MBC201 BEL

Document overview

  • PDF pages: 7

Technical content

Page 1

PDF page 1 diagram

The MBC201 Series of open-frame medical power supplies, with its wide universal input range 90-264 VAC, is available at 200 W of output power with forced-air cooling and a variety of single output voltages.

The MBC Series is designed and approved to the latest medical standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I & Class II applications.

PDF page 1 diagram

These power supplies are ideal for medical, telecom, datacom, industrial equipment and other applications.

5 x 3 x 1.5 Inch Form Factor

200 W with Forced-Air Cooling

Approved to EN / IEC 60601-1

High Efficiency > 88%

12 V Fan Output

5 V Standby Output

  • Remote Sense
  • IEC Protection Class Options: − Class I: Earthing Tab J4 (no suffix) − Class II: No Earthing Tab (-2 suffix)
  • Medical Safety Agency Approvals
  • CE Marked
  • Diagnostic
  • Dialysis
  • Drug Pump
  • Home Health Care
  • Monitoring
  • Portable Equipment
PDF page 1 diagram

Page 2

2

PDF page 2 diagram

MBC201-1012G

JST (Header)

0.0 A

12 VDC

200 W

13.33 A

16.67 A

MBC201-1T12G

Screw Terminal

MBC201-1015G

JST (Header)

0.0 A

15 VDC

200 W

10.67 A

13.33 A

MBC201-1T15G

Screw Terminal

MBC201-1024G

JST (Header)

0.0 A

24 VDC

200 W

6.67 A

8.33 A

MBC201-1T24G

Screw Terminal

MBC201-1030G

JST (Header)

0.0 A

30 VDC

200 W

5.33 A

6.67 A

MBC201-1T30G

Screw Terminal

MBC201-1048G

JST (Header)

0.0 A

48 VDC

200 W

3.33 A

4.17 A

MBC201-1T48G

Screw Terminal

COVER-201-XBC 3 Metal Cover Kit (accessory)

1 Class I products have an Earthing tab. For Class II version (without Earthing tab) add suffix -2 (e.g.: MBC201-1012G-2). 2 Combined output power from V1, VSTBY and VFAN should not exceed the total output power rating. 3 When used in Cover Kit, de-rate output power to 70 % under all operating conditions.

Specifications are for nominal input voltage, 25°C unless otherwise stated.

PDF page 2 diagram
PDF page 2 diagram

tech.support@psbel.com

Page 3

PARAMETER

DESCRIPTION / CONDITION

Output Voltage

Refer to Model Selection table

Output Power 4

Peak Power: 250 W for 2 s

Fan Output 5

@ 5 A 120 VAC:

Efficiency

230 VAC:

Hold Up Time

120 VAC / 230 VAC: 120 VAC:

Power Factor

230 VAC:

Line Regulation

Load Regulation

Transient Response

50 to 100% load change, 50 Hz, 50% duty cycle, 0.1A / µs

Ripple 6

All outputs

Rise Time

Set Point Accuracy

Main output

Output Voltage Adjustment

V1

Over Current Protection

Over Voltage Protection

V1

Short Circuit Protection

Short term, Automatic recovery

Convection

Cooling

300 LFM

4 Derate output power linearly to 80% from 90 VAC to 80 VAC input. 5

3

SPECIFICATION

5 V to 48 V

160 to 200 W

12 V 84% typical 86% typical 10 ms 0.99 0.95 ± 0.5%

± 2.0% < 10%, recovery time < 5 ms 1.0 % max

< 100 ms

± 1%

± 3 %

110% typical above rating

110 to 150%

83 W (5 V model), 160 W (other models)

175 W (5 V model), 200 W (other models)

Fan output voltage tolerance is +/-20%. During V1 full load, VFAN needs min. 20 mA load to be within regulation band. Peak current for fan output is 1 A. 6 Ripple is 2% up to 20% load and < 1% above 20% load. Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Electrolytic capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges.

Conducted Emissions

EN 55011-B, CISPR22-B, FCC PART15-B

Radiated Emissions

EN 55011 B

Input Current Harmonics

EN 61000-3-2

Voltage Fluctuation and Flicker

EN 61000-3-3

ESD Immunity

EN 61000-4-2

Radiated Field Immunity

EN 61000-4-3

Electrical Fast Transient Immunity

EN 61000-4-4

Surge Immunity

EN 61000-4-5

Conducted Immunity

EN 61000-4-6

Magnetic Field Immunity

EN 61000-4-8

Voltage Dips, Interruptions

EN 61000-4-11

PDF page 3 diagram

Pass

Pass

Class D

Pass

Level 4, Criterion A

Level 3, Criterion A

Level 3, Criterion A

Level 3, Criterion A

Level 3, Criterion A

Level 4, Criterion A

Criterion A & B

Asia-Pacific Europe, Middle East

North America +1 408 785 5200

+86 755 298 85888

+353 61 225 977

© 2021 Bel Power Solutions & Protection

BCD.00289_AG

Page 4

4

PDF page 4 diagram

Agency Approvals

Nemko, UL, C-UL

CE mark

Complies with LVD Directive

PARAMETER

DESCRIPTION / CONDITION

SPECIFICATION

-20 to 70°C

Operating Temperature 7

-20 to 0°C start-up is guaranteed

-20 to 0°C

Storage Temperature

-40 to 85° C

PDF page 4 diagram

7 Refer to derating curves

PDF page 4 diagram
PDF page 4 diagram

Figure 1. Derating Curves

PDF page 4 diagram

Convection load: 83 W up to 40 °C De-rate above 40 °C @ 1.89% per °C

Forced air cooled load: 130 W up to 50°C (for screw terminal version) De-rate above 50 °C @ 1.63% per °C

Forced air cooled load: 175 W up to 50°C (for JST connector version) De-rate above 50 °C @ 2.5% per °C

Convection load: 160 W up to 40 °C De-rate above 40 °C @ 2% per °C

Forced air cooled load: 200 W up to 50°C De-rate above 50 °C @ 2.5% per °C

tech.support@psbel.com

Page 5

5

PARAMETER

DESCRIPTION / CONDITION

TTL signal goes high after main output is in regulation band. Delay is 0.1 to 0.3 sec.

Power Good Signal

TTL signal goes low 1 ms advance before output goes out of regulation due to mains failure

Power Fail Signal

Remote Sense

Compensates for 200 mV cable drop

Remote ON / OFF

To turn on PSU short remote pin to ground

PDF page 5 diagram

8 PSU is supplied with J3 housing, pin-4 and pin-6 shorted to enable main output without remote on/off feature 9 Class I product only

PARAMETER

DESCRIPTION / CONDITION

Weight

325 g (0.72 lbs.)

Dimensions

127.0 x 76.2 x 38.1 mm (5 x 3 x 1.5 in)

PDF page 5 diagram

North America

Asia-Pacific Europe, Middle East

+1 408 785 5200

+86 755 298 85888

+353 61 225 977

© 2021 Bel Power Solutions & Protection

BCD.00289_AG

Page 6

6

PDF page 6 diagram

Figure 2. Mechanical Drawing

NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following: 1 Stand off, used to mount PCB has OD of 5.4 mm max. 2 Screws, used to fix PCB on stand off, have head dia of 6.0 mm max. 3 Washer, if used, to have dia of 6.5 mm max.

PDF page 6 diagram

tech.support@psbel.com

Page 7

7

PDF page 7 diagram

Figure 3. Mechanical Drawing with Cover Kit

NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice.

PDF page 7 diagram

North America

Asia-Pacific Europe, Middle East

+1 408 785 5200

+86 755 298 85888

+353 61 225 977

© 2021 Bel Power Solutions & Protection

BCD.00289_AG