CPS11D EH

Document overview

  • Manufacturer or author: Provided By www.digicamel.com(free datasheet download site)
  • PDF pages: 16

Technical content

Digital with Memosens technology or analog Application

  • Long-term monitoring and limit control in processes with stable process conditions
  • Chemical industry: strong acids/bases, plastic, pulp and paper industry
  • Power stations (e.g. flue gas scrubbing), oil and gas
  • Incinerator plants
  • Water and wastewater treatment
  • Boiler feedwater and cooling water
  • Well water and drinking water
  • All industrial and municipal treatment plants With ATEX, IECEx, FM, CSA, TIIS and NEPSI approval for use in hazardous areas. Your benefits
  • Low-maintenance and robust thanks to large, dirt-repellent PTFE ring junction
  • Can be used at pressures up to 17 bar (246.5 psi) (absolute)
  • Process glass also for very alkaline applications (BA and BT versions)
  • Process glass for applications in media containing hydrofluoric acid (FA version)
  • For media with low conductivity (AS version)
  • Integrated NTC 30K (Memosens) temperature sensor for effective temperature compensation
  • Optional: poison-resistant reference with ion trap Other advantages provided by Memosens technology
  • Maximum process safety
  • Data security thanks to digital data transmission
  • Very easy to use as sensor data are saved in the sensor
  • Predictive maintenance can be performed with the Memobase Plus CYZ71D by recording sensor load data in the sensor Products Solutions Services Technical Information Orbisint CPS11D and CPS11 pH sensor for standard applications in process technology and environmental engineering TI00028C/07/EN/15.20 71526304 2020-03-23

2 Endress+Hauser

Function and system design Measuring principle pH measurement The pH value is used as a unit of measurement for the acidity or alkalinity of a medium. The membrane glass of the electrode delivers an electrochemical potential that depends on the pH value of the medium. This potential is generated by the selective accumulation of H+ ions on the outer layer of the membrane. As a result, an electrochemical boundary layer with an electrical potential difference forms at this point. An integrated Ag/AgCl reference system serves as the required reference electrode. The measured voltage is converted to the corresponding pH value using the Nernst equation. Measuring system A complete measuring system comprises at least:

  • pH sensor CPS11D or CPS11
  • Transmitter, e.g. Liquiline CM42, CM44x, Mycom S CPM153, Liquisys M CPM2x3
  • Memosens data cable CYK10 for Memosens sensors or CPK9 for analog sensors
  • Assembly
  • Immersion assembly, e.g. Dipfit CPA111
  • Flow assembly, e.g. Flowfit CPA250
  • Retractable assembly, e.g. Cleanfit CPA871
  • Permanent installation assembly, e.g. Unifit CPA842 Other options are available depending on the application: automatic cleaning and calibration system, e.g. Liquiline Control CDC90 A0025757  1 Example of a measuring system for pH measurement

1 Retractable assembly Cleanfit CPA871

3 Memosens data cable CYK10

4 Liquiline M CM42 two-wire transmitter for hazardous areas

Communication with the transmitter Always connect digital sensors with Memosens technology to a transmitter with Memosens technology. Data transmission to a transmitter for analog sensors is not possible.

Digital sensors can store measuring system data in the sensor. These include the following:

  • Manufacturer data
  • Serial number
  • Order code
  • Date of manufacture
  • Calibration data
  • Calibration date
  • Slope at 25 °C (77 °F)
  • Zero point at 25 °C (77 °F)
  • Offset of integrated temperature sensor
  • Number of calibrations
  • Calibration history
  • Serial number of the transmitter used to perform the last calibration or adjustment
  • Application data
  • Temperature application range
  • pH application range
  • Date of initial commissioning
  • Maximum temperature value
  • Hours of operation under extreme conditions
  • Number of sterilizations
  • Resistance of glass membrane
  • CIP counter The data listed above can be displayed with Liquiline CM42, CM44x, and Memobase Plus CYZ71D. Dependability Reliability Easy handling Sensors with Memosens technology have integrated electronics that store calibration data and other information (e.g. total hours of operation or operating hours under extreme measuring conditions). Once the sensor has been connected, the sensor data are transferred automatically to the transmitter and used to calculate the current measured value. As the calibration data are stored in the sensor, the sensor can be calibrated and adjusted independently of the measuring point. The result:
  • Easy calibration in the measuring lab under optimum external conditions increases the quality of the calibration.
  • Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the availability of the measuring point.
  • Thanks to the availability of the sensor data, maintenance intervals can be accurately defined and predictive maintenance is possible.
  • The sensor history can be documented with external data carriers and evaluation programs, e.g. Memobase Plus CYZ71D.
  • Thus, the current application of the sensors can be made to depend on their previous history. Integrity Data security thanks to digital data transmission Memosens technology digitizes the measured values in the sensor and transmits the data to the transmitter via a non-contact connection that is free from potential interference. The result:
  • If the sensor fails or there is an interruption in the connection between the sensor and transmitter, this is reliably detected and reported.
  • The availability of the measuring point is reliably detected and reported. Security Maximum process safety With inductive transmission of the measured value using a non-contact connection, Memosens guarantees maximum process safety and offers the following benefits:
  • All problems caused by moisture are eliminated:
  • Plug-in connection free from corrosion
  • Measured values cannot be distorted by moisture
  • Can even be connected under water
  • The transmitter is galvanically decoupled from the medium. Issues concerning "symmetrical high- impedance" or "asymmetry" or an impedance converter are a thing of the past.
  • EMC safety is guaranteed by screening measures for the digital transmission of measured values.
  • Intrinsically safe electronics mean operation in hazardous areas is not a problem.

4 Endress+Hauser

Measured variables pH value Temperature Measuring range AA and AS versions

  • pH: 1 to 12
  • Temperature: –15 to 80 °C (5 to 176 °F) BA version
  • pH: 0 to 14
  • Temperature: 0 to 135 °C (32 to 275 °F) FA version
  • pH: 0 to 10
  • Temperature: 0 to 70 °C (30 to 158 °F) BT version with ion trap
  • pH: 0 to 14
  • Temperature: 0 to 135 °C (32 to 275 °F) Pay attention to the operating conditions in the process. Power supply Electrical connection Memosens sensors GN/YE YE GN BN WH GND Com A Com B GY A0024019  2 Measuring cable CYK10 or CYK20 Sensors with TOP68 plug-in head BN n.c. BK WH GN WH YE Temp. Temp. Temp. GY A0028048  3 Measuring cable CPK9 Sensors with GSA plug-in head BK Ref. WH pH/ORP BN PM PM A0028051  4 Measuring cable CPK1 ‣ Please follow the connection instructions contained in the Operating Instructions for the transmitter.

CPS11D: Memosens plug-in head for digital, non-contact data transmission CPS11: ESA: Threaded plug-in head Pg 13.5, TOP68 for electrodes with and without a temperature sensor, 17 bar (246 psi)(abs.) overpressure protection (threefold), Ex GSA: Threaded plug-in head Pg 13.5 for electrodes without a temperature sensor Performance characteristics Reference system AA, BA, FA versions: Ag/AgCl reference lead with advanced gel 3M KCl, AgCl-free AS version: Ag/AgCl reference lead with advanced gel, saturated KCl (> 3M KCl) with salt rings, AgCl-free The following are indicative of used salt rings (fixed supply of KCl) under constant process conditions (e.g. stable temperature and flow):

  • a continuous upward trend in the pH value (to alkaline pH values)
  • a continuous downward trend in the zero point (to acidic pH values) following adjustment during calibration BT version: Ag/AgCl reference lead with ion trap and advanced gel 3M KCl Installation Orientation • Do not install the sensors upside-down.
  • The installation angle from the horizontal must be at least 15°. An installation angle <15° is not permitted, as otherwise an air bubble will form. Contact between the membrane glass and the reference lead will then no longer be guaranteed. 15° 15° A B A0028039  5 Installation angle at least 15° from the horizontal A Permitted orientation B Forbidden orientation Installation instructions For detailed installation instructions for the assembly: refer to the Operating Instructions of the assembly used. 1. Before screwing in the sensor, make sure the assembly thread, the O-rings and the sealing surface are clean and undamaged and that the thread runs smoothly. 2. Screw in the sensor and tighten by hand with a torque of 3 Nm (2.21 lbf ft) (specifications only apply if installing in Endress+Hauser assemblies).

6 Endress+Hauser

Ambient temperature range NOTICE Risk of damage from frost! ‣ Do not use the sensor at temperatures below –15 °C (5 °F) . Storage temperature 0 to 50 °C (32 to 122 °F) Degree of protection IP 68: Memosens plug-in head (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl) IP 68: TOP68 plug-in head, autoclavable up to 135 °C (275 °F), 1 m (3.3 ft) water column, IP 67: GSA plug-in head (with closed connector system) Electromagnetic compatibility (EMC) Interference emission and interference immunity as per EN 61326: 2012 Process Process temperature range AA, AS versions: –15 to 80 °C (5 to 176 °F) BA, BT versions: 0 to 135 °C (32 to 275 °F) FA version: 0 to 70 °C (32 to 158 °F) Process pressure range AA, AS, FA versions: 1 to 7 bar (14.5 to 101.5 psi) (absolute) BA, BT versions: 1 to 17 bar (14.5 to 246.5 psi) (absolute) LCAUTION Pressurization of sensor due to prolonged use under increased process pressure Possibility of sudden rupture and injury from glass splinters! ‣ Avoid fast heating of these pressurized sensors if they are used under reduced process pressure or under atmospheric pressure. ‣ When handling these sensors, always wear protective goggles and appropriate protective gloves. Conductivity AA, BA, BT, FA versions: 50 μS/cm minimum (minimized flow; pressure and temperature must be stable) AS version: 0.1 μS/cm minimum (stainless steel flow assembly with grounding; stable and minimized flow; pressure and temperature must be stable)

x 212 0 100 [°C] T [°F] 20 40 60 80 120 24832 68 104 140 1765 A [psi] p [bar] 116 145 174 203 0 0 -15 16232 B A0025761  6 Pressure/temperature chart A BA, BT versions B FA version x Atmospheric pressure x 212 0 100 [°C] T [°F] 20 40 60 80 120 24832 68 104 140 1765 [psi] p [bar] 116 145 174 203 0 0 -15 16232 A0042300  7 Pressure/temperature chart A AA, AS versions x Atmospheric pressure

8 Endress+Hauser

Design, dimensions 812(0.47) 120 / 225 / 360 50(1.97) A0025726  8 CPS11D with Memosens plug-in head, temperature sensor 1 Memosens plug-in head, Pg 13.5

2 Viton O-ring with thrust collar

3 Ag/AgCl reference lead - reference

4 "Advanced gel" electrolyte

5 Ag/AgCl reference lead - pH

6 PTFE junction

8 Temperature sensor NTC 30K

12(0.47) 50(1.97) 120 / 225 / 360 A0025725  9 CPS11D-7BTxx, with ion trap 1 Memosens plug-in head, Pg 13.5 4 "Advanced gel" electrolyte

5 Ion trap

6 Ag/AgCl reference lead - pH

7 PTFE junction

9 Temperature sensor NTC 30K

12 (0.47) A0025733  10 CPS11 with GSA plug-in head 1 GSA plug-in head, Pg 13.5 4 "Advanced Gel" electrolyte

6 PTFE diaphragm

12 (0.47) A0025729  11 CPS11 with TOP68 plug-in head, temperature sensor 1 TOP68 plug-in head, Pg 13.5 4 "Advanced Gel" electrolyte

8 Temperature sensor Pt100

Weight 0.1 kg (0.2 lbs) Materials Sensor shaft: Glass to suit process pH membrane glass: Type A, B, F Metal lead: Ag/AgCl Open aperture: Ring-shaped PTFE junction, sterilizable, not cytotoxic Temperature sensor CPS11D: NTC30K CPS11: Pt100, Pt1000 Process connections Pg 13.5 Certificates and approvals  mark The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EU directives. The manufacturer confirms successful testing of the product by affixing to it the  mark.

10 Endress+Hauser

  • ATEX II 1G Ex ia IIC T3/T4/T6 Ga
  • IECEx 1G Ex ia IIC T3/T4/T6 Ga
  • FM/CSA Class I Div. 2, in conjunction with the Liquiline M CM42 and Mycom S CPM153 transmitters Hazardous area versions of the digital sensors with Memosens technology are indicated by a red/orange ring in the plug-in head. CPS11 (TOP68)
  • ATEX II 1G Ex ia IIC T3/T4/T6 Ga
  • FM Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters Marine approvals Marine approvals A selection of the devices and sensors have type approval for marine applications, issued by the following classification societies: ABS (American Bureau of Shipping), BV (Bureau Veritas), DNV-GL (Det Norske Veritas-Germanischer Lloyd) and LR (Lloyd’s Register). Details of the order codes of the approved devices and sensors, and the installation and ambient conditions, are provided in the relevant certificates for marine applications on the product page on the Internet. Additional certification Biocompatibility Cytotoxicity verified in accordance with: USP 2009, chapter <88> (USP Class VI) for diaphragm TÜV certificate for Memosens plug-in head Pressure resistance 16 bar (232 psi) relative, minimum three times the safety pressure TÜV certificate for TOP68 plug-in head Pressure resistance 16 bar (232 psi) relative, minimum three times the safety pressure EAC The product has been certified according to guidelines TP TC 004/2011 and TP TC 020/2011 which apply in the European Economic Area (EEA). The EAC conformity mark is affixed to the product.

Ordering information

Product page www.endress.com/cps11d www.endress.com/cps11 Product Configurator On the product page there is a Configure button to the right of the product image. 1. Click this button.  The Configurator opens in a separate window. 2. Select all the options to configure the device in line with your requirements.  In this way, you receive a valid and complete order code for the device. 3. Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right above the selection window. For many products you also have the option of downloading CAD or 2D drawings of the selected product version. Click the CAD tab for this and select the desired file type using picklists. Scope of delivery The delivery comprises:

  • Sensor in the version ordered
  • Operating Instructions
  • Safety instructions for the hazardous area (for sensors with Ex approval)

The following are the most important accessories available at the time this documentation was issued. ‣ For accessories not listed here, please contact your Service or Sales Center. Assemblies Cleanfit CPA871

  • Flexible process retractable assembly for water, wastewater and the chemical industry
  • For applications with standard sensors with 12 mm diameter
  • Product Configurator on the product page: www.endress.com/cpa871 Technical Information TI01191C Cleanfit CPA875
  • Retractable process assembly for sterile and hygienic applications
  • For in-line measurement with standard sensors with 12 mm diameter, e.g. for pH, ORP, oxygen
  • Product Configurator on the product page: www.endress.com/cpa875 Technical Information TI01168C Cleanfit CPA472D
  • Robust retractable assembly for pH, ORP and other industrial sensors
  • Heavy-duty version made of durable materials
  • For manual or pneumatic, remote-controlled operation
  • Product Configurator on the product page: www.endress.com/cpa472d Technical Information TI00403C Cleanfit CPA450
  • Manual retractable assembly for installing sensors with a diameter of 12 mm and a length of 120 mm in tanks and pipes
  • Product Configurator on the product page: www.endress.com/cpa450 Technical Information TI00183C Cleanfit CPA473
  • Stainless steel process retractable assembly with ball valve shutoff for particularly reliable separation of the medium from the environment
  • Product Configurator on the product page: www.endress.com/cpa473 Technical Information TI00344C Cleanfit CPA474
  • Plastic process retractable assembly with ball valve shutoff for particularly reliable separation of the medium from the environment
  • Product Configurator on the product page: www.endress.com/cpa474 Technical Information TI00345C Unifit CPA442
  • Installation assembly for food, biotechnology and pharmaceutics
  • With EHEDG and 3A certificate
  • Product Configurator on the product page: www.endress.com/cpa442 Technical Information TI00306C Dipfit CPA111
  • Immersion and installation assembly made of plastic for open and closed vessels
  • Product Configurator on the product page: www.endress.com/cpa111 Technical Information TI00112C Dipfit CPA140
  • pH/ORP immersion assembly with flange connection for very demanding processes
  • Product Configurator on the product page: www.endress.com/cpa140 Technical Information TI00178C

12 Endress+Hauser

  • pH/ORP flow assembly for processes with stringent requirements
  • Product Configurator on the product page: www.endress.com/cpa240 Technical Information TI00179C Flowfit CPA250
  • Flow assembly for pH/ORP measurement
  • Product Configurator on the product page: www.endress.com/cpa250 Technical Information TI00041C Ecofit CPA640
  • Set comprising adapter for 120 mm pH/ORP sensors and sensor cable with TOP68 coupling
  • Product Configurator on the product page: www.endress.com/cpa640 Technical Information TI00246C Flexdip CYA112
  • Immersion assembly for water and wastewater
  • Modular assembly system for sensors in open basins, channels and tanks
  • Material: PVC or stainless steel
  • Product Configurator on the product page: www.endress.com/cya112 Technical Information TI00432C Buffer solutions High-quality buffer solutions from Endress+Hauser - CPY20 The secondary buffer solutions have been referenced to primary reference material of the PTB (German Federal Physico-technical Institute) or to standard reference material of NIST (National Institute of Standards and Technology) according to DIN 19266 by a laboratory accredited by the DAkkS (German accreditation body) according to DIN 17025. Product Configurator on the product page: www.endress.com/cpy20 Measuring cable Memosens data cable CYK10
  • For digital sensors with Memosens technology
  • Product Configurator on the product page: www.endress.com/cyk10 Technical Information TI00118C Measuring cable CPK9
  • Terminated measuring cable for connecting analog sensors with TOP68 plug-in head
  • Selection in accordance with product structure
  • Product Configurator on the product page: www.endress.com/cpk9 Technical Information TI00118C CPK1
  • For pH/ORP sensors with GSA plug-in head
  • Product Configurator on the product page: www.endress.com/cpk1 For ordering information, contact your sales office or see www.endress.com.

www.addresses.endress.com *71526304* 71526304