L1314 HAMAMATSU
Document overview
- PDF pages: 2
Technical content
Page 1

High Intensity, 150 W Type,
Vacuum UV (from 115 nm) Light Source - L1835
The L1314 and L1835 are water-cooled 150 W deuterium lamps that deliver a radiant intensity 3 to 4 times higher than standard 30 W deuterium lamps. The lamp bulb is enclosed in a cylindrical metal jacket specially designed for water cooling. The L1314 has a synthetic silica window for an efficient emission of UV radiation, and the L1835 em- ploys a MgF2 (magnesium fluoride) window which even allows emitting vacuum UV radiation. Select the lamp that matches wavelengths required by your application. Vacuum flanges are also available as op- tions in the L1835 for easy mounting to a vacuum chamber.
APPLICATIONS
GSpectrophotometer, Fluorescence Spectrophotometer GRemoval of static electricity from the semiconductor wafer GPID (Photo Ionization Detector) GSolar Simulator GOptical CVD GOptical Chemical Reaction GExcitation Light Source
SPECIFICATIONS
GENERAL
| Parameter | L1314 | L1835 | Unit |
|---|---|---|---|
| Spectral Distribution | 160 to 400 | 115 to 400 | nm |
| Window Material | Synthetic silica | MgF2 | — |
| Aperture Size | 2.5 | mm dia. | |
| Cooling Method | Water cooling A | — | |
| Weight (Approx.) | 720 | 950 | g |
| Installation to Vacuum Port | No | Yes | — |
RECOMMENDED OPERATING CONDITION / CHARACTERISTICS (at 25 °C)
| Parameter | Value | Unit | |
|---|---|---|---|
| Warm-up | Heater Voltage (DC, AC) | 10 ± 1 | V |
| Heater Current (DC, AC) | 1.2 | A Typ. | |
| Warm-up Time | 20 | s | |
| Operation | Heater Voltage (DC) | 5 | V |
| Heater Current (DC) | 1 | A Typ. | |
| Discharge Starting Voltage (DC) | 500 | V Min. | |
| Anode Current | 1.2 | A | |
| Tube Drop Voltage (DC) | 120 ± 10 | V | |
| Output Stability | Drift (Max.) | ±1.0 | %/h |
| Fluctuation (Max.) | 0.5 | % p-p | |
| Water Flow Rate | 1.5 | L/min | |
| Estimated Life | 300 | h | |
NOTE: ATypes L1314 and L1835 cannot be operated without cooling water passing Flow of cooling water through them. Care should be taken that

the lamp is positioned properly so that the two nozzles are aligned vertically, with the water flowing into the jacket at the bottom nozzle and leaving the jacket at the top nozzle. If this arrangement is not observed, there is a danger of
TLSOC0016EC
damaging the lamp due to overheating.

Figure 1: Spectral Distribution L1314: UV range

WAVELENGTH (nm)
L1835: Vacuum UV range

WAVELENGTH (nm)
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2006 Hamamatsu Photonics K.K.
Page 2
WATER-COOLED TYPE DEUTERIUM LAMPS L1314, L1835
Figure 2: Dimensional Outline

\* L1314 cannot be used with the vacuum chamber. The main body might be damaged when using it by mistake.
OPTION (sold separately)
G POWER SUPPLY C3150
The C3150 power supply is specifically designed for water-cooled 150

CATHODE
IC
1 4 2 3 ANODE HEATER
PIN CONNECTION ( Connector : Miyama Elec . MC-032 )
TLSOA0028EC
Filament (heater) output

| Parameter | Description/Value | |
|---|---|---|
| Output Current (DC) | 1.2 A | |
| Output Voltage | Normal Operation (DC) | 120 V ± 25 V |
| No Load (DC) | 250 V Typ. | |
| Trigger Voltage | 600 V peek | |
| Output Fluctuation | Input Fluctuation (±10 %) | ±0.1 % Max. |
| Load Fluctuation (normal operation range | ) ±0.1 % Max. | |
| Drift | ±0.1 %/h Max. | |
| Ripple | 0.1 % p-p Max. | |
| Over-load Protection | 1.5 A Fuse | |
G VACUUM FLANGE E3444 SERIES (for L1835)
Since the L1835 vacuum UV deuterium lamp is often used while installed onto a vacuum chamber, Hamamatsu provides the E3444 series vacuum flanges specially designed for this purpose. Among these, the E3444-02 has a flange conforming to ICF114 specifications, allowing easy installation onto most vacuum chamber ports. The E3444 series also includes a general-purpose "N" flange and JIS (Japanese Industrial Standards) specification flange. Select the desired type depending on the vacuum equipment to be used.
SPECIFICATIONS
| Type No. | Sealing Method | Flange | Mount Flange | Sealing Force Retention |
|---|---|---|---|---|
| E3444 E3444-01 E3444-02 | O-Ring | Regular JIS VF50 | — JIS VG50 | 1.33 × 10-4 Pa L/s or less (1 × 10-6 Torr L/s) |
| ICF114 | ICF114 |
| Parameter | Description/Value |
|---|---|
| Output Voltage for Warm-up (DC) | 10 V, 1.2 A |
| Warm-up Time | Approx. 30 s |
| Output Voltage for Operation (DC) | 5 V ± 0.5 V, 1.0 A |
| Input Fluctuation (±10 %) | ±1.0 % or less |
| Over-load Protection | 2 A slow-blow or by electronic circuit |
| Input Source Voltage (AC) | 100 V / 118 V / 230 V (±10 %) |
| Apparent Power | Approx. 330 VA |
| Operating Ambient temperature | 0 °C to 40 °C |
| Performance Guaranteed Temperature | +5 °C to +35 °C |
| Cooling Method | Forced air cooling |
| Cooling Water Detection Method | Flow rate switch (1.5 L/min) |
| Protective Functions (LED on when malfunction) | Less cooling water, excessive temperature in the C3150, short circuit, disconnection |
| External Dimensions (mm) | 215(W) × 125(H) × 265(D) |
| Weight | Approx. 10 kg |
Dimensional Outline E3444-02 (Unit: mm)


TLSOA0053EA
Attachment Reference
It is designed to be simply inserted over the lamp housing.
VACUUM FLANGE
TLSOC0017EB
WEB SITE www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Electron Tube Division 314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: info@hamamatsu.co.uk North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se
TLSO1012E03
Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it
JUL. 2006 IP