Space-saving transmitter for monitoring and controlling processes in industry and the environmental sector Application The CM72 transmitter can be used in all sectors and by plant manufacturers in these sectors and supports sensors with the blue Memosens plug-in head:
- pH sensors
- ORP sensors
- Contacting conductivity sensors
- Oxygen sensors Direct connection to PLC via: 4 to 20 mA Your benefits
- Space-saving installation:
- The two-wire device fits into an assembly and does not require a separate power supply.
- Minimum inventory
- Fast commissioning and maintenance:
- Thanks to its permanent configuration, the CM72 does not need to be commissioned and can start measurement immediately.
- All of the benefits of Memosens technology: lab-calibrated sensors, hot plug & play
- The status of the transmitter and the connected sensor is indicated by a red/ green LED.
- Suitable for all locations Regardless of whether your measuring point is exposed to dust, steam, rain, snow, heat or cold, the CM72 is always exactly the transmitter you need! Products Solutions Services Technical Information Liquiline Compact CM72 Single-parameter transmitter for Memosens sensors TI01409C/07/EN/03.22-00 71601178 2022-08-05
2 Endress+Hauser
Function and system design Measuring system The overview shows examples of measuring systems. Other sensors and assemblies can be ordered for conditions specific to your application (www.endress.com/products). A complete measuring system comprises the following components:
- Liquiline compact transmitter
- Sensors with Memosens technology
- Assemblies to suit the sensors used A0036844 1 Example of a measuring system
1 Measuring point with assembly and Memosens sensor
2 Liquiline Compact CM72
3 Junction box (optional)
4 PLC (programmable logic controller)
Sensors with Memosens protocol Sensor types Sensors Digital sensors with inductive Memosens plug-in head • pH sensors
- ORP sensors
- Oxygen sensors
- Conductivity sensors Communication and data processing Communication protocols: 4 to 20 mA For configuration with the measured value and the current output turndown, select the option in the order structure when ordering. This cannot be changed at a later stage.
Memosens makes your measuring point safer and more reliable:
- Non-contact, digital signal transmission enables optimum galvanic isolation
- No contact corrosion
- Completely watertight
- Sensor can be calibrated in a lab, thus increasing the availability of the measuring point in the process
- Predictive maintenance thanks to recording of sensor data, e.g.:
- Total hours of operation
- Hours of operation with very high or very low measured values
- Hours of operation at high temperatures
- Number of steam sterilizations
- Sensor condition A0035116 2 Plug & Play with Memosens technology The status of the transmitter and the connected sensor is indicated by a red/green LED. A0036843 3 LED display Security Measured value compensation pH: Temperature Oxygen:
- Temperature
- Air pressure Conductivity: Temperature Compensation of temperature dependency is linear.
4 Endress+Hauser
Measured variable The transmitter is designed for digital Memosens sensors with an inductive plug-in head:
- pH
- ORP
- Conductive conductivity
- Dissolved oxygen Depending on the order version, the measuring range is configured to suit the sensor type:
- pH sensor: 0 to 14 pH
- ORP: -1500 mV to +1500 mV
- Conductivity: 0 to 20 μS/cm
- Conductivity: 0 to 500 μS/cm
- Conductivity: 0 to 20 mS/cm
- Conductivity: 0 to 500 mS/cm
- Oxygen: 0 to 200 μg/l
- Oxygen: 0 to 20 mg/l Measuring range → Documentation of the connected sensor Type of input Digital sensor inputs for Memosens-sensors Output Output signal 4 ... 20 mA, galvanically isolated from the sensor circuits Linearization Linear Transmission behavior Linear Power supply Supply voltage 12.6 to 30 VDC (when failure current setting > 20 mA) 14 to 30 VDC (when failure current setting < 4 mA) V 12 13 14 15 16 17 18 19 20 100 300 500 800 700 600 12,6 Ω 200 400 21 22 23 24 25 26 27 28 29 30 A0036752 4 Supply voltage and load
The lower voltage value in each case applies only to a load resistance of 0 Ohm. NOTICE The device does not have a power switch ‣ At the supply point, the power supply must be isolated from dangerous live cables by double or reinforced insulation in the case of devices with a 24 V power supply. Cable specification Cable length:
- Max. 3 m (10 ft)
- Max. 7 m (23 ft)
- Max. 15 m (49 ft) Overvoltage protection IEC 61 000-4-4 and IEC 61 000-4-5 with +/- 1 kV Performance characteristics Resolution Current output < 5 µA Repeatability → Documentation of the connected sensor Response time Current output t90 = max. 500 ms for an increase from 0 to 20 mA Tolerance Current output Typical measuring tolerances: < ±20 µA (if current value = 4 mA) < ±50 µA (for current values 4 to 20 mA) at 25 ˚C (77˚ F) each additional tolerance depending on the temperature: < 1.5 µA/K Environment Ambient temperature range –20 to 85 °C (–4 to 185 °F) The maximum ambient temperature depends on the process temperature and the transmitter's installation position. ‣ Make sure that the ambient temperature at the transmitter does not exceed 85 °C (185 °F). Example for ambient conditions in Endress+Hauser assemblies:
- for open installation (without protective cover, i.e. free convection at the transmitter), e.g. CPA442, CPA842
- for enclosed installation (with protective cover), e.g. CPA871, CPA875, CPA842 Tambient = max. 60 °C (140 °F) Tprocess = max. 100 °C (212 °F), in continuous operation Tprocess = max. 140 °C (284 °F), < 2h (for sterilization)
6 Endress+Hauser
5 Installation position of transmitter with or without protective cover
1 Ambient temperature Tambient
2 Process temperature Tprocess
Storage temperature -40 to +85 °C (-40 to 185 °F) Relative humidity 5 to 95 % Degree of protection IP 67 IP 68 NEMA Type 6 Electromagnetic compatibility (EMC)
- EN 61326-1
- EN 61326-2-3
- NAMUR NE 21 Electrical safety EN 61010-1 Operating height < 2000 m (< 6562 ft) above MSL Pollution degree Complete device: Pollution level 4 Internal: Pollution level 2
110.8 (4.33) 99.8 (3.92) 32.75 (1.29) 114.7 (4.52) 124.6 (4.90) Ø18.95 (0.75) Ø5.5 (0.22) A0033272 6 Dimensions in mm (inch) Materials Components Material Housing, cover Peek 151 Strain relief EPDM (peroxide crosslinked) Axial ring Peek 450 G Optical waveguide PC transparent Impact loads The product is designed for mechanical impact loads of 1 J (IK06) as per the requirements of EN 61010-1. Weight without cable Approx. 42 g (1.5 oz) 3 m (9 f) cable Approx. 190 g (7 oz) 7 m (23 f) cable Approx. 380 g (13 oz) 15 m (49 f) cable Approx. 760 g (27 oz) For every 1 m (3 f) of cable Approx. 48 g (2 oz) Operability Operating concept For configuration with the measured value and the current output turndown, select the option in the order structure when ordering. This cannot be changed at a later stage. Certificates and approvals Current certificates and approvals that are available for the product can be selected via the Product Configurator at www.endress.com: 1. Select the product using the filters and search field. 2. Open the product page.
8 Endress+Hauser
- Select Configuration.
Ordering information
Product page www.endress.com/CM72 Product Configurator 1. Configure: Click this button on the product page. 2. Select Extended selection. The Configurator opens in a separate window. 3. Configure the device according to your requirements by selecting the desired option for each feature. In this way, you receive a valid and complete order code for the device. 4. Apply: Add the configured product to the shopping cart. For many products, you also have the option of downloading CAD or 2D drawings of the selected product version. 5. Show details: Open this tab for the product in the shopping cart. The link to the CAD drawing is displayed. If selected, the 3D display format is displayed along with the option to download various formats. Scope of delivery The scope of delivery includes:
- CM72
- Brief Operating Instructions Accessories Device-specific accessories Sensors pH glass electrodes Memosens CPS11E
- pH sensor for standard applications in process and environmental engineering
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cps11e Technical Information TI01493C Memosens CPS31E
- pH sensor for standard applications in drinking water and swimming pool water
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cps31e Technical Information TI01574C Memosens CPS41E
- pH sensor for process technology
- With ceramic junction and KCl liquid electrolyte
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cps41e Technical Information TI01495C Memosens CPS71E
- pH sensor for chemical process applications
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cps71e Technical Information TI01496C
- pH electrode for bio-fermenters with digital Memosens technology
- Product Configurator on the product page: www.endress.com/cps171d Technical Information TI01254C Memosens CPS91E
- pH sensor for heavily polluted media
- With open aperture
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cps91e Technical Information TI01497C Memosens CPF81E
- pH sensor for mining operations, industrial water and wastewater treatment
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cpf81e Technical Information TI01594C Enamel pH electrodes Ceramax CPS341D
- pH electrode with pH-sensitive enamel
- Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
- Product Configurator on the product page: www.endress.com/cps341d Technical Information TI00468C ORP sensors Memosens CPS12E
- ORP sensor for standard applications in process and environmental engineering
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cps12e Technical Information TI01494C Memosens CPS42E
- ORP sensor for process technology
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cps42e Technical Information TI01575C Ceragel CPS72D
- ORP electrode with reference system including ion trap
- Product Configurator on the product page: www.endress.com/cps72d Technical Information TI00374C Memosens CPF82E
- ORP sensor for mining operations, industrial water and wastewater treatment
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cpf82e Technical Information TI01595C Orbipore CPS92D
- ORP electrode with open aperture for media with high dirt load
- Product Configurator on the product page: www.endress.com/cps92d Technical Information TI00435C
10 Endress+Hauser
- Sterilizable ISFET sensor for low-conductivity media
- Liquid KCl electrolyte
- Product Configurator on the product page: www.endress.com/cps441d Technical Information TI00352C Tophit CPS471D
- Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process engineering
- Water treatment and biotechnology
- Product Configurator on the product page: www.endress.com/cps471d Technical Information TI00283C Tophit CPS491D
- ISFET sensor with open aperture for media with high dirt load
- Product Configurator on the product page: www.endress.com/cps491d Technical Information TI00377C Conductivity sensors with conductive measurement of conductivity Memosens CLS15E
- Digital conductivity sensor for measurements in pure and ultrapure water
- Conductive measurement
- With Memosens 2.0
- Product Configurator on the product page: www.endress.com/cls15e Technical Information TI01526C Memosens CLS16E
- Digital conductivity sensor for measurements in pure and ultrapure water
- Conductive measurement
- With Memosens 2.0
- Product Configurator on the product page: www.endress.com/cls16e Technical Information TI01527C Memosens CLS21E
- Digital conductivity sensor for media with medium or high conductivity
- Conductive measurement
- With Memosens 2.0
- Product Configurator on the product page: www.endress.com/cls21e Technical Information TI01528C Memosens CLS82E
- Hygienic conductivity sensor
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cls82e Technical Information TI01529C Oxygen sensors Memosens COS22E
- Hygienic amperometric oxygen sensor with maximum measurement stability over multiple sterilization cycles
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cos22e Technical Information TI01619C
- Amperometric oxygen sensor for water, wastewater and utilities
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cos51e Technical Information TI01620C Memosens COS81E
- Hygienic optical oxygen sensor with maximum measurement stability over multiple sterilization cycles
- Digital with Memosens 2.0 technology
- Product Configurator on the product page: www.endress.com/cos81e Technical Information TI01558C Software Memobase Plus CYZ71D
- PC software to support laboratory calibration
- Visualization and documentation of sensor management
- Sensor calibrations stored in database
- Product Configurator on the product page: www.endress.com/cyz71d Technical Information TI00502C DeviceCare SFE100 Configuration tool for HART, PROFIBUS and FOUNDATION Fieldbus field devices DeviceCare is available for download at www.software-products.endress.com. You need to register in the Endress+Hauser software portal to download the application. Technical Information TI01134S Other accessories Cable junction with Velcro strip Cable junction with Velcro strip
- 4 pieces, for sensor cable
- Order No. 71092051 System components RIA15
- Process display unit, Digital display unit for integration into 4-20 mA circuits
- Panel mounting
- With optional HART communication Technical Information TI01043K
www.addresses.endress.com *71601178* 71601178