Digital with Memosens 2.0 technology Robust pH sensor with plastic shaft Application Measurements in drinking water, wastewater and process water Your benefits

  • Simple to use, extremely versatile
  • Physical protection of glass membrane
  • Minimum risk of contamination thanks to open aperture diaphragm
  • Fast response time even in cold, ion-weak water
  • Integrated temperature sensor for automatic temperature compensation
  • Suitable for use with Liquiline Mobile, Liquiline To Go and Memobase Plus Other advantages of Memosens technology
  • Maximum analysis safety with non-contact, inductive signal transmission
  • Data security thanks to digital data transmission
  • Very easy to use as sensor data saved in the sensor Products Solutions Services Technical Information Memosens CPL51E pH sensor for laboratory measurements and random sampling in the field TI01672C/07/EN/01.20-00 71557011 2022-01-03

2 Endress+Hauser

Function and system design Measuring principle pH measurement The pH value is used as a unit of measurement for the acidity or alkalinity of a medium. The membrane glass of the electrode delivers an electrochemical potential that depends on the pH value of the medium. This potential is generated by the selective accumulation of H+ ions on the outer layer of the membrane. As a result, an electrochemical boundary layer with an electrical potential difference forms at this point. An integrated Ag/AgCl reference system serves as the required reference electrode. The measured voltage is converted to the corresponding pH value using the Nernst equation. Measuring system A0047842  1 Measuring system

1 Transmitter CML18

2 Smartphone with Smartblue app (optional)

3 Memosens CPL51E

Communication with the handheld device Always connect digital laboratory sensors with Memosens technology to a handheld device with Memosens technology, e.g. CML18. Digital laboratory sensors can store measuring system data in the sensor, including:

  • Manufacturer data
  • Serial number
  • Order code
  • Date of manufacture
  • Calibration data
  • Calibration date
  • Number of calibrations
  • Serial number of the handheld device used to perform the last calibration or adjustment
  • Application data
  • Temperature application range
  • pH application range
  • Date of initial commissioning Input Measured variable pH value Temperature

Measuring range • pH value: 0 to 14 pH

  • Temperature: 0 to 80 °C (32 to 176 °F) Performance characteristics Reference system Ag/AgCl lead, bridging electrolyte: gel KCl, 3M, AgCl-free Environment Ambient temperature range NOTICE Risk of damage from frost! ‣ Do not use the sensor at temperatures below 0 °C (32 °F). Storage temperature 0 to 50 °C (32 to 122 °F), non-freezing Degree of protection IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl) Electromagnetic compatibility (EMC) Interference emission and interference immunity as per EN 61326-1: 2013 Process Process temperature range 0 to 80 °C (32 to 176 °F) Mechanical construction Design, dimensions Ø 11.9 (0.47) 120 (4.7) A0047294  2 Engineering unit: mm (in)

4 Endress+Hauser

Weight 33 g (1.2 oz) Materials Sensor shaft PEEK Metal lead Ag/AgCl Nameplate Ceramic metal oxide Junction Open junction Temperature sensor NTC 30K Plug-in head Memosens laboratory plug-in head for digital, non-contact data transmission Process connections Pg 13.5 Accessories The following are the most important accessories available at the time this documentation was issued. ‣ For accessories not listed here, please contact your Service or Sales Center. Device-specific accessories Memosens data cable CYK10

  • For digital sensors with Memosens technology
  • Product Configurator on the product page: www.endress.com/cyk10 Technical Information TI00118C Memosens laboratory cable CYK20
  • For digital sensors with Memosens technology
  • Product Configurator on the product page: www.endress.com/cyk20 Liquiline Mobile CML18
  • Multiparameter mobile device for laboratory and field
  • Reliable transmitter with display and app connection
  • Product Configurator on the product page: www.endress.com/CML18 Operating Instructions BA02002C Memobase Plus CYZ71D
  • PC software to support laboratory calibration
  • Visualization and documentation of sensor management
  • Sensor calibrations stored in database
  • Product Configurator on the product page: www.endress.com/cyz71d Technical Information TI00502C High-quality buffer solutions from Endress+Hauser - CPY20 The secondary buffer solutions have been referenced to primary reference material of the PTB (German Federal Physico-technical Institute) or to standard reference material of NIST (National Institute of Standards and Technology) according to DIN 19266 by a laboratory accredited by the DAkkS (German accreditation body) according to DIN 17025. Product Configurator on the product page: www.endress.com/cpy20

www.addresses.endress.com *71557011* 71557011