L7901-01 HAMAMATSU

Document overview

  • PDF pages: 4

Technical content

Page 1

100 kV

PDF page 1 diagram

MICROFOCUS X-RAY SOURCE

L7901-01

PDF page 1 diagram

FEATURES

G Focal Spot Size: 5 µm (at 4 W)

Providing 5µm focal spot size, for the sealed-off type X-ray source with built-in power supply. It gives a clear clairvoyance image despite of higher magnification ratio.

G Easy Handling

Convenient handling (with no high voltage wiring) since the air cooled X-ray source and the high voltage power supply are in one package.

G Continuous Operation

It allows continuous maximum output, 100kV and 100µA, operation.

G Built-in External Control Function

It is automatically operable from external control since it has a built-in RS-232C interface.

APPLICATIONS

G X-ray Non-destructive Test

  • Printed Circuits
  • Electronic Components
  • Semiconductor Devices
  • Ceramics
  • Soldered Parts
  • Rubbers
  • Combinated Materials
  • Plastics
  • New materials
  • Die casting

G Multilayer Printed Circuit Board

X-ray Inspection•Drilling Instrument

G Industrial On-line Process Control

and Measurement

G Industrial X-ray Computer

Tomography System

Page 2

SPECIFICATIONS

GGENERAL

Parameter Value/Description Unit
Target Voltage 20 to 100 kV
Target Current 0 to 250a µA
Maximum Output Power 10 (isowatts) W
Focal Spot Size 7 (5 µm at 4 W) µm
Beam Angle 39 degree
Minimum Distance Focus/Object 9.5 mm
Environmental X-ray Dose Leakage 1 mSv/h Max.
Operation Continuous —

NOTE: a In the range of output (= Tube voltage × Tube current) less than 10 W

IX-RAY TUBE UNIT

IX-RAY CONTROL UNIT

Parameter Description
X-ray Tube Sealed Type
Cooling Method Forced Air Cooling
Window Material Beryllium (150 µmt)
Target Material Tungsten
Cathode Material Dispenser Cathode
Window Position End-window
High Voltage Power Supply Built-in Type
Operation /Storage Temperature +10 to +40 °C / 0 to +50 °C
Operation/Storage Humidity 85 % RH Max.a
Weight Approx. 6 kg
Parameter Description
Function Target Voltage/Current Pre-set, Auto Warm-up
Protection Safety Interlock
External Control RS-232C
Operation /Storage Temperature 0 to +40 °C
Operation/Storage Humidity 85 % RH Max.a
Power Consumption 200 W Max.
Inter Voltage 24 Vdc
Installation Method Designed for both bench-top and rack mounting
Weight Approx. 5 kg

NOTE: a Without moisture condensation.

Page 3

100 KV MICROFOCUS X-RAY SOURCE

DIMENSIONAL OUTLINE (Unit: mm)

X-RAY TUBE UNIT

PDF page 3 diagram

TLSOA0063EA

X-RAY CONTROL UNIT

PDF page 3 diagram

X - R AY 1 0 0 C O N T R O L U N I T

VOLTAGE CURRENT

X-RAY MODE

kV µA

PDF page 3 diagram

482 ± 2

X-RAY REMOTE AGING SIGNAL INTER LOCK 1 INTER LOCK 2 OUTPUT RS-232C
I D E N T I F I C AT I O N P L AT E

LABEL

TLSOA0064EA

Page 4

100 KV MICROFOCUS X-RAY SOURCE

RELATED PRODUCTS

PDF page 4 diagram

X-RAY IMAGE INTENSIFIER CAMERA UNIT

C7716, C7716-10

The C7716 series X-ray image intensifier (I. I.) camera unit now offers greatly improved X-ray detection efficiency even at low energy levels. This improvement stems primarily from a built-in X-ray image intensifier having an extremely thin aluminum input window. Its thickness is only 0.3mm or less, virtually at the limit of present technology providing excellent X-ray transmit-tance and low scattering. The results are sharp, clear, high-quality images taken at low energy X-ray levels down to several keV which penetrate plastic (PET) materials.

PDF page 4 diagram

X-RAY IMAGE INTENSIFIER CAMERA UNIT

C7876, C7876-10

The C7876 is an X-ray image intensifier camera unit ideal for non-destructive inspection of light-element materials and radiation imaging at low-energy X-ray levels. The C7876 remarkably improves X-ray transmittance at low-energy X-ray levels by using a beryllium window instead of the aluminum window currently used for most X-ray image intensifiers. The result is a sharp and clear, high-contrast image taken in real time even at low energy X-ray levels down to a few keV.

PDF page 4 diagram
PDF page 4 diagram

\* This product(s) conforms to the EMC directive (89/336/EEC) of the European Union. \* To meet FDA standards, the Interlock-1 circuit incorporates a relay that shuts off the X-ray generation power supply circuit if the circuit is open.

Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2000 Hamamatsu Photonics K.K

HOMEPAGE URL http://www.hamamatsu.com

HAMAMATSU PHOTONICS K.K., Electron Tube Center 314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: infos@hamamatsu.fr

United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: info@hamamatsu.co.uk North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se

TLSO1038E03

Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it

DEC. 2001 IP Printed in Japan (1000)