Digital Storage Oscilloscopes TDS1000B Series Data Sheet Features & Benefits
40 MHz, 60 MHz, and 100 MHz Bandwidths
S a m p l eR a t e su pt o1G S / sR e a lT i m e
2 Channels
Removable Data Storage using the Front-panel USB Port Seamless PC Connectivity through the USB Device Port, with OpenChoice® and NI SignalExpress® PC Software Advanced Triggers including Pulse Width Trigger and Line-selectable Video Trigger FFT Standard on All Models
12 Automatic Measurements
Multiple-language User Interface and Context-sensitive Help Direct Print to All PictBridge®-compatible Printers through the USB Device Port Lifetime Warranty*1
Applications
Manufacturing Test and Quality Control Service and Repair *1 Limitations apply. For terms and conditions, visit www.tektronix.com/lifetimewarranty.
Quickly and easily capture waveforms. TDS1000B Series Oscilloscopes Instantly Productive. Incredibly Easy. The TDS1000B Series digital storage oscilloscopes deliver an unbeatable combination of performance and ease of use, at a price you can afford. Affordable Digital Precision With up to 100 MHz bandwidth and 1 GS/s maximum sample rate, no other digital storage oscilloscope offers as much bandwidth and sample rate for the price. The TDS1000B Series oscilloscopes provide accurate real-time acquisition up to their full bandwidth, the same record length at all time base settings, advanced triggers to isolate signals of interest, and 12 standard automatic measurements on all models. Their Fast Fourier Transform (FFT) and waveform add, subtract, and multiply math functions allow you to analyze, characterize, and troubleshoot circuits. Quick and Easy Waveform Capture The simple user interface with classic analog-style controls makes these instruments easy to use, reducing learning time and increasing efficiency. Innovative features such as the Autoset Menu, Probe Check Wizard, and Context-sensitive Help Menu optimize instrument setup and operation. Flexib l eD a t aT r a n s f e r With USB host and device ports which enable removable data storage, seamless PC connectivity, and direct printing, no other digital storage oscilloscope offers as muchflexibility and ease of data transfer for the price. Simple Documentation and Analysis Easily capture, save, and analyze measurement results with OpenChoice PC Communications Software. Simply pull screen images and waveform data into the stand-alone desktop application or directly into Microsoft Word and Excel. To complement OpenChoice, National Instruments Conveniently use your USBflash drive to store screenshots and waveform data. Easily capture, save, and analyze measurement results with OpenChoice PC Communications Software. SignalExpress Tektronix Edition Software provides you with extended capabilities, including advanced analysis, remote oscilloscope control, and live waveform analysis. Alternatively, if you prefer not to use the PC, you can simply print your image directly to any PictBridge-compatible printer using the USB device port. Performance You Can Count On Depend on Tektronix to provide you with performance you can count on. In addition to industry-leading service and support, every TDS1000B Series oscilloscope comes backed with a Lifetime Warranty* 1 as standard. *1 Limitations apply. For terms and conditions, visit www.tektronix.com/lifetimewarranty. 2 www.tektronix.com
Digital Storage Oscilloscopes — TDS1000B Series Characteristics TDS1000B Series Digital Storage Oscilloscopes Characteristic TDS1001B TDS1002B TDS1012B Display (1/4 VGA LCD) Mono Bandwidth*2 40 MHz 60 MHz 100 MHz Channels 2 External Trigger Input Included on all models Sample Rate on Each Channel 500 MS/s 1.0 GS/s 1.0 GS/s Record Length 2.5K points at all time bases on all models Vertical Resolution 8b i t s Vertical Sensitivity 2 mV to 5 V/div on all models with calibratedfine adjustment DC Vertical Accuracy ±3% on all models Vertical Zoom Vertically expand or compress a live or stopped waveform Maximum Input Voltage
300 VRMS CAT II; derated at 20 dB/decade above
100 kHz to 13 Vp-p AC at 3 MHz Position Range 2m Vt o200 mV/div +2 V; >200 mV to 5 V/div +50 V Bandwidth Limit 20 MHz for all models Input Coupling AC, DC, GND on all models Input Impedance 1M Ω in parallel with 20 pF Time Base Range 5n st o50 s/div Time Base Accuracy 50 ppm Horizontal Zoom Horizontally expand or compress a live or stopped waveform I/O Interfaces USB Ports USB host port on front panel supports USB flash drives USB device port on back of instrument supports connection to PC and all PictBridge-compatible printers GPIB Optional Nonvolatile Storage Reference waveform display (2) 2.5K point reference waveforms Waveform storage without USB flash drive (2) 2.5K point Maximum USB flash drive size 64 GB Waveform storage with USB flash drive 96 or more reference waveforms per 8 MB Setups without USB flash drive 10 front-panel setups Setups with USB flash drive 4000 or more front-panel setups per 8 MB Screen images with USBflash drive 128 or more screen images per 8 MB (the number of images depends onfile format selected) Save All with USB flash drive 12 or more Save All operations per 8 MB A single Save All operation creates 3 to 9files (setup, image, plus onefile for each displayed waveform) *2 Bandwidth is 20 MHz at 2 mV/div, all models. Acquisition Modes Mode Description Peak Detect High-frequency and random-glitch capture. Captures glitches as narrow as 12 ns (typical) at all time base settings from 5μs / d i vt o5 0s / d i v Sample Sample data only Average Waveform averaged, selectable: 4, 16, 64, 128 Single Sequence Use the Single Sequence button to capture a single triggered acquisition sequence Roll At acquisition time base settings of >100 ms/div Trigger System Characteristic Description Trigger Modes Auto, Normal, Single Sequence Trigger Types Trigger Description Edge (Rising/Falling) Conventional level-driven trigger. Positive or negative slope on any channel. Coupling selections: AC, DC, Noise Reject, HF Reject, LF Reject Video Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC, PAL, SECAM) Pulse Width (or Glitch) Trigger on a pulse width less than, greater than, equal to, or not equal to, a selectable time limit ranging from 33 ns to 10 s Trigger Source C H 1 ,C H 2 ,E x t ,E x t / 5 ,A CL i n e . Trigger View Displays trigger signal while Trigger View button is depressed. Trigger Signal Frequency Readout Provides a frequency readout of the trigger source. Cursors Characteristic Description Types Amplitude, Time Measurements ΔT, 1/ΔT, ΔV Automatic Waveform Measurements Period, Frequency, +Width, –Width, Rise Time, Fall Time, Max, Min, Peak-to-Peak, Mean, RMS, Cycle RMS. Waveform Math Characteristic Description Operators Add, Subtract, Multiply, FFT FFT Windows: Hanning, Flat Top, Rectangular; 2048 sample points Sources CH1 – CH2, CH2 – CH1, CH1 + CH2, CH1 × CH2 www.tektronix.com 3
Single-button, automatic setup of all channels for vertical, horizontal, and trigger systems, with undo Autoset. Signal Type Autoset Menu C hoices Square Wave Single Cycle, Mult icycle, Rising or Falling Edge Sine Wave Single Cycle, Multicycle, FFT Spectrum Video (NTSC, PAL, SECAM) Field: All, Odd, or Even Line: All or Selectable Line Number Autorange Automatically adjust vertical and/or horizontal oscilloscope settings when probe is moved from point to point, or when the signal exhibits large changes. Display Characteristics Characteristic Description Display ¼ VGA backlit passive LCD with adjustable multilevel contrast and inverse video selectable from front panel Interpolation Sin (x)/x Display Types Dots, vectors Persistence O f f ,1s ,2s ,5s ,i nfinite Format YT and XY Multiple-language User Interface and Context-sensitive Help Characteristic Description Languages Available English, French, German, Italian, Japanese, Korean, Portuguese, Russian*3,S i m p l ified Chinese, Spanish, Traditional Chinese *3 Requires Russianfirmware, indicated by "RUS" suffix. Environmental and Safety Characteristic Description Temperature Operating 0 to +50 ºC Nonoperating – 4 0t o+ 7 1º C Humidity Operating and Nonoperating Up to 80% RH at or below +40 ºC Operating and Nonoperating Up to 45% RH up to +50 ºC Altitude Operating and Nonoperating Up to 3,000 m Electromagnetic Compatibility Meets Directive 2004/108/EC, EN 61326-2-1 Class A; Australian EMC Framework Safety UL61010-1:2004, CSA22.2 No. 61010-1:2004, EN61010-1:2001, IEC61010-1:2001 Physical Characteristics Instrument Dimensions mm in. Width 326.3 12.85 Height 158.0 6.22 Depth 124.2 4.89 Weight kg lb. Instrument Only 2.0 4.4 With accessories 2.2 4.9 Instrument Shipping Package Dimensions mm in. Width 476.2 18.75 Height 266.7 10.5 Depth 228.6 9.0 RM2000B Rackmount mm in. Width 482.6 19.0 Height 177.8 7.0 Depth 108.0 4.25 4 www.tektronix.com
Digital Storage Oscilloscopes — TDS1000B Series
Ordering Information
TDS1001B 40 MHz, 2 Ch, 500 MS/s, Monochrome DSO TDS1002B 60 MHz, 2 Ch, 1 GS/s, Monochrome DSO TDS1012B 100 MHz, 2 Ch, 1 GS/s, Monochrome DSO Standard Accessories Accessory Description Passive Probes 200 MHz (one per channel) Power Cord (Please specify plug option) NIM/NIST Traceable Certi ficate of Calibration Documentation User manual (please specify preferred language option) OpenChoice PC Communications Software Enables fast and easy communication between a Windows PC and the TDS1000B Series using USB. Transfer and save settings, waveforms, measurements, and screen images National Instruments SignalExpress Tektronix Edition Interactive Measurement Software – Base Version A fully interactive measurement software environment optimized for the TDS1000B Series. Enables you to instantly acquire, generate, analyze, compare, import, and save measurement data and signals using intuitive drag-and-drop user interface that does not require any programming. Standard TDS1000B Series support for acquiring, controlling, viewing, and exporting your live signal. A 30-day trial period of the Professional Version provides additional signal processing, advance analysis, mixed signal, sweeping,limit testing, and user-defined step capabilities. Order SIGEXPTE for permanent Professional Version capability Limited Lifetime Warranty* Covers labor and parts for defects in materials and workmanship for a minimum of 10 years, excluding probes and accessories* *4 Lifetime is defined as 5 years after Tektronix discontinues manufacturing the product, but the warranty length shall be at least 10 years from date of original purchase. Lifetime warranty is nontransferable, proof of original purchase is required. Limitations apply. For terms and conditions visit www.tektronix.com/lifetimewarranty. 5 Probes and accessories are not covered by the oscilloscope warranty and service offerings. Refer to the data sheet of each probe and accessory model for its unique warranty and calibration terms. Power Plug Options Option Description A0 North America power A1 Universal Euro power A2 United Kingdom power A3 Australia power A5 Switzerland power A6 Japan power A10 China power A11 India power A99 No power cord or AC adapter User Manual Options Translated front-panel overlays included with their respective user manuals. Option Description L0 English manual L1 French manual L2 Italian manual L3 German manual L4 Spanish manual L5 Japanese manual L6 Portuguese manual L7 Simple Chinese manual L8 Standard Chinese manual L9 Korean manual L10 Russian manual Recommended Accessories Accessory Description TEK-USB-488 GPIB-to-USB Converter SIGEXPTE National Instruments SignalExpress Tektronix Edition Interactive Measurement Software – Professional Version AC2100 Soft Carrying Case for instrument HCTEK4321 Hard Plastic Carrying Case for instrument (requires AC2100) RM2000B Rackmount Kit 071-1075-xx Programmer's Manual – English Only 071-1828-xx Service Manual – English Only 174-4401-xx USB Host-to-Device Cable, 3 ft. long Recommended Probes Probe Description P2220 10X to 1X Switchable Passive probe (200 MHz when 10X is selected) P6101B 1X Passive probe (15 MHz, 300 VRMS CAT II rating) P6015A 1000X High-voltage Passive probe (75 MHz) P5100 100X High-voltage Passive probe (250 MHz) P5200 High-voltage Active Differential probe (25 MHz) P6021 15 A, 60 MHz AC Current probe P6022 6 A, 120 MHz AC Current probe A621 2 0 0 0A ,5t o5 0k H zA CC u r r e n tp r o b e A622 100 A, 100 kHz AC/DC Current probe/BNC TCP303/TCPA300 150 A, 15 MHz AC/DC Current probe/ampli fier TCP305/TCPA300 50 A, 50 MHz AC/DC Current probe/ampli fier TCP312/TCPA300 30 A, 100 MHz AC/DC Current probe/ampli fier TCP404XL/TCPA400 500 A, 2 MHz AC/DC Current probe/amplifier www.tektronix.com 5
Service Options*5 Option Description D1 Calibration Data Report CA1 Provides a single calibration event or coverage for the designated calibration interval, whichever comesfirst *5 Probes and accessories are not covered by the oscilloscope warranty and service offerings. Refer to the data sheet of each probe and accessory model for its unique warranty and calibration terms. Service Offerings (Available after purchase) Option Description TDSxxxxB-CA1 Provides a single calibration event or coverage for the designated calibration interval, whichever comesfirst The Complete Measurement Solution The AFG3000 Series arbitrary function generator pairs with the TDS1000B Series digital storage oscilloscopes to deliver the two elements of a complete measurement solution – stimulus and acquisition. This instrument combines the capabilities of a function generator with the power of an arbitrary waveform generator, offering the performance needed to accurately verify, validate, and characterize designs with ease and confid e n c ea tap r i c ey o uc a na f f o r d . The Tektronix Customer Service Advantage You can trust Tektronix to offer unequalled engineering expertise and a customer-centric approach to ensure the optimal performance of your Tektronix products and maximize the lifetime value of your Tektronix investment. With service from Tektronix you get: Access to the source of product knowledge; unsurpassed technical expertise Your challenges solved by front-line technical experts, design engineering reinforcement, and online support tools Comprehensive and thorough support provided worldwide, including software andfirmware updates, data reports, and adjustments Efficiency and convenience; no-hassle service from initial service call to turnaround and delivery Flexible repair and calibration service with access to the best on-call technical troubleshooting staff in the industry, with over 20 years of training per support engineer Customer-centric approach dedicated to serving your needs everyday with services designed to optimize your product performance, increase productivity and ROI by delivering afixed cost of ownership, and efficient management of service Get checked by Tektronix. Visit www.tektronix.com/serviceandsupport. Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar. Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats. 6 www.tektronix.com
Digital Storage Oscilloscopes — TDS1000B Series www.tektronix.com 7
Data Sheet Contact Tektronix: ASEAN / Australasia (65) 6356 3900 Austria 00800 2255 4835* Balkans, Israel, South Africa and other ISE Countries+41 52 675 3777 Belgium 00800 2255 4835* Brazil + 5 5( 1 1 )3 7 5 97 6 2 7 Canada 1 800 833 9200 Central East Europe and the Baltics+ 4 15 26 7 53 7 7 7 Central Europe & Greece+41 52 675 3777 Denmark + 4 58 08 81 4 0 1 Finland + 4 15 2675 3777 France 00800 2255 4835* Germany 00800 2255 4835* Hong Kong400 820 5835 India 000 800 650 1835 Italy 00800 2255 4835* Japan 81 (3) 6714 3010 Luxembourg + 4 15 26 7 53 7 7 7 Mexico, Central/South America & Caribbean52 (55) 56 04 50 90 Middle East,Asia, and North Africa+41 52 675 3777 The Netherlands00800 2255 4835* Norway 800 16098 People’s Republic of China400 820 5835 Poland +41 52 675 3777 Portugal 80 08 12370 Republic ofKorea 001 800 8255 2835 Russia & CIS+7 (495) 7484900 South Africa+ 4 15 26 7 53 7 7 7 Spain 00800 2255 4835* Sweden 00800 2255 4835* Switzerland 00800 2255 4835* Taiwan 886 (2) 2722 9622 United Kingdom & Ireland00800 2255 4835* USA 1 800 833 9200 * European toll-free number. If not accessible, call: +41 52 675 3777 Updated 10 February 2011 For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visitwww.tektronix.com Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies.
16 Mar 2012 3GW-25644-2