CPA250 EH

Document overview

  • Manufacturer or author: Provided By www.digicamel.com(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 8

Technical content

TI041C/07/en/04.09 71093745 Technical Information Flowfit CPA250 Flow assembly for pH and ORP sensors Application The CPA250 assembly is totally made of corrosion resistant plastic. It is designed for pipeline installation of Pg 13.5 pH or ORP sensors with installation length 120 mm (4.72 inch). Inlet and outlet of the fluid lie on one axis. Due to especially designed internal fluid deviation and siphon, the sensors are kept wet even when the flow is interrupted. Main areas of application are:

  • Water treatment
  • Drinking water
  • Process applications
  • Cooling water cycle Your benefits
  • Mounting places for max. three analog or digital sensors with Pg 13.5 thread (pH, ORP, temperature)
  • Threaded cap: – easy installation and removal of the sensors – protection of the sensor connection compartment
  • Potential matching pin selectable: – stainless steel 1.4571 (AISI 316Ti) or titane – no potential matching pin for Memosens sensors
  • Extension options: – Connection to the electrolyte vessel CPY7 – Adaption of the Chemoclean system
  • Economic flow assembly concept for in-line installation
  • Easy calibration by removable calibration vessel

2 Endress+Hauser

Function and system design Measuring system A complete measuring system comprises:

  • Flowfit CPA250
  • pH or ORP sensor, e.g. CPS71D
  • Measuring cable, e.g. CYK10
  • Transmitter, e.g. Liquiline M CM42 Optional:
  • up to two more pH/ORP sensors or sing le electrodes or temperature sensors
  • Junction box for cable extension, e.g. RM junction box a0007329 Fig. 1: Measuring system (bypass installation)

1 Process line with bypass and valves

2 Supply line of the transmitter

3 Transmitter Liquiline M CM42

4 Measuring cable CYK10

5 Flowfit CPA250

Installation instructions To get a flow through the by-pass, pressure p1 has to be higher than pressure p2. Therefore, you have to install an aperture or a throttle in the main conduit (→ å2). a0007334 Fig. 2: Installation example with by-pas s and aperture in the main conduit Alternatively, you can install a booster pump in the bypass conduit to produce the required pressure (→ å3). a0007335 Fig. 3: Installation example with open outlet p1 p> p12p2

4 Endress+Hauser

In case of an open outlet installation, no pressure increasing procedure is needed (→ å4). a0007336 Fig. 4: Installation example with open outlet Environment Ambient temperature range 0 to 50 °C (32 to 120 °F) Storage temperature 0 to 50 °C (32 to 120 °F) Process Process temperature range 0 to 80 °C (32 to 170 °F, 6 bar (87 psi) at 20 °C (68 °F) and pressureless at 80 °C (176 °F) Process pressure max. 6 bar (87 psi) at 20 °C (68 °F) p

Fig. 5: Temperature-Pressure diagram Mechanical construction Design, dimensions a0007327 Fig. 6: Dimensions Weight 0.5 to 0.8 kg (1.1 to 1.8 lbs), depending on process connection 85 (3.35) 247 (9.72) 124 (4.88) 86 (3.39) 38 (1.5) Ø 9 (0.35) mm (inch)

6 Endress+Hauser

Materials In contact with medium:

  • Assembly body: PP-H
  • O-rings: Viton/FPM Potential matching pin:
  • CPA250-A00/01/02/03/30: titane or stainless steel
  • CPA250-A04/05: no potential matching pin Process connection depending on the version: G1, NPT 1" Suitable sensors Glass electrodes or ISFET sensors, analog or digital with Memosens technology, with Pg 13.5 thread and shaft length 120 mm (4.72 inch), with or without liquid KCl reference electrode

Ordering information

  • Material: PP
  • Process connection: NPT ½"
  • Modification number: C-PA050720-41 Note! Use the product root CPA240 to find stainless steel versions. Request a technical special product for special material versions. Scope of delivery The scope of delivery comprises:
  • Flowfit assembly (ordered version)
  • Operating Instructions (English) Design A Mounting places for 3 sensors Process connection, Material, Potential matching pin (PMP) 00 G1, PP, PMP 1.4571 (316 Ti)

01 G1, PP, PMP titanium

02 NPT 1", PP, PMP 1.4571 (316 Ti)

03 NPT 1", PP, PMP titanium

04 NPT 1", PP, no PMP

05 G1, PP, no PMP

30 G1, PP LABS free, PMP titanium

CPA250- complete order code

Note! In the following sections, you find the accessories available at the time of issue of this documentation. For information on accessories that are not listed here, please contact your local service. Accessories kits •N P – 2 nipples for adaption of CPA250-A* to PP pipe without thread – OD 32 mm (1.26 inch) – order no. 50003450

  • N V – Adapter for adaption of CPA250-A* to PVC pipe – OD 32 mm (1.26 inch), order no. 50003454 – OD 25 mm (0.98 inch), order no. 50003456
  • B F – Wall mounting kit for CPA250-A* – 2 long V4A screws, incl. screw anchors – order no. 50001130 Sensors Orbisint CPS11/11D
  • pH electrode for process applications with dirt-repellent PTFE diaphragm
  • Optional Memosens technology (CPS11D)
  • Ordering acc. to product structure, see Technical Information (TI028C/07/en) Ceragel CPS71/CPS71D
  • pH electrode with double junction refere nce system and integrated bridge electrolyte
  • Optional with Memosens technology (CPS71D)
  • Ordering acc. to product structure, see Technical Information (TI245C/07/en) Ceragel CPS72/CPS72D
  • Redox sensor with double junction refere nce system and integrated bridge electrolyte
  • Optional with Memosens technology (CPS72D)
  • Ordering acc. to product structure, see Technical Information (TI374C/07/de) Cleaning systems Chemoclean CPR31 / CPR3
  • Automatic spray cleaning system for sensors
  • CPR31 is installed in one of the three sensor mounting places of the assembly
  • order no. on request Note! When installing CPR31 remove the metal pin and align the spray nozzles. Chemoclean
  • Injector CYR10 and program sequencer CYR20
  • Ordering acc. to product structure, see Technical Information (TI046C/07/en)

Instruments International AG Kaegenstrasse 2

4153 Reinach

Tel.+41 61 715 81 00 Fax+41 61 715 25 00 www.endress.com info@ii.endress.com TI041C/07/en/04.09 71093745 FM+SGML 6.0