MBC201 BEL
Document overview
- PDF pages: 7
Technical content
Page 1

The MBC201 Series of open-frame medical power supplies, with its wide universal input range 90-264 VAC, is available at 200 W of output power with forced-air cooling and a variety of single output voltages.
The MBC Series is designed and approved to the latest medical standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I & Class II applications.

These power supplies are ideal for medical, telecom, datacom, industrial equipment and other applications.
5 x 3 x 1.5 Inch Form Factor
200 W with Forced-Air Cooling
Approved to EN / IEC 60601-1
High Efficiency > 88%
12 V Fan Output
5 V Standby Output
- Remote Sense
- IEC Protection Class Options: − Class I: Earthing Tab J4 (no suffix) − Class II: No Earthing Tab (-2 suffix)
- Medical Safety Agency Approvals
- CE Marked
- Diagnostic
- Dialysis
- Drug Pump
- Home Health Care
- Monitoring
- Portable Equipment

Page 2
2

MBC201-1012G
JST (Header)
0.0 A
12 VDC
200 W
13.33 A
16.67 A
MBC201-1T12G
Screw Terminal
MBC201-1015G
JST (Header)
0.0 A
15 VDC
200 W
10.67 A
13.33 A
MBC201-1T15G
Screw Terminal
MBC201-1024G
JST (Header)
0.0 A
24 VDC
200 W
6.67 A
8.33 A
MBC201-1T24G
Screw Terminal
MBC201-1030G
JST (Header)
0.0 A
30 VDC
200 W
5.33 A
6.67 A
MBC201-1T30G
Screw Terminal
MBC201-1048G
JST (Header)
0.0 A
48 VDC
200 W
3.33 A
4.17 A
MBC201-1T48G
Screw Terminal
COVER-201-XBC 3 Metal Cover Kit (accessory)
1 Class I products have an Earthing tab. For Class II version (without Earthing tab) add suffix -2 (e.g.: MBC201-1012G-2). 2 Combined output power from V1, VSTBY and VFAN should not exceed the total output power rating. 3 When used in Cover Kit, de-rate output power to 70 % under all operating conditions.
Specifications are for nominal input voltage, 25°C unless otherwise stated.


tech.support@psbel.com
Page 3
PARAMETER
DESCRIPTION / CONDITION
Output Voltage
Refer to Model Selection table
Output Power 4
Peak Power: 250 W for 2 s
Fan Output 5
@ 5 A 120 VAC:
Efficiency
230 VAC:
Hold Up Time
120 VAC / 230 VAC: 120 VAC:
Power Factor
230 VAC:
Line Regulation
Load Regulation
Transient Response
50 to 100% load change, 50 Hz, 50% duty cycle, 0.1A / µs
Ripple 6
All outputs
Rise Time
Set Point Accuracy
Main output
Output Voltage Adjustment
V1
Over Current Protection
Over Voltage Protection
V1
Short Circuit Protection
Short term, Automatic recovery
Convection
Cooling
300 LFM
4 Derate output power linearly to 80% from 90 VAC to 80 VAC input. 5
3
SPECIFICATION
5 V to 48 V
160 to 200 W
12 V 84% typical 86% typical 10 ms 0.99 0.95 ± 0.5%
± 2.0% < 10%, recovery time < 5 ms 1.0 % max
< 100 ms
± 1%
± 3 %
110% typical above rating
110 to 150%
83 W (5 V model), 160 W (other models)
175 W (5 V model), 200 W (other models)
Fan output voltage tolerance is +/-20%. During V1 full load, VFAN needs min. 20 mA load to be within regulation band. Peak current for fan output is 1 A. 6 Ripple is 2% up to 20% load and < 1% above 20% load. Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Electrolytic capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges.
Conducted Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
Radiated Emissions
EN 55011 B
Input Current Harmonics
EN 61000-3-2
Voltage Fluctuation and Flicker
EN 61000-3-3
ESD Immunity
EN 61000-4-2
Radiated Field Immunity
EN 61000-4-3
Electrical Fast Transient Immunity
EN 61000-4-4
Surge Immunity
EN 61000-4-5
Conducted Immunity
EN 61000-4-6
Magnetic Field Immunity
EN 61000-4-8
Voltage Dips, Interruptions
EN 61000-4-11

Pass
Pass
Class D
Pass
Level 4, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 4, Criterion A
Criterion A & B
Asia-Pacific Europe, Middle East
North America +1 408 785 5200
+86 755 298 85888
+353 61 225 977
© 2021 Bel Power Solutions & Protection
BCD.00289_AG
Page 4
4

Agency Approvals
Nemko, UL, C-UL
CE mark
Complies with LVD Directive
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
-20 to 70°C
Operating Temperature 7
-20 to 0°C start-up is guaranteed
-20 to 0°C
Storage Temperature
-40 to 85° C

7 Refer to derating curves


Figure 1. Derating Curves

Convection load: 83 W up to 40 °C De-rate above 40 °C @ 1.89% per °C
Forced air cooled load: 130 W up to 50°C (for screw terminal version) De-rate above 50 °C @ 1.63% per °C
Forced air cooled load: 175 W up to 50°C (for JST connector version) De-rate above 50 °C @ 2.5% per °C
Convection load: 160 W up to 40 °C De-rate above 40 °C @ 2% per °C
Forced air cooled load: 200 W up to 50°C De-rate above 50 °C @ 2.5% per °C
tech.support@psbel.com
Page 5
5
PARAMETER
DESCRIPTION / CONDITION
TTL signal goes high after main output is in regulation band. Delay is 0.1 to 0.3 sec.
Power Good Signal
TTL signal goes low 1 ms advance before output goes out of regulation due to mains failure
Power Fail Signal
Remote Sense
Compensates for 200 mV cable drop
Remote ON / OFF
To turn on PSU short remote pin to ground

8 PSU is supplied with J3 housing, pin-4 and pin-6 shorted to enable main output without remote on/off feature 9 Class I product only
PARAMETER
DESCRIPTION / CONDITION
Weight
325 g (0.72 lbs.)
Dimensions
127.0 x 76.2 x 38.1 mm (5 x 3 x 1.5 in)

North America
Asia-Pacific Europe, Middle East
+1 408 785 5200
+86 755 298 85888
+353 61 225 977
© 2021 Bel Power Solutions & Protection
BCD.00289_AG
Page 6
6

Figure 2. Mechanical Drawing
NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following: 1 Stand off, used to mount PCB has OD of 5.4 mm max. 2 Screws, used to fix PCB on stand off, have head dia of 6.0 mm max. 3 Washer, if used, to have dia of 6.5 mm max.

tech.support@psbel.com
Page 7
7

Figure 3. Mechanical Drawing with Cover Kit
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice.

North America
Asia-Pacific Europe, Middle East
+1 408 785 5200
+86 755 298 85888
+353 61 225 977
© 2021 Bel Power Solutions & Protection