CCA250 EH

Document overview

  • Manufacturer or author: Provided By www.digicamel.com(free datasheet download site)
  • PDF pages: 8

Technical content

TI062C/07/en/07.09 71099077 Technical Information Flowfit CCA250 Flow assembly for chlorine and pH/ORP sensors Application The flow assembly CCA250 has been specially designed to hold the chlorine resp. chlorine dioxide sensors CCS120, CCS140, CCS141, CCS240, CCS241 or CCS142D. Additionally there are two mounting positions for sensors with Pg 13.5 thread and 120 mm (4.72 inch) installation length, e.g. pH or ORP sensors. Main areas of application are:

  • Drinking water
  • W a s h w a t e r
  • Process applications
  • Swimming pool water Your benefits
  • Needle valve to adjust the flow and flow meter for optical control
  • Inductive proximity switch in combination with Liquisys M CCM223/253: a "Flow alarm" can be activated
  • Potential matching pin for pH measurement
  • Cap screwed from below allows calibration of the pH and ORP sensors without sensor removal

2 Endress+Hauser

Function and system design Measuring system A complete measuring system comprises:

  • Flowfit CCA250
  • Chlorine sensor, e.g. CCS142D
  • Measuring cable, e.g. CYK10
  • Transmitter, e.g. Liquiline M Optional:
  • up to two pH sensors, e.g. Orbisint CPS11D
  • Junction box for cable extension, e.g. RM junction box a0007341 Fig. 1: Measuring system

1 Transmitter Liquiline M CM44 wi th weather protection cover

2 Supply line of the transmitter

3 Chlorine sensor CCS142D

4 Flowfit CCA250

5 Assembly inlet (outlet at backsid e, not to be seen in figure)

7 Measuring cable CYK10

Installation instructions Bypass operation To achieve a flow through the bypass, pressure p1 has to be higher than pressure p2. Therefore, you have to install an orifice plate or a throttle valve in the main pipe (→ å, pos. 4). a0011520 Installation example with bypass and orifice plate in the main pipe

1 Stop valve (to be provided by customer)

2 Pressure reducer (if p1 > 4 bar (58 psi)) (to be provided by customer)

3 Stop valve (to be provided by customer)

4 Orifice plate in the main pipe (to be provided by customer)

Caution! p2 must not exceed the permissible operating pressure of the assembly of 4 bar (58 psi). Open outlet operation a0011521 Installation example with open outlet

2 Pressure reducer (if p > 4 bar (58 ps i)) (to be provided by customer)

p1 p2 p 1>p 2 1 2 3 p

4 Endress+Hauser

Ambient temperature range 0 to 50 °C (32 to 120 °F) Storage temperature 0 to 50 °C (32 to 120 °F) Process Process temperature range 0 to 45 °C (32 to 110 °F), non-freezing Process pressure max. medium pressure: 4 bar (58 psi) at 40 °C (104 °F) Pressure-Temperature diagram a0007413 Fig. 2: Temperature-Pressure diagram Flow optimum 30 l/h (7.9 gal/h) 30 to 120 l/h (7.9 to 31.7 gal/h), adjustable

Design, dimensions a0007328 Fig. 3: Dimensions a depending on version: G1/2, NPT 1/2" or NPT 1/4" Weight 0.5 to 0.8 kg (1.1 to 1.8 lbs), depending on process connection Materials In contact with medium:

  • Assembly body: PMMA
  • Mounting parts: PVC, stainless steel 1.4571 (AISI 316 Ti), EPDM Process connection G1/2, NPT 1/2" or NPT 1/4" Suitable sensors • Chlorine sensors: Chlorine or Chlorine dioxide sensors with Ø25 mm (0.98 inch) and shaft length up to 80 mm (3.15 inch)
  • pH/ORP sensors: Glass electrodes with Pg 13.5 thread and shaft length 120 mm (4.72 inch)

6 Endress+Hauser

Ordering information

Scope of delivery The scope of delivery comprises:

  • Flowfit assembly (ordered version)
  • a PML adapter
  • Operating Instructions (English) Accessories Note! In the following sections, you find the accessories available at the time of issue of this documentation. For information on accessories that are not listed here, please contact your local service. Accessories kits For proce ss connection G1/2 only! NV 1/2
  • 2 PVC nipples for connection to PVC pipework
  • for pipes with OD 16 mm (0.63 inch)
  • order no. 50003228 SV 1/2
  • 2 PVC nipples
  • different adaptersfor hose connection – with Ø 6/12 mm (0.24/0.47 inch) in let and Ø 16 mm (0.63 inch) outlet – order no. 50003232
  • identical adapters – for hose connection with Ø 6/12 mm (0.24/0.47 inch) inlet and outlet – order no. 50003230 Sensors CCS120
  • Membrane covered amperometric sensor for total chlorine
  • Measuring range 0.1 to 10 mg/l
  • Ordering acc. to product structure, see Technical Information (TI388C/07/en) CCS140
  • Membrane covered amperometric sensor for free chlorine
  • Measuring range 0.05 to 20 mg/l
  • Ordering acc. to product structure, see Technical Information (TI058C/07/en) CCS141
  • Membrane covered amperometric tr ace sensor for free chlorine
  • Measuring range 0.01 to 5 mg/l
  • Ordering acc. to product structure, see Technical Information (TI058C/07/en) CCS142D
  • Membrane covered amperometric sensor for free chlorine
  • Memosens technology
  • Measuring range 0.01 to 20 mg/l
  • Ordering acc. to product structure, see Technical Information (TI419C/07/en) Process connection, adapter A G1/2, none B G1/2, 2x NV 1/2" C G1/2, 2x D 6/12 D G1/2, D 6/12 + D 16 M NPT 1/2", none Q NPT 1/4", none Proximity switch 0n o n e 1 with inductive proximity switch CCA250- complete order code
  • Membrane covered amperometric sensor for chlorine dioxide
  • Measuring range 0.05 to 20 mg/l
  • Ordering acc. to product structure, see Technical Information (TI114C/07/en) CCS241
  • Membrane covered amperometric tr ace sensor for chlorine dioxide
  • Measuring range 0.01 to 5 mg/l
  • Ordering acc. to product structure, see Technical Information (TI114C/07/en) Orbisint CPS11/11D
  • pH electrode for process applications with dirt-repellent PTFE diaphragm
  • Optional Memosens technology (CPS11D)
  • Ordering acc. to product structure, see Technical Information (TI028C/07/en) Ceragel CPS71/CPS71D
  • pH electrode with double junction refere nce system and integrated bridge electrolyte
  • Optional with Memosens technology (CPS71D)
  • Ordering acc. to product structure, see Technical Information (TI245C/07/en) Ceragel CPS72/CPS72D
  • Redox sensor with double junction refere nce system and integrated bridge electrolyte
  • Optional with Memosens technology (CPS72D)
  • Ordering acc. to product structure, see Technical Information (TI374C/07/de) Ceratex CPS31
  • pH electrode especially suited for swimming pools, with 3 ceramic diaphragms
  • Ordering according to product structure, see Technical Information (TI030C/07/en)

Instruments International AG Kaegenstrasse 2

4153 Reinach