Ultrapure water is used in sensitive areas of the pharmaceutical and food industries and in process engineering. Calibration of process conductivity measurement is absolutely necessary for these applications. Conducal is a reference unit that permits the calibration of process measuring devices by means of a certified comparison measurement. It can be used as follows:
- Calibration of quality-relevant process measurements in the ultrapure and pure water range up to 20 μS/cm
- Calibration of in-line measurements after inspections or interruptions in operation
- Calibration of measuring systems in the pharmaceutical and food industry
- Calibration to ensure product quality, such as in the semi-conductor industry, for example Your benefits
- Factory calibration traceable to NIST and PTB
- Factory calibration of the Conducal system according to ASTM D-5399–93
- Flow assembly with adjustment and monitoring functions according to ASTM D-5391
- Flexible use in the field
- Easy to clean thanks to polished surfaces
- Battery-operated system (independent of power network)
- Robust IP67 case with external sensor connection Products Solutions Services Technical Information Conducal CLY421 Conductivity calibration kit for ultrapure water
applications
TI00496C/07/EN/02.22-00 71600839 2022-09-29
2 Endress+Hauser
Function and system design Measuring system Comparison measurement with bypass (preferred) A0050828 1 Measuring arrangement for comparison measurement in the bypass
1 Process transmitter
2 Process conductivity sensor
3 Calibration kit transmitter
4 Calibration kit conductivity sensor
5 Bypass outlet
6 Calibration kit flow assembly
7 Bypass inlet
8 Sterile main pipe
With this measuring arrangement, the sensor is not removed from the process. Please ensure that the composition of the medium and the temperature at the process measurement point and comparison measurement point are the same. This is ensured by:
- Using short hose connections
- Waiting until the temperature in the flow assembly adjusts to match the process temperature. Advantage: Process remains hygienic Disadvantage: Waiting time for temperature harmonization
Direct comparison measurement 2 4 A0050829 2 Measuring arrangement for direct comparison measurement
3 Calibration kit conductivity sensor
4 Calibration kit transmitter
5 Outlet
7 Inlet
The flow assembly included with the calibration kit has two installation slots for conductivity sensors and therefore enables a direct comparison measurement. Advantages:
- No temperature difference, therefore no waiting time
- Absolutely identical medium Disadvantage: As the process must be opened, the medium can become contaminated Dependability Flow assembly The Conducal calibration kit is fitted with a special flow assembly for the Memosens CLS15E conductivity sensor.
4 Endress+Hauser
3 Flow assembly
1 Switch from bypass (1) to direct (2)
2 Installation slot for calibration kit conductivity sensor
3 Installation slot for process conductivity sensor
4 Inlet for bypass measurement
5 Inlet for direct comparison measurement
6 Flow control valve
7 Outlet
8 Flow monitoring
The medium must be bubble-free to prevent false conductivity measurements. Outgassing can occur when the medium releases tension, i.e. also in the bypass arrangement described. Thanks to the bubble trap integrated in the flow assembly, the medium around the conductivity sensor stays bubble-free. A0050890 4 Medium supply in case of comparison measurement with bypass
1 Vent hole
2 Inlet
3 Second installation slot
4 Lateral bore hole
When the flow assembly is used with the bypass setting, the second installation slot is used as a hydrocyclone. Inflow is tangential, producing a turbulent flow, so any bubbles can escape through a vent opening at the top. The medium is conveyed to the actual measurement chamber via a lateral bore hole.
Controlled conditions with flow monitoring When calibrating with Conducal according to ASTM D-5391, a minimum flow specified by the sensor manufacturer must be observed. The flow assembly in the Conducal calibration kit is designed for conductivity calibration according to the standards:
- Bubble trap (for bypass measurement)
- Float-type flow measurement
- Flow adjustment with control valve
- Temperature stability up to 100 °C (210 °F) 2 3 4 5 A0050892 5 Schematic of bypass arrangement
1 G½ inlet
2 Vent
4 Flow control valve
5 Flow monitoring
6 G½ outlet
The flow assembly is equipped with a flexible holder for mounting on round and square pipes. This makes comparison measurements in the field safe and easy. A0050893 6 Holder
6 Endress+Hauser
Calibration of Conducal CLY421 (flowchart) Traceable to international NIST and PTB standards according to ASTM D-5391 Standards NIST U.S. metrology authority SRM (Standard Reference Material) No. 3191 from NIST (100 μS/cm) DAkkS German metrology authority Calibration of working standards for temperature and resistance → Ohmic resistors as secondary standard ↓ ↓ ↓ Working standard at Endress+Hauser Sensor with cell constant k = 1 cm as secondary standard Temperature sensor Pt100 as secondary standard Calibration of transmitter inputs (linearity) ↓ ↓ Conducal CLY421 Calibration of the cell constant in a solution with approx. 5 μS/cm Calibration of the temperature sensor in a water bath ↓ ↓ Sensor and transmitter in the customer's facility Calibration of the cell constant of the customer's measuring system at 0.04 to 20 μS/cm, and of the temperature sensor Input Measured variable Conductivity [μS/cm] or [MΩcm]; configurable Power supply Supply voltage Wide range power supply 100 to 240 VAC, 47 to 63 Hz, Class II equipment with functional grounding Battery Integrated lithium-ion battery 14.4 V; 2.4 Ah The fully charged battery enables a calibration kit operating time of over 80 hours. External sensor cable connection Buccaneer plug, 6-pin, IP 68 Performance characteristics Error calculation Reference system adjustment with standard NIST reference material Uncertainty of reference solution 0.2 % Uncertainty of temperature measurement << 0.1 % Uncertainty of reference system display 0.2 % Total uncertainty of reference system adjustment 0.3 % Conducal adjustment with 5 μS/cm (or 200 kΩcm) Uncertainty of reference system adjustment 0.3 % Uncertainty of reference system measurement at 5 μS/cm 0.6 % Uncertainty of Conducal display at 5 μS/cm 0.6 %
Total uncertainty of Conducal adjustment at 5 μS/cm 0.9 % (Only corresponds to the uncertainty of Conducal. The adjustment of measuring points with Conducal requires an additional uncertainty analysis.) The change of the Memosens CLS15E cell constant in the conductivity range between the standard reference material and 5 μS/cm is not considered. Reference devices Reference measuring device used Liquiline CM42 Reference measuring cell used Condumax CLS15E Environment Ambient temperature +5 to +40 °C (41 to 104 °F) Relative humidity Max. 80 % Operating altitude Up to 2000 m Degree of protection IP 30 with case open IP 67 with case closed without power cable Indoor use (Pollution degree II) Process Process temperature 0 to 100 °C (32 to 210 °F) Process pressure Max. 6 bar (87 psi) Minimum flow 30 l/h (8 gal/h) Mechanical construction Dimensions L x W x H (case) 530 x 442 x 215 mm (20.9" x 17.4" x 8.5") Weight Approx. 12.7 kg (28 lb) Materials Flow assembly: PVDF Clamp seal: EPDM Adapter PVDF Process connection Inlet: G½ or Clamp ½" outlet Outlet G½ or Clamp ½" Vent G½
8 Endress+Hauser
Certificates and approvals Current certificates and approvals that are available for the product can be selected via the Product Configurator at www.endress.com: 1. Select the product using the filters and search field. 2. Open the product page. 3. Select Configuration.
Ordering information
Scope of delivery The scope of delivery includes:
- Calibration kit in the version ordered
- Operating Instructions Conducal CLY421
- Calibration certificate If you have any questions, please contact your supplier or your local sales center. Product page www.endress.com/CLY421 Product Configurator 1. Configure: Click this button on the product page. 2. Select Extended selection. The Configurator opens in a separate window. 3. Configure the device according to your requirements by selecting the desired option for each feature. In this way, you receive a valid and complete order code for the device. 4. Apply: Add the configured product to the shopping cart. For many products, you also have the option of downloading CAD or 2D drawings of the selected product version. 5. Show details: Open this tab for the product in the shopping cart. The link to the CAD drawing is displayed. If selected, the 3D display format is displayed along with the option to download various formats. Accessories The following are the most important accessories available at the time this documentation was issued. Listed accessories are technically compatible with the product in the instructions. 1. Application-specific restrictions of the product combination are possible. Ensure conformity of the measuring point to the application. This is the responsibility of the operator of the measuring point. 2. Pay attention to the information in the instructions for all products, particularly the technical data. 3. For accessories not listed here, please contact your Service or Sales Center. Device-specific accessories Memosens CLS15E
- Digital conductivity sensor for measurements in pure and ultrapure water
- Conductive measurement
- With Memosens 2.0
- Product Configurator on the product page: www.endress.com/cls15e Technical Information TI01526C
- Universal assembly for analysis systems in industrial utilities
- Product Configurator on the product page: www.endress.com/CYA21 Technical Information TI01441C Service-specific accessories Calibration service Recalibration
- The conductivity calibration kit must be calibrated regularly onsite at the manufacturer's depending on the frequency of use and operating conditions.
- Recommended period: 1 year