PX409_V03 ETC

Document overview

  • Manufacturer or author: Provided By www.digicamel.com(free datasheet download site)
  • PDF pages: 2

Technical content

FoR BARomETRIC PRESSURE mEASUREmENT B-133 SpecificationS output: Millivolt: 10 mV/V (100 mV @ 10 Vdc) amplified Voltage: 0 to 5 Vdc or 0 to 10 Vdc current Loop: 4 to 20 mA power Requirements: Millivolt: 5 to 10 Vdc (5 mA @ 10 Vdc) amplified Voltage: 0 to 5 Vdc Supply Voltage: 10 to 30 Vdc @ 10 mA 0 to 10 Vdc Supply Voltage: 15 to 30 Vdc @ 10 mA current Loop: 9 to 30 Vdc [max loop res = (Vs-9) x 50]; [9 to 20 Vdc above 105°C (229°F)] accuracy (combined Linearity, Hysteresis and Repeatability): ±0.08% BSL Zero Balance: ±0.5% FS typical, 1% max Span Setting: ±0.5% FS typical, 1% max; calibrated in vertical direction with fitting down Barometric Pressure Absolute Pressure outputs: mV/V, 0 to 5V, 0 to 10V or 4 to 20 mA 0 to 32, 16 to 32 or 26 to 32 inHg 0 to 1100, 550 to 1100, 880 to 1100 hPa U High accuracy ±0.08% BSL includes Linearity, Hysteresis, and Repeatability U Broad temperature compensated Range -18 to 85°c (0 to 185°f) U premium temperature performance Span: ±0.5% over compensated Range U 5-point niSt traceable calibration included U all Stainless Steel Wetted parts U fast Response time U Solid State Reliability and Stability U 400% proof pressure OMEGA’s PX409 series electronic barometers are high-accuracy barometers available in three ranges: 16 to 32 inHg, 26 to 32 inHg, or 0 to 32 inHg (absolute pressure) for laboratory or industrial test applications. Models with integral cables (PX409) are IP67 while connector styles (PX419 and PX429) are IP65 rated. They feature all stainless steel construction, which makes them ideal for harsh industrial environments. Their high operating temperature, broad compensated range, and excellent temperature compensation allow stable readings in applications with fluctuating temperatures commonly found where barometric pressure must be monitored. Models are available with millivolt, voltage or current outputs. ip65 rated pX429-26Bi ip67 rated pX409-32BV ip65 rated pX419-16B5V Product Label Hex 22 (0.87) AF ¼-18 NPT Pressure Adaptor (0.88) 81 (3.2) 81 (3.2) ¼-18 NPT Pressure Adaptor Hex 2.5 (1.0) AF Conduit back end PXM409C Integral cable back end PXM409 (1.1) (0.2) Twist lock back end PXM429 Vent with porous plug (gage units only) mini DIN back end PXM419 (0.2) M1 2x1 Ø 20.2 0.8 (0.2) (0.88) Vent with porous plug (gage units only) Glass to medal sealed pins PXM449 (0.2) Integral cable back end PXM409 (0.354) Dimensions: mm (inch) Standard cable style mini Din style Metric threads available m12 Connector also Available See To order Chart PX409 Series all shown actual size. twist-lock style

B VoLtaGe oUtpUt pReSSURe tRanSDUceRS B-134 operating temperature Range: -45 to 121°C (-49 to 250°F) [-45 to 115°C (-49 to 240°F) for voltage or current outputs] compensated temperature: Ranges > 5 psi: -29 to 85°C (-20 to 185°F) Ranges ≤ 5 psi: -18 to 85°C (0 to 185°F) thermal effects Span (over compensated Range): Ranges > 5 psi: ±0.5% span Ranges ≤ 5 psi: ±1% span Long term Stability (1-Year): ±0.1% FS typical Shock: 50 g, 11 mS half sine, vertical and horizontal axis Vibration: 5-2000-5 Hz, 30 minute cycle, Curve L, Mil-Spec 810 figure 514-2-2, vertical and horizontal axis Response time: <1 ms Bandwidth: DC to 1 kHz typical proof pressure: 400% of span Burst pressure: 1000 psia electrical termination: pX409: Integral 2 m (6') cable pX419: mini DIN pX429: Twist-lock pX419-26B5V, shown smaller than actual size. HIGH ACCURACY, mICRo-mACHINED SIlICoN TRANSDUCERS mini Din style MoDeL no. DeScRiption pt06f10-6S Mating connector for PX429 series, sold separately cX5302 Spare mating connector for PX419 series; one included with each unit M12c-pVc-4-S-f-5 PVC cable, straight 4-pin M12 female connector one end, flying leads one end, 5 m (16.4') long, fits PX459 M12c-pVc-4-R-f-5 PVC cable, right angled 4-pin M12 female connector one end, flying leads one end, 5 m (16.4') long, fits PX459 -MB Mounting bracket, factory installed acceSSoRieS To Order caLiBRateD RanGe mV/V oUtpUt

5 Vdc**

BaRoMetRic RanGeS (absolute pressure) 0 to 32 inHg — pX4[*]9-32BV pX4[*]9-32B5V pX4[*]9-32Bi 16 to 32 inHg — pX4[*]9-16BV pX4[*]9-16B5V pX4[*]9-16Bi 26 to 32 inHg — pX4[*]9-26BV pX4[*]9-26B5V pX4[*]9-26Bi — 0 to 1100 hPa pX4[*]9-32HBV pX4[*]9-32HB5V pX4[*]9-32HBi — 550 to 1100 hPa pX4[*]9-16HBV pX4[*]9-16HB5V pX4[*]9-16HBi — 880 to 1100 hPa pX4[*]9-26HBV pX4[*]9-26HB5V pX4[*]9-26HBi ** For 0 to 10 Vdc output replace “5V” with “10V” in model number. [*] Select Electrical Termination (no extra charge). Insert “0” for 2 m (6') cable with flying leads. For ½ NPT conduit connector order PX409CS. Insert “1” for Mini DIN connector (mating connector included). Insert “2” for Twist-lock connector (mating connector PT06F10-6S sold separately). Insert “5” for M12 connector, M12 cables sold separately. Custom configurations also available. PX459-32HB5V, M12 connector, 0 to 1100 hPa barometric range, 0 to 5 Vdc output. PX419-16HBV mini DIN termination, 550 to 1100 hPa barometric range, mV/V output. pX429 tWiSt LocK pinoUt pin mV 5/10V ma a + eXc + eXc + Supply B – eXc common – Supply c + SiG + output nc D – SiG nc nc e nc nc nc f nc nc nc pX419 and pX459 pin oUt pin mV 5/10V ma 1 + eXc + eXc + Supply 2 – eXc common – Supply 3 + SiG + output nc 4 – SiG nc nc pX409 and pX409c caBLe connection coLoR mV 5/10V ma Black – eXc common – Supply White + SiG + output nc Green – SiG nc nc Red + eXc + eXc + Supply ip67 rated pX409-32BV ip65 rated pX429-26Bi ip65 rated pX419-16B5V Mating connectors: pX419: CX5302 (included) pX429: PT06F10-6S (sold separately) environmental protection: pX409: IP67 pX419: IP65 pX429: IP65 pX459: IP65 Wetted parts: 316 SS pressure port: 1⁄4-18 NPT male Weight: 115 to 200 g (4 to 7 oz) depending upon configuration all models shown smaller than actual size.