DATASHEET SEARCH SITE
Document overview
- PDF pages: 2
Technical content
Page 1
P562
Medical Silicon
Pressure Die
Description
NovaSensor’s P562 piezoresistive pressure sensor is specifically designed for medical applications. Low linearity errors and low input and output impedance make the P562 an industry standard for disposable pressure sensor die. The sensor die is radiation tolerant for newer sterilization methods. Wire bonds are located on one side to minimize package width for DPT’s and IUP’s. Choose an open bridge wirebond configuration depending on your compensation circuit . Sensitivity can be easily scaled with external circuitry to meet the AAMI standard of 5 μV/V/mmHg sensitivity.
Applications
- Disposable pressure transducer (DPT)
- Intrauterine pressure (IUP)
- Biomedical instruments
- Infusion pumps
Amphenol
Advanced Sensors
Features
- Highly reliable, silicon pressure sensor
- Low-cost for disposable applications
- 0 to 100 mmHg pressure range
- Radiation tolerant design
- Linearity errors less than 1% in physiological pressure range
Page 2
P562 Specifications
Parameter
Value
Units
Notes
General Pressure Range
100
mmHg
-1.9 psi (-0.13 bar)
Maximum Overpressure
125 (9 bar)
psi
Topside pressure
Electrical @ 72°F (25°C) unless noted Excitation
3.0
mA
3 VDC Maximum
Ω
Input Impedance
305 to 400
Ω
Output Impedance
305 to 400
Environmental Temperature Range Operating
32 to 158
°F
(0°C to 70°C)
Storage
-13 to 185
°F
(-25°C to 85°C)
Mechanical Weight
0.00004
lb
(0.02 grams)
Media Compatibility
Clean dry air, non-corrosive gases
Performance Parameters (1) Zero Offset
±4
mV/V
2
Sensitivity
44 ±11
µV/V/mmHg
Linearity
1.5
%FSO
3
Pressure Hysteresis
0.1
%FSO
Temperature
30
µV/V/°C
4
Coefficient of Zero Temperature
0.16
%/°C
4
Coefficient of Resistance Temperature Coefficient of -0.11
%FSO/°C
4
Sensitivity Thermal Hysteresis of Zero 0.1
%FSO
4
Long Term Stability of FSO 0.1
%FSO
5
- All values measured at 72°F (25°C) and 1 mA excitation, unless otherwise noted. 2. Zero psig. 3. Measured as the deviation at 300 mmHg from a line drawn from the output at zero pressure through the output at 100 mmHg. 4. Between 32°F and 158°F (0°C and 70°C). 5. Typical value over one year.
Shipping and Handling
All wafers are shipped in protective containers. The wafers are sawn on sticky tape with plastic rings. All wafers are electrically probed and visually inspected. Samples from each wafer verify offset , full-scale output , and linearity. Electrical rejects are inked with red dots. Visual rejects are inked with blue dots. Each wafer will have the following information: Lot number, wafer number, device number, and the number of good dice.
Warranty
NovaSensor warrants its products against defects in material and workmanship for 12 months from the date of shipment . Products not subjected to misuse will be repaired or replaced. NovaSensor reserves the right to make changes without further notice to any products herein. NovaSensor makes no warranty, representation or guarantee regarding the suitability of its products for any particular application. Nova- Sensor does not assume any liability arising out of the application or use of any product or circuit and specifically disclaims, and all liability, without limitation consequential or incidental damages. The foregoing warranties are exclusive and in lieu of all other war-
ranties, whether written, oral, implied or statutory. No implied statutory warranty of merchantability or fitness for a particular purpose shall apply.

Die dimensions: 0.08 in x 0.08 in x 0.01 in (2.1 mm x 2.1 mm x 0.4 mm)

Ordering Information
Part Number:
51279 100mmHg Sensor
Minimum release quantity: Two wafers or approximately 2,000 die.
Amphenol
P562 schematic and wirebond diagram
www.amphenol-sensors.com
© 2014 Amphenol Corporation. All Rights Reserved. Specifications are subject to change without notice. Other company names and product names used in this document are the registered trademarks or
Advanced Sensors
trademarks of their respective owners.