CPM223 EH
Document overview
- Manufacturer or author: Provided By www.digicamel.com(free datasheet download site)
- PDF pages: 24
Technical content
Transmitter for analog and digital glass and ISFET sensors Application
- Wastewater treatment
- Neutralization
- Detoxification (electroplating)
- Water treatment Your benefits
- Memosens technology
- Field or panel-mounted housing
- Universal application, Easy to use
- Safe to operate – Manual contact control and user-defined alarm configuration – Calibration plausibility check The basic device can be extended with:
- 2 or 4 contacts for use as – Limit contacts (also for temperature) – P(ID) controller – Timer for simple rinse processes or Chemoclean – Current input
- Plus package: – Configurable current output characteristic – Cleaning started automatically – Sensor check system – Sensor signal live check – Special neutralization controller
- HART or PROFIBUS-PA/-DP
- 2nd current output for temperature, pH/ORP or actuating variable Products Solutions Services Technical Information Liquisys M CPM223/253 pH/ORP measurement TI00194C/07/EN/15.14 71271003
2 Endress+Hauser
Function and system design Measuring system A complete measuring system comprises:
- Transmitter Liquisys M CPM223 or CPM253
- pH or ORP sensor with or without an integrated temperature sensor
- Immersion, flow or retractable assembly
- pH measuring cable (e.g. CPK9) Optionally:
- Extension cable, junction box VBA or VBM
- Weather protection cover CYY101 for field housing 1 2 3 A0024634 1 Complete measuring systems
1 Flow assembly CPA250
2 Junction box VBA
3 Liquisys M CPM253
4 Measuring cable e.g. CPK9
5 Liquisys M CPM223
6 Retractable assembly Cleanfit W CPA450
7 Electrode, e.g. Orbisint CPS11
8 Immersion assembly CPA111
9 Extension cable
Block diagram With Memosens sensors
15 V85
N 23 + 24 - mA ~ = pH mA31 mA33 A B C D E F G H J K L M N S IA B BN WH GN YE A0007754 2 Block circuit diagram with Memosens sensors A Shielding H Alarm (current-free contact position) B Sensor I Relay 1 (current-free contact position) C Signal output 1 pH/ORP J Relay 2 (current-free contact position) D Signal output 2 temperature, pH/ORP or controller K Relay 3 (current-free contact position) E Binary input 1 (hold) L Relay 4 (current-free contact position) F Binary input 2 (Chemoclean) M Current input 4 to 20 mA G Auxiliary voltage output N Power supply * Auxiliary voltage, terminal 85/86 can be used
4 Endress+Hauser
Ref. PA/ PM S 10-50 V 10-50 V N 23 + 24 - mA ~ = pH mA31 mA33 A C D E F G H I J K L M N O P Q SRC DRN. B A0007753 3 Block circuit diagram with analog sensors A Standard sensor J Auxiliary voltage output B ISFET sensor K Alarm (current-free contact position) C Outer shield connection with glass electrodes L Relay 1 (current-free contact position) D Potential equalization M Relay 2 (current-free contact position) E Temperature sensor N Relay 3 (current-free contact position) F Signal output 1 pH/ORP O Relay 4 (current-free contact position) G Signal output 2 temperature, pH/ORP or controller P Current input 4 to 20 mA H Binary input 1 (hold) Q Power supply I Binary input 2 (Chemoclean) * Auxiliary voltage, terminal 85/86 can be used
Reliability Sensor check system (SCS (Plus package)) The Sensor Check System (SCS) monitors the high impedance of the pH glass. An alarm is issued if a minimum impedance value is undershot or a maximum impedance is exceeded.
- Glass breakage is the main reason for a drop in high impedance values
- The reasons for increasing impedance values include: – Dry sensor – Worn pH glass membrane Sensor live check (process check system, PCS (Plus package)) The process check system (PCS) checks the measuring signal for stagnation. An alarm is triggered if the measuring signal does not change over a specific period (several measured values). The main causes of stagnating measured values are:
- Contaminated sensor, or sensor outside of medium
- Sensor defective
- Process error (e.g. through control system) Plausibility check pH electrodes are usually always calibrated with the same pH values. The transmitter therefore provides the setting for the last calibration as a default value when the next calibration is performed. If the buffer solutions are swapped by mistake during the calibration (e.g. pH 4 buffer first, then pH 7 buffer instead of pH 7 buffer first, then pH 4 buffer), the plausibility check ensures that the calibration is accepted nevertheless. Current output configuration (Plus package) In order to display wide measuring ranges while still achieving a high resolution in specific ranges, the current output can be configured as required via a table. This permits bilinear and quasi- logarithmic curves etc. Current input The transmitter current input permits two different applications:
- Flow monitoring with controller switch-off if flow falls below lower flow level in the main flow
- Feedforward control to the controller The two functions can also be combined. Neutralization controller (Plus package) The neutralization of solutions requires a special control response which cannot be provided satisfactorily with a simple P(ID) response. The transmitter provides the control response of a special neutralization controller combined with two P(ID) controllers. Safety Process safety Different alarms are required depending on the application and operator. The transmitter therefore permits the independent configuration of the fault-signaling contact and the error current for every individual error. Unnecessary or undesired alarms can be suppressed in this way. Up to four contacts can be used as limit contacts (also for temperature), as a P(ID) controller and for cleaning functions. The direct manual operation of the contacts (bypassing the menu) enables quick access to limit value, control or cleaning contacts. Any deviations can be quickly corrected in this way.
6 Endress+Hauser
Measuring ranges pH 0 to 14 ORP –1500 to +1500 mV / 0 to 100 % Temperature Pt 100 -50 to +150 °C (-60 to +300 °F) Pt 1000 (CPM2x3-IS) -50 to +150 °C (-60 to +300 °F) NTC 30K (CPM2x3-IS) -20 to +100 °C (0 to +212 °F) Input impedance > 1012 Ω (under rated operating conditions) for standard sensors Binary inputs Voltage 10 to 50 V Current consumption Max. 10 mA Current input 4 to 20 mA, galvanically isolated Load: 260 Ω for 20 mA (voltage drop 5.2 V) Output Output signal 0/4 to 20 mA, galvanically isolated, active HART Signal encoding Frequency Shift Keying (FSK) + 0.5 mA via current output signal Data transmission rate 1200 baud Galvanic isolation Yes PROFIBUS PA Signal encoding Manchester Bus Powered (MBP) Data transmission rate 31.25 kBit/s, voltage mode Galvanic isolation Yes (IO modules) PROFIBUS DP Signal encoding RS485 Galvanic isolation Yes (IO modules) Signal on alarm 2.4 or 22 mA in the event of an error Load Max. 500 Ω
Transmission range pH Configurable, min. Δ 1 pH ORP Absolute Configurable, min. Δ 50 mV Relative Fixed, 0 to 100 % Temperature Configurable, Δ 10 to Δ 100 % of end of measuring range Signal resolution Max. 700 digits/mA Minimum spread of output signal 10 % of the measuring range span Separation voltage Max. 350 VRMS / 500 V DC Auxiliary voltage output Output voltage 15 V ± 0.6 V Output current Max. 10 mA Contact outputs Switching current with ohmic load (cos φ = 1) Max. 2 A Switching current with inductive load (cos φ = 0.4) Max. 2 A Switching voltage Max. 250 V AC, 30 V DC Switching power with ohmic load (cos φ = 1) Max. 500 VA AC, 60 W DC Switching power with inductive load (cos φ = 0.4) Max. 500 VA AC, 60 W DC Limit contactors Pickup/dropout delay 0 to 2000 s Controller Function (configurable) Pulse length/pulse frequency controller, continuous controller Controller behavior P, PI, PD, PID, basic load dosing Control gain Kp 0.01 to 20.00 Integral action time Tn 0.0 to 999.9 min Derivative action time Tv 0.0 to 999.9 min Period length for pulse length controller 0.5 to 999.9 s Frequency for pulse frequency controller 60 to 180 min-1 Basic load 0 to 40 % of max. actuating variable Alarm Function (switchable) Latching/momentary contact Alarm threshold adjustment range pH/temperature: entire measuring range Alarm delay 0 to 2000 s Monitoring time for lower limit violation 0 to 2000 min Monitoring time for upper limit violation 0 to 2000 min Protocol-specific data HART Manufacturer ID 11h Device type 0091h Transmitter-specific revision 0001h HART version 5.0 Device description files (DD) www.endress.com/hart HART load (communication resistor) 250 Ω Device variables None (only dynamic variables PV and SV) Supported features -
8 Endress+Hauser
Profile version 2.0 GSD files www.endress.com/profibus GSD version Output values Primary value, temperature Input variables PCS display value Supported features Device lock: The device can be locked using the hardware or software. PROFIBUS DP Manufacturer ID 11h Device type 1520h Profile version 2.0 GSD files www.endress.com/profibus GSD version Output values Primary value, temperature Input variables PCS display value Supported features Device lock: The device can be locked using the hardware or software. Power supply Supply voltage Depending on order version:
- 100/115/230 V AC +10/-15 %, 48 to 62 Hz Power supply via fieldbus HART Supply voltage Not applicable, active current outputs Reverse polarity protection Not applicable, active current outputs PROFIBUS PA Supply voltage 9 V to 32 V, max. 35 V Sensitivity to reverse polarity No FISCO/FNICO compliant according to IEC 60079-27 No
Supply voltage 9 V to 32 V, max. 35 V Sensitivity to reverse polarity Not applicable FISCO/FNICO compliant according to IEC 60079-27 No Power consumption Max. 7.5 VA Mains fuse Fine-wire fuse, semi-delay 250 V/3.15 A Circuit breaker NOTICE The device does not have a power switch ‣ The customer must provide a protected circuit breaker in the vicinity of the device. ‣ The circuit breaker must be a switch or power switch, and you must label it as the circuit breaker for the device. ‣ At the supply point, the power supply for the 24 V versions must be isolated from dangerous live cables by double or reinforced insulation. Cable specification Cable length Memosens Max. 100 m (330 ft) Cable length analog sensors Max. 50 m (160 ft) Overvoltage protection According to EN 61000-4-5 Sensor connection The pH and ORP sensors are connected via multi-core, pre-terminated, shielded special cables. Use a junction box and an extension cable to extend the measuring cable. Cable termination instructions are provided with the measuring cables. S CYK10 CPS11D BN WH GN YE 87 88 96 97 A0007773 4 Connecting the digital sensor CPS11D with CYK10
10 Endress+Hauser
5 Connecting the glass electrode CPS11 with CPK9 (left) and ISFET sensor CPS471 with CPK12 (right) A Panel-mounted device B Field device C Potential matching PM for symmetrical connection A0024644 6 Asymmetrical (without PML) and symmetrical (with PML) connection of ORP electrodes A Panel-mounted device B Field device C Potential matching PM for symmetrical connection
Performance characteristics Reference operating conditions Reference temperature: 25 ˚C (77 ˚F) Measured value resolution pH value 0.01 pH ORP 1 mV/0.1 % Temperature 0.1 °C Maximum measured error Display pH Max. 0.5 % of measuring range ORP Max. 0.5 % of measuring range Temperature Max. 1.0 % of measuring range Signal output pH Max. 0.75 % of measuring range ORP Max. 0.75 % of measuring range Temperature Max. 1.25 % of measuring range Measured errors in accordance with DIN IEC 746 Part 1, at rated operating conditions Repeatability Max. 0.2 % of measuring range Zero point shift Glass electrode pH 5.0 to 9.0 (nominal pH 7.00) Antimony electrode pH -1.0 to 3.0 (nominal pH 1.00) ISFET sensor -500 to +500 mV Slope adjustment Glass electrode 38.00 to 65.00 mV/pH (nominal 59.16 mV/ pH) Antimony electrode 25.00 to 65.00 mV/pH (nominal 59.16 mV/ pH) ISFET sensor 38.00 to 65.00 mV/pH (nominal 59.16 mV/ pH) Offset pH ±2 pH units ORP ±120 mV/±50 % Temperature ±5 °C for adjusting the temperature display
12 Endress+Hauser
Installation instructions Field device wall mounting 154 (6.06) Ø 6 (0.24) A0024638 7 Field device wall mounting
1 Fixing bore holes
2 Plastic caps
Field device post mounting Ø max. 60 (2.36) A0024635 8 Field device on horizontal or vertical pipes
1 Securing screws
2 Fixing screws
3 Securing plate
9 Field device with universal post and weather protection cover
1 Bore holes in the weather protection cover to secure to the upright post
2 Bore holes in the upright post to secure the weather protection cover
3 Bore holes in the weather protection cover to secure the field device
92 (3.62 +0.5 +0.02 92 (3.62 +0.5 +0.02 90 (3.54) 57 (2.24) 33 (1.30) max. 45 (1.77) 6 (0.24) 96 (3.78) ~ 165 (6.50)* 92 (3.62) A0024639 10 Dimensions in mm (inch)
1 Mounting plate
2 Seal
3 Tensioning screws
- Necessary installation depth
14 Endress+Hauser
Ambient temperature range -10 to +55 °C (+10 to +130 °F) Storage temperature –25 to +65 °C (-10 to +150 °F) Electromagnetic compatibility Interference emission and interference immunity as per EN 61326-1:2006, EN 61326-2-3:2006 Degree of protection Field device IP 65 / integrity according to NEMA 4X Panel-mounted device IP 54 (front), IP 30 (housing) Electrical safety As per EN/IEC 61010-1:2010, overvoltage category II for installations up to 2000 m (6500 ft) above MSL CSA Device versions with CSA General Purpose approval are certified for indoor use. Relative humidity 10 to 95%, not condensing Degree of contamination The product is suitable for pollution degree 2. Mechanical construction Design A0024640 11 View into the field device housing
1 Removable electronics box
2 Terminals
3 Partition plate
4 Fuse
11 (0.43) Ø 6 (0.24) 115 (4.53) 157 (6.18) 70 (2.76) 154 (6.06) 170 (6.69) Pg 13.5 A0024637 12 Field device: dimensions in mm (inch) 90 (3.54) 139 (5.47) 96 (3.78) 6 (0.24) 92 (3.62) 149.5 (5.89) A0024641 13 Panel-mounted device: dimensions in mm (inch) Weight Panel-mounted device Max. 0.7 kg (1.54 lbs.) Field device Max. 2.3 kg (5.07 lbs.)
16 Endress+Hauser
Materials Panel-mounted device housing Polycarbonate Field housing ABS PC FR Front membrane Polyester, UV-resistant Terminals Cable cross-section Max. 2.5 mm2 (14 AWG) Operability Operating concept All the device's operating functions are arranged in a clear menu structure. The individual parameters can be selected and modified once the access code has been entered. Display and operating elements The display shows the current measured value and the temperature simultaneously, which means you have an overview of the most important process data at once. Help text in the configuration menu helps users configure the device parameters. A0024629-EN 14 Operating elements
1 LC display for displaying the measured values and configuration data
2 Key to switch relays in manual mode and to display the active contact
3 LED for alarm function
4 Changeover switch for automatic/manual mode
5 LEDs for limit contactor relay (switch status)
6 Main operating keys for calibration and device configuration
7 Field for user-defined information
Certificates and approvals mark The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EC directives. The manufacturer confirms successful testing of the product by affixing to it the mark. CSA General Purpose The following device versions meet the requirements of CSA and ANSI/UL for Canada and the US:
Ordering information
Product page www.endress.com/cpm223 www.endress.com/cpm253 Product Configurator The navigation area is located on the right of the product page. 2. Under "Device support" click "Configure your selected product". The Configurator opens in a separate window. 3. Select all the options to configure the device in line with your requirements. In this way, you receive a valid and complete order code for the device. 4. Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of the screen. Scope of delivery The delivery of the field device comprises:
- 1 transmitter CPM253
- 1 plug-in screw terminal, 3-pin
- 1 cable gland Pg 7
- 1 cable gland Pg 16 reduced
- 2 cable glands Pg 13.5
- 1 set of Operating Instructions
- For versions with HART communication: 1 set of Operating Instructions: Field communication with HART
- For versions with PROFIBUS interface: 1 set of Operating Instructions: Field communication with PROFIBUS PA/DP The delivery of the panel-mounted device comprises:
- 1 transmitter CPM223
- 1 set of plug-in screw terminals
- 2 tensioning screws
- 1 BNC connector (solder-free measuring cable connection)
- 1 set of Operating Instructions
- For versions with HART communication: 1 set of Operating Instructions: Field communication with HART
- For versions with PROFIBUS interface: 1 set of Operating Instructions: Field communication with PROFIBUS PA/DP
18 Endress+Hauser
The following are the most important accessories available at the time this documentation was issued. For accessories not listed here, please contact your service or sales office. Sensors pH glass electrodes Orbisint CPS11D/ CPS11
- pH electrode for process technology
- Optional SIL version for connecting to SIL transmitter
- With dirt-repellent PTFE diaphragm
- Product Configurator on the product page: www.endress.com/cps11d or www.endress.com/cps11 Technical Information TI00028C Ceraliquid CPS41D/ CPS41
- pH electrode with ceramic junction and KCl liquid electrolyte
- Product Configurator on the product page: www.endress.com/cps41d or www.endress.com/cps41 Technical Information TI00079C Ceragel CPS71D/ CPS71
- pH electrode with double-chamber reference system and integrated bridge electrolyte
- Product Configurator on the product page: www.endress.com/cps71d or www.endress.com/cps71 Technical Information TI00245C Orbipore CPS91D/ CPS91
- pH electrode with open aperture diaphragm for media with high dirt load
- Product Configurator on the product page: www.endress.com/cps91d or www.endress.com/cps91 Technical Information TI00375C pH ISFET sensors Tophit CPS441D/ CPS441
- Sterilizable ISFET sensor for low-conductivity media
- Liquid KCl electrolyte
- Product Configurator on the product page: www.endress.com/cps441d or www.endress.com/cps441 Technical Information TI00352C Tophit CPS471D/ CPS471
- Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process engineering
- Water treatment and biotechnology
- Product Configurator on the product page: www.endress.com/cps471d or www.endress.com/cps471 Technical Information TI00283C Tophit CPS491D/ CPS491
- ISFET sensor with open aperture diaphragm for media with high dirt load
- Product Configurator on the product page: www.endress.com/cps491d or www.endress.com/cps491 Technical Information TI00377C ORP sensors Orbisint CPS12D/ CPS12
- ORP sensor for process technology
- Product Configurator on the product page: www.endress.com/cps12d or www.endress.com/cps12 Technical Information TI00367C
- ORP electrode with ceramic junction and KCl liquid electrolyte
- Product Configurator on the product page: www.endress.com/cps42d or www.endress.com/cps42 Technical Information TI00373C Ceragel CPS72D/ CPS72
- ORP electrode with double-chamber reference system and integrated bridge electrolyte
- Product Configurator on the product page: www.endress.com/cps72d or www.endress.com/cps72 Technical Information TI00374C Orbipore CPS92D/ CPS92
- ORP electrode with open aperture diaphragm for media with high dirt load
- Product Configurator on the product page: www.endress.com/cps92d or www.endress.com/cps92 Technical Information TI00435C Sensor simulators Memocheck Plus CYP01D / Memocheck CYP02D / Memocheck Sim CYP03D
- Testing tools for analysis measuring points
- Simple, fast and reliable sensor simulation
- Product Configurator on the product page: – www.endress.com/cyp01d – www.endress.com/cyp02d – www.endress.com/cyp03d Technical Information TI00481C Connection accessories CYK10 Memosens data cable
- For digital sensors with Memosens technology
- Product Configurator on the product page: www.endress.com/cyk10 Technical Information TI00118C Memosens data cable CYK11
- Extension cable for digital sensors with Memosens protocol
- Product Configurator on the product page: www.endress.com/cyk11 Technical Information TI00118C CPK9 For pH/ORP electrodes with TOP68 plug-in head CPK1 For pH/ORP electrodes with GSA plug-in head CPK2 For pH/ORP electrodes with GSA plug-in head, with three electrode connectors CPK12 For pH glass electrodes and ISFET sensors with TOP68 plug-in head Ordering information is available from your sales office or at www.endress.com. VBM
- Junction box for cable extension
- 10 terminal strips
- Cable entries: 2 x Pg 13.5 or 2 x NPT ½"
- Material: aluminum
- Degree of protection: IP 65
- Order numbers – Cable entries Pg 13.5 : 50003987 – Cable entries NPT ½": 51500177 VBA
- Junction box for cable extension
- 10 terminal strips
- Cable entries: 2 x Pg 13.5, 2 x Pg 16
20 Endress+Hauser
- Material: polycarbonate
- Degree of protection: IP 65
- Order number: 50005276 M12 socket
- Four-pin metal socket for mounting on transmitter
- For connecting sensor cables with an M12 connector
- Length of connecting cable for wiring in transmitter: 150 mm (5.9 inch).
- Order number: 51502184 n.c. n.c. A (PA-) A (PA-) S S B (PA+) B (PA+) A (PA-) S (PE) B (PA+) 1.77” 0.28” 5.91” / 11.81” A0024657 Installation accessories CYY101
- Weather protection cover for field devices
- Absolutely essential for field installation
- Material: stainless steel 1.4301 (AISI 304)
- Order No. CYY101-A 320 (12.6) A0024627 15 Dimensions in mm (inch)
- For securing the field housing to horizontal and vertical posts and pipes
- Material: stainless steel 1.4301 (AISI 304)
- Order No. 50086842 70 (2.76) 90 (3.54) Ø 6 (0.24) Ø 8 (0.31) 70 (2.76)90 (3.54) 60 (2.36) 90 (3.54) 70 (2.76) Ø max. 60 (2.36) A0024660 16 Dimensions in mm (inch) Universal post CYY102
- Square pipe for mounting transmitters
- Material: stainless steel 1.4301 (AISI 304)
- Order No. CYY102-A 1495 (58.86) 60 x 60 150 (5.91) A0024659 17 Dimensions in mm (inch)