C9619 HAMAMATSU

Document overview

  • PDF pages: 2

Technical content

Page 1

PDF page 1 diagram

The C9619 series are compact PC–board mountable high voltage power supplies, especially designed for photomultiplier tubes. The design offers better performance and completed fail–safe protection. The C9619 and –01 output negative polarity and the C9619–50 and –51 output positive polarity high voltages.

PDF page 1 diagram

FEATURES

GCompact and Lightweight GHigh Stability GHigh Output Power (2 kV / 2 mA) GCompleted Fail-safe Functions GSix-plane Metal Shielded

SPECIFICATIONS

Parameter C9619 C9619-01 C9619-50 C9619-51 Unit
Input Voltage +14 to +16 +11 to +13 +14 to +16 +11 to +13 V
Input Current at Maximum Output Voltage with no load Typ. 120 100 120 100 mA
with full load Typ. 380 460 380 460
Variable Output Voltage Range 0 to -2000 0 to +2000 V
Specification Guaranteed Output Voltage Range -320 to -2000 +320 to +2000 V
Output Current Max. 2 mA
Input Regulation against ± 1 V Input Change Typ. ±0.01 ±0.03 ±0.01 ±0.03 %
Load Regulation against 0 % to 100 % Load Change Typ. ±0.03 %
Ripple Noise (p-p) Typ. 0.003 %
Output Voltage Controlling Modes By external controlling voltage (0 V to +5 V) or external potentiometer (50 kΩ) —
Controlling Voltage Input Impedance Typ. 110 97 kΩ
Output Voltage Setting (Controlling voltage × (-400)) ± 1 % (Controlling voltage × (+400)) ± 1 % V
Output Voltage Rise Time (0 % / 99 %) Typ. 150 ms
Temperature Coefficient Typ. ±0.01 %/°C
Operating Ambient Temperature / Humidity 0 to +40 / 85 °C / %
Storage Temperature / Humidity -20 to +60 / 90 °C / %
Weight 100 g
Protective Functions Units protected against reversed power input, reversed / excessive controlling voltage input, continuous overloading / short circuit in output —

Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2007 Hamamatsu Photonics K.K.

Page 2

HIGH VOLTAGE POWER SUPPLY UNIT C9619 SERIES

Figure 1: Output Voltage Controlling

BY EXTERNAL VOLTAGE

BY EXTERNAL POTENTIOMETER

PDF page 2 diagram

voltage should be minimized as it directly affects the output voltage.

Figure 2: Output Voltage Controlling Characteristics

PDF page 2 diagram

CONTROLLING VOLTAGE (V)

Figure 4: Dimensional Outline (Unit: mm)

TACCC0139EA

Figure 3: Example of Ripple Noise Reduction Circuit

PDF page 2 diagram

+12 V/+15 V

\*The ripple noise can be reduced to approx. 1/10 by adding capacitors as illustrated above.

TACCC0140EA

PDF page 2 diagram

TACCA0291EA

WEB SITE www.hamamatsu.com

HAMAMATSU PHOTONICS K.K., Electron Tube Division 314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: infos@hamamatsu.fr

United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: info@hamamatsu.co.uk North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se

TACC1039E01

Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it

APR. 2007. IP