F6492 HAMAMATSU
Document overview
- PDF pages: 2
Technical content
Page 1
LONG SCALE
MICROCHANNEL PLATE
F6492
Very Long (Effective Area: 127 4mm)
This new long-scale MCP allows for continuous detection of wide range position information without dead space correction. This new configuration promises reliable measurement, and is well suited for double-focusing mass spectrometers. Assembled configurations are also available upon request.
FEATURES
Long Scale (Effective Area: 127mm 4mm) Simultaneous Multielement Measurement Possible Sensitive to Ions, Electrons, UV Radiation, Soft X-rays, -rays and High Energy Particles High Gain (1 104)
APPLICATIONS
SIMS (Secondary Ions Mass Spectrometers) UV Spectrometers X-ray Spectrometers Electron Spectrometers
GENERAL
| Parameter | Description/Value | Unit |
|---|---|---|
| Outer Dimension | 140 9 | mm |
| Electrode Dimension | 138 8 | mm |
| Effective Area Dimension | 127 4 | mm |
| Thickness | 0.48 | mm |
| Channel Diameter | 12 | m |
| Channel Pitch | 15 | m |
| Bias Angle | 8 | |
| Open Area Ratio | 60 | % |
| Electrode Material | Inconel | – |
CHARACTERISTICS (at 1000V, 1.3 10-4 Pa (1 10-6 Torr), 25 )
| Parameter | Discription/Value | Unit |
|---|---|---|
| Current Gain Min. | 1 104 | – |
| Plate Resistance | 10 to 100 | M |
| Maximum Dark Current | 5 10-13 | A/cm2 |
| Maximum Linear Output Current | up to 7% of the strip current | – |
MAXIMUM RATINGS (Absolute Values)
| Parameter | Discription/Value | Unit |
|---|---|---|
| Supply Voltage | 1000 | V |
| Ambient Temperature | -50 to +50 |
NOTE:
: Strip current is current flowing through the channel walls, which supplies the current released from the channel walls. It is given by: Supply voltage / Plate resistance. : At a vacuum of 1.3 10-4 Pa (1 10-6 Torr) or better.

Figure 1: Typical Current Gain
TMCPB0036EA

SUPPLY VOLTAGE (V)
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. lnformation furnished by HAMAMATSU is believed to be reliabIe. However, no responsibility is assumed for possibIe inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. © 1996 Hamamatsu Photonics K.K.
Page 2
LONG SCALE MICROCHANNEL PLATE F6492
Figure 2: Dimensional Outline (Unit: mm)
EFFECTIVE AREA
ELECTRODE
INDICATOR

Figure 3: Application Example to SIMS1)
SECONDARY ION
PRIMARY ION SIGNAL
MASS ANALYZER
IRRADIATION DISPLAY
SYSTEM
SYSTEM SYSTEM
SECTOR ELECTRIC FIELD

+ LINEAR IMAGE SENSOR
TMCPC0038EA
REFERENCE
1) Japan Academic Promotion Society, 141th Committee on Micro Beam Analysis: “Micro beam analysis” Asakura Shoten, 293 (1985)
HAMAMATSU PHOTONICS K.K., Electoron Tube Center
PRECAUTIONS FOR USE
Avoid touching the MCP or MCP assembly with bare hands.
Handle the MCP only in a clean room since dust and humidity may adversely affect MCP characteristics.
The MCP should be operated in vacuum below 1.33 10-4 Pa (1 10-6 torr).
The MCP should be kept in vacuum or dry nitrogen gas atmosphere during long periods of storage.
When outgassing from the MCP occurs, baking the MCP at 350 C maximum in a vacuum system is recommended. In addition electron bombarding may be effective.
314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 United Kingdom: Hamamatsu Photonics UK Limted: Lough Point, 2 Gladbeck Way, Windmill Hill, Enfield, Middlesex EN2 7JA, United Kingdom, Telephone: (44)181-367-3560, Fax: (44)181-367-6384
North Europe: Hamamatsu Photonics Norden AB: Färögatan 7, S-164-40 Kista Sweden, Telephone: (46)8-703-29-50, Fax: (46)8-750-58-95
TMCP1009E05
Italy: Hamamatsu Photonics Italia: S.R.L.: Via Della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)2-935 81 733, Fax: (39)2-935 81 741
JAN. 1997 SI (9609) Printed in Japan