L8601-01 HAMAMATSU
Document overview
- PDF pages: 4
Technical content
Page 1
90 kV

MICROFOCUS X-RAY SOURCE
L8601-01

FEATURES
G Focal Spot Size: 5 µm (at 4 W)
Providing 5µm focal spot size, for the sealed-off type X-ray source with built-in power supply. It gives a clear clairvoyance image despite of higher magnification ratio.
G Easy Handling
Convenient handling (with no high voltage wiring) since the air cooled X-ray source and the high voltage power supply are in one package.
G Continuous Operation
It allows continuous maximum output operation.
G Built-in External Control Function
It is automatically operable from external control since it has a built-in RS-232C interface.
APPLICATIONS
G X-ray Non-destructive Test
- Printed Circuits
- Electronic Components
- Semiconductor Devices
- Ceramics
- Soldered Parts
- Rubbers
- Combinated Materials
- Plastics
- New materials
- Die casting
G Multilayer Printed Circuit Board
X-ray Inspection•Drilling Instrument
G Industrial On-line Process Control
and Measurement
G Industrial X-ray Computer
Tomography System
Page 2
SPECIFICATIONS
GGENERAL
| Parameter | Value/Description | Unit |
|---|---|---|
| Target Voltage | 20 to 90 | kV |
| Target Current | 0 to 250a | µA |
| Maximum Output Power | 10 (isowatts) | W |
| Focal Spot Size | 7 (5 µm at 4 W) | µm |
| Beam Angle | 39 | degree |
| Minimum Distance Focus/Object | 9.5 | mm |
| Environmental X-ray Dose Leakage | 1 | mSv/h Max. |
| Operation | Continuous | — |
NOTE: a In the range of output (= Tube voltage × Tube current) less than 10 W
IX-RAY TUBE UNIT
IX-RAY CONTROL UNIT
| Parameter | Description |
|---|---|
| X-ray Tube | Sealed Type |
| Cooling Method | Convection Cooling |
| Window Material | Beryllium (150 µmt) |
| Target Material | Tungsten |
| Cathode Material | Dispenser Cathode |
| Window Position | End-window |
| High Voltage Power Supply | Built-in Type |
| Operation /Storage Temperature | +10 to +40 °C / 0 to +50 °C |
| Operation/Storage Humidity | 85 % RH Max.a |
| Weight | Approx. 6 kg |
| Parameter | Description |
|---|---|
| Function | Target Voltage/Current Pre-set, Auto Warm-up |
| Protection | Safety Interlock |
| External Control | RS-232C |
| Operation /Storage Temperature | 0 to +40 °C |
| Operation/Storage Humidity | 85 % RH Max.a |
| Power Consumption | 200 W Max. |
| Input Voltage | 24 Vdc |
| Installation Method | Designed for both bench-top and rack mounting |
| Weight | Approx. 5 kg |
NOTE: a Without moisture condensation.
Page 3
90 KV MICROFOCUS X-RAY SOURCE
DIMENSIONAL OUTLINE (Unit: mm)
X-RAY TUBE UNIT

TLSOA0063EA
X-RAY CONTROL UNIT

X - R AY 9 0 C O N T R O L U N I T
VOLTAGE CURRENT
X-RAY MODE
kV µA

482 ± 2
| X-RAY REMOTE AGING SIGNAL INTER LOCK 1 INTER LOCK 2 OUTPUT RS-232C | |
| I D E N T I F I C AT I O N P L AT E |
LABEL
TLSOA0076EA
Page 4
90 KV MICROFOCUS X-RAY SOURCE
RELATED PRODUCTS

X-RAY IMAGE INTENSIFIER CAMERA UNIT
C7716, C7716-10
The C7716 series X-ray image intensifier (I. I.) camera unit now offers greatly improved X-ray detection efficiency even at low energy levels. This improvement stems primarily from a built-in X-ray image intensifier having an extremely thin aluminum input window. Its thickness is only 0.3mm or less, virtually at the limit of present technology providing excellent X-ray transmit-tance and low scattering. The results are sharp, clear, high-quality images taken at low energy X-ray levels down to several keV which penetrate plastic (PET) materials.

X-RAY IMAGE INTENSIFIER CAMERA UNIT
C7876, C7876-10
The C7876 is an X-ray image intensifier camera unit ideal for non-destructive inspection of light-element materials and radiation imaging at low-energy X-ray levels. The C7876 remarkably improves X-ray transmittance at low-energy X-ray levels by using a beryllium window instead of the aluminum window currently used for most X-ray image intensifiers. The result is a sharp and clear, high-contrast image taken in real time even at low energy X-ray levels down to a few keV.


\* This product(s) conforms to the EMC directive (89/336/EEC) of the European Union. \* To meet FDA standards, the Interlock-1 circuit incorporates a relay that shuts off the X-ray generation power supply circuit if the circuit is open.
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2001 Hamamatsu Photonics K.K
HOMEPAGE URL http://www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Electron Tube Center 314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: info@hamamatsu.co.uk North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it
TLSO1045E01 MAY. 2001 IP