (–40 to +85 °C (–40 to +185 °F)) with restrictions in optical properties, such as display speed and contrast 3) Ambient temperature range and process temperature range are mutually dependent - see "Heat insulation" section → 110 For high-temperature applications, either a PMP75 with a temperature isolator or with a capillary can be used. If vibrations also occur in the application, Endress+Hauser recommends using a PMP75 with a capillary. If a PMP75 with a temperature isolator or capillary is used, we recommend a suitable bracket for mounting (see "Wall and pipe mounting" section)→ 43. Hazardous areas
- For devices for use in hazardous areas, see Safety Instructions, Installation or Control Drawing → 122.
- Pressure measuring devices that have the usual explosion protection certificates (e.g. ATEX/ FM/ CSA/ IEC Ex,...) can be used in hazardous areas at ambient temperatures down to –50 °C (–58 °F). The functionality of the explosion protection is also guaranteed for ambient temperatures down to –50 °C (–58 °F).
- The nameplate specification is limited to an ambient temperature of –40 °C (–40 °F), as all measurement-specific testing of the device is performed only down to –40 °C (–40 °F). If the device is operated at an ambient temperature below –40 °C (–40 °F), the technical data in this document are no longer valid. Functional restrictions can be expected. Storage temperature range • –40 to +90 °C (–40 to +194 °F)
- Local display: –40 to +85 °C (–40 to +185 °F)
- Separate housing –40 to +60 °C (–40 to +140 °F)
- Devices with PVC-sheathed capillary: –25 to +90 °C (–13 to +194 °F) Degree of protection Depends on the deployed
- Housing → 54
- Separate housing → 45 Climate class Class 4K4H (air temperature: –20 to +55 °C (–4 to +131 °F), relative humidity: 4 to 100 %) fulfilled as per DIN EN 60721-3-4 (condensation possible. With PMC71, avoid condensate in the device.) Electromagnetic compatibility
- Electromagnetic compatibility as per EN 61326 and NAMUR recommendation EMC (NE21).
- With enhanced immunity against electromagnetic fields as per EN 61000-4-3: 30 V/m with closed cover (for devices with T14 housing)
- Maximum deviation: < 0.5 % of span
- All EMC measurements were performed with a turn down (TD) = 2:1.
- Class E3 as per OIML R75-2
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 49 For further details refer to the Declaration of Conformity. Vibration resistance Device/accessory Test standard Vibration resistance PMC71 1) GL Guaranteed for 3 to 25 Hz: ±1.6 mm (0.063 in); 25 to 100 Hz: 4 g in all 3 planes PMP71 PMP75 2) 3) With mounting bracket IEC 61298-3 Guaranteed for 10 to 60 Hz: ±0.15 mm (0.0059 in); 60 to 500 Hz: 2 g in all 3 planes PMP71 with MID parts certificate OIML R117-1 Class M3 1) Not for high-temperature version with Ex d[ia], CSA XP or FM XP 2) With aluminum T14 housing only 3) For applications with very high temperatures, a PMP75 with either a temperature isolator or a capillary can be used. If vibrations also occur in the application, Endress+Hauser recommends using a PMP75 with a capillary. If a PMP75 with a temperature isolator or capillary is used, it must be mounted with a mounting bracket
Cerabar S PMC71, PMP71, PMP75
50 Endress+Hauser
Process temperature limits For oxygen applications → 47 PMC71 (with ceramic process isolating diaphragm)
- High-temperature version: –25 to +150 °C (–13 to +302 °F); Product Configurator, order code for "Additional options 1", version "T".
- For saturated steam applications, use a device with a metal process isolating diaphragm, or provide a siphon for temperature isolation when installing.
- Observe the process temperature range of the seal in the following table. Seal Notes Process temperature range Option 1) FKM Viton — –25 to +125 °C (–13 to +257 °F)/ 150 °C (302 °F) 2) A, L EPDM FDA 21CFR177.2600; 3A Class II; USP Class VI DVGW (KTW, W270, W534), WRAS, ACS, NSF61 150 °C (302 °F) 2) B EPDM –20 to +150 °C (–4 to +302 °F) B FFKM Perlast G75LT — –20 to +125 °C (–4 to +257 °F)/ 150 °C (302 °F) 2) C Kalrez, Compound 4079 — +5 to +125 °C (+41 to +257 °F)/ 150 °C (302 °F) 2) D, M Chemraz, Compound 505 — –10 to +125 °C (+14 to +257 °F)/ 150 °C (302 °F) 2) E HNBR FDA 21CFR177.2600; 3A Class II; KTW; AFNOR; BAM –25 to +125 °C (–13 to +257 °F) F 3) NBR — –10 to +100 °C (+14 to +212 °F) F FKM Viton FDA 21CFR177.2600 –5 to +125 °C (+23 to +257 °F) G FKM Viton cleaned of oil and grease –10 to +125 °C (+14 to +257 °F)/ 150 °C (302 °F) 2) FKM Viton cleaned for oxygen service –10 to +60 °C (+14 to +140 °F) 2 or A 4) The process temperature ranges specified here refer to permanent application of the PMC71. They may be exceeded for a short time (e.g. for cleaning). 1) Product Configurator, order code for "Seal" 2) 150 °C (302 °F) for high-temperature version 3) These seals are used for devices with 3A-approved process connections. 4) with option "HB", see Product Configurator, order code for "Service" Applications with jumps in temperature Extreme jumps in temperature can result in temporary measuring errors. Temperature compensation takes effect after several minutes. Internal temperature compensation is faster the smaller the jump in temperature and the longer the time interval involved. For further information please contact your local Endress+Hauser Sales Center. PMP71 (with metallic process isolating diaphragm) Description Limits Process connections with internal process isolating diaphragm –40 to +125 °C (–40 to +257 °F) (150 °C (302 °F) for max. one hour) Process connections with flush-mounted process isolating diaphragm, G ½ A, G 1 A, G 1 ½ A, G 2 A, 1 NPT, 1 ½ NPT, 2 NPT, M20x1.5, M 44x1.25, EN/DIN, ASME and JIS flanges Process connections with flush-mounted process isolating diaphragm –20 to +85 °C (–4 to +185 °F)
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 51 PMP71 (with metallic process isolating diaphragm) with MID parts certificate PMP75 (with diaphragm seal)
- Depending on the design and depending on the diaphragm seal and filling oil: –70 °C (–94 °F) up to +400 °C (+752 °F). Please note the temperature application limits of the diaphragm seal oil → 109.
- Please observe the maximum gauge pressure and maximum temperature. Devices with PTFE-coated process isolating diaphragm The non-stick coating has excellent gliding properties and is used to protect the process isolating diaphragm against abrasive media. NOTICE Incorrect use of the PTFE foil will destroy the device! ‣ The PTFE foil used is designed to protect the unit against abrasion. It does not provide protection against corrosive media. For the range of application of the 0.25 mm (0.01 in) PTFE foil on an AISI 316L (1.4404/1.4435) process isolating diaphragm, see the following diagram: [°F] [psi p ]g 725 1450 2175 2900 3625 -40 77 122 167 212 257 302 347 392 A0026949-EN For vacuum applications: pabs ≤ 1 bar (14.5 psi) to 0.05 bar (0.725 psi) up to max. +150 °C (302 °F). Process temperature limits of capillary armoring: PMP75
- 316L: No restrictions
- PTFE: No restrictions
- PVC: See the following diagram Tp Ta +32 +8040 +185113 200 100 −70 300 392 392 +32 −94 572 Ta [°C] [°F] [°C][°F] − −40 25 − −40 22 Tp A0028220
Cerabar S PMC71, PMP71, PMP75
52 Endress+Hauser
Pressure specifications LWARNING The maximum pressure for the measuring device depends on the lowest-rated element with regard to pressure. ‣ For pressure specifications, see the "Measuring range" section and the "Mechanical construction" section. ‣ The measuring device must be operated only within the specified limits! ‣ The MWP (maximum working pressure) is specified on the nameplate of the individual sensor module. This value refers to a reference temperature of +20 °C (+68 °F), or +38 °C (+100 °F) for ASME flanges, and can be present at the device for an unlimited period. Observe the pressure- temperature dependency of the MWP. ‣ Please refer to the standards EN 1092-1: 2001 Tab. 18 for the pressure values permitted in the case of higher temperatures. (With regard to their stability-temperature property, the materials 1.4435 and 1.4404 are grouped together under 13E0 in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical). // ASME B 16.5a – 1998 Tab. 2-2.2 F316 // ASME B 16.5a – 1998 Tab. 2.3.8 N10276 // JIS B 2220 ‣ The test pressure corresponds to the over pressure limit of the individual sensors (OPL = 1.5 x MWP (formula does not apply to the PMP71 or PMP75 with a 40 bar (600 psi) or 100 bar (1500 psi) measuring cell)) and may be applied only for a limited period of time to prevent any lasting damage. ‣ The Pressure Equipment Directive (2014/68/EU) uses the abbreviation "PS". The abbreviation "PS" corresponds to the MWP (maximum working pressure) of the measuring device. ‣ In the case of sensor range and process connections where the over pressure limit (OPL) of the process connection is smaller than the nominal value of the sensor, the device is set at the factory, at the very maximum, to the OPL value of the process connection. If you want to use the entire sensor range, select a process connection with a higher OPL value (1.5 x PN; MWP = PN) ‣ In oxygen applications, the values for "pmax and Tmax for oxygen applications must not be exceeded → 47. ‣ Steam hammering must be avoided. Steam hammering can cause zero point drifts. Recommendation: Residue (water droplets or condensation) may remain on the process isolating diaphragm following SIP cleaning and can result in local steam hammering the next time steam cleaning takes place. In practice, drying the process isolating diaphragm (e.g. by blowing) has proved to prevent steam hammering.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 53 Mechanical construction Device height The device height is calculated from
- the height of the housing
- the height of optional mounted parts such as temperature isolators or capillaries
- the height of the relevant process connection. The individual heights of the components can be found in the following sections. To calculate the device height, simply add up the individual heights of the components. If necessary, the installation space (the space used to install the device) must also be taken into account. You can use the following table for this: Section Page Height Example Height of housing → 54 ff. (A) A I A H B H A0021437 Optional mounted parts → 58 (B) Process connections → 58 (H) Installation space - (I) Device height
Cerabar S PMC71, PMP71, PMP75
54 Endress+Hauser
T14 housing, optional display on the side 152 (5.98) 129 (5.08) A0019997 Engineering unit mm (in). Front view, left-hand side view, top view Material Degree of protection Cable entry Weight in kg (lb) Option 1) Housing Cover seal with display without display Aluminum EPDM IP66/67 NEMA 6P M20 gland A IP66/67 NEMA 6P G ½" thread B IP66/67 NEMA 6P NPT ½" thread C IP66/67 NEMA 6P M12 plug D IP66/67 NEMA 6P 7/8" plug E IP65 NEMA 4 HAN7D plug 90 degrees F 316L EPDM IP66/67 NEMA 6P M20 gland IP66/67 NEMA 6P G ½" thread 2 IP66/67 NEMA 6P NPT ½" thread 3 IP66/67 NEMA 6P M12 plug 4 IP66/67 NEMA 6P 7/8" plug 5 IP65 NEMA 4 HAN7D plug 90 degrees 6 FVMQ IP66/67 NEMA 6P M20 gland 7 FVMQ IP66/67 NEMA 6P NPT ½" thread 8 1) Product Configurator, order code for "Housing, cover seal, cable entry, degree of protection"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 55 T17 housing (hygienic), optional display on the side 94 (3.7) 132 (5.2) 115 (4.53) 147 (5.79) A0020000 Engineering unit mm (in). Front view, left-hand side view, top view Material Degree of protection 1) Cable entry Weight in kg (lb) Option 2) Housing Cover seal with display without display 316L EPDM IP66/68 NEMA 6P M20 gland R IP66/68 NEMA 6P G ½" thread S IP66/68 NEMA 6P NPT ½" thread T IP66/68 NEMA 6P M12 plug U IP66/68 NEMA 6P 7/8" plug V 1) Degree of protection IP 68: 1.83 mH 2O for 24 h 2) Product Configurator, order code for "Housing, cover seal, cable entry, degree of protection"
Cerabar S PMC71, PMP71, PMP75
56 Endress+Hauser
PMC71 with internal process isolating diaphragm Threaded connection ISO 228 G A ø6 (0.24) ø3 (0.12) ø8 (0.31) G ½" 3 (0.12) H 17 (0.67)20 (0.79) A0020935 B ø17.5 (0.24) G ¼" G ½" 20 (0.79) H 17 (0.67)13 (0.51) A0020936 C ø17.5 (0.24) ø11.4 (0.45) G ½" H 20 (0.79)17 (0.67) A0020937 Engineering unit mm (in). Height H → 58 Item Description Material Weight Approval 1) Option 2) kg (lb) A Thread ISO 228 G ½" A EN 837 AISI 316L 0.63 (1.39) CRN GA Alloy C276 (2.4819) CRN GB Monel (2.4360) - GC PVDF
- Only mount with a mounting bracket (included)
- MWP 10 bar (150 psi), OPL max. 15 bar (225 psi)
- Process temperature range: –10 to +60 °C (+14 to +140 °F) - GD B Thread ISO 228 G ½" A, G ¼" (internal) AISI 316L CRN GE Alloy C276 (2.4819) CRN GF Monel (2.4360) - GG C Thread ISO 228 G ½" A, Hole11.4 mm (0.45 in) AISI 316L CRN GH Alloy C276 (2.4819) CRN GJ Monel (2.4360) - GK 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 57 Threaded connection ANSI A NPT ¼" NPT ½" H 25 (0.98) A0020938 B ø11.4 (0.45) NPT ½" H 25 (0.98) A0020939 C ø3 (0.12) NPT ½" H 25 (0.98) A0020940 D ø26.5 (1.04) H 25 (0.98) NPT ½" A0020943 Engineering unit mm (in). Height H → 58 Item Description Material Weight Approval 1) Option 2) kg (lb) A ANSI ½" MNPT, ¼" FNPT AISI 316L 0.63 (1.39) CRN RA Alloy C276 (2.4819) CRN RB Monel (2.4360) - RC B ANSI ½" MNPT, Hole11.4 mm (0.45 in) AISI 316L CRN RD Alloy C276 (2.4819) CRN RE Monel (2.4360) - RF C ANSI ½" MNPT, Hole3 mm (0.12 in) PVDF
- Only mount with a mounting bracket (included)
- MWP 10 bar (150 psi), OPL max. 15 bar (225 psi)
- Process temperature range: +10 to +60 °C (+14 to +140 °F) - RG D ANSI ½" FNPT Hole11.4 mm (0.45 in) AISI 316L CRN RH Alloy C276 (2.4819) CRN RJ Monel (2.4360) - RK 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
58 Endress+Hauser
PMC71 with internal process isolating diaphragm Threaded connection JIS A ø5 (0.2) ø3 (0.12) ø8 (0.31) G ½" 3 (0.12) H 20 (0.79) A0021444 B ø11.4 (0.45) R ½" H A0021445 Engineering unit mm (in). Height H → 58 Item Description Material Weight Option 1) kg (lb) A JIS B0202 G ½" (male) AISI 316L 0.63 (1.39) GL B JIS B0203 R ½" (male) RL 1) Product Configurator, order code for "Process connection" Threaded connection DIN 13 ø6 (0.24) ø3 (0.12) ø8 (0.31) M20x1.5 3 (0.12) H 17 (0.67)20 (0.79) A0020945 Engineering unit mm (in). Height H → 58 Description Material Weight Option 1) kg (lb) DIN 13 M20 x 1.5, EN 837 3 mm (0.12 in) AISI 316L 0.63 (1.39) GP Alloy C276 (2.4819) GQ 1) Product Configurator, order code for "Process connection" Process connections for PMC71 with internal process isolating diaphragm - height H Description Height H Standard height 26 mm (1.02 in) Device with Ex d[ia], CSA XP or FM XP 96 mm (3.78 in) High-temperature version 106 mm (4.17 in) High-temperature version with Ex d[ia], CSA XP or FM XP 176 mm (6.93 in)
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 59 Process connections for PMC71 with flush-mounted process isolating diaphragm Threaded connection DIN 13 H 11 (0.43) M44x1.25 22 (0.87) ø40.5 (1.59) A0020950 Engineering unit mm (in). Height H → 60 Description Material Weight Option 1) kg (lb) DIN 13 M44 x 1.25 AISI 316L 0.63 (1.39) Alloy C276 (2.4819) 1S 1) Product Configurator, order code for "Process connection" Threaded connection ISO 228 G A ø55 (2.17) H 25 (0.98) G 1½" 28 (1.1) A0020946 B ø68 (2.68) G 2" H A0020947 Engineering unit mm (in). Height H → 60 Item Description Material Weight Option 1) kg (lb) A Thread ISO 228 G 1 ½" A AISI 316L 0.63 (1.39) Alloy C276 (2.4819) 1H Monel (2.4360) 1J B Thread ISO 228 G 2" A AISI 316L 1K Alloy C276 (2.4819) 1L Monel (2.4360) 1M 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
60 Endress+Hauser
process isolating diaphragm Threaded connection ANSI A NPT 1½" H 27 (1.06) A0020948 B ø68 (2.68) H NPT 2" A0020949 Engineering unit mm (in). Height H → 60 Item Description Material Weight Approval 1) Option 2) kg (lb) A ANSI 1 ½" MNPT AISI 316L (CRN) 0.63 (1.39) CRN 2D Alloy C276 (2.4819) (CRN) CRN 2E Monel (2.4360) - 2F B ANSI 2" MNPT AISI 316L (CRN) CRN 2G Alloy C276 (2.4819) (CRN) CRN 2H Monel (2.4360) - 2J 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection" Process connections for PMC71 with flush-mounted process isolating diaphragm - height H Description Height H High-temperature version 86 mm (3.39 in) High-temperature version with Ex d[ia], CSA XP or FM XP 151 mm (5.94 in)
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 61 Process connections for PMC71 with flush-mounted process isolating diaphragm EN/DIN flanges, connection dimensions in accordance with EN 1092-1/DIN 2527 D k b 62 g A0026336 Engineering unit mm Flange Boltholes Weight Option 1) Material Nominal diameter Nominal pressure Shape 2) D Thickness Raised face Number g2 Hole circle b g k mm mm mm mm mm kg (lb) AISI 316L DN 25 PN 10-40 B1 (D) 115 18 68 4 14 85 1.4 (3.09) BA AISI 316L DN 32 PN 10-40 B1 (D) 140 18 78 4 18 100 2.0 (4.41) CP AISI 316L DN 40 PN 10-40 B1 (D) 150 18 88 4 18 110 2.4 (5.29) CQ AISI 316L DN 50 PN 10-40 B1 (D) 165 20 102 4 18 125 3.2 (7.06) B3 PVDF DN 50 PN 10-16 B1 (D) 165 21.4 102 4 18 125 0.6 (1.32) BR AISI 316L DN 50 PN 63 B2 (E) 180 26 102 4 22 135 4.6 (10.14) C3 PVDF DN 80 PN 10-16 B1 (D) 200 21.4 138 8 18 160 1.0 (2.21) BS AISI 316L DN 80 PN 10-40 B1 (D) 200 24 138 8 18 160 5.4 (11.91) B4 1) Product Configurator, order code for "Process connection" 2) Description as per DIN 2527 provided in brackets
Cerabar S PMC71, PMP71, PMP75
62 Endress+Hauser
ASME flanges, connection dimensions in accordance with ASME B 16.5, raised face RF D k b 2.44 0.08 g 1.024 A0026337 Engineering unit in. Flange Boltholes Weight Approval 1) Option 2) Material Nominal diameter Class D Thickness Raised face Number g2 Hole circle b g k [in] [lb./sq.in] [in] [in] [in] [in] [in] [kg (lb)] 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection" 3) Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated) 4) Screws must be 15 mm (0.59 in) longer than the standard flange screws 5) ECTFE coating on AISI 316/316L. When operating in hazardous areas, avoid electrostatic charge on the plastic surfaces.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 63 JIS flanges, connection dimensions in accordance with JIS B 2220 BL, raised face RF D k b 62 g A0026336 Engineering unit mm Flange Boltholes Weight Option 1) Material Nominal diameter Nominal pressure D Thickness Raised face Number g2 Hole circle b g k mm mm mm mm mm kg (lb) AISI 316L (1.4435) 50 A 10 K 155 16 96 4 19 120 2.0 (4.41) KF 80 A 10 K 185 18 127 8 19 150 3.3 (7.28) KL 100 A 10 K 210 18 151 8 19 175 4.4 (9.7) KH 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
64 Endress+Hauser
Hygienic process connections for PMC71 with flush- mounted process isolating diaphragm Many process connections with an EPDM or HNBR seal are approved for the PMC71 in accordance with the guidelines of the 3A Sanitary Standard. To ensure that the 3A approval is valid for the PMC71 version, a 3A‑approved process connection together with an EPDM or HNBR seal must be selected when ordering (Product Configurator, order code for "Seal"). A 7.5 (0.3) 10 (0.39) 52 (2.05) 61 (2.4) 38 (1.5) 56 (2.2) 32 (1.26) 70 (2.76) 48 (1.89) A0021061 B 52 (2.05) 61 (2.4) 68 (2.68) 70 (2.76) 42 (1.65) A0021063 C 42 (1.65) 64 (2.52) 61 (2.4) 56,5 (2.22) 52 (2.05) 7 (0.28) 19,8 (0.78) 48 (1.89) A0021058 D 42 (1.65) 52 (2.05) 76.7 (3.02) 91 (3.58) 83.5 (3.29) 48 (1.89) A0020952 Engineering unit mm (in) Item Description Nominal pressure Material 1) Weight Approval 2) Option 3) kg (lb) A DIN 11851 DN 40 PN 25, with HNBR or EPDM seal PN 25 AISI 316L (1.4435) 0.7 (1.54) EHEDG, 3A, CRN MP 4) B DIN 11851 DN 50 PN 25, with HNBR or EPDM seal PN 25 0.9 (1.98) EHEDG, 3A, CRN MR 4) C Tri-Clamp ISO 2852 DN 51 (2"), with HNBR or EPDM seal PN 40 5) 0.7 (1.54) EHEDG, 3A, CRN TD 4) D Tri-Clamp ISO 2852 DN 76.1 (3"), with NBR or EPDM seal PN 40 5) 0.9 (1.98) EHEDG, 3A, CRN TF 4) 2) CSA approval: Product Configurator, order code for "Approval" 3) Product Configurator, order code for "Process connection" 4) Endress+Hauser supplies these slotted nuts in stainless steel AISI 304 (DIN/EN material number 1.4301) or in AISI 304L (DIN/EN material number 1.4307). 5) Restricted nominal pressure (13.8 bar (200 psi)) for the following approvals: Product Configurator, order code for "Approval", option "E", "U" and "V".
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 65 E 16 (0.63) 42 (1.65) 52 (2.05) 61 (2.4) 84 (3.31) 105 (4.13) 65 (2.56) 48 (1.89) A0023740 F 42 (1.65) 75 (2.95) 71 (2.8) 84 (3.31) 52 (2.05) 48 (1.89) A0021060 Engineering unit mm (in) Item Description Nominal pressure Material 1) Weight Approval 2) Option 3) kg (lb) E DRD DN50 (65 mm) with HNBR or EPDM seal PN 25 AISI 316L (1.4435) 0.9 (1.98) EHEDG TK 4) F Varivent type N for pipes 40 – 162, with HNBR or EPDM seal PN 40 1 (2.21) EHEDG, 3A, CRN TR 2) CSA approval: Product Configurator, order code for "Approval" 3) Product Configurator, order code for "Process connection" 4) Endress+Hauser supplies these slotted nuts in stainless steel AISI 304 (DIN/EN material number 1.4301) or in AISI 304L (DIN/EN material number 1.4307).
Cerabar S PMC71, PMP71, PMP75
66 Endress+Hauser
PMP71 with internal process isolating diaphragm Threaded connection ISO 228 G A 3 (0.12) G ½ A " ø3.2 (0.13) ø6 (0.24) ø17.5 (0.69) 20 (0.79) 3 (0.12) 36 (1.42) A0021588 B G ½ A " G ¼" 17 (0.67) 36 (1.42) A0021591 C ø11.4 (0.45) G ½ A " ø17.5 (0.69) 20 (0.79) 17 (0.67) 36 (1.42) A0021592 Engineering unit mm (in) Item Description Material Weight Option 1) kg (lb) A Thread ISO 228 G ½" A EN 837 AISI 316L 0.63 (1.39) GA Alloy C276 (2.4819) GB B Thread ISO 228 G ½" A, G ¼" (internal) AISI 316L GE Alloy C276 (2.4819) GF C Thread ISO 228 G ½" A, Hole11.4 mm (0.45 in) AISI 316L GH Alloy C276 (2.4819) GJ 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 67 Process connections for PMP71 with internal process isolating diaphragm Threaded connection ANSI A 15 (0.59) ø21.4 (0.84) 25 (0.98) NPT ½" NPT ¼" 36 (1.42) A0021112 B ø21.4 (0.84) NPT ½" ø11.4 (0.45) 25 (0.98) 36 (1.42) A0021113 C ø38 (1.5) 36 (1.42) NPT ½" 7 (0.28) 8 (0.31)20 (0.79) 10 (0.39) A0020933 Engineering unit mm (in) Item Description Material Weight Approval 1) Option 2) kg (lb) A ANSI ½" MNPT, ¼" FNPT AISI 316L 0.63 (1.39) CRN RA Alloy C276 (2.4819) CRN RB B ANSI ½" MNPT, Hole 11.4 mm (0.45 in) = 400 bar (6 000 psi) Hole 3.2 mm (0.13 in) = 700 bar (10 500 psi) AISI 316L CRN RD Alloy C276 (2.4819) CRN RE C ANSI ½" FNPT AISI 316L 0.7 (1.54) CRN RH Alloy C276 (2.4819) CRN RJ 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
68 Endress+Hauser
PMP71 with internal process isolating diaphragm Threaded connection JIS A G ½" ø3 (0.12) ø5 (0.2) 23 (0.91)3 (0.12) 36 (1.42)ø8 (0.31) A0021607 B ø21.4 (0.84) R ½" ø11.4 (0.45) 23 (0.91) 36 (1.42) A0021608 Engineering unit mm (in) Item Description Material Weight Option 1) kg (lb) A JIS B0202 G ½" (male) AISI 316L 0.6 (1.32) GL B JIS B0203 R ½" (male) RL 1) Product Configurator, order code for "Process connection" Threaded connection DIN 13 A ø11.4 (0.45) M20x1.5 ø17.5 (0.69) 20 (0.79) 17 (0.67) 36 (1.42) A0021609 B 3 (0.12) M20x1.5 ø3.2 (0.13) ø6 (0.24) ø17.5 (0.69) 20 (0.79) 3 (0.12) 36 (1.42) A0021588 Engineering unit mm (in) Item Description Material Weight Option 1) kg (lb) A DIN 13 M20 x 1.5 11.4 mm (0.45 in) AISI 316L 0.6 (1.32) GP Alloy C276 (2.4819) GQ B DIN 13 M20 x 1.5, EN 837 3 mm (0.12 in) AISI 316L GR Alloy C276 (2.4819) GS 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 69 Process connections for PMP71 with flush-mounted process isolating diaphragm Threaded connection ISO 228 G A ø17.5 (0.69) ø18.4 (0.72) ø26.5 (1.04) ø23.6 (0.93) G ½" A 34 (1.34)3 (0.12) A0021613 B ø30 (1.18) ø28 (1.1) ø36 (1.42) ø39 (1.54) G 1" A 38 (1.5)3 (0.12) A0021614 C 34 (1.34)8 (0.31) 2 (0.08)26 (1.02) ø28 (1.1) G 1½" A ø55 (2.17) ø36 (1.42) A0021615 D 2 (0.08)30 (1.18) ø28 (1.1) G 2" A ø68 (2.68) ø56.3 (2.22) 33 (1.3)1 (0.04) A0021616 Engineering unit mm (in) Item Description Material Weight Option 1) kg (lb) A Thread ISO 228 G ½" A, DIN 3852 FKM seal (item 1) pre-installed AISI 316L 0.4 (0.88) Alloy C276 (2.4819) 1B B Thread ISO 228 G 1" A AISI 316L 0.7 (1.54) Alloy C276 (2.4819) 1E C Thread ISO 228 G 1 ½" A AISI 316L 1.1 (2.43) Alloy C276 (2.4819) 1H D Thread ISO 228 G 2" A AISI 316L 1.5 (3.31) Alloy C276 (2.4819) 1L 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
70 Endress+Hauser
A ø17.5 (0.69) ø17.7 (0.7) ø26.5 (1.04) ø23.6 (0.93) M20x1.5 34 (1.34)3 (0.12) A0021622 B 41 (1.61)8 (0.31) 2 (0.08)26 (0.91) ø17.5 (0.69) M44x1.25 ø55 (2.17) ø40 (1.57) A0021623 Engineering unit mm (in) Item Description Material Weight Option 1) kg (lb) A Thread DIN 16288 M20 AISI 316L 0.4 (0.88) Alloy C276 (2.4819) 1P B Thread DIN 13 M44 x 1.25 AISI 316L 1.1 (2.43) Alloy C276 (2.4819) 1S 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 71 Process connections for PMP71 with flush-mounted process isolating diaphragm Threaded connection ANSI A 33 (1.3) 23 (0.91) NPT 1" ø28 (1.1) A0021624 B 36 (1.42) 27 (1.06) NPT 1 " ½ ø28 (1.1) A0021625 C 30 (1.18) 31 (1.22) NPT 2" ø28 (1.1) A0021626 Engineering unit mm (in) Item Description Material Weight Approval 1) Option 2) kg (lb) A ANSI 1" MNPT AISI 316L 0.7 (1.54) CRN 2A Alloy C276 (2.4819) CRN 2B B ANSI 1 ½" MNPT AISI 316L 1 (2.21) CRN 2D Alloy C276 (2.4819) CRN 2E C ANSI 2" MNPT AISI 316L 1.3 (2.87) CRN 2G Alloy C276 (2.4819) CRN 2H 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
72 Endress+Hauser
process isolating diaphragm EN/DIN flanges, connection dimensions in accordance with EN 1092-1/DIN 2527 D k b g 2 d =M ø28 m A0022643 Engineering unit mm Flange 1) Boltholes Weight Flange Option 2) Nominal diameter Nominal pressure Shape 3) D Thickness Raised face Width of Raised face Number g2 Hole circle b g m k [mm] [mm] [mm] [mm] [mm] [mm] [kg] DN 25 PN 10-40 B1 (D) 115 18 68 4) 4 4 14 85 1.2 (2.65) CN DN 32 PN 10-40 B1 (D) 140 18 78 4) 9 4 18 100 1.9 (4.19) CP DN 40 PN 10-40 B1 (D) 150 18 88 4) 14 4 18 110 2.2 (4.85) CQ DN 50 PN 10-40 B1 (D) 165 20 102 - 4 18 125 3.0 (6.62) B3 DN 80 PN 10-40 B1 (D) 200 24 138 - 8 18 160 5.3 (11.69) B4 1) Material: AISI 316L 2) Product Configurator, order code for "Process connection" 3) Name as per DIN 2527 provided in brackets 4) With these process connections the raised face is smaller than described in the standard. Due to a smaller raised face a special seal must be used. For further information, please contact a seal manufacturer or your local Endress+Hauser Sales Center.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 73 Process connections for PMP71 with flush-mounted process isolating diaphragm ASME flanges, connection dimensions in accordance with ASME B 16.5, raised face RF* D k b g 2.36 c d =M ø1.1 1.61 m A0022645 Engineering unit in. (*all apart from AN, see table below) Flange 1) Boltholes Weight Approval 2) Option 3) Nominal diameter Class/ nominal pressure D Thickness Raised face Thickness of raised face Width of Raised face Number g2 Hole circle b g c m k [in] lb./sq.in [in] [in] [in] [in] [in] [in] [in] [kg] (25.88) CRN AT 1) Material: AISI 316/316L; Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated) 2) CSA approval: Product Configurator, order code for "Approval" 3) Product Configurator, order code for "Process connection" 4) With these process connections the raised face is smaller than described in the standard. Due to a smaller raised face a special seal must be used. For further information, please contact a seal manufacturer or your local Endress+Hauser Sales Center.
Cerabar S PMC71, PMP71, PMP75
74 Endress+Hauser
process isolating diaphragm JIS flanges, connection dimensions in accordance with JIS B 2220 BL, raised face RF D k b g 2 d =M ø28 m A0022643 Engineering unit mm Flange Boltholes Weight Flange Option 1) Material Nominal diameter Class/ Nominal pressure D Thickness Raised face Width of Raised face Number g2 Hole circle b g m k [mm] [mm] [mm] [mm] [mm] [mm] [kg] AISI 316L 25 A 20 K 125 16 67 2) 3.5 4 19 90 1.5 (3.31) KA AISI 316L 50 A 10 K 155 16 96 - 4 19 120 2.0 (4.41) KF AISI 316L 80 A 10 K 185 18 127 - 8 19 150 3.3 (7.28) KL AISI 316L 100 A 10 K 210 18 151 - 8 19 175 4.4 (9.7) KH 1) Product Configurator, order code for "Process connection" 2) With these process connections the raised face is smaller than described in the standard. Due to a smaller raised face a special seal must be used. For further information, please contact a seal manufacturer or your local Endress+Hauser Sales Center.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 75 Process connections for PMP71 Oval flange UNF7/16-20 NPT ¼" -18 13 (0.51) 41.3 (1.63) A0021632 Engineering unit mm (in) Material Description Weight Approval 1) Option 2) kg (lb) AISI 316L (1.4404) Oval flange adapter 1/4-18 NPT as per IEC 61518 Mounting: 7/16-20 UNF 1.9 (4.19) CRN UR 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection" Process connections for PMP71 Prepared for diaphragm seal mount ø7.95 (0.31) ø2.5 (0.1) 60 (2.36) 5 (0.2) 1 2 3 A0025596 1 Setscrew with hexagonal recess4 mm (0.16 in), material A2-70 2 Bearing DIN 5401 (1.3505)
3 Hole for filling fluid
Engineering unit mm (in) Material Description Weight in kg (lb) Approval 1) Option 2) AISI 316L (1.4404) Prepared for diaphragm seal mount 1.9 (4.19) CRN U1 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
76 Endress+Hauser
40 (1.57) 60 (2.36)60 (2.36) 60 (2.36) max. 115 (4.53) A0021634
1 PMP75 basic device
2 Diaphragm seal, here, e.g. flange diaphragm seal
3 Diaphragm seal with temperature isolator
4 Process connections with capillary lines are 40 mm (1.57 in) higher than process connections without capillary lines Engineering unit mm (in). PMP75 basic device with diaphragm seal.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 77 Process connections for PMP75 with flush-mounted process isolating diaphragm
- The weights of the diaphragm seals are given in the tables. For the weight of the housing, see
- The following drawings are drawings that illustrate how the system works in principle. In other words, the dimensions of a diaphragm seal supplied can deviate from the dimensions given in this document.
- With the use of high-temperature oils the design can deviate strongly.
- Note "Planning instructions, diaphragm seal systems" section→ 107
- For further information please contact your local Endress+Hauser Sales Center. Diaphragm seal cell structure dM D b A0021635 Engineering unit mm (in) Flange Diaphragm seal Approval 1) Option 2) Material Nominal diameter Nominal pressure 3) D Thickness Max. diameter of the process isolating diaphragm Weight b dM [kg (lb)] [mm] [mm] [mm] AISI 316L DN 50 PN 16-400 102 20 59 1.3 (2.87) - UI DN 80 PN 16-400 138 20 89 2.3 (5.07) - UJ DN 100 PN 16-400 162 20 89 3.1 (6.84) - UK [in] [lb/sq.in] [in (mm)] [in (mm)] [in (mm)] 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection" 3) The specified nominal pressure applies to the diaphragm seal. The maximum pressure for the measuring device is dependent on the lowest-rated element, with regard to pressure, of the selected components → 52.
Cerabar S PMC71, PMP71, PMP75
78 Endress+Hauser
process isolating diaphragm Thread ISO 228 and ANSI G NPT SW/AFSW/AF dM x h x h dM A0021639 Engineering unit mm (in). Left: thread ISO 228, right: thread ANSI Threaded connection Diaphragm seal Approval 1) Option 2) Material Thread Nominal pressure d1 d2 Screw-in length Across flats Max. diameter of the process isolating diaphragm Height Weight PN SW/AF h [mm] [mm] [mm] [mm] [mm] [kg (lb)] AISI 316L G 1" A 400 30 39 21 3) 32 30 19 0.4 (0.88) - 1D Alloy C276 0.5 (1.1) - 1E AISI 316L G 1 ½" A 400 44 55 30 50 42 20 0.9 (1.98) - 1G Alloy C276 1.0 (2.21) - 1H AISI 316L G 2" 400 56 68 30 65 50 20 1.9 (4.19) - 1K Alloy C276 2.1 (4.63) - 1L AISI 316L 1" MNPT 400 – 48 28 41 24 37 0.6 (1.32) CRN 2A Alloy C276 0.7 (1.54) CRN 2B AISI 316L 1 ½" MNPT 400 – 60 30 41 36 20 0.9 (1.98) CRN 2D Alloy C276 1.0 (2.21) CRN 2E AISI 316L 2" MNPT 400 – 78 30 65 38 35 1.8 (3.97) CRN 2G Alloy C276 2.0 (4.41) CRN 2H 1) CSA approval: Product Configurator, order code for "Approval" 2) Product Configurator, order code for "Process connection" 3) 28 mm (1.1 in) in conjunction with high-temperature oil
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 79 Process connections for PMP75 with flush-mounted process isolating diaphragm Tri-Clamp ISO 2852 h dM A0021644 Engineering unit mm (in) Material 1) Nominal diameter ISO 2852 Nominal diameter DIN 32676 Nominal diameter Diameter Max. diameter of the process isolating diaphragm Height Weight Approval 2) Option 3) Standard with TempC membrane C7 dM dM h [in] [mm] [mm] [mm] [mm] [kg (lb)] AISI 316L ND 25 / 33.7 DN 25 1 50.5 24 - 37 0.32 (0.71) EHEDG, 3A, CRN TB ND 38 DN 40 1 ½ 50.5 36 36 30 1 (2.21) EHEDG, 3A, CRN TC 4) 5) ND 51 / 40 DN 50 2 64 48 41 30 1.1 (2.43) EHEDG, 3A, CRN TD 4) 5) ND 63.5 DN 50 2 ½ 77.5 61 61 30 0.7 (1.54) EHEDG, 3A TE 6) ND 76.1 – 3 91 73 61 30 1.2 (2.65) EHEDG, 3A, CRN TF 5) 1) Surface roughness of the wetted surfaces R a < 0.76 µm (29.9 µin) as standard. Lower surface roughness on request. 2) CSA approval: Product Configurator, order code for "Approval" 3) Product Configurator, order code for "Process connection" 4) Optionally available as an ASME-BPE-compliant diaphragm seal version for use in biochemical processes, surfaces in contact with medium R a < 0.38 µm (15 µin)) , electropolished; order using order code for "Additional options", option "P". 5) Alternatively available with TempC membrane. 6) With TempC membrane PN max. = 40 bar (580 psi). The maximum PN depends on the clamp used.
Cerabar S PMC71, PMP71, PMP75
80 Endress+Hauser
process isolating diaphragm Tri-Clamp pipe diaphragm seal ISO 2852 d d h L D A0021671 Engineering unit mm (in) Material Nominal diameter ISO 2852 Nominal diameter Nominal pressure D d2 d3 Height Installati on length Weight Approval Option 3) [mm] L [in] [in] [mm] [mm] [mm] [mm] [kg (lb)] AISI 316L DN 25 1 PN 40 22.5 50.5 54 67 126 1.7 (3.75) 3A, CRN SB DN 38 1 ½ PN 40 35.5 50.5 69 67 126 1.0 (2.21) 3A, CRN SC 4) DN 51 2 PN 40 48.6 64 78 79 100 1.7 (3.75) 3A, CRN SD 4) 1) Surface roughness of the wetted surfaces R a < 0.8 µm (31.5 µin) as standard. 2) CSA approval: Product Configurator, order code for "Approval" 3) Product Configurator, order code for "Process connection" 4) incl. 3.1 and pressure test in accordance with the Pressure Equipment Directive, category II
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 81 Hygienic process connections for PMP75 with flush- mounted process isolating diaphragm Varivent for pipes 30 (1.18) D dM A0021672 Engineering unit mm (in) Material 1) Description Nominal pressure D Max. diameter of the process isolating diaphragm Weight Approval Option 2) Standard with TempC membrane dM dM [mm] [mm] [mm] [kg (lb)] AISI 316L Type F for pipes DN 25 - DN PN 40 50 34 36 0.4 (0.88) EHEDG, 3A TU 3) AISI 316L Type N for pipes DN 40 - DN 162 PN 40 68 58 61 0.8 (1.76) EHEDG, 3A TR 4) 5) 1) Surface roughness of the wetted surfaces R a < 0.76 µm (29.9 µin) as standard. 2) Product Configurator, order code for "Process connection" 3) With TempC membrane 4) Optionally available as an ASME-BPE-compliant diaphragm seal version for use in biochemical processes, surfaces in contact with medium R a < 0.38 µm (15 µin)) , electropolished; order using order code for "Additional options", option "P". Reduced surface roughness on request. 5) Alternatively available with TempC membrane.
Cerabar S PMC71, PMP71, PMP75
82 Endress+Hauser
Hygienic process connections for PMP75 with flush- mounted process isolating diaphragm DRD DN50 (65 mm) ø84 (3.31) 4 x ø11.5 (0.45) dM A0021673 Engineering unit mm (in) Material 1) Nominal pressure Max. diameter of the process isolating diaphragm Weight Option 2) Standard with TempC membrane dM dM [mm] [mm] [kg (lb)] AISI 316L PN 25 50 48 0.75 (1.65) TK 3) 1) Surface roughness of the wetted surfaces R a < 0.76 µm (29.9 µin) as standard. 2) Product Configurator, order code for "Process connection" 3) Alternatively available with TempC membrane. SMS nozzles with coupling nut f h G D m dM A0021674 Engineering unit mm (in) Material 1) Nominal diameter Nominal pressure D Nozzle height Thread Height Height max. diameter of the diaphragm Weight Approval Option 2) f G m h dM [mm] [mm] [mm] [mm] [mm] [kg (lb)] AISI 316L 1 PN 25 54 3.5 Rd 40 – 1/6 20 42.5 24 0.25 (0.55) 3A, EHEDG TG 1 ½ PN 25 74 4 Rd 60 – 1/6 25 57 36 0.65 (1.43) 3A, EHEDG TH 3) 2 PN 25 84 4 Rd 70 – 1/6 26 62 48 1.05 (2.32) 3A, EHEDG TI 3) 1) Surface roughness of the wetted surfaces R a < 0.76 µm (29.9 µin) as standard. 2) Product Configurator, order code for "Process connection" 3) Alternatively available with TempC diaphragm.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 83 APV-RJT nozzles with coupling nut f h G D m dM A0021674 Engineering unit mm (in) Material 1) Nominal diameter Nominal pressure D Nozzle height Thread Height Height Max. diameter of the process isolating diaphragm Weight Option 2) PN f G m h dM [in] [bar] [mm] [mm] [mm] [mm] [mm] [kg (lb)] AISI 316L 1 PN 40 77 6.5 1 13/16 – 1/8" 22 42.6 21 0.45 (0.99) TL 1 ½ PN 40 72 6.4 2 5/16 – 1/8" 22 42.6 28 0.75 (1.65) TM 2 PN 40 86 6.4 2 7/8 – 1/8" 22 42.6 38 1.2 (2.65) TN 1) Surface roughness of the wetted surfaces R a < 0.8 µm (31.5 µin) as standard. 2) Product Configurator, order code for "Process connection" APV-ISS nozzles with coupling nut G D dM 4 (0.16) 50 (1.97) 30 (1.18) SW/AF A0021677 Engineering unit mm (in) Material 1) Nominal diameter Nominal pressure D Thread Across flats Max. diameter of the process isolating diaphragm Weight Option 2) PN G AF dM [in] [bar] [mm] [mm] [kg (lb)] AISI 316L 1 PN 40 54.1 1 ½" – 1/8" 46.8 24 0.4 (0.88) TP 1 ½ PN 40 72 2" – 1/8" 62 34 0.6 (1.32) TQ 2 PN 40 89 2 ½" – 1/8" 77 45 1.1 (2.43) TS 1) Surface roughness of the wetted surfaces R a < 0.8 µm (31.5 µin) as standard. 2) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
84 Endress+Hauser
Hygienic process connections for PMP75 with flush- mounted process isolating diaphragm Taper adapter with coupling nut, DIN 11851 40 (1.57) D G dM f m A0021678 Engineering unit mm (in) Material 1) Taper adapter Slotted nut Diaphragm seal Approval Option 2) Max. diameter of the process isolating diaphragm Weight Description Nominal pressure D Nozzle height Thread Height Standard with TempC membrane PN f G m dM dM [bar] [mm] [mm] [mm] [mm] [mm] [kg (lb)] AISI 316L DN 32 PN 40 50 10 Rd 58 x 1/6" 21 32 28 0.45 (0.99) 3A, EHEDG MI 3) DN 40 PN 40 56 10 Rd 65 x 1/6" 21 38 36 0.45 (0.99) 3A, EHEDG MZ 3) DN 50 PN 25 68.5 11 Rd 78 x 1/6" 19 52 48 1.1 (2.43) 3A, EHEDG MR 4) DN 65 PN 25 86 12 Rd 95 x 1/6" 21 66 61 2.0 (4.41) 3A, EHEDG MS 4) DN 80 PN 25 100 12 Rd 110 x 1/4" 26 81 61 2.55 (5.62) 3A, EHEDG MT 4) 1) Surface roughness of the wetted surfaces R a < 0.76 µm (29.9 µin) as standard. 2) Product Configurator, order code for "Process connection" 3) With TempC diaphragm 4) Alternatively available with TempC membrane.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 85 NEUMO BioControl Process temperature range: –10 to +200 °C (+14 to +392 °F) D d dM h A0023435 Engineering unit mm (in) Material 1) NEUMO BioControl Diaphragm seal Approv- al Option 2) Max. diameter of the process isolating diaphragm Weight Nominal of the diaphragm Nominal pressure Diameter Height Standard with TempC membrane PN D d d2 d3 d1 m dM dM [bar] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [kg (lb)] AISI 316L DN 50 PN 16 90 17 4 x Ø 9 50 70 27 40 36 1.1 (2.43) 3A S4 3) DN 80 PN 16 140 25 4 x Ø 11 87.4 115 37 61 61 2.6 (5.73) 3A S6 3) 1) Surface roughness of the wetted surfaces R a < 0.76 µm (29.9 µin) as standard. 2) Product Configurator, order code for "Process connection" 3) With TempC membrane Universal process adapter ø43.5 (1.71) = ø36 (1.42) 68.5 (2.7) dM A0020013 Engineering unit mm (in)
Cerabar S PMC71, PMP71, PMP75
86 Endress+Hauser
Description Material 1) Weight Approval Option 2) [kg (lb)] Universal process adapter with pre-installed silicone molded seal (item 3) (spare part no.: 52023572) FDA 21CFR177.2600/USP Class VI-70C
- Item 1: Top section AISI 316L (1.4404)
- Item 2: Bottom section AISI 316L (1.4435) 0.8 (1.76) 3A, EHEDG 00 3) 4) 1) Surface roughness of the wetted surfaces R a < 0.76 µm (29.9 µin) as standard. Surface quality Ra< 0.38 µm (15 µin) electropolished (wetted) Ordering information: Product Configurator, order code for "Additional options 2", option "P". 2) Product Configurator, order code for "Process connection" 3) Endress+Hauser supplies these slotted nuts in stainless steel AISI 304 (DIN/EN material number 1.4301) or in AISI 304L (DIN/EN material number 1.4307). 4) Alternatively available with TempC membrane.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 87 Process connections for PMP75 with flush-mounted process isolating diaphragm EN/DIN flanges, connection dimensions in accordance with EN 1092-1/DIN 2527 and DIN 2501-1 dM k D g b f A0021680 Engineering unit mm Flange 1) 2) 3) Boltholes Diaphragm seal Option 4) Nominal diameter Nominal pressure Shape 5) D Thickness Raised face Number g2 Hole circle max. diameter process isolating diaphragm Weight b g f k [mm] [mm] [mm] [mm] [mm] [mm] dM [kg (lb)] DN 25 PN 10-40 B1 (D) 115 18 68 3 4 14 85 32 2.1 (4.63) CN DN 25 PN 63-160 B2 (E) 140 24 68 2 4 18 100 28 2.5 (5.51) DN DN 25 PN 250 B2 (E) 150 28 68 2 4 22 105 28 3.7 (8.16) EN DN 25 PN 400 B2 (E) 180 38 68 2 4 26 130 28 7.0 (15.44) E1 DN 32 PN 10-40 B1 (D) 140 18 77 2.6 4 18 100 34 1.9 (4.19) CP DN 40 PN 10-40 B1 (D) 150 18 87 2.6 4 18 110 48 2.2 (4.85) CQ DN 50 PN 10-40 B1 (D) 165 20 102 3 4 18 125 59 3.0 (6.62) B3 DN 50 PN 63 B2 (E) 180 26 102 3 4 22 135 59 4.6 (10.14) C3 DN 50 PN 100-160 B2 (E) 195 30 102 3 4 26 145 59 6.2 (13.67) EF DN 50 PN 250 B2 (E) 200 38 102 3 8 26 150 59 7.7 (16.98) ER DN 50 PN 400 B2 (E) 235 52 102 3 8 30 180 59 14.7 (32.41) E3 DN 80 PN 10-40 B1 (D) 200 24 138 3.5 8 18 160 89 5.3 (11.69) B4 DN 80 PN 100 B2 (E) 230 32 138 4 8 24 180 89 8.9 (19.62) C4 DN 100 PN 100 B2 (E) 265 36 175 5 8 30 210 89 13.7 (30.21) C5 1) material: AISI 316L 2) The roughness of the surface in contact with the medium including the raised face of the flanges (all standards) made of Alloy C276, Monel, tantalum, rhodium-gold or PTFE is Ra< 0.8 µm (31.5 µin). Reduced surface roughness on request. 3) The raised face of the flange is made of the same material as the process isolating diaphragm. 4) Product Configurator, order code for "Process connection" 5) Description as per DIN 2527 provided in brackets
Cerabar S PMC71, PMP71, PMP75
88 Endress+Hauser
EN/DIN flanges with barrel (extended diaphragm seal), connection dimensions in accordance with EN 1092-1/DIN 2527 and DIN 2501-1 dM k D g b A0023914 Engineering unit mm Flange 1) 2) Boltholes Diaphragm seal Option 3) Nominal diameter Nominal pressure Shape 4) D Thickness Raised face Number g2 Hole circle max. diameter of the process isolating diaphragm Weight b g k [mm] [mm] [mm] [mm] [mm] dM [kg (lb)] DN 50 PN 10-40 B1 (D) 165 20 102 4 18 125 47 5) D3 5) DN 80 PN 10-40 B1 (D) 200 24 138 8 18 160 72 5) D4 5) 1) Material: AISI 316L 2) In the case of process isolating diaphragms made of Alloy C276, Monel or tantalum, the raised face of the flange and the barrel pipe are made of 316L 3) Product Configurator, order code for "Process connection" 4) Description as per DIN 2527 provided in brackets 5) Available with 50 mm (1.97 in), 100 mm (3.94 in) and 200 mm (7.87 in) barrel, (extended diaphragm seal), for barrel (extended diaphragm seal) diameter and weight see the following table Option 1) Nominal diameter Nominal pressure Barrel (extended diaphragm seal) length Barrel (extended diaphragm seal) diameter Weight (L) d3 [mm] [mm] [kg (lb)] (17.2) 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 89 Process connections for PMP75 with flush-mounted process isolating diaphragm ASME flanges, in accordance with connection dimensions ASME B 16.5, raised face RF dM k D g b 1.57 f A0023913 Engineering unit in Flange 1) 2) 3) Boltholes Diaphragm seal Approval 4) Option 5) Nominal diameter Class D Thickness Raised face Number g2 Hole circle max. diameter of the process isolating diaphragm Weight b g f k dM [in] [lb./sq.in] [in] [in] [in] [in] [in] [in] [in] [kg (lb)] 1) Material AISI 316/316L: Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated) 2) The roughness of the surface in contact with the medium including the raised face of the flanges (all standards) made of Alloy C276, Monel, tantalum, rhodium-gold or PTFE is Ra< 0.8 µm (31.5 µin). Lower surface roughness on request. 3) The flange raised face is made of the same material as the process isolating diaphragm. 4) CSA approval: Product Configurator, order code for "Approval" 5) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
90 Endress+Hauser
ASME flanges with barrel (extended diaphragm seal), connection dimensions in accordance with ASME B 16.5, raised face RF dM k D g b 1.57 f A0021683 Engineering unit in Flange 1) 2) Boltholes Diaphragm seal Approval 3) Option 4) Nominal diameter Class D Thickness Raised face Number g2 Hole circle max. Ø membrane Weight b g f k dM [in] [lb./sq.in] [in] [in] [in] [in] [in] [in] [in] [kg (lb)] 1) Material: AISI 316/316L. Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated) 2) In the case of process isolating diaphragms made of Alloy C276, Monel or tantalum, the raised face of the flange and the barrel pipe are made of 316L. 3) CSA approval: Product Configurator, order code for "Approval" 4) Product Configurator, order code for "Process connection" 5) Choice of 2", 4", 6" or 8" barrel (extended diaphragm seal), for diameter and weight of barrel (extended diaphragm seal) see the following table Option 1) Nominal diaphragm Class Barrel (extended diaphragm seal) length Barrel (extended diaphragm seal) process isolating diaphragm Weight (L) d3 [in] [lb./sq.in] in (mm) in (mm) [kg (lb)] (9.7) 7.8 (17.2) 9.6 (21.2) 12.4 (27.3) 16.9 (37.3) 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 91 JIS flanges, connection dimensions in accordance with JIS B 2220 BL, raised face RF dM k D g b f A0021680 Engineering unit mm Flange 1) 2) 3) Boltholes Diaphragm seal Option 4) Nominal diameter Nominal pressure D Thickness Raised face Number g2 Hole circle max. Ø membrane Weight b g f k dM [mm] [mm] [mm] [mm] [mm] [mm] [mm] [kg (lb)] 25 A 10 K 125 14 67 1 4 19 90 32 1.5 (3.31) KC 50 A 10 K 155 16 96 2 4 19 120 59 2.3 (5.07) KF 80 A 10 K 185 18 127 2 8 19 150 89 3.3 (7.28) KL 100 A 10 K 210 18 151 2 8 19 175 89 4.4 (9.7) KH 1) Material: AISI 316L 2) The roughness of the surface in contact with the medium including the raised face of the flanges (all standards) made of Alloy C276, Monel, tantalum, rhodium-gold or PTFE is Ra< 0.8 µm (31.5 µin). Lower surface roughness on request. 3) The flange raised face is made of the same material as the process isolating diaphragm. 4) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
92 Endress+Hauser
Thread ½ NPT and 1 NPT, separator 76 (2.99) 94.5 (3.72) 26 (1.02) ø118 (4.65) ø30 (1.18) ø52 (2.05) ø118 (4.65) 19 (0.75)12 (0.47) NPT ½" NPT 1" A0021684 Engineering unit mm (in) Material Description Measuring range Nominal pressure Weight Option 1) [bar (psi)] [kg (lb)] AISI 316L Threaded, ½" NPT with Viton seal (200 °C (392 °F)) ≤ 250 (3625) PN 250 4.75 (10.47) UG Threaded, 1" NPT with Viton seal (200 °C (392 °F)) 5.0 (11.03) UH 1) Product Configurator, order code for "Process connection" Thread ISO 228 G ½ A and ANSI ½ MNPT, separator ø11.4 (0.45) M20x1.5 ø17.5 (0.69) NPT ½" 14/c180 ø6 (0.24) ø4 (0.16) G ½"A 20 (0.79)17 (0.67) 3 (0.12) ø18.4 (0.72) ø72.6 (2.86) ø min 4 (0.16) A0021685 Engineering unit mm (in) Material Description Measuring range Nominal pressure Approval Weight Option 1) [bar (psi)] [kg (lb)] AISI 316L Welded, ISO 228 G ½ A EN837 ≤ 160 (2320) PN 160 1.43 (3.15) UA Welded, ANSI ½ MNPT CRN 2) UB Welded, thread DIN13 M20x1.5 - UF 1) Product Configurator, order code for "Process connection" 2) CSA approval: Product Configurator, order code for "Approval"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 93 100 (3.94) A B 100 (3.94) NPT ½" G ½" ø6 (0.24) A0027135 A Option "UC" with threaded connection ISO 228 G ½ B B Option "UD" with threaded connection ANSI ½ MNPT 1 PTFE seal as standard max. 260 °C (500 °F) (higher temperatures on request) Engineering unit mm (in) Material Description Measuring range Nominal pressure Weight Option 1) [bar (psi)] [kg (lb)] AISI 316L (1.4404), screws made of A2 ISO 228 G ½ B EN837 ≤ 40 (580) PN 40 1.43 (3.15) UC ANSI ½ MNPT UD 1) Product Configurator, order code for "Process connection" ø115 (4.53) NPT ½" 84 (3.31) 19 (0.75) ø2.5 (0.1) 100 (3.94) 20 (0.79) 17 (0.67)3 (0.12) ø115 (4.53) G ½" ø2.5 (0.1) A0021688 Engineering unit mm (in) Material Description Measuring range Nominal pressure Weight Option 1) [bar (psi)] [kg (lb)] AISI 316L (1.4404), screws made of A2 Threaded, ISO 228 G ½ B EN837, with integrated seal lip > 40 bar (580) PN 400 4.75 (10.47) UC Threaded, ANSI ½ MNPT, with integrated seal lip UD 1) Product Configurator, order code for "Process connection"
Cerabar S PMC71, PMP71, PMP75
94 Endress+Hauser
Separate housing: Wall and pipe mounting with mounting bracket 122 (4.8) 52 (2.05) 86 (3.39)70 (2.76) 84 (3.31) 129 (5.08) 147 (5.79) 140 (5.51) A B 76 (2.99) 6 (0.24) ø6 (0.24) A0020017 Engineering unit mm (in) Item Description Weight in kg (lb) Option 1) Housing (T14 or T17) Mounting bracket A Dimensions with T14 housing, optional side display → 54 0.5 (1.10) U B Dimensions with T17 housing, optional side display 1) Product Configurator, order code for "Additional options 2", version "G" Also available for order as a separate accessory: Part number 71102216
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 95 Materials not in contact with process Transmitter housing T14 1011 A0020019 Item number Component part Material
1 T14 housing, RAL 5012 (blue)
- Die-cast aluminum with protective powder-coating on polyester base
- Thread coating: Heat-curing lubricant varnish 1 T14 housing • Precision casting AISI 316L (1.4435)
- Thread coating: Heat-curing lubricant varnish
2 Cover, RAL 7035 (gray) Die-cast aluminum with protective powder-coating
3 Cover seal EPDM
4 Nameplates • AISI 316L (1.4404), if T14 housing is precision- cast
- Anodized aluminum, if housing T14/T15 of die- cast aluminum 5 Pressure compensation filter AISI 316L (1.4404) and PBT-FR
6 Pressure compensation filter, O-
7 Sight glass Mineral glass
8 Sight glass seal Silicone (VMQ)
9 Screw A4
10 Sealing ring EPDM
11 Snap ring PA66-GF25
12 Snap ring for nameplates AISI 304 (1.4301)/AISI 316 (1.4401) 13 External ground terminal AISI 304 (1.4301) 14 Cover clamp Clamp AISI 316L (1.4435), screw A4
15 Cable entry Polyamide (PA) or CuZn nickel-plated
16 Seal of cable entry and plug Silicone (VMQ)
17 Blind plug PBT-GF30 FR, for dust ignition-proof: AISI 316L
(1.4435)
18 External operation (keys and key
cover), RAL 7035 (gray) Polycarbonate PC-FR, screw A4 Devices with MID parts certificate Seal wire DIN 1367-0 St/Zn (soft galvanized steel) Devices with MID parts certificate Seals Pb (lead)
Cerabar S PMC71, PMP71, PMP75
96 Endress+Hauser
Item number Component part Material
1 T17 housing
AISI 316L (1.4404)
2 Cover
4 Nameplates Lasered on
5 Pressure compensation filter AISI 316L (1.4404) and PBT-FR
6 Pressure compensation filter, O-ring VMQ or EPDM
7 Sight glass for non-hazardous area, ATEX Ex
ia, NEPSI Zone 0/1 Ex ia, IECEx Zone 0/1 Ex ia, FM NI, FM IS, CSA IS Polycarbonate (PC)
8 Sight glass for ATEX 1/2 D, ATEX 1/3 D,
ATEX 1 GD, ATEX 1/2 GD, ATEX 3 G, FM DIP, CSA dust ignition-proof Mineral glass
9 Sight glass seal EPDM
10 Screw A2-70
11 Sealing ring EPDM
12 Snap ring PA6
13 Screw A4-50
Thread coating: Heat-curing lubricant varnish 14 External ground terminal AISI 304 (1.4301)
15 Cable entry Polyamide PA, for dust ignition-proof:
17 Blind plug PBT-GF30 FR, for dust ignition-proof:
AISI 316L (1.4435) Devices with MID parts certificate Seal wire DIN 1367-0 St/Zn (soft galvanized steel) Devices with MID parts certificate Seals Pb (lead)
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 97 Connecting parts 1 1 2 3 4 4 6591 A0028222 Item number Component part Material
1 Connection between the housing and
AISI 316L (1.4404) 2 Mounting bracket Bracket AISI 316L (1.4404)
3 Screw and nuts A4-70
4 Half-shells: AISI 316L (1.4404)
5 Seal for cable from separate housing EPDM
6 Gland for cable from separate housing AISI 316L (1.4404)
7 PE cable for separate housing abrasion-proof cable with strain-relief Dynema
members; shielded using aluminum-coated foil; insulated with polyethylene (PE-LD), black; copper wires, twisted, UV‑resistant
8 FEP cable for separate housing Abrasion-proof cable; shielded using galvanized
steel wire netting; insulated with fluorinated ethylene propylene (FEP), black; copper cores, twisted, UV-resistant
9 Process connection adapter for separate
AISI 316L (1.4404) 10 Cell body AISI 316L (1.4404)
11 Connection between body of measuring cell
AISI 316L (1.4404)
12 Heat-shrink tube
(available only if capillary has PTFE or PVC sheath) Polyolefin
Cerabar S PMC71, PMP71, PMP75
98 Endress+Hauser
B PVC-coated Armor for capillary C PTFE hose Armor for capillary 1 Capillary AISI 316 Ti (1.4571) AISI 316 Ti (1.4571) AISI 316 Ti (1.4571) 2 Flexible armor for capillary AISI 316L (1.4404) 1) AISI 316L (1.4404) AISI 316L (1.4404)
3 Coating/armor - PVC 2) PTFE 3)
4 Single-ear clamp - - 1.4301
5 Shrink tubing at capillary junction - Polyolefin -
1) Product Configurator, order code for "Armor for capillary:" option "SA" 2) Product Configurator, order code for "Capillary armoring:" option "SB" 3) Product Configurator, order code for "Capillary armoring:" option "SC" Weight Component part Weight Housing See "Housing" section Process connection See "Process connections" section Capillary with armoring made of AISI 316L (PVC) 0.21 kg/m (0.46 lb/m) + 0.35 kg (0.77 lb) Capillary with armoring made of AISI 316L (PTFE) 0.29 kg/m (0.64 lb/m) + 0.35 kg (0.77 lb) Materials in contact with process NOTICE ‣ The device components in contact with the process are specified in the "Mechanical construction" → 53 and "Ordering information" → 119 sections. Delta-ferrite content A delta-ferrite content of ≤ 3% can be guaranteed and certified for the wetted parts if option "8" is selected in the "Additional options 1" or "Additional options 2" order code in the Product Configurator. If the PMC71 with hygienic process connections is selected, a delta-ferrite content of ≤ 1% can be guaranteed and certified if option "8" is selected in the "Additional options 1" or "Additional options 2" order code in the Product Configurator. TSE Certificate of Suitability (Transmissible Spongiform Encephalopathy) The following applies to all device components in contact with the process:
- They do not contain any materials derived from animals.
- No additives or operating materials derived from animals are used in production or processing.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 99 Process connections
- "Clamp connections" and "Hygienic process connections": AISI 316L (DIN/EN material number 1.4435)
- Endress+Hauser supplies process connections with a threaded connection as well as DIN/ EN flanges made of stainless steel as per AISI 316L (DIN/EN material number 1.4404 or 1.4435). With regard to their stability-temperature property, the materials 1.4404 and 1.4435 are grouped together under 13E0 in EN 1092-1: 2001 Tab. 18. The chemical composition of the two materials can be identical.
- Some process connections are also available in alloy C276 (DIN/EN material number 2.4819). For this purpose see the information in the "Mechanical construction" section. Process isolating diaphragm Sensor Description Option 1) PMC71 Al2O3 aluminum-oxide ceramic FDA, ultra-pure 99.9 % 2) Ceraphire® (see also www.endress.com/ceraphire) Standard PMP71 AISI 316L 1 AISI 316L with gold-rhodium coating 6 Alloy C276 (2.4819) 2 PMP75 AISI 316L 1 AISI 316L, TempC E AISI 316L with gold-rhodium coating 6 AISI 316L with 0.25 mm (0.01 in)PTFE coating 8 Alloy C276 (2.4819) 2 3) Monel (2.4360) 3 3) Tantalum (UNS R05200) 5 3) 1) Product Configurator, order code for "Membrane material" 2) The US Food & Drug Administration (FDA) has no objections to the use of ceramics made of aluminum oxide as a surface material in contact with foodstuffs. This declaration is based on the FDA certificates of our ceramic suppliers 3) The material of the flange raised face is the same material as is used for the process isolating diaphragm. For devices with a barrel (extended diaphragm seal), the flange raised face and the barrel pipe are made of 316L. Seals Device Description Option 1) PMC71 FKM Viton A FKM Viton, FDA G EPDM B FFKM Perlast G75LT C Kalrez D Chemraz E NBR (FDA)/3A: HNBR (FDA) F FKM Viton, cleaned for silicone-free applications L Kalrez, cleaned for silicone-free applications M FKM Viton, cleaned from oil and grease 1 FKM Viton, cleaned for oxygen service, note pressure and temperature application limits 2 1) Product Configurator, order code for "Seal"
Cerabar S PMC71, PMP71, PMP75
100 Endress+Hauser
Description Option 1) Silicone oil A Inert oil F Inert oil, cleaned from oil and grease K Inert oil, cleaned for oxygen service (observe application limits pressure/temperature) N 1) Product Configurator, order code for "Fill fluid" PMP75 Description Option 1) 2) Silicone oil (food-safe FDA 21 CFR 175.105) A ...m capillary, inert oil B ...ft capillary, inert oil C Vegetable oil (food-safe FDA 21 CFR 172.856) D Inert oil F High-temperature oil, temperature isolator G Silicone oil, temperature isolator (food-safe FDA 21 CFR 175.105) H Vegetable oil, temperature isolator J Inert oil, cleaned from oil and grease K Inert oil, cleaned for oxygen service N ...m capillary, silicone oil (food-safe FDA 21 CFR 175.105) 1 ...ft capillary, silicone oil (food-safe FDA 21 CFR 175.105) 2 ...m capillary, high-temperature oil 3 ...ft capillary, high-temperature oil 4 ...m capillary, vegetable oil (food-safe FDA 21 CFR 172.856) 5 ...ft capillary, vegetable oil (food-safe FDA 21 CFR 172.856) 6 ...m capillary, low-temperature oil 7 ...ft capillary, low-temperature oil 8 1) Product Configurator, order code for "Fill fluid" 2) For diaphragm seal devices with 3-A and EHEDG certificates, only select filling oils with FDA approval!
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 101 Operability Operating concept Operator-oriented menu structure for user-specific tasks
- Commissioning
- Operation
- Diagnosis Fast and safe commissioning Guided menus for applications Reliable operation
- Local operation possible in several languages
- Standardized operation at the device and in the operating tools
- Parameters relating to measured values can be locked/unlocked using the device's write protection switch, using the device software or via remote operation Efficient diagnostics increase measurement availability
- Remedial measures are integrated in plain text
- Diverse simulation options Local operation Functions Function External operation (operating keys, optional, not T17 housing) Internal operation (electronic insert) Onsite display (optional) Position adjustment (zero point correction) Setting lower-range value and upper-range value - reference pressure present at the device (HART only) (HART only) Device reset Locking and unlocking parameters relevant to the measured value ¾ Value acceptance indicated by green LED Switching damping on and off (only if display is connected) (HART and PA only) Configuring the bus address of the device (PA) ¾ Switching simulation mode on and off (FOUNDATION Fieldbus) ¾ Operating the device using onsite display (optional) A 4-line liquid crystal display (LCD) is used for display and operation. The onsite display shows measured values, dialog text as well as fault and notice messages in plain text, thereby supporting the user in every stage of operation. The display can be removed for easy operation. The device display can be turned in 90° steps. Depending on the installation position of the device, this makes it easy to operate the device and read the measured value.
Cerabar S PMC71, PMP71, PMP75
102 Endress+Hauser
Functions:
- 8-digit measured value display incl. leading sign and decimal point, bar graph for – 4 to 20 mA HART as current display – PROFIBUS PA as graphic display of the standardized value of the AI block – FOUNDATION Fieldbus as graphic display of the transducer output.
- Simple and complete menu guidance due to breakdown of parameters into several levels and groups
- Menu guidance in up to 8 languages
- Each parameter is given a 3-digit ID number for easy navigation.
- Option for configuring the display according to individual requirements and preferences, such as language, alternating display, display of other measured values such as sensor temperature, contrast setting.
- Comprehensive diagnostic functions (fault and warning message, peak-hold indicators, etc.).
- Rapid and safe commissioning with the Quick Setup menus. Overview E+– 4 5 A0016498
1 Operating keys
2 Bargraph
3 Symbol
4 Header
5 Parameter ID number
Operating keys on the exterior of the device With the aluminum housing (T14), the operating keys are located either outside on the housing, under the protection cap or inside on the electronic insert. With the stainless steel housing (T17), the operating keys are always located inside the housing on the electronic insert. Zero A B A0020030 A 4 to 20 mA HART B PROFIBUS PA and FOUNDATION Fieldbus The operating keys located externally on the device work on the Hall sensor principle. As a result, no additional openings are required in the device. This guarantees:
- Complete protection against environmental influences such as moisture and contamination.
- Simple operation without any tools.
- No wear. Ordering information:
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 103 Product Configurator, order code for "Output, operation" Operating keys and elements located internally on the electronic insert Ordering information: Product Configurator, order code for "Output, operation" HART 2 1 PC 3 4 5 61 2 /c116 on off A0020031
1 DIP switch for locking/unlocking parameters relevant to the measured value
2 DIP switch for switching damping on/off
3 Green LED to indicate value being accepted
4 Operating keys
5 Slot for optional display
6 Slot for optional HistoROM®/M-DAT
/c116 on off 1 2 A0020032
4 Key for position adjustment and device reset
5 DIP switch for bus address
6 Slot for optional display
7 Slot for optional HistoROM®/M-DAT
Cerabar S PMC71, PMP71, PMP75
104 Endress+Hauser
Sim. A0020033
2 DIP switch for simulation mode on/off
Remote operation All software parameters are accessible depending on the position of the write protection switch on the device. Hardware and software for remote operation HART PROFIBUS PA FOUNDATION Fieldbus FieldCare FieldXpert SFX100 ¾ NI-FBUS Configurator ¾ ¾ HistoROM®/M-DAT FieldCare FieldCare is an Endress+Hauser asset management tool based on FDT technology. With FieldCare, you can configure all Endress+Hauser devices as well as devices from other manufacturers that support the FDT standard. FieldCare supports the following functions:
- Configuration of transmitters in offline and online mode
- Loading and saving device data (upload/download)
- HistoROM®/M-DAT analysis
- Documentation of the measuring point Connection options:
- HART via Commubox FXA195 and the USB interface of a computer
- PROFIBUS PA via segment coupler and PROFIBUS interface card
- Service interface with Commubox FXA291 and ToF adapter FXA291 (USB). For further information please contact your local Endress+Hauser Sales Center. Field Xpert SFX100 Field Xpert is an industrial PDA with integrated 3.5" touchscreen from Endress+Hauser based on Windows Mobile. It offers wireless communication via the optional VIATOR Bluetooth modem from Endress+Hauser. Field Xpert also works as a stand-alone device for asset management applications. For details, refer to BA00060S/04/EN. Commubox FXA195 For intrinsically safe HART communication with FieldCare via the USB interface. For details refer to TI00404F/00/EN.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 105 Commubox FXA291 The Commubox FXA291 connects Endress+Hauser field devices with a CDI interface (=Endress +Hauser Common Data Interface) to the USB interface of a personal computer or a notebook. For details refer to TI00405C/07/EN. For the following Endress+Hauser devices you need the "ToF adapter FXA291" as an additional accessory:
- Cerabar S PMC71, PMP7x
- Deltabar S PMD7x, FMD7x
- Deltapilot SFMB70 ToF adapter FXA291 The ToF adapter FXA291 connects the Commubox FXA291 with devices of the ToF platform, pressure equipment and Gammapilot via the USB interface of a personal computer or a notebook. For details refer to KA00271F. Profiboard For connecting a PC to PROFIBUS. Proficard For connecting a laptop to PROFIBUS FF configuration program FF configuration program, such as NI-FBUS Configurator, to
- connect devices with "FOUNDATION Fieldbus signal" into an FF network
- set FF-specific parameters Remote operation via the NI-FBUS Configurator: The NI-FBUS Configurator is an easy-to-use graphical environment for creating linkages, field-based control loops and schedules based on the FOUNDATION Fieldbus concept. You can use the NI-FBUS Configurator to configure a fieldbus network as follows:
- Set block and device tags
- Set device addresses
- Create and edit function block control strategies (function block applications)
- Configure vendor-defined function and transducer blocks
- Create and edit schedules
- Read and write to function block control strategies (function block applications)
- Invoke Device Description (DD) methods
- Display DD menus
- Download a configuration
- Verify a configuration and compare it to a saved configuration
- Monitor a downloaded configuration
- Replace a virtual device by a real device
- Save and print a configuration
Cerabar S PMC71, PMP71, PMP75
106 Endress+Hauser
HistoROM®/M-DAT (optional) HistoROM®/M-DAT is a memory module which can be attached to every electronic insert. HistoROM®/M-DAT can be retrofitted at any stage (order number: 52027785). Your benefits
- Quick and safe commissioning of the same measuring points by copying the configuration data of one transmitter to another transmitter.
- Reliable process monitoring thanks to cyclical recording of pressure and sensor temperature measured values.
- Simple diagnosis by recording diverse events such as alarms, configuration changes, counters for measuring range undershoot and overshoot for pressure and temperature as well as user limit overshoot and undershoot for pressure and temperature etc.
- Analysis and graphic evaluation of the events and process parameters via software (contained in scope of supply). A CD with an Endress+Hauser operating program is also included in the scope of delivery. You can copy data from one transmitter to another transmitter when operating a FOUNDATION Fieldbus device via an FF configuration program. You need the Endress+Hauser FieldCare operating program, the Commubox FXA291 service interface and the ToF adapter FXA291 to be able to access the data and events saved in the HistoROM®/M-DAT. Ordering information: Product Configurator, order code for "Additional options:", version "N" or Product Configurator, order code for "Application package:", option "EN" or as a separate accessory (part no.: 52027785). For further information please contact your local Endress+Hauser Sales Center. System integration The device can be given a tag name (max. 8 alphanumeric characters) Description Option 1) Measuring point (TAG), see additional spec. Z1 Bus address, see additional spec. Z2 1) Product Configurator, order code for "Identification"
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 107 Planning instructions, diaphragm seal systems NOTICE Incorrect sizing/ordering of diaphragm seal systems The performance and the permitted range of application of a diaphragm seal system depend on the process isolating diaphragm used, the filling oil, the coupling, the unit design and on the process and ambient conditions present in the individual application. ‣ To help you select the right diaphragm seal system for your applications, Endress+Hauser provides its customers with the free "Applicator Sizing Diaphragm Seal" tool, which is available on DVD or at "www.endress.com/applicator". A0021695-EN
1 My Applicator - configuration of the Applicator settings
2 Applicator Help
3 Mouse-over help - hover the cursor over these fields for brief information
For more detailed information or the layout of the optimum diaphragm seal solution for your application, please contact your local Endress+Hauser Sales Center. Applications Diaphragm seal systems should be used if the process and the device should be separated. Diaphragm seal systems offer clear advantages in the following instances:
- In the case of extreme process temperatures
- For aggressive media
- In the case of process media that crystallize
- In the case of corrosive or highly various process media or process media with solids content
- In the case of heterogeneous and fibrous process media
- If extreme measuring point cleaning is necessary, or for very humid mounting locations
- If the measuring point is exposed to severe vibrations
- For mounting locations that are difficult to access
Cerabar S PMC71, PMP71, PMP75
108 Endress+Hauser
Design and operation mode Diaphragm seals are separating equipment between the measuring system and the process. A diaphragm seal system consists of:
- A diaphragm seal
- If necessary, a capillary tube or temperature isolator
- Fill fluid
- A pressure transmitter The process pressure acts via the process isolating diaphragm of a diaphragm seal on the liquid-filled system, which transfers the process pressure onto the sensor of the pressure transmitter. Endress+Hauser delivers all diaphragm seal systems as welded versions. The system is hermetically sealed, which ensures the highest reliability. The diaphragm seal determines the application range of the system by:
- The process isolating diaphragm diameter
- The process isolating diaphragm stiffness and material
- The design (oil volume) diameter of the process isolating diaphragm The greater the diameter of the process isolating diaphragm (less stiff), the smaller the temperature effect on the measurement result. Stiffness of the process isolating diaphragm The stiffness depends on the diameter of the process isolating diaphragm, the material, any existing coating and the thickness and shape of the process isolating diaphragm. The process isolating diaphragm thickness and the shape are determined by the design. The stiffness of a process isolating diaphragm of a diaphragm seal influences the temperature operating range and the measuring error caused by temperature effects. The new TempC membrane: Highest accuracy and process safety in pressure and differential pressure measurement with diaphragm seals To measure even more precisely in such applications and to increase process safety, Endress+Hauser has developed the TempC membrane based on a completely revolutionary technology. This diaphragm guarantees the highest level of accuracy and process safety in diaphragm seal applications.
- The very low temperature effect minimizes the influence of fluctuations of both process and ambient temperatures, thus guaranteeing accurate and stable measurements. Measurement inaccuracies caused by temperature are reduced to a minimum.
- The TempC membrane can be used at temperatures between –40 °C (–40 °F) and +250 °C (+482 °F). This guarantees maximum process safety even for very long sterilization and cleaning cycles (SIP/CIP) in tanks and pipes at high temperatures.
- Thanks to the TempC membrane, smaller dimension process connections are possible. With a smaller process connection, the new diaphragm measures at least as accurately as a conventional diaphragm with a larger diameter.
- Short recovery times after temperature shocks mean less downtime during batch processes and therefore a far higher level of availability of the production facilities.
- In addition, the TempC membrane excels in terms of its improved hygienic cleanability and its insensitivity to substantial changes in the pressure load. Ordering information: See the Product Configurator for the individual process connection and the choice of process isolating diaphragm. Selection in the Applicator: Under "Transmitter data" in the "Diaphragm material" field. Capillary As standard, capillaries with an internal diameter of 1 mm (0.04 in) are used. The capillary tube influences the thermal change, the ambient temperature operating range and the response time of a diaphragm seal system as a result of its length and internal diameter.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 109 Filling oil When selecting the filling oil, the media and ambient temperature as well as the operating pressure are of crucial importance. Observe the temperatures and pressures during commissioning and cleaning. A further selection criterion is the compatibility of the filling oil with the requirements of the process media. For this reason, only filling oils that are harmless to health may be used in the food industry, such as vegetable oil or silicone oil (see also the following section on "diaphragm seal filling oils"). The filling oil used influences the thermal change, the temperature operating range of a diaphragm seal system and the response time. A temperature change results in a volume change of the filling oil. The volume change is dependent on the expansion coefficient and on the volume of the filling oil at calibration temperature (constant in range: +21 to +33 °C (+70 to +91 °F)). The application range can be extended by a filling oil with a lower expansion coefficient and a shorter capillary. For example, the filling oil expands in the event of a temperature increase. The additional volume presses against the process isolating diaphragm of a diaphragm seal. The stiffer a diaphragm is, the greater its return force, which counteracts a volume change and acts on the measuring cell together with the operating pressure, thus shifting the zero point. Pressure transmitter The pressure transmitter influences the temperature operating range, the TK zero point and the response time as a result of its volume change. The volume change is the volume that has to be shifted to pass through the complete measuring range. Pressure transmitters from Endress+Hauser are optimized with regard to minimum volume change. Diaphragm seal filling oils Filling oil Permitted temperature range 1) at 0.05 bar (0.725 psi) ≤ pabs ≤ 1 bar (14.5 psi) Permitted temperature range 1) at pabs ≥1 bar (14.5 psi) Density [g/cm3] / [SGU] Viscosity [mm²/s] / [cSt] at 25 °C (77 °F) Expansion coefficient 2) [1/K] Note 3) Option 4) Silicone oil –40 to +180 °C (–40 to +356 °F) –40 to +250 °C (–40 to +482 °F) 0.96 100 0.00096 (Suitable for use in food FDA 21 CFR 175.105) A, H, 1 or 2 High- temperature oil –10 to +200 °C (+14 to +392 °F) –10 to +400 °C 1.00 150 0.00096 (High temperatures) G, 3 or 4 Inert oil –40 to +80 °C (–40 to +176 °F) –40 to +175 °C (–40 to +347 °F) 1.87 27 0.000876 (For ultrapure gas and oxygen applications) F or N Vegetable oil –10 to +120 °C (+14 to +248 °F) –10 to +200 °C (+14 to +392 °F) 0.94 9.5 0.00101 (Suitable for use in food FDA 21 CFR 172.856) D, 5 or 6 Low- temperature oil –70 to +80 °C (–94 to +176 °F) –70 to +180 °C (–94 to +356 °F) 0.92 4.4 0.00108 (Low temperatures) 7 or 8 1) Observe temperature limits of the device and the system. 2) Please refer to the "Applicator Sizing Diaphragm Seal" selection tool for the thermal change of the diaphragm seal and other important technical characteristics. 3) For diaphragm seal devices with 3-A and EHEDG certificates, only select filling oils with FDA approval! 4) Product Configurator, order code for "Fill fluid" 5) 325 °C (617 °F) at ≥1 bar (14.5 psi) absolute pressure. 6) 350 °C (662 °F) at ≥1 bar (14.5 psi) absolute pressure (max. 200 hours). 7) 400 °C (752 °F) at ≥1 bar (14.5 psi) absolute pressure (max. 10 hours).
Cerabar S PMC71, PMP71, PMP75
110 Endress+Hauser
Information on cleaning • Endress+Hauser provides flushing rings as an accessory to enable cleaning of the process isolating diaphragm without removing the transmitter from the process. For further information please contact your local Endress+Hauser Sales Center.
- We recommend you perform CIP (cleaning in place (hot water)) before SIP (sterilization in place (steam)) for pipe diaphragm seals. A frequent use of sterilization in place (SIP) will increase the stress on the process isolating diaphragm. Under unfavorable circumstances in the long term view we cannot exclude that a frequent temperature change could lead to a material fatigue of the process isolating diaphragm and possibly to a leakage. Installation instructions Diaphragm seal systems
- The diaphragm seal together with the transmitter form a closed, calibrated system, which is filled through ports in the diaphragm seal and in the measuring system of the transmitter. These ports are sealed and must not be opened.
- For devices with a temperature isolator or capillary, a suitable fastening device (mounting bracket) is recommended.
- When mounting, sufficient strain relief must be provided for the capillary line to prevent the capillary from bending (capillary bending radius ≥ 100 mm (3.94 in)
- For more detailed installation instructions, Endress+Hauser provides its customers with the free "Applicator Sizing Diaphragm Seal" tool, which is available on CD or can be downloaded online at "www.endress.com/applicator". Capillary In order to obtain more precise measurement results and to avoid a defect in the device, mount the capillaries as follows:
- vibration-free (in order to avoid additional pressure fluctuations)
- not in the vicinity of heating or cooling lines
- insulate if the ambient temperature is below or above the reference temperature
- with a bending radius ≥ 100 mm (3.94 in)
- When using diaphragm seal systems with a capillary, sufficient strain relief must be provided to prevent the capillary from bending (capillary bending radius ≥ 100 mm (3.94 in)).
- In the case of devices with diaphragm seals and capillaries, the zero point shift caused by the hydrostatic pressure of the filling liquid column in the capillaries must be taken into account when selecting the measuring cell. If a measuring cell with a small measuring range is selected, a position adjustment can cause range violation. Heat insulation The PMP75 must only be insulated up to a certain height. The maximum permitted insulation height is indicated on the devices and applies to an insulation material with a heat conductivity ≤ 0.04 W/(m x K) and to the maximum permitted ambient and process temperature. The data were determined under the most critical application "quiescent air". Maximum permitted insulation height, here indicated on a PMP75 with a flange: A B 1 2 A0020474 A Ambient temperature ≤ 70 °C (158 °F) B Process temperature 400 °C (752 °F), depends on the diaphragm seal filling oil used
1 Maximum insulation height
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 111 Mounting with temperature isolator Endress+Hauser recommends the use of temperature isolators in the event of constant extreme fluid temperatures which cause the maximum permitted electronics temperature of +85 °C (+185 °F) to be exceeded. Diaphragm seal systems with temperature isolators can be used with maximum temperatures up to +400 °C (+752 °F) depending on the filling oil used → 111, see the "Diaphragm seal filling oils" section. To minimize the influence of rising heat, Endress+Hauser recommends the device be mounted horizontally or with the housing pointing downwards. The additional installation height also brings about a zero point shift of maximum 21 mbar (0.315 psi) due to the hydrostatic column in the temperature isolator. You can correct this zero point shift at the device. The temperature restrictions are at their lowest with an insulation height of 30 mm (1.18 in). Complete insulation behaves in practically the same way as no insulation! The graphic below illustrates the temperature limits with an insulation height of 30 mm (1.18 in): Tp Ta +400 +350 +300 +250 +200 +150 +100 +50 -50 -100 +752 +662 +572 +482 +392 +302 +212 +122 +32 -58 -148 [°C] [°F] [°C][°F] A B A D E FG C B A0031354 A Ambient temperature: ≤ 85 °C (185 °F) B Process temperature: max. 400 °C (752 °F), depends on the diaphragm seal filling oil used C Device with temperature isolator, material 316L (1.4404) D No insulation E Maximum insulation F 30 mm (1.18 in)Insulation G No insulation, maximum insulation, 30 mm (1.18 in) insulation 1 Insulation height30 mm (1.18 in)
Cerabar S PMC71, PMP71, PMP75
112 Endress+Hauser
Vacuum applications Installation instructions For applications under vacuum, Endress+Hauser recommends mounting the pressure transmitter below the diaphragm seal. This prevents vacuum loading of the diaphragm seal caused by the presence of fill fluid in the capillary. When the pressure transmitter is mounted above the diaphragm seal, the maximum height difference H1 in accordance with the illustration below must not be exceeded. The following graphic depicts mounting above the lower diaphragm seal: FIELD TERMINALS A0020472 The maximum height difference depends on the density of the filling oil and the smallest ever pressure that is permitted to occur at the diaphragm seal (empty container), see the following diagram. The following diagram depicts the maximum installation height above the diaphragm seal for vacuum applications. A B 0.0 6.6 [ft] A0023986-EN A Height difference H1 B Pressure at diaphragm seal
1 Low-temperature oil
2 Vegetable oil
3 Silicone oil
4 High-temperature oil
5 Inert oil
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 113 Certificates and approvals CE mark The device meets the legal requirements of the relevant EC directives. Endress+Hauser confirms that the device has been successfully tested by applying the CE mark. RCM-Tick marking The supplied product or measuring system meets the ACMA (Australian Communications and Media Authority) requirements for network integrity, interoperability, performance characteristics as well as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic compatibility are met. The products are labelled with the RCM- Tick marking on the name plate. A0029561 Ex approvals • ATEX
- FM
- CSA
- NEPSI
- IECEx
- TIIS
- Also combinations of different approvals All explosion protection data are given in separate documentation which is available upon request. The Ex documentation is supplied as standard with all Ex-systems → 122. EAC conformity The measuring system meets the legal requirements of the applicable EAC guidelines. These are listed in the corresponding EAC Declaration of Conformity together with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the EAC mark. Suitable for hygiene
The device is available with hygienic process connections (overview: see order code). The materials of the hygienic process connections in contact with foodstuffs comply with framework Regulation (EC) 1935/2004. LCAUTION Contamination in the process! Risk of contamination if incorrect seals and parts are used! ‣ To avoid the risk of contamination, when installing the device comply with the design principles of EHEDG, Guideline 37 "Hygienic Design and Application of Sensors" and Guideline 16 "Hygienic Pipe Couplings". ‣ Suitable assemblies and seals must be used to ensure hygienic design in accordance with 3-A SSI and EHEDG specifications. ‣ The leak-proof connections can be cleaned with the cleaning methods typical of this industry (CIP and SIP). Attention must be paid to the pressure and temperature specifications of the sensor and process connections for CIP and SIP processes (clean in place/sterilize in place). ‣ For diaphragm seal devices with 3-A and EHEDG certificates, only select filling oils with FDA approval!
Cerabar S PMC71, PMP71, PMP75
114 Endress+Hauser
The gap-free connections can be cleaned of all residue using the usual cleaning methods within this industry. Functional safety SIL/ IEC 61508 Declaration of Conformity (optional) The Cerabar S devices with a 4 to 20 mA output signal have been developed in accordance with the IEC 61508 standard. These devices can be used to monitor the process level and pressure up to SIL 3. For a detailed description of the safety functions with Cerabar S, settings and functional safety data, see the "Functional safety manual - Cerabar S" SD00190P/00. For devices up to SIL 3 / IEC 61508 Declarations of Conformity see: Ordering information: Product Configurator, order code for "Additional options 1" and "Additional options 2", version "E". Overfill protection WHG (see document ZE00260P/00/DE) Ordering information: Product Configurator, order code for "Approval", option "6". CRN approval Some device versions have CRN approval. A CRN-approved process connection with a CSA approval must be ordered for a CRN-approved device. These devices are fitted with a separate nameplate with the registration number CRN 0F10525.5C . PMP75 devices with a capillary are not CRN-approved. Ordering information: Product Configurator, order code for "Process connection; material" and Product Configurator, order code for "Approval" (only in conjunction with an approved process connection) Other standards and guidelines The applicable European guidelines and standards can be found in the relevant EU Declarations of Conformity. The following were also applied: DIN EN 60770 (IEC 60770): Transmitters for use in industrial-process control systems. Part 1: Methods for operating performance evaluation DIN 16086: Electrical pressure measuring instruments, pressure sensors, pressure transmitters, pressure measuring instruments, concepts, specifications on data sheets EN 61326-X: EMC product family standard for electrical equipment for measurement, control and laboratory use. EN 60529: Degrees of protection provided by enclosures (IP code)
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 115 WELMEC guide 8.8: General and Administrative Aspects of the Voluntary System of Modular Evaluation of Measuring instruments under the MID. OIML R117-1 Edition 2007 (E): Dynamic measuring systems for liquids other than water. EN 12405-1/A1 Edition 2006: Gas meters – Conversion devices – Part 1: Volume conversion Pressure Equipment Directive 2014/68/EU (PED) Pressure equipment with allowable pressure ≤ 200 bar (2 900 psi) Pressure equipment (having a maximum allowable pressure PS ≤ 200 bar (2 900 psi)) can be classified as pressurized equipment in accordance with Pressure Equipment Directive 2014/68/EU. If the maximum allowable pressure is ≤ 200 bar (2 900 psi) and the pressurized volume of the pressure equipment is ≤ 0.1 l, the pressure equipment is subject to the Pressure Equipment Directive (cf. Pressure Equipment Directive 2014/68/EU, Article 4, point 3). The Pressure Equipment Directive only requires that the pressure equipment shall be designed and manufactured in accordance with the "sound engineering practice of a Member State". Reasons:
- Pressure Equipment Directive (PED) 2014/68/EU Article 4, point 3
- Pressure equipment directive 2014/68/EU, Commission's Working Group "Pressure", Guideline A-05 + A-06 Note: A partial examination shall be performed for pressure instruments that are part of safety equipment for the protection of a pipe or vessel from exceeding allowable limits (equipment with safety function in accordance with Pressure Equipment Directive 2014/68/EU, Article 2, point 4). Pressure equipment with allowable pressure > 200 bar (2 900 psi) Pressure equipment designated for application in every process fluid having a pressurized volume of <0.1 l and a max. allowable pressure PS > 200 bar (2 900 psi) shall satisfy the essential safety requirements set out in Annex I of the Pressure Equipment Directive 2014/68/EU. According to Article 13 pressure equipment shall be classified by category in accordance with Annex II. The conformity assessment of the pressure equipment shall be determined by the category I under consideration of the above-mentioned low pressurized volume. These devices shall be provided with CE marking. Reasons:
- Pressure Equipment Directive 2014/68/EU, Article 13, Annex II
- Pressure equipment directive 2014/68/EU, Commission's Working Group "Pressure", Guideline A-05 Note: A partial examination shall be performed for pressure instruments that are part of safety equipment for the protection of a pipe or vessel from exceeding allowable limits (equipment with safety function in accordance with Pressure Equipment Directive 2014/68/EU, Article 2, point 4). The following also applies:
- PMP71 with threaded connection and internal process isolating diaphragm PN > 200 as well as oval flange adapter PN > 200: Suitable for stable gases in group 1, category I, module A
- PMP75 with pipe diaphragm seal ≥ 1.5"/PN40: Suitable for stable gases in group 1, category II, module A2
- PMP75 with barriers PN > 200 ≥ 1.5"/PN40: Suitable for stable gases in group 1, category I, module A
- PMP75 with threaded connection PN > 200: Suitable for stable gases in group 1, category I, module A
Cerabar S PMC71, PMP71, PMP75
116 Endress+Hauser
Manufacturer declarations Depending on the desired configuration, the following documents can be ordered additionally with the device:
- FDA conformity
- TSE-free: materials free from animal origin
- Regulation (EC) No. 2023/2006 (GMP)
- Regulation (EC) No. 1935/2004 on materials and articles intended to come into contact with food Downloading the Declaration of Conformity http://www.endress.com/en/download A0027319-EN 1. Select "Approvals & Certificates" 2. Select "Manufact. Declaration" 3. Enter the required product code 4. Click "Search" The available downloads are displayed. Marine approval • GL (German Lloyd)
- ABS Ordering information: Product Configurator, order code for "Additional options 1" or "Additional options 2", version "S". Drinking water approval PMC71/PMP71: NSF 61 approval PMC71/PMP71: UBA/W270 approval (ordering information as for NSF approval: Product Configurator, order code for "Additional options 1" or "Additional options 2" option "F".) Ordering information: Product Configurator, order code for "Additional options 1" or Additional options 2", version "F". Approvals for custody transfer All aspects of OIML R117-1 Edition 2007 (E) and EN 12405-1/A1 Edition 2006 are fulfilled. MID Parts Certificate TC7975 Classification of process sealing between electrical systems and (flammable or combustible) process fluids in accordance with ANSI/ISA 12.27.01 Endress+Hauser devices are designed in accordance with ANSI/ISA 12.27.01. allowing the user to waive the use and save the cost of installing external secondary process seals in the conduit as required by the process sealing sections of ANSI/NFPA 70 (NEC) and CSA 22.1 (CEC). These instruments comply with the North American installation practice and provide a very safe and cost-
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 117 saving installation for pressurized applications with hazardous fluids. Please refer to the following table for the seal class assigned (single seal or dual seal): Device Approval Remark Single seal MWP Dual seal MWP PMC71 CSA C/US IS, XP except separate housing - 60 bar (900 psi) CSA C/US IS With separate housing 40 bar (600 psi) - PMP71 CSA C/US XP, XP+IS except separate housing 400 bar (6 000 psi) - CSA C/US IS except separate housing >200 to 400 bar (3 000 to 6 000 psi) ≤200 bar (3 000 psi) CSA C/US IS With separate housing 400 bar (6 000 psi) - PMP75 XP, XP+IS except separate housing 400 bar (6 000 psi) - CSA C/US IS except separate housing >200 to 400 bar (3 000 to 6 000 psi) ≤200 bar (3 000 psi) CSA C/US IS With separate housing 400 bar (6 000 psi) - Further information can be found in the control drawings of the relevant devices. Inspection certificate Description PMC71 PMP71 PMP75 Option Material test certificate for wetted components, inspection certificate as per EN10204-3.1 acc. to specification 52005759 B 1) 3) Declaration of Conformity NACE MR0175, wetted metal parts ¾ C 1) Material test certificate for wetted components as per EN 10204 3.1 and NACE MR0175 material, inspection certificate as per EN10204-3.1 acc. to specification 52010806 ¾ D 1) 3) Individual testing with test certificate, inspection certificate as per EN10204-3.1 3 1) Overpressure testing with test certificate, inspection certificate as per EN10204-3.1 4 1) Helium leak test EN 1518 with test certificate, inspection certificate as per EN10204-3.1 ¾ 5 1) EN10204-3.1 material wetted parts +Ra, Ra= surface roughness, dimensional check, inspection certificate ¾ ¾ 6 1) EN10204-3.1 measurement of delta ferrite content, inspection certificate ¾ ¾ 8 1) 3.1 Material documentation, wetted metal parts, EN10204-3.1 inspection certificate JA 2) 3) Declaration of Conformity NACE MR0175, wetted metal parts ¾ JB 2) 3) Declaration of Conformity NACE MR0103, wetted metal parts JE 3) PMI test (XRF), internal procedure, metal parts in contact with the medium KG Welding documentation, wetted/pressurized seams ¾ ¾ KS 1) Product Configurator, order code for "Additional options 1" and "Additional options 2" 2) Product Configurator, order code for "Test, Certificate" 3) The choice of this feature for coated process isolating diaphragms/process connections refers to the metal base material. Calibration Description PMC71 PMP71 PMP75 Option 1) Sensor range; mbar/bar 1 Sensor range; kPa/MPa 2 Sensor range; mmH2O/mH2O 3 Sensor range; inH2O/ftH2O 4 Sensor range; psi 6 Factory calibration certificate, 5-point; see additional specification C
Cerabar S PMC71, PMP71, PMP75
118 Endress+Hauser
Description PMC71 PMP71 PMP75 Option 1) DKD/DAkkS certificate; see additional specification D Customized pressure; see additional specification E Customized level; see additional specification F Customized pressure + 5-point factory calibration certificate; see additional specification H Customized level + 5-point factory calibration certificate; see additional specification I Platinum; see additional specification ¾ K Platinum + factory calibration certificate 5-point; see additional specification ¾ L Platinum + DKD/DAkkS certificate; see additional specification ¾ M 1) Product Configurator, order code for "Calibration; unit" Service Description Option 1) Cleaned of oil+grease 2) HA Cleaned for oxygen service 2) HB Cleaned of PWIS (paint-wetting impairment substances) 2) HC 1) Product Configurator, order code for "Service" 2) Device only, not accessories or enclosed accessories.
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 119
Ordering information
Detailed ordering information is available from the following sources:
- In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate" -> Select your country -> Click "Products" -> Select the product using the filters and search field -> Open product page -> The "Configure" button to the right of the product image opens the Product Configurator.
- From your Endress+Hauser Sales Center: www.addresses.endress.com Product Configurator - the tool for individual product configuration
- Up-to-the-minute configuration data
- Depending on the device: Direct input of measuring point-specific information such as measuring range or operating language
- Automatic verification of exclusion criteria
- Automatic creation of the order code and its breakdown in PDF or Excel output format
- Ability to order directly from the Endress+Hauser Online Shop Scope of delivery • Measuring device
- Optional accessories
- Brief Operating Instructions
- Certificates of calibration
- Optional certificates Configuration data sheet Pressure The following configuration data sheet must be completed and included with the order if the option "E" or "H" has been selected in the Product Configurator, order code for "Calibration; Unit". Pressure unit mbar bar psi mmH2O 1) mH2O 1) ftH2O 1) inH2O 1) mmHg 2) inHg 2) gf/cm2 kgf/cm2 Pascal hPa kPa MPa torr g/cm2 kg/cm2 lb/ft2 atm 1) The conversion factor for the pressure unit is based on a reference temperature of 4 °C (39.2 °F). 2) The conversion factor of the pressure unit refers to a reference temperature of 0 °C (32 °F). Calibration range / Output Lower range value (LRV): Upper range value (URV): [pressure unit] [pressure unit] Display Display of the content of the main line (option depends on sensor and communication variant) Primary value [PV] (default) Primary value [%] Pressure Current [mA] (HART only) Temperature Error number Alternating display Damping
Cerabar S PMC71, PMP71, PMP75
120 Endress+Hauser
Smallest span (factory calibration) Level The following configuration data sheet must be completed and included with the order if the option "F" or "I" has been selected in the Product Configurator, order code for "Calibration; Unit". Pressure unit Output unit (scaled unit) Mass Lengths Volume Volume Percent mbar bar psi mmH2O 1) mH2O 1) ftH2O 1) inH2O 1) mmHg 2) inHg 2) gf/cm2 kgf/cm2 Pascal hPa kPa MPa torr g/cm2 kg/cm2 lb/ft2 atm kg t lb m dm cm mm ft inch l hl ft3 USgal impgal USbblPE TR % Example B A A0020477 A 0 mbar/0 m B 300 mbar (4.5 psi) / 3 m (9.8 ft) Empty calibration [a]: Low pressure value (empty) [pressure unit] Empty calibration [a]: Low measured value (empty) [Scaled unit] Full calibration [b]: High pressure value (full) [pressure unit] Full calibration [b]: High level value (full) [Scaled unit] 1) The conversion factor for the pressure unit is based on a reference temperature of 4 °C (39.2 °F). 2) The conversion factor of the pressure unit refers to a reference temperature of 0 °C (32 °F). Display Display of the content of the main line (option depends on sensor and communication variant) Primary value [PV] (default) Primary value [%] Pressure Current [mA] (HART only) Temperature Level before lin. Tank content Error number Alternating display Damping
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 121 Accessories HistoROM®/M-DAT HistoROM®/M-DAT is a memory module which can be attached to every electronic insert. Ordering information: Product Configurator, order code for "Additional options 1" or Additional options 2", version "N" or as a separate accessory (part no.: 52027785). Welding flanges and welding neck For details refer to TI00426F/00/EN "Weld-in adapter and flanges". Manifolds For details see SD01553P/00/EN "Mechanical accessories for pressure measuring devices". Additional mechanical accessories Oval flange adapters, pressure gauge valves, shutoff valves, siphons, condensate pots, cable shortening kits, adapter test, flushing rings, block&bleed valves, protective roofs. For details see SD01553P/00/EN "Mechanical accessories for pressure measuring devices".
Cerabar S PMC71, PMP71, PMP75
122 Endress+Hauser
Supplementary documentation Field of Activities Pressure measurement, powerful instruments for process pressure, differential pressure, level and flow: FA00004P/00/EN Technical Information • Deltabar S: TI00382P/00/EN
- Deltapilot S: TI00416P/00/EN
- EMC test procedures: TI00241F/00/EN
- Weld-in adapters, process adapters and flanges: TI00426F/00/EN Special Documentation Mechanical accessories for pressure measuring devices: SD01553P/00/EN Operating Instructions 4 to 20 mA HART:
- Cerabar S: BA00271P/00/EN
- Description of Device Functions Cerabar S/Deltabar S/Deltapilot S: BA00274P/00/EN 4 to 20 mA HART with MID parts certificate:
- BA00412P/00/EN
- Description of Device Functions: BA00413P/00/EN PROFIBUS PA:
- Cerabar S: BA00295P/00/EN
- Description of Device Functions Cerabar S/Deltabar S/Deltapilot S: BA00296P/00/EN FOUNDATION Fieldbus:
- Cerabar S: BA00302P/00/EN
- Description of Device Functions Cerabar S/Deltabar S/Deltapilot S: BA00303P/00/EN Brief Operating Instructions • 4 to 20 mA HART, Cerabar S: KA01019P/00/EN
- PROFIBUS PA, Cerabar S: KA01022P/00/EN
- FOUNDATION Fieldbus, Cerabar S: KA01025P/00/EN Functional safety manual (SIL) Cerabar S (4 to 20 mA): SD00190P/00/EN Overfill protection WHG: ZE00260P/00/DE Safety Instructions (XA) Depending on the approval, the following Safety Instructions (XA) are supplied with the device. They are an integral part of the Operating Instructions. Directive Device Electronics Documentation Option 1) ATEX II 1/2 G Ex ia IIC T6 PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00244P 1 ATEX II 1/2 D Ex ia PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus
- XA00246P
- XA00289P ATEX II 1/2D Ex ia IIIC PMC71 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus
- XA00247P
- XA00290P ATEX II 1/3D Ex tD PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus
- XA00248P
- XA00291P ATEX II 2G Ex d IIC T6 Gb PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00249P 5 ATEX II 2G Ex d [ia] IIC T6 Gb PMC71 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00250P 5 ATEX II 1/2G Ex ia IIC T6, WHG (German Water Resources Act) PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00244P 6 ATEX II 3 G Ex nA II T6 PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00251P 7
Cerabar S PMC71, PMP71, PMP75 Endress+Hauser 123 Directive Device Electronics Documentation Option 1) ATEX II 1/2G Ex ia + II 1/2D Ex iaD PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00253P 3 ATEX II 1G Ex ia + II 1D Ex iaD PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00276P 8 ATEX II 1/2G Ex ia IIC T6+II 2G Ex d IIC T6 PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00252P B ATEX II Ex ia + FM IS + CSA IS ATEX II 1/2G Ex ia IIC T6 + FM/CSA IS Cl.I,II,III Div. 1 Gr.A-G, FM/CSA: Zone 0,1,2 PMC71 • 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus
- HART
- PROFIBUS PA, FOUNDATION Fieldbus
- XA00244P
- XA00593P + XA01059P
- XA00596P + XA01060P E ATEX II Ex ia / Ex d + FM/CSA IS + XP ATEX II 1/2G Ex ia IIC T6+ ATEX II 2G Ex d IIC T6+ FM/CSA IS + XP Cl.I.II Div.1 Gr.A- G/B-G FM: Zone 1.2/CSA: Zone 1,2 PMP71, PMP75 • 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus
- HART
- PROFIBUS PA, FOUNDATION Fieldbus
- XA00252P
- XA00592P + XA01197P
- XA00590P + XA01198P F 1) Product Configurator, order code for "Approval" Directive Device Electronics Documentation Option 1) IECEx Zone 0/1 Ex ia IIC T6 PMC71, PMP71, PMP75 4 to 20 mA HART XB00005P I IEC Ex d[ia] IIC T6 Gb PMC71 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00511P B IEC Ex d IIC T6 Gb PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00510P M 1) Product Configurator, order code for "Approval" Directive Device Electronics Documentation Option 1) NEPSI Ex ia IIC T6 PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00549P H NEPSI Ex d IIC T6 PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00551P G NEPSI Ex d[ia] IIC T6 PMC71 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA00551P G 1) Product Configurator, order code for "Approval" Directive Device Electronics Documentation Option 1) TIIS Ex d [ia] IIC T6 PMC71 4 to 20 mA HART TC17436 L TIIS Ex d [ia] IIC T4 PMC71 4 to 20 mA HART TC17398, TC17399 M TIIS Ex d IIC T6 PMP71 (700 bar version) 4 to 20 mA HART TC17445 L TIIS Ex d IIC T6 PMP71, PMP75 4 to 20 mA HART TC17446 L 1) Product Configurator, order code for "Approval" Directive Device Electronics Documentation Option 1) INMETRO Ex ia IIC T6 Ga/Gb PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA01315P J INMETRO Ex d IIC T6 Gb PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA01279P O INMETRO Ex ta IIIC Da/Db PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA01313P Z INMETRO Ex d ia IIC T6 Gb PMC71 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA01280P P INMETRO Ex ia IIIC Da/Db PMC71 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA01314P Z 1) Product Configurator, order code for "Approval"
Cerabar S PMC71, PMP71, PMP75 Installation/Control Drawings Directive Device Electronics Documentation Option 1) FM IS Class I, II, III, Division 1, Groups A – G; NI, Class I Division 2, Groups A – D; AEx ia PMC71, PMP71, PMP75 • 4 to 20 mA HART
- PROFIBUS PA, FOUNDATION Fieldbus
- XA01059P
- XA01060P S CSA IS Class I, II, III, Division 1, Groups A – G; Class I Division 2, Groups A – G PMC71, PMP71, PMP75 • 4 to 20 mA HART
- PROFIBUS PA, FOUNDATION Fieldbus
- XA00593P
- XA00596P U FM IS + XP Class I, Division 1, Groups A – D PMP71, PMP75 • 4 to 20 mA HART
- PROFIBUS PA, FOUNDATION Fieldbus
- XA01197P
- XA01198P C CSA IS + XP Class I Division 1, Groups A – D PMP71, PMP75 • 4 to 20 mA HART
- PROFIBUS PA, FOUNDATION Fieldbus
- XA00592P
- XA00590P D FM/CSA IS + XP Class I Division 1, Groups A – D PMP71, PMP75 • 4 to 20 mA HART
- PROFIBUS PA, FOUNDATION Fieldbus
- XA00592P + XA01197P
- XA01198P + XA00590P E FM NI Cl.I Div.2 Groups A - D, Zone 2 PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA01063P R FM XP Cl.I Div.1 Groups A - D, AEx d, Zone 1,2 PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus XA01070P T FM DIP Cl.II,III Div.1 Gr.E-G, Zone 21,22 PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus FM3017778 Q CSA C/US XP Cl.I Div.1 Gr.B-D, Ex d, Zone 1,2 PMC71, PMP71, PMP75 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus ZD00230P + XA00599P V CSA C/US General Purpose PMD75, FMD77, FMD78 4 to 20 mA HART, PROFIBUS PA, FOUNDATION Fieldbus - X 1) Product Configurator, order code for "Approval" www.addresses.endress.com *71331990* 71331990