Transmitter for conductive and inductive sensors Application
- Ultrapure water
- Water treatment
- Ion exchanger and reverse osmosis
- Cooling water desalination
- Wastewater Your benefits
- Field or panel-mounted housing
- Easy to use – Simple menu structure – Calibration using the CAL key
- Manual contact control and user-defined alarm configuration The basic device can be extended with:
- 2 or 4 contacts for use as – Limit contacts (also for temperature) – P(ID) controller – Timer for simple rinse processes or Chemoclean
- Plus package: – Configurable current output characteristic – Ultrapure water monitoring (USP/EP, conductive) – Concentration Measurement – Process check system (PCS): Sensor signal live check
- HART or PROFIBUS-PA/-DP
- 2nd current output: temperature, main measured value, actuating variable
- Current input for flow monitoring or for feedforward control Products Solutions Services Technical Information Liquisys M CLM223/253 Conductivity/resistivity measurement TI00193C/07/EN/15.14 71273532
2 Endress+Hauser
Function and system design Measuring system A complete measuring system comprises:
- Transmitter Liquisys M CLM223 or CLM253
- sensor with or without an integrated temperature sensor
- If necessary, measuring cable: CYK71 or CPK9 (conductive measurement) or CLK5 (inductive measurement) Optionally:
- Immersion assembly, e.g. CLA111
- Extension cable, junction box VBM
- Weather protection cover CYY101 for field housing ENDRESS+HAUSER LIQUISYS S ENDRESS+HAUSER LIQUISYS S A0024642 1 Complete measuring systems
1 Conductive sensor CLS15
2 Liquisys M CLM253
3 Liquisys M CLM223
4 Inductive sensor CLS54
5 Inductive sensor CLS50
6 Conductive sensor CLS21
7 Immersion assembly CLA111
S PE L–N ~ = PE 23 + 24 - mA 85+ 86 _ A B C D E F G H I J K L M N O A0024647 2 Block diagram A Sensor (conductive) I Alarm (current-free contact position) B Sensor (inductive) J Relay 1 (current-free contact position) C Temperature sensor K Relay 2 (current-free contact position) D Signal output 1, conductivity L Relay 3 (current-free contact position) E Signal output 2, user-definable variable M Relay 4 (current-free contact position) F Binary input 1 (hold) N Current input 4 to 20 mA G Binary input 2 (Chemoclean) O Power supply H Auxiliary voltage output * Auxiliary voltage, terminal 85/86 can be used
4 Endress+Hauser
Reliability Sensor live check (process check system, PCS (Plus package)) The process check system (PCS) checks the measuring signal for stagnation. An alarm is triggered if the measuring signal does not change over a specific period (several measured values). The main causes of stagnating measured values are:
- Contaminated sensor, or sensor outside of medium
- Sensor defective
- Process error (e.g. through control system) Current output configuration (Plus package) In order to display wide measuring ranges while still achieving a high resolution in specific ranges, the current output can be configured as required via a table. This permits bilinear and quasi- logarithmic curves etc. Second current output The second current output can be flexibly configured to output the temperature, the main measured value (conductivity, resistance, concentration) or the controller actuating variable. Current input The transmitter current input permits two different applications:
- Flow monitoring with controller switch-off if flow falls below lower flow level in the main flow
- Feedforward control to the controller The two functions can also be combined. Integrity Temperature compensation The temperature compensation options are: linear, NaCl curve according to IEC 746, ultrapure water NaCl (neutral compensation) or ultrapure water HCl (acid compensation, also for ammonia). The reference temperature is user-definable, the standard value is 25 °C (77 °F). Polarization detection (Plus package) Polarization effects in the boundary layer between the sensor and the medium limit the measuring range of conductive conductivity sensors. The transmitter can detect polarization effects using a smart signal evaluation process. Adaptive calibration for determining the installation factor (inductive sensor, Plus package) The measured value changes if inductive sensors must be installed in narrow pipes at a short distance from the pipe wall. The built-in adaptive calibration using the installation factor allows you to compensate for this once the sensor is installed. Safety Process safety Different alarms are required depending on the application and operator. The transmitter therefore permits the independent configuration of the fault-signaling contact and the error current for every individual error. Unnecessary or undesired alarms can be suppressed in this way. Up to four contacts can be used as limit contacts (also for temperature), as a P(ID) controller and for cleaning functions. The direct manual operation of the contacts (bypassing the menu) enables quick access to limit value, control or cleaning contacts. Any deviations can be quickly corrected in this way. Ultrapure water monitoring according to USP (United States Pharmacopoeia) and EP (European Pharmacopoeia) (Plus package) Ultrapure water monitoring according to USP <645> or EP means that the uncompensated conductivity and the temperature are measured and the results are compared against a table. The transmitter (conductive with additional contacts) has the following functions:
- Monitoring of "Water for Injection" (WFI) according to USP and EP
- Monitoring of "Highly Purified Water" (HPW) according to EP
- Monitoring of "Purified Water" (PW) according to EP
The user-configurable pre-alarm indicates undesired operating states in good time. To comply with USP or EP requirements, it is necessary to use a sensor that measures with highest precision, e.g. CLS16. Input Measured values Conductivity Resistivity Temperature Measuring ranges Conductivity (conductive) 0 to 600 mS/cm (uncompensated) Conductivity (inductive) 0 to 2000 mS/cm (uncompensated) Resistance 0 to 200 MΩcm Concentration 0 to 9999 (%, ppm, mg/l, TDS) Temperature -35 to +250 °C (can also be displayed in °F) Cell constant Configurable cell constant k = 0.0025 to 99.99 cm-1 Measuring frequency Conductivity, resistance (conductive) 170 Hz to 2 kHz Conductivity (inductive) 2 kHz Connectable temperature sensors Pt 100, Pt 1000, NTC 30K Binary inputs Voltage 10 to 50 V Current consumption Max. 10 mA Current input 4 to 20 mA, galvanically isolated Load: 260 Ω for 20 mA (voltage drop 5.2 V) Output Output signal 0/4 to 20 mA, galvanically isolated, active HART Signal encoding Frequency Shift Keying (FSK) + 0.5 mA via current output signal Data transmission rate 1200 baud Galvanic isolation Yes PROFIBUS PA Signal encoding Manchester Bus Powered (MBP) Data transmission rate 31.25 kBit/s, voltage mode Galvanic isolation Yes (IO modules)
6 Endress+Hauser
Galvanic isolation Yes (IO modules) Signal on alarm 2.4 or 22 mA in the event of an error Load Max. 500 Ω Transmission range Conductivity Configurable Resistivity Configurable Concentration Configurable actuating variable Configurable Temperature Configurable Signal resolution Max. 700 digits/mA Minimum spread of output signal Conductivity Measured value 0 to 1.999 μS/cm 0.2 μS/cm Measured value 0 to 19.99 μS/cm 2 μS/cm Measured value 20 to 199.9 μS/cm 20 μS/cm Measured value 200 to 1999 μS/cm 200 μS/cm Measured value 2 to 19.99 mS/cm 2 mS/cm Measured value 20 to 2000 mS/cm 20 mS/cm Resistance Measured value 0 to 199.9 kΩcm 20 kΩcm Measured value 200 to 1999 kΩcm 200 kΩcm Measured value 2 to 19.99 MΩcm 2.0 MΩcm Measured value 20 to 200 MΩcm 20 MΩcm Concentration No minimum turndown Temperature 15 °C Separation voltage Max. 350 VRMS / 500 V DC Auxiliary voltage output Output voltage 15 V ± 0.6 V Output current Max. 10 mA Contact outputs Switching current with ohmic load (cos φ = 1) Max. 2 A Switching current with inductive load (cos φ = 0.4) Max. 2 A Switching voltage Max. 250 V AC, 30 V DC Switching power with ohmic load (cos φ = 1) Max. 500 VA AC, 60 W DC Switching power with inductive load (cos φ = 0.4) Max. 500 VA AC, 60 W DC Limit contactors Pickup/dropout delay 0 to 2000 s Controller Function (configurable) Pulse length/pulse frequency controller, continuous controller Controller behavior P, PI, PD, PID, basic load dosing Control gain Kp 0.01 to 20.00 Integral action time Tn 0.0 to 999.9 min Derivative action time Tv 0.0 to 999.9 min Period length for pulse length controller 0.5 to 999.9 s Frequency for pulse frequency controller 60 to 180 min-1 Basic load 0 to 40 % of max. actuating variable
Alarm Function (switchable) Latching/momentary contact Alarm threshold adjustment range Conductivity/resistance/concentration/ temperature/USP/EP: entire range Alarm delay 0 to 2000 s Monitoring time for lower limit violation 0 to 2000 min Monitoring time for upper limit violation 0 to 2000 min Protocol-specific data HART Manufacturer ID 11h Device type 0092h (inductive measurement), 0093h (conductive measurement) Transmitter-specific revision 0001h HART version 5.0 Device description files (DD) www.endress.com/hart HART load (communication resistor) 250 Ω Device variables None (only dynamic variables PV and SV) Supported features - PROFIBUS PA Manufacturer ID 11h Device type 1515h Device revision 0001h Profile version 2.0 GSD files www.endress.com/profibus GSD version Output values Primary value, temperature Input variables PCS display value Supported features Device lock: The device can be locked using the hardware or software. PROFIBUS DP Manufacturer ID 11h Device type 1521h Profile version 2.0 GSD files www.endress.com/profibus GSD version Output values Primary value, temperature Input variables PCS display value Supported features Device lock: The device can be locked using the hardware or software.
8 Endress+Hauser
Supply voltage Depending on order version:
- 100/115/230 V AC +10/-15 %, 48 to 62 Hz Power supply via fieldbus HART Supply voltage Not applicable, active current outputs Reverse polarity protection Not applicable, active current outputs PROFIBUS PA Supply voltage 9 V to 32 V, max. 35 V Sensitivity to reverse polarity No FISCO/FNICO compliant according to IEC 60079-27 No PROFIBUS DP Supply voltage 9 V to 32 V, max. 35 V Sensitivity to reverse polarity Not applicable FISCO/FNICO compliant according to IEC 60079-27 No Power consumption Max. 7.5 VA Mains fuse Fine-wire fuse, semi-delay 250 V/3.15 A Circuit breaker NOTICE The device does not have a power switch ‣ The customer must provide a protected circuit breaker in the vicinity of the device. ‣ The circuit breaker must be a switch or power switch, and you must label it as the circuit breaker for the device. ‣ At the supply point, the power supply for the 24 V versions must be isolated from dangerous live cables by double or reinforced insulation. Cable specification Cable length (conductive) Conductivity: max. 100 m (330 ft) (CYK71) Resistance: max. 15 m (49 ft) (CYK71) Cable length (inductive) Max. 55 m (180 ft) (CLK5) CYK71 cable resistance 165 Ω/km (conductivity measurement) Overvoltage protection According to EN 61000-4-5
Sensor connection You require shielded special measuring cables to connect conductivity sensors to the transmitter. Use a junction box and an extension cable to extend the measuring cable (see Accessories). A0024755 3 Connection of conductive sensors RD WH CLK 5 J 83 16 15 8411 12 13 GN WH YE BU RD RD BU S A0024754 4 Connection of inductive sensors Performance characteristics Reference operating conditions Reference temperature: 25 ˚C (77 ˚F) Measured value resolution Conductivity Depends on measured value; 0.001 μS/cm up to a measured value of 1.999 μS/cm and k ≤ 0.5 cm-1 Temperature 0.1 °C Maximum measured error Display Conductivity Max. 0.5 % of measured value ± 4 digits Resistance Max. 0.5 % of measured value ± 4 digits Temperature Max. 1.0 % of measuring range Signal output Conductivity Max. 0.75 % of current output range Resistance Max. 0.75 % of current output range Temperature Max. 1.25 % of measuring range Measured errors in accordance with DIN IEC 746 Part 1, at rated operating conditions Repeatability Max. 0.2% of measured value ± 2 digits Temperature compensation Range -35 to +250 °C (-30 to 480 °F) Types of compensation Uncompensated, linear, NaCl, table Conductive only: ultrapure water NaCl, ultrapure water HCl Offset Temperature ±5 °C for adjusting the temperature display
10 Endress+Hauser
Installation instructions Field device wall mounting 154 (6.06) Ø 6 (0.24) A0024638 5 Field device wall mounting
1 Fixing bore holes
2 Plastic caps
Field device post mounting Ø max. 60 (2.36) A0024635 6 Field device on horizontal or vertical pipes
1 Securing screws
2 Fixing screws
3 Securing plate
7 Field device with universal post and weather protection cover
1 Bore holes in the weather protection cover to secure to the upright post
2 Bore holes in the upright post to secure the weather protection cover
3 Bore holes in the weather protection cover to secure the field device
92 (3.62 +0.5 +0.02 92 (3.62 +0.5 +0.02 90 (3.54) 57 (2.24) 33 (1.30) max. 45 (1.77) 6 (0.24) 96 (3.78) ~ 165 (6.50)* 92 (3.62) A0024639 8 Dimensions in mm (inch)
1 Mounting plate
2 Seal
3 Tensioning screws
- Necessary installation depth
12 Endress+Hauser
Ambient temperature range -10 to +55 °C (+10 to +130 °F) Storage temperature –25 to +65 °C (-10 to +150 °F) Electromagnetic compatibility Interference emission and interference immunity as per EN 61326-1:2006, EN 61326-2-3:2006 Degree of protection Field device IP 65 / integrity according to NEMA 4X Panel-mounted device IP 54 (front), IP 30 (housing) Electrical safety As per EN/IEC 61010-1:2010, overvoltage category II for installations up to 2000 m (6500 ft) above MSL CSA Device versions with CSA General Purpose approval are certified for indoor use. Relative humidity 10 to 95%, not condensing Degree of contamination The product is suitable for pollution degree 2. Mechanical construction Design A0024640 9 View into the field device housing
1 Removable electronics box
2 Terminals
3 Partition plate
4 Fuse
11 (0.43) Ø 6 (0.24) 115 (4.53) 157 (6.18) 70 (2.76) 154 (6.06) 170 (6.69) Pg 13.5 A0024637 10 Field device: dimensions in mm (inch) 90 (3.54) 139 (5.47) 96 (3.78) 6 (0.24) 92 (3.62) 149.5 (5.89) A0024641 11 Panel-mounted device: dimensions in mm (inch) Weight Panel-mounted device Max. 0.7 kg (1.54 lbs.) Field device Max. 2.3 kg (5.07 lbs.)
14 Endress+Hauser
Materials Panel-mounted device housing Polycarbonate Field housing ABS PC FR Front membrane Polyester, UV-resistant Terminals Cable cross-section Max. 2.5 mm2 (14 AWG) Operability Operating concept All the device's operating functions are arranged in a clear menu structure. The individual parameters can be selected and modified once the access code has been entered. Display and operating elements The display shows the current measured value and the temperature simultaneously, which means you have an overview of the most important process data at once. Help text in the configuration menu helps users configure the device parameters. A0024632-EN 12 Operating elements
1 LC display for displaying the measured values and configuration data
2 Key to switch relays in manual mode and to display the active contact
3 LED for alarm function
4 Changeover switch for automatic/manual mode
5 LEDs for limit contactor relay (switch status)
6 Main operating keys for calibration and device configuration
7 Field for user-defined information
Certificates and approvals mark The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EC directives. The manufacturer confirms successful testing of the product by affixing to it the mark. CSA General Purpose The following device versions meet the requirements of CSA and ANSI/UL for Canada and the US:
Ordering information
Product page www.endress.com/clm223 www.endress.com/clm253 Product Configurator The navigation area is located on the right of the product page. 2. Under "Device support" click "Configure your selected product". The Configurator opens in a separate window. 3. Select all the options to configure the device in line with your requirements. In this way, you receive a valid and complete order code for the device. 4. Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of the screen. Scope of delivery The delivery of the field device comprises:
- 1 transmitter CLM253
- 1 plug-in screw terminal, 3-pin
- 1 cable gland Pg 7
- 1 cable gland Pg 16 reduced
- 2 cable glands Pg 13.5
- 1 set of Operating Instructions
- For versions with HART communication: 1 set of Operating Instructions: Field communication with HART
- For versions with PROFIBUS interface: 1 set of Operating Instructions: Field communication with PROFIBUS PA/DP The delivery of the panel-mounted device comprises:
- 1 transmitter CLM223
- 1 set of plug-in screw terminals
- 2 tensioning screws
- 1 set of Operating Instructions
- For versions with HART communication: 1 set of Operating Instructions: Field communication with HART
- For versions with PROFIBUS interface: 1 set of Operating Instructions: Field communication with PROFIBUS PA/DP
16 Endress+Hauser
The following are the most important accessories available at the time this documentation was issued. For accessories not listed here, please contact your service or sales office. Sensors Sensors with conductive measurement of conductivity Condumax CLS12
- Conductive conductivity sensor
- For pure water, Ex and high-temperature applications
- Product Configurator on the product page: www.endress.com/CLS12 Technical Information TI00082C Condumax CLS13
- Conductive conductivity sensor
- For pure water, Ex and high-temperature applications
- Product Configurator on the product page: www.endress.com/CLS13 Technical Information TI00083C Condumax CLS15
- Conductive conductivity sensor
- For pure water applications, ultrapure water applications and Ex applications
- Product Configurator on the product page: www.endress.com/CLS15 Technical Information TI00109C Condumax CLS16
- Hygienic, conductive conductivity sensor
- For pure water applications, ultrapure water applications and Ex applications
- With EHEDG and 3A approval
- Product Configurator on the product page: www.endress.com/CLS16 Technical Information TI00227C Condumax CLS19
- Cost-effective, conductive conductivity sensor
- For applications with pure and ultrapure water
- Product Configurator on the product page: www.endress.com/CLS19 Technical Information TI00110C Condumax CLS21
- Two-electrode sensor in plug-in head and fixed cable version
- Product Configurator on the product page: www.endress.com/CLS21 Technical Information TI00085C Sensors with inductive measurement of conductivity Indumax CLS50
- High-durability inductive conductivity sensor
- For standard, hazardous and high-temperature applications
- Product Configurator on the product page: www.endress.com/CLS50 Technical Information TI182C Indumax CLS52
- Inductive conductivity sensor
- Short response times for the food industry
- Product Configurator on the product page: www.endress.com/CLS52 Technical Information TI00167C
- Inductive conductivity sensor
- For standard and hazardous area applications, available with hygienic design for food, beverages, pharmaceuticals and biotechnology
- Product Configurator on the product page: www.endress.com/CLS54 Technical Information TI00400C Connection accessories Measuring cable CYK71
- Unterminated cable for connecting analog sensors and for extending sensor cables
- Sold by the meter, order numbers: – Non-Ex version, black: 50085333 – Ex-version, blue: 50085673 Measuring cable CLK6
- Extension cable for inductive conductivity sensors, for extension via VBM junction box
- Sold by the meter, order number: 71183688 VBM
- Junction box for cable extension
- 10 terminal strips
- Cable entries: 2 x Pg 13.5 or 2 x NPT ½"
- Material: aluminum
- Degree of protection: IP 65
- Order numbers – Cable entries Pg 13.5 : 50003987 – Cable entries NPT ½": 51500177 M12 socket
- Four-pin metal socket for mounting on transmitter
- For connecting sensor cables with an M12 connector
- Length of connecting cable for wiring in transmitter: 150 mm (5.9 inch).
- Order number: 51502184 n.c. n.c. A (PA-) A (PA-) S S B (PA+) B (PA+) A (PA-) S (PE) B (PA+) 1.77” 0.28” 5.91” / 11.81” A0024657 Installation accessories CYY101
- Weather protection cover for field devices
- Absolutely essential for field installation
- Material: stainless steel 1.4301 (AISI 304)
- Order No. CYY101-A 320 (12.6) A0024627 13 Dimensions in mm (inch)
18 Endress+Hauser
- For securing the field housing to horizontal and vertical posts and pipes
- Material: stainless steel 1.4301 (AISI 304)
- Order No. 50086842 70 (2.76) 90 (3.54) Ø 6 (0.24) Ø 8 (0.31) 70 (2.76)90 (3.54) 60 (2.36) 90 (3.54) 70 (2.76) Ø max. 60 (2.36) A0024660 14 Dimensions in mm (inch) Universal post CYY102
- Square pipe for mounting transmitters
- Material: stainless steel 1.4301 (AISI 304)
- Order No. CYY102-A 1495 (58.86) 60 x 60 150 (5.91) A0024659 15 Dimensions in mm (inch)